Ambu® aScope™ 4 Cysto is a sterile, single-use, flexible cystoscope intended to be used for endoscopic access to and examination of the lower urinary tract. The Ambu® aScope™ 4 Cysto is intended to provide visualization via Ambu® displaying unit and can be used with endoscopic accessories. Ambu® aScope™ 4 Cysto is intended for use in a hospital environment or medical office environment. Ambu® aScope™ 4 Cysto is designed for use in adults.
Device Story
Sterile, single-use, flexible cystoscope; provides visualization of urethra and bladder. Input: optical signals from distal tip camera/LED light source. Transformation: transmits video signal to external reusable Ambu displaying unit for real-time visualization. Operation: physician-controlled maneuverable tip (up/down) via handle; includes working channel for endoscopic accessories (graspers/instruments) and Luer lock for irrigation/aspiration. Used in hospital or medical office. Output: visual image on monitor. Benefit: enables diagnostic/interventional access to lower urinary tract while eliminating cross-contamination risk via single-use design.
Clinical Evidence
Bench testing only. No clinical data provided. Performance verified via ISO 8600-1/3/4, IEC 60601-1, IEC 60601-2-18, and ISO 10993-1 (biocompatibility).
Technological Characteristics
Flexible endoscope; distal tip camera/LED illumination; maneuverable tip; working channel; Luer lock connector. Sterilized via Ethylene Oxide (ISO 11135:2014). Complies with ISO 8600-1, 8600-3, 8600-4, IEC 60601-1, IEC 60601-2-18, and IEC 60601-1-2. Single-use.
Indications for Use
Indicated for endoscopic access to and examination of the lower urinary tract in adults. Used in hospital or medical office settings.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
K240227 — LiNA CystoVu HD (CYV-100-5; CYV-101-5); LiNA ScopeVu (SCP-100-1) · Lina Medical Aps · Aug 22, 2024
K231702 — WiScope® Digital Cystoscope System, WiScope® Single-Use Digital Flexible Cystoscope, WiScope® Medical Tablet · OTU Medical, Inc. · Sep 29, 2023
K222162 — Flexible Cystoscope (Model: CY50H-20EU, CY50H-20US, CY55H-24EU, CY55H-24US) · Shanghai AnQing Medical Instrument Co., Ltd. · Nov 9, 2022
K252176 — Single-Use Video Flexible Cysto-Nephroscope (RP-U-C01F, RP-U-C01FS) · Guangzhou Red Pine Medical Instrument Co., Ltd. · Aug 8, 2025
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 2, 2020
Ambu Inc. Sanjay Parikh Director, QA/RA 6230 Old Dobbin Lane, Suite 250 Columbia, MD 21045
Re: K193095
> Trade/Device Name: Ambu aScope 4 Cysto Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: FAJ Dated: March 2, 2020 Received: March 4, 2020
Dear Sanjay Parikh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Martha W. Betz, Ph.D. Acting Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K193095
Device Name Ambu® aScone™ 4 Cysto
#### Indications for Use (Describe)
Ambu® aScope™ 4 Cysto is a sterile, single-use, flexible cystoscope intended to be used for endoscopic access to and examination of the lower urinary tract. The Ambu® aScope™ 4 Cysto is intended to provide visualization via Ambu® displaying unit and can be used with endoscopic accessories.
Ambu® aScope™ 4 Cysto is intended for use in a hospital environment or medical office environment. Ambu® aScope™ 4 Cysto is designed for use in adults.
Type of Use (Select one or both, as applicable)
× Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW! *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Section 5 510(k) Summary
This 510(k) summary has been prepared in accordance with 21 CFR 807.87(h) and the content and format of the 510(k) summery has been prepared in accordance with 21 CFR 807.92.
| Submitter | Ambu A/S<br>Baltorpbakken 13<br>DK-2750 Ballerup<br>Denmark<br>Tel.: +45 7225 2000<br>Fax: +45 7225 2050 |
|---------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person | Name: Anita Fjaestad<br>Job Title: Senior Regulatory Affairs Professional<br>Address: Ambu A/S, Baltorpbakken 13, DK-2750 Ballerup<br>Telephone number: +45 7225 2544<br>Fax number: +45 7225 2050 |
| Date Summary<br>Prepared | April 1, 2020 |
| Device Trade<br>Name | Ambu® aScope™TM 4 Cysto |
| Device Common<br>Name | Flexible Cystoscope |
| Device<br>Classification | Cystoscope and Accessories, Flexible/Rigid<br>Product Codes: FAJ<br>21 CFR 876.1500<br>Class II |
| Legally<br>Marketed<br>devices to which<br>the device is<br>substantially<br>equivalent | Predicate Device:<br>Ambu® USR, Ambu® M, (K160766), Ambu A/S<br>Reference Device:<br>VISERA Cystovideoscope CYF-V2/VA2 (K133538), Olympus Medical<br>Systems Corporation |
| Description of<br>the Device | Ambu® aScope™ 4 Cysto is a sterile, single-use, flexible cystoscope<br>intended to be used for endoscopic access to and examination of the<br>lower urinary tract. Visualization will be achieved by Ambu® displaying<br>unit. The Ambu® aScope™ 4 Cysto is intended to be used with a<br>reusable Ambu® displaying unit to visualize the urethra and the bladder.<br>The Ambu® aScope™ 4 Cysto can be operated by either the left or right<br>hand. The optical module in the distal tip consists of a camera housing<br>containing camera and LED light sources |
| | Ambu® aScope™ 4 Cysto has the following physical and performance<br>characteristics: |
| | Maneuverable tip controlled by the user Flexible insertion cord Working channel that can be used for graspers or other<br>instrumentation Camera and LED light source at the distal tip Luer lock connector for irrigation and aspiration Sterilized by Ethylene Oxide sterilization For single use |
| Indications for<br>Use | Ambu® aScope™ 4 Cysto is a sterile, single-use, flexible cystoscope<br>intended to be used for endoscopic access to and examination of the<br>lower urinary tract. The Ambu® aScope™ 4 Cysto is intended to provide<br>visualization via Ambu® displaying unit and can be used with endoscopic<br>accessories. |
| | Ambu® aScope™ 4 Cysto is intended for use in a hospital environment or<br>medical office environment. |
| | Ambu® aScope™ 4 Cysto is designed for use in adults. |
| Summary of the<br>technological<br>characteristics<br>in comparison to<br>the predicate<br>device | Ambu® aScope™ 4 Cysto is similar to the predicate device in the<br>following areas:<br>It is a flexible endoscope with a manoeuvrable tip It has a handle with a control button giving the operator ability to<br>manoeuvre the tip of the endoscope up and down It is a video endoscope with a camera located in the distal tip to<br>provide an image on a separate monitor |
| | It's provide illumination from the distal tip It is connected to a monitor by a cable It allows for irrigation It is single use and delivered sterile |
| | Ambu® aScope™ 4 Cysto is different to the predicate devices in the<br>following area: |
| | Ambu® aScope™ 4 Cysto has a working channel that can be used<br>for graspers or other instrumentation. The predicate device has<br>a working channel with an integrated grasper for stent removal The bending angle is larger in aScope 4 cysto than in the<br>predicate, which is because the predicate has an integrated<br>grasper |
| | Insertion cord working length is 390 mm for Ambu® aScope™ 4 Cysto while predicate device insertion cord working length is 380 mm. Distal end outer diameter is 5.4 mm for Ambu® aScope™ 4 Cysto while predicate device distal end outer diameter is 4.8 mm. Depth of field for Ambu® aScope™ 4 Cysto is 3-100 mm while for predicate device it is 3-50 mm. |
| Performance<br>Data -Bench | The following data has been submitted in the premarket notification:<br><b>Declaration of conformity</b> to the following recognized consensus standards applicable for Ambu® aScope™ 4 Cysto: ISO 8600-1, ISO 8600-3 and ISO 8600-4 Optics and optical instruments - Medical endoscopes and certain accessories. IEC 60601-2-18 ED 3.0:2009 Medical electrical equipment - Part 2-18: Particular requirements for the basic safety and essential performance of endoscopic equipment Result: All tests were passed.<br><b>Performance test</b> reports to document the following properties of the Ambu® aScope™ 4 Cysto: Insertion cord working length Insertion cord outer diameter Maximum diameter of the insertion portion Distal end outer diameter Maximum Diameter of the insertion portion Minimum diameter of the working channel Irrigation possible Luer Lock Connector to working channel Angulations range (Distal bending section) Field of view Depth of field Direction of View Result: All tests were passed.<br><b>Sterilization</b><br>The sterilization method used is Ethylene Oxide (EO) with devices in a fixed chamber, in accordance with ISO 11135:2014.<br>Sterilization is conducted in a facility certified to EN ISO 13485:2016 with respect to sterilization in accordance with ISO 11135:2014.<br>Result: The Ambu® aScope™ 4 Cysto is sterile with a determination of lethal rate of the sterilization process to demonstrate achievement of the required SAL of 10-6 is in accordance to half cycle overkill approach as described in Annex B of ISO 11135:2014.<br><b>Shelf life</b><br>Performance test report to document shelf life. Tests were performed on finished, sterilized and aged products: Performance test Sterile Packaging Integrity |
| | Biocompatibility tests reports to document that Ambu® aScope™ 4 Cysto complies with the requirements of ISO 10993-1. |
| | Cytotoxicity (ISO 10993-5) Sensitization (ISO 10993-10) Intracutaneous reactivity test (ISO 10993-10) Result: All tests were passed. |
| | Test reports that verify the <b>Electromagnetic Compatibility and Electrical Safety</b> . |
| | Electromagnetic Compatibility in compliance with IEC 60601-1-2. Electrical Safety in compliance with IEC 60601-1 and IEC 60601-2-18. Result: All tests were passed. |
| Performance Data - Clinical | Not applicable. |
| Conclusion | Based on the indication for use, technological characteristics, performance data, and comparison to predicate device it has been concluded that the functionality and intended use of Ambu® aScope™ 4 Cysto is equivalent to the predicate and reference devices.<br><br>It is concluded that Ambu® aScope™ 4 Cysto is as safe and effective and performs as well as the chosen legally marketed predicate device. |
{4}------------------------------------------------
{5}------------------------------------------------
{6}------------------------------------------------
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.