AnQing Medical Single Use Flexible Cystoscope is used to provide visualization and operative access during diagnostic and therapeutic endoscopic procedures of the lower urinary tract, including the bladder and urethra. The Flexible Cystoscope is intended for use in a hospital environment or medical office environment. The Flexible Cystoscope is designed for use in adults.
Device Story
Single-use flexible cystoscope for lower urinary tract visualization and operative access; hospital or medical office use. Device comprises handle with deflection lever, insertion section, distal tip, and connector. Distal tip contains CMOS sensor and LED illumination. User operates deflection lever to maneuver tip; inputs include visual data from body cavity. CMOS captures images; RGB video signals transmitted to external video processor (EOS-H-01) via wiring. Processor outputs video to monitor and controls LED brightness. Luer port on handle allows accessory insertion and irrigation. Physician uses real-time visualization to guide diagnostic or therapeutic interventions. Benefits include sterile, single-use design eliminating reprocessing requirements.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing, including electrical safety (ANSI/AAMI ES:60601-1, IEC 60601-2-18, IEC 60601-1-2), photobiological safety (IEC 62471), mechanical/optical performance (ISO 8600), biocompatibility (ISO 10993), and sterilization validation (ISO 11135).
Technological Characteristics
Sterile, single-use flexible endoscope. CMOS imaging sensor; LED illumination. Materials compliant with ISO 10993-1. Sterilized via Ethylene Oxide (EO) per ISO 11135. Dimensions: 5.0-5.5mm outer diameter; 2.0-2.4mm working channel. Connectivity: Wired connection to external video processor. Energy source: Electricity.
Indications for Use
Indicated for adult patients undergoing diagnostic and therapeutic endoscopic procedures of the lower urinary tract, including the bladder and urethra.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION".
November 9, 2022
Shanghai AnQing Medical Instrument Co., Ltd Shuwen Fan RA Manager 3 & 4 Floor, No.2 Building, 366 Huiging Road, East Zhangjiang High-Tech Park Shanghai, 201201 China
Re: K222162
> Trade/Device Name: Flexible Cystoscope (Model: CY50H-20EU, CY50H-20US, CY55H-24EU, CY55H-24US) Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: FAJ Dated: October 11, 2022 Received: October 11, 2022
Dear Shuwen Fan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Mark J. Antonino -S
Mark J. Antonino, M.S. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K222162
Device Name
Flexible Cystoscope (Model: CY50H-20EU, CY50H-20US, CY55H-24EU, CY55H-24US)
Indications for Use (Describe)
AnQing Medical Single Use Flexible Cystoscope is used to provide visualization and operative access during diagnostic and therapeutic endoscopic procedures of the lower urinary tract, including the bladder and urethra.
The Flexible Cystoscope is intended for use in a hospital environment or medical office environment. The Flexible Cystoscope is designed for use in adults.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|
| <span style="font-size: large;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size: large;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image contains the logo for Innovex Anqing Medical. The logo features a purple graphic on the left side, with the word "INNOVEX" in a sans-serif font to the right of the graphic. Below "INNOVEX" is the text "Anqing Medical" in a larger, bolder font. A curved line is above the word "INNOVEX".
Device Class:
Review Panel:
510(k) Premarket Notification Submission Flexible Cystoscope
| | 510(k) Summary |
|-------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date Prepared: | July 8, 2022 |
| Manufacturer: | Shanghai AnQing Medical Instrument Co., Ltd.<br>3 & 4 Floor, No.2 Building, 366 Huiqing Rd, East<br>Zhangjiang High-Tech Park, 201201<br>Shanghai, China |
| Contact Person: | Shuwen Fan<br>RA Manager<br>Shanghai AnQing Medical Instrument Co., Ltd.<br>Tel: +86-21-61117375<br>Fax: +86-21-61117374<br>ra_dept@anqing-sh.com |
| Identification of the Device: | |
| Proprietary/Trade Name: | Flexible Cystoscope |
| Model: | CY50H-20EU, CY50H-20US, CY55H-24EU,<br>CY55H-24US |
| Common name: | Flexible Cystoscope |
| Classification Name: | Endoscope and Accessories |
| Regulatory Number: | 21 CFR Part 876.1500 |
| Product Code: | FAJ |
# rketed Predicate Devic
| Identification of the Legally Marketed Predicate Device: | | | | |
|-----------------------------------------------------------------|--------------------------------------------|--|--|--|
| Trade Name: | Ambu aScope 4 Cysto | | | |
| Common name | Flexible Cystoscope | | | |
| Classification Name: | Cystoscope And Accessories, Flexible/Rigid | | | |
| Regulatory Number: | 21 CFR Part 876.1500 | | | |
| Product Code: | FAJ | | | |
| Device Class: | Class II | | | |
| Review Panel: | Gastroenterology/Urology | | | |
| Submitter/510(k) Holder: | Ambu A/S. | | | |
| Clearance: | K193095 (cleared April 2, 2020) | | | |
| This predicate has not been subject to a design-related recall. | | | | |
Class II
Gastroenterology/Urology
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image contains a logo for Innovex Anqing Medical. The logo features the word "INNOVEX" in a bold, sans-serif font, with the registered trademark symbol to the right. Below that is the text "ANQING MEDICAL" in a slightly smaller, bold font. To the left of the text is a purple graphic with a white leaf design.
## Device Description:
The Flexible Cystoscope (Model: CY50H-20EU, CY50H-20US, CY55H-24EU, is a sterile single-use endoscope which is used with the video CY55H-24US) processor (Model: EOS-H-01, FDA cleared #K211169) produced by AnQing for providing endoscopic imaging of the lower urinary tract, including the bladder and urethra for the purpose of diagnosis and treatment.
The CY50H-20 and CY55H-24 difference is the size of Distal End Outer Diameter and Working Channel Inner Diameter. The US and EU model are identical except the deflection versions, which is opposite from each other (EU version or US version).
The Flexible Cystoscope is a single-use endoscope, which consists of Handle, Insertion Section, Distal Tip, and Endoscope Connector.
The handle includes a deflection lever, a lever lock, a push button for picture taking/video recording and a Luer port for insertion of accessory devices and irrigation to the working channel. The insertion contains one working channel and wiring to transmit the image signals to the Video Processor. The distal bending section of the insertion section is controlled by the user via the deflection lever on the handle. The distal end of the insertion section contains a CMOS sensor for capturing image and transmitting it to the Video Processor, LEDs for illumination, and the distal opening of the working channel. The endoscope connector connects the endoscope handle to the video processor, which provides power and processes video signals from the endoscope.
# Mechanism of action:
The light emitted by the LED cold light source at the distal tip of the disposable Flexible Cystoscope is irradiated into the body cavity, and the light reflected from the cavity enters the optical system and is captured by the CMOS image sensor. The CMOS acquisition image is controlled by the CMOS drive circuit, and the RGB video signal is output to the Video Processor via the VI circuit. The Video Processor receives video signals from the endoscope, processes the video signals, and outputs the processed video signal to the attached monitor. The video processor also controls the brightness of the LEDs on the endoscope.
Flexible Cystoscope has the following physical and performance characteristics:
- O Maneuverable tip controlled by the user
- Flexible insertion cord
- Camera and LED light source at the distal tip
- O Sterilized by Ethylene Oxide
- For single use O
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image contains a logo for a company called "Innovex Anqing Medical". The logo features a stylized purple graphic on the left, with the text "Innovex" in a larger font size and "Anqing Medical" in a smaller font size below it. There is a curved line above the text "Innovex".
## Indications for Use:
AnQing Medical Single Use Flexible Cystoscope is used to provide visualization and operative access during diagnostic and therapeutic endoscopic procedures of the lower urinary tract, including the bladder and urethra.
The Flexible Cystoscope is intended for use in a hospital environment or medical office environment.
The Flexible Cystoscope is designed for use in adults.
# Comparison with Predicate Device:
The Flexible Cystoscope and its predicate device, the Ambu aScope 4 Cysto (K193095), have the same intended use, sterilization Methods and similar physical characteristics, optical characteristics.
# Substantial Equivalence:
The Flexible Cystoscope employs the same fundamental scientific technology as its predicate device, as below table:
| | Subject Device<br>(CY50H-20EU,<br>CY50H-20US,<br>CY55H-24EU,<br>CY55H-24US) | Predicate Device,<br>Ambu aScope 4 Cysto<br>(K193095) | Comparison |
|-------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------|
| Indications For Use | | | |
| Indications<br>For Use: | AnQing Medical Single<br>Use Flexible Cystoscope is<br>used to provide<br>visualization and operative<br>access during diagnostic<br>and therapeutic endoscopic<br>procedures of the lower<br>urinary tract, including the<br>bladder and urethra.<br><br>The Flexible Cystoscope is<br>intended for use in a<br>hospital environment or<br>medical office<br>environment. | Ambu® aScope™ 4<br>Cysto is a sterile,<br>single-use, flexible<br>cystoscope intended to<br>be used for endoscopic<br>access to and<br>examination of the<br>lower urinary tract.<br><br>The Ambu® aScope™<br>4 Cysto is intended to<br>provide visualization<br>via Ambu®<br>displaying unit and<br>can be used with<br>endoscopic | Equivalent<br>See Note1. |
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for Innovex Anqing Medical. The logo consists of a purple square with a white leaf design on the left, and the word "INNOVEX" in purple on the right. Below the word "INNOVEX" is the phrase "Anqing Medical" also in purple. There is a registered trademark symbol next to the word Innovex.
| 510(k) Premarket Notification Submission |
|------------------------------------------|
| Flexible Cystoscope |
| Subject Device<br>(CY50H-20EU,<br>CY50H-20US,<br>CY55H-24EU,<br>CY55H-24US)<br>The Flexible Cystoscope is<br>designed for use in adults. | Predicate Device,<br>Ambu aScope 4 Cysto<br>(K193095)<br>accessories.<br>Ambu® aScope™ 4<br>Cysto is intended for<br>use in a hospital<br>environment or<br>medical office<br>environment.<br>Ambu® aScope™ 4<br>Cysto is designed for<br>use in adults. | Comparison | |
|------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------|-----------------------|
| Physical Characteristics | | | |
| Type of<br>Scope | Flexible | Flexible | Same |
| Outer<br>diameter<br>(mm) | Max. 5.0 mm<br>(CY50H-20EU,<br>CY50H-20US)<br>Max. 5.5 mm<br>(CY55H-24EU,<br>CY55H-24US) | Max. 6.0 mm | Similar<br>See Note2. |
| Inner<br>diameter<br>(mm) | Min. 2.0 mm<br>(CY50H-20EU,<br>CY50H-20US)<br>Min. 2.4 mm<br>(CY55H-24EU,<br>CY55H-24US) | Min. 2.2 mm | Similar<br>See Note2. |
| Working<br>length | 380 mm | 390 mm (15.4") | Similar<br>See Note2. |
| Deflection | 210°up, 180°down | 210° up, 120° down | Similar<br>See Note2. |
| Optical Characteristics | | | |
| Type of<br>Imager | CMOS | CMOS | Same |
| Field of View | 110° | 120° | Similar<br>See Note2. |
| Direction of | 0° | 0° | Same |
| | Subject Device<br>(CY50H-20EU,<br>CY50H-20US,<br>CY55H-24EU,<br>CY55H-24US) | Predicate Device,<br>Ambu aScope 4 Cysto<br>(K193095) | Comparison |
| View | | | |
| Depth of Field | 5mm~100mm | 3mm~100mm | Similar<br>See Note2. |
| Light Source | Internal LED | Internal LED | Same |
| Patient Contacting Materials | | | |
| General material type of main patient-contact part | Compliance with<br>ISO10993-1 | Compliance with<br>ISO10993-1 | Similar<br>See Note2. |
| Duration and type of contact | "Surface –Mucosal Membrane" with a contact duration of “Limited (< 24 hours)” | "Surface -Mucosal Membrane" with a contact duration of “Limited (< 24 hours)” | Same |
| Sterilization Methods | | | |
| Number of Users | Single-Use | Single-Use | Same |
| Sterilization | EO Sterilized,<br>SAL 10-6 | EO Sterilized,<br>SAL 10-6 | Same |
| Technological Characteristics | | | |
| Environment of use | Healthcare facility/hospital | Healthcare facility/hospital | Same |
| Energy source | Electricity | Electricity | Same |
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for Innovex Anqing Medical. The logo is purple and white. The logo has a stylized image on the left and the words "INNOVEX" and "ANQING MEDICAL" on the right.
#### 510(k) Premarket Notification Submission Flexible Cystoscope
Note:
1. The indications for use statement for the subject Cystoscope is very similar to that of the predicate device. A slightly different is wording description. The differences do not alter the intended use of the device nor do they raise different questions of safety and effectiveness of the device relative to the predicate.
2. The subject and predicate device have the same fundamental technology,
type of scope, type of imager, light source, Sterilization method, number of users and environment.
The subject Cystoscope differs from the predicate in Outer diameter, Inner diameter,
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image contains the logo for Anqing Medical. The logo features a purple square with a white leaf design inside. To the right of the square, the word "INNOVEX" is written in a sans-serif font, with a registered trademark symbol next to it. Below the logo, the words "Anqing Medical" are written in a larger, bolder font.
510(k) Premarket Notification Submission Flexible Cystoscope
| Subject Device | Predicate Device, | |
|-----------------------------------------------------------|----------------------------------|------------|
| (CY50H-20EU,<br>CY50H-20US,<br>CY55H-24EU,<br>CY55H-24US) | Ambu aScope 4 Cysto<br>(K193095) | Comparison |
Working length, Deflection, Field of View, Depth of Field, and patient-contacting materials. These differences do not raise different questions of safety and effectiveness as compared to the predicate, and can be evaluated through performance testing.
# Summary of Testing:
# Summary of Non-Clinical Tests:
Electrical Safety and Electromagnetic Compatibility Summary The electrical safety and EMC data included in the submission is in compliance with the following FDA recognized standards:
- •ANSI/AAMI ES:60601-1:2005/A2:2010
- ·IEC 60601-2-18:2009: Part 2-18
- •IEC 60601-1-2:2014
# Bench Testing Summary
Photobiological safety
The LEDs in submitted Cystoscope were tested according to the following FDA recognized standards:
· IEC 62471:2006 Medical electrical equipment, Photobiological safety of lamps and lamp systems
#### Mechanical and Optical Performance
The Flexible Cystoscope was designed to comply with applicable parts of ISO 8600. Optical measurements were performed according to applicable part of ISO 8600 standard.
Mechanical characteristics were tested and include leakage tightness, bending, deflection endurance, withstand of channel.
In addition, comparative testing related to image quality parameters was performed for submitted Flexible Cystoscope and the predicate device to support substantial equivalence.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image contains the logo for "Innovax Anqing Medical". The logo is purple and features a stylized leaf design on the left. The text "ANQING MEDICAL" is written in a larger, bolder font below the logo.
#### Biocompatibility Summary
The biocompatibility evaluation for the patient contacting components of the Flexible Cystoscope was performed according to ISO 10993-1 and FDA Guidance. The following tests were conducted based contact category of "Surface -Mucosal Membrane" with a contact duration of "Limited (< 24 hours):
- · Cytotoxicity: ISO 10993-5:2009/(R) 2014
- · Sensitization, Intracutaneous reactivity/irritation: ISO 10993-10:2010
- · Material-mediated pyrogenicity: ISO 10993-11:2017
- · Acute systemic toxicity: ISO 10993-11:2017
#### Sterilization and shelf life testing
The sterilization method has been validated to ISO 11135:2014. which has thereby determined the routine control and monitoring parameters.
EO/ECH residual test was performed according to ISO 10993-7:2008.
The shelf life of the Flexible Cystoscope is determined based on stability study which includes ageing test according to ASTM F1980-16, Standard Guide for Accelerated Aging of Sterile Barrier.
#### Package Validation
Package validation was conducted according to ISO 11607-1:2019 and ISO 11607-2:2019, and F88/F88M-15, ASTM F 1929-15 Transport and shipping testing as per ASTM D4169-16.
#### Summary of Clinical Tests:
The subject of this premarket submission, did not require clinical studies to support substantial equivalence.
#### Conclusion:
The conclusions drawn from the nonclinical tests demonstrate that the subject device, the Flexible Cystoscope is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.