Snowden-Pencer Ergonomic Take-Apart In-Line and Pistol Grip Laparoscopic Instruments

K151036 · Carefusion 2200, Inc. · GCJ · Feb 1, 2016 · Gastroenterology, Urology

Device Facts

Record IDK151036
Device NameSnowden-Pencer Ergonomic Take-Apart In-Line and Pistol Grip Laparoscopic Instruments
ApplicantCarefusion 2200, Inc.
Product CodeGCJ · Gastroenterology, Urology
Decision DateFeb 1, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Snowden-Pencer Ergonomic Take-Apart In-Line and Pistol-Grip Laparoscopic Instruments are designed to transmit cutting, clamping, grasping, suturing and dissecting force to working tips in minimally invasive surgeries such as plastic and general laparoscopic procedures.

Device Story

Reusable laparoscopic instruments; handle, shaft, and insert assembly; end-effector jaw patterns. Manual operation by surgeons in OR. Transmits mechanical force for cutting, clamping, grasping, suturing, and dissecting. Take-apart design facilitates cleaning and sterilization. Sold non-sterile; steam sterilized by end-user. Benefits include ergonomic handling and improved reprocessing access compared to single-piece designs.

Clinical Evidence

No clinical tests were conducted for this submission. Substantial equivalence is supported by non-clinical bench testing, including instrument strength, pull strength, assembly/disassembly force, reliability, cleaning/sterilization validation per AAMI/ISO standards, and biocompatibility per ISO 10993.

Technological Characteristics

Reusable manual laparoscopic instruments. Materials: Stainless Steel, Radel-R, Halar, Tungsten Carbide. Configurations: In-line and pistol-grip handles; rotatable/non-rotatable shafts (5mm, 10mm diameters). Sterilization: Steam (Prevacuum). Biocompatibility: ISO 10993. Cleaning: Manual and automatic.

Indications for Use

Indicated for patients undergoing minimally invasive plastic or general laparoscopic surgical procedures requiring cutting, clamping, grasping, suturing, or dissection.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized design of three human profiles facing right, with flowing lines above them. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the design. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 February 1, 2016 CareFusion 2200 Incorporated Ms. Jane Weber, MS Regulatory Affairs Manager 75 North Fairway Drive Vernon Hills, Illinois 60061 Re: K151036 Trade/Device Name: Snowden-Pencer Ergonomic Take-Apart In-Line and Pistol-Grip Laparoscopic Instruments Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: GCJ Dated: December 14, 2015 Received: December 15, 2015 Dear Ms. Weber: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {1}------------------------------------------------ forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. # Jennifer R. Stevenson -S For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K151036 ### Device Name Snowden-Pencer Ergonomic Take-Apart In-Line and Pistol-Grip Laparoscopic Instruments #### Indications for Use (Describe) Snowden-Pencer Ergonomic Take-Apart In-Line and Pistol-Grip Laparoscopic Instruments are designed to transmit cutting, clamping, grasping, suturing and dissecting force to working tips in minimally invasive surgeries such as plastic and general laparoscopic procedures. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) ] Over-The-Counter Use (21 CFR 801 Subpart C) ### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. ### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) {3}------------------------------------------------ This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ ## 510(k) SUMMARY – K151036 Image /page/4/Picture/1 description: The image shows the CareFusion logo. The logo features an orange circle with a white shield inside, with a yellow accent on the top left of the shield. To the right of the circle is the text "CareFusion" in a gray sans-serif font. Below the text is the phrase "has joined BD" in a smaller, lighter gray font. A summary of 510(k) safety and effectiveness information in accordance with 21 CFR 807.92. | SUBMITTER INFORMATION | | |------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Name | CareFusion 2200 Inc | | Address | 75 N. Fairway Dr.<br>Vernon Hills, IL 60061 | | Phone number | (847) 362-8094 (Jane Weber) | | Fax number | (312) 949-0272 (Jane Weber) | | Establishment<br>Registration Number | 1423507 | | Name of contact person | Jane Weber, MS, Regulatory Affairs Manager<br>jane.weber@carefusion.com | | Date prepared | 27-JAN-2016 | | DEVICE INFORMATION | | | Trade or proprietary<br>name | Snowden-Pencer Ergonomic Take-Apart In-Line and Pistol-Grip<br>Laparoscopic Instruments | | Common or usual name | Reusable Laparoscopic Instruments | | Classification name | Endoscopic and Accessories | | Classification panel | 78 Gastroenterology and Urology | | Regulation | Class II per 21CFR 876.1500, Product code GCJ | | Product Code(s) | Multiple devices | | Legally marketed<br>device(s) to which<br>equivalence is claimed | Snowden-Pencer Inc. - Reusable Laparoscopic Instruments - K930667,<br>GCJ | | Reason for 510(k)<br>submission | Design modification to allow device to be taken apart | | Device description | Snowden-Pencer Ergonomic Take-Apart In-Line and Pistol-Grip<br>Laparoscopic Instruments consists of a handle, shaft and insert. The<br>insert contains an end-effector (jaw pattern). There are several device<br>models that encompass various lengths, diameters and jaw patterns<br>based on the surgeon's needs. | | Intended use of the<br>device | Cutting, clamping, grasping, suturing and dissecting | | Indications for use | Snowden-Pencer Ergonomic Take-Apart In-Line and Pistol-Grip<br>Laparoscopic Instruments are designed to transmit cutting, clamping,<br>grasping, suturing and dissecting force to working tips in minimally<br>invasive surgeries such as plastic and general laparoscopic procedures. | {5}------------------------------------------------ ### 510(k) SUMMARY – K151036 ### SUMMARY OF THE TECHNOLOGICAL CHARACTERISTICS OF THE DEVICE COMPARED TO THE PREDICATE DEVICE | Characteristic | Predicate Device | New Device | |--------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------| | Type of Device | Reusable | Reusable | | Intended for direct patient<br>contact? | Yes | Yes | | Intended for extended<br>use? (more than 24 hrs) | No | No | | Cleaning Method | Manual and Automatic | Manual and Automatic | | Sterilization Method | Product is sold non-sterile and<br>sterilized by the end user via the<br>following steam sterilization options:<br>• Gravity Displacement<br>• Prevacuum | Product is sold non-sterile and<br>sterilized by the end user via<br>the following steam sterilization<br>options:<br>• Prevacuum | | Materials | Stainless Steel<br>Aluminum<br>Radel-R<br>Halar<br>Tungsten Carbide | Stainless Steel<br>Radel-R<br>Halar<br>Tungsten Carbide | | Handle Configuration | Ring Handle, In-Line handle, and Pistol Grip (bent) | In-Line handle and Pistol Grip<br>(bent) | | Shaft/End Effector<br>Rotation Capability | Rotatable and Non Rotatable | Rotatable and Non Rotatable | | Instrument Length | 24cm, 33cm, 36cm, 45cm | 24cm, 33cm, 36cm, 45cm | | Instrument Diameter | 3.5mm, 5mm, 10mm | 5mm and 10mm | | End Effectors (Jaw<br>Patterns) | Scissors, Graspers, Dissectors,<br>Clamps and Needle Holders | Scissors, Graspers, Dissectors,<br>Clamps and Needle Holders | | Device Configuration | Single-piece design | Take-Apart design | {6}------------------------------------------------ ### PERFORMANCE DATA ### SUMMARY OF NON-CLINICAL TESTS CONDUCTED FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE ### Performance Test Summary - New Device | Characteristic | Standard / Test / FDA Guidance | Results Summary | |--------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|-----------------| | Instrument strength | Strength Tests | PASS | | Inadvertent disassembly | Pull Strength Test | PASS | | Ease of disassembly/reassembly | Force Tests | PASS | | Handle must be reusable | Reliability Test | PASS | | Handle compatibility with existing<br>shafts and inserts | Handle ability to connect to worst case<br>shafts and inserts | PASS | | Device is able to be cleaned and<br>sterilized | AAMI TIR12, AAMI TIR30, ANSI AAMI<br>ST79, ANSI AAMI ST81, ISO 11138, ISO<br>17664, ISO 17665 | PASS | | Device materials are<br>biocompatible | ISO 10993 | PASS | | SUMMARY OF CLINICAL TESTS CONDUCTED FOR DETERMINATION OF<br>SUBSTANTIAL EQUIVALENCE AND/OR OF CLINICAL INFORMATION | | | N/A - No clinical tests were conducted for this submission. ### CONCLUSIONS DRAWN FROM NON-CLINICAL AND CLINICAL DATA The results of the non-clinical tests demonstrate the Snowden-Pencer Ergonomic Take-Apart In-Line and Pistol-Grip Laparoscopic Instruments meet all performance requirements, and are substantially equivalent to the predicate devices.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...