K991814 · Dexide, Inc. · GEI · Jun 23, 1999 · General, Plastic Surgery
Device Facts
Record ID
K991814
Device Name
MULTACHOICE INSTRUMENTS
Applicant
Dexide, Inc.
Product Code
GEI · General, Plastic Surgery
Decision Date
Jun 23, 1999
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The MultAchoice™ Instruments may be used in a variety of endoscopic procedures. The MultAchoice™ Scissors are designed for cutting or transecting tissue. The MultAchoice™ Grasper is designed for grasping and manipulating tissue. The MultAchoice™ Dissector is designed for blunt dissection, temporary grasping, and manipulation of tissue.
Device Story
MultAchoice™ Instruments are reusable, configurable laparoscopic tools comprising a handle, insulated shaft, and instrument insert. Available as scissors, graspers, or dissectors. Used in endoscopic procedures by surgeons to cut, transect, grasp, manipulate, or blunt-dissect tissue. Features include 5mm OD, 31cm working length, 360° shaft rotation for improved visibility/access, and a standard banana plug for optional monopolar electrosurgery. Devices are fully autoclavable for reuse. Clinical benefit derived from providing surgeons with versatile, reusable tools for tissue management during minimally invasive surgery.
Clinical Evidence
bench testing only
Technological Characteristics
Reusable laparoscopic instruments (scissors, graspers, dissectors). Features 5mm OD insulated shaft, 31cm working length, 360° rotation mechanism, and standard banana plug for monopolar electrosurgery. Fully autoclavable.
Indications for Use
Indicated for use in endoscopic procedures for cutting, transecting, grasping, manipulating, or blunt dissection of tissue in patients requiring such surgical intervention.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
## Attachment 3
## SUMMARY OF SAFETY AND EFFECTIVENESS 991814 SPECIAL 510(k) SUMMARY
| 1) | Submitter: | Dexide, Inc.<br>7509 Flagstone Drive<br>Fort Worth, TX 76118-6995<br>Phone No.: (817) 589-1454 |
|----|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Contact Person: | John Corzine<br>Vice President, Research and Development |
| | Date Prepared: | May 26, 1999 |
| 2) | Name of Device: | MultAchoice™ Instruments |
| | Common Name: | Laparoscopic Scissors, Graspers, and Dissectors |
| | Classification Name: | Laparoscopic, General & Plastic Surgery |
| 3) | Predicate Device: | Dexide Inc., Laparoscopic Accessories, K923845 |
| 4) | Description of Device: | The MultAchoice™ Instruments are a set of reusable and<br>configurable instruments consisting of a Handle, Insulated Shaft,<br>and Instrument Insert that comprise a Laparoscopic Scissor,<br>Grasper, and Dissector. |
| 5) | Intended Use: | The MultAchoice™ Instruments may be used in a variety of<br>endoscopic procedures. The MultAchoice™ Scissors are<br>designed for cutting or transecting tissue. The MultAchoice™<br>Grasper is designed for grasping and manipulating tissue. The<br>MultAchoice™ Dissector is designed for blunt dissection,<br>temporary grasping, and manipulation of tissue. |
- 6) The candidate device is identical to the Predicate Device except for the additional capability of being reusable. They have the same intended use and operating principle. They have the same basic design principles and incorporate the same basic materials except for those required to accomplish the reusable specification.
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, which is a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the caduceus. The logo is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN 23 1999
Mr. John Corzine Vice President, Research and Development Dexide, Inc. 7509 Flagstone Drive Fort Worth, Texas 76118
Re: K991814
> Trade Name: MultAchoice™ Instruments Regulatory Class: II Product Code: GEI Dated: May 26, 1999 Received: May 27, 1999
Dear Mr. Corzine:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition. FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{2}------------------------------------------------
Page 2 - Mr. John Corzine
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits vour device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Celia M. Witten, Ph.D., M.D.
Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Attachment 8
1. 11
| Page | 1 | Of | 1 |
|------|---|----|---|
|------|---|----|---|
K 991814 510(k) Number (if known): _
Device Name: MultAchoice™ Instruments
Indications For Use:
MultAchoice Scissor Instrument:
The MultAchoice Scissor may be used in a variety of endoscopic procedures. This device is designed for cutting or transecting tissue. It is fully autoclavable. The insulated shaft of this device has a 5mm OD and a 31cm working length. This shaft may be rotated 360° independent of the instrument handle, thereby permitting increased visibility and access to desired structures. The MultAchoice Scissor has been fitted with a standard banana plug to accept a monopolar connecting cable (not included) for electrosurgery.
MultAchoice Grasper Instrument:
The MultAchoice Grasper may be used in a variety of endoscopic procedures. This device is designed for grasping and manipulating tissue. It is fully autoclavable. The insulated shaft of this device has a 5mm OD and a 31cm working length. This shaft may be rotated 360° independent of the instrument handle, thereby permitting increased visibility and access to desired structures. The MultAchoice Grasper has been fitted with a standard banana plug to accept a monopolar connecting cable (not included) for electrosurgery.
MultAchoice Dissector Instrument:
The MultAchoice Dissector may be used in a variety of endoscopic procedures. This device is designed for blunt dissection, temporary grasping, and manipulation of tissue. It is fully autoclavable. The insulated shaft of this device has a 5mm OD and a 31cm working length. This shaft may be rotated 360° independent of the instrument handle, thereby permitting increased visibility and access to desired structures. The MultAchoice Dissector has been fitted with a standard banana plug to accept a monopolar connecting cable (not included) for electrosurgery.
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
| | Concurrence of CDRH, Office of Device Evaluation (ODE) |
|-----------------------------------------|--------------------------------------------------------|
| (Division Sign-Off) | |
| Division of General Restorative Devices | |
| 510(k) Number | K991814 |
Prescription Use
OR
Over-The-Counter Use _____
(Per 21 CFR 801.109)
(Optional Format 1-2-96)