Electric nasal aspirator (BC026, BC025, BC023, BC028, BC029, BC036)

K260147 · Shenzhen Kingboom Technology Co., Ltd. · BTA · Mar 13, 2026 · General, Plastic Surgery

Device Facts

Record IDK260147
Device NameElectric nasal aspirator (BC026, BC025, BC023, BC028, BC029, BC036)
ApplicantShenzhen Kingboom Technology Co., Ltd.
Product CodeBTA · General, Plastic Surgery
Decision DateMar 13, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4780
Device ClassClass 2
AttributesTherapeutic, Pediatric

Intended Use

The Electric nasal aspirator is intended for intermittent removal of nasal secretions and mucus from children (age 2-12 years old). This device is used in a home environment.

Device Story

Portable electric nasal aspirator; motor-driven pump creates negative pressure to remove nasal secretions/mucus. Operated by user (parent/caregiver) in home environment. Device consists of main unit with rechargeable Li-ion battery, buttons, LED display, and suction portion with silicone nozzles. User selects airflow levels via buttons; some models include light function for distraction or pause function. Output is physical removal of nasal mucus; healthcare provider/caregiver observes output to assess clearance. Benefits include improved nasal hygiene and breathing for children.

Clinical Evidence

No clinical test data was used to support the decision of substantial equivalence. Bench testing only.

Technological Characteristics

Powered suction pump; rechargeable 3.7V Li-ion battery. Materials: ABS, PC, Silicone. Dimensions vary by model. Connectivity: None. Sterilization: Not stated. Software: Complies with ANSI AAMI IEC 62304. Standards: ES60601-1 (Safety), IEC 60601-1-2 (EMC), IEC 60601-1-11 (Home healthcare), ISO 10993-1 (Biocompatibility), IEC 62133-2 (Battery).

Indications for Use

Indicated for intermittent removal of nasal secretions and mucus in children aged 2-12 years. Used in home environment.

Regulatory Classification

Identification

A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD &amp; DRUG ADMINISTRATION March 13, 2026 Shenzhen Kingboom Technology Co., Ltd. % Reanny Wang General Manager Shenzhen Reanny Medical Devices Management Consulting Co.Ltd Rm. 1509, Jingting Bldg., Dongzhou Community Guangming St., Guangming District Shenzhen, Guangdong 518107 China Re: K260147 Trade/Device Name: Electric nasal aspirator (BC026, BC025, BC023, BC028, BC029, BC036) Regulation Number: 21 CFR 878.4780 Regulation Name: Powered Suction Pump Regulatory Class: Class II Product Code: BTA Dated: January 20, 2026 Received: January 20, 2026 Dear Reanny Wang: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" U.S. Food &amp; Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K260147 - Reanny Wang Page 2 (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). {2} K260147 - Reanny Wang Page 3 Sincerely, **SHUCHEN PENG -S** Shu-Chen Peng, Ph.D. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} | Indications for Use | | | | --- | --- | --- | | Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K260147 | ? | | Please provide the device trade name(s). | | ? | | Electric nasal aspirator (BC026, BC025, BC023, BC028, BC029, BC036) | | | | Please provide your Indications for Use below. | | ? | | The Electric Nasal Aspirator is intended for intermittent removal of nasal secretions and mucus from children (age 2-12 years old). This device is used in a home environment. | | | | Please select the types of uses (select one or both, as applicable). | ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | ? | {4} Traditional 510(k) Summary 510(k) number: K260147 1. Information of Submitter and Correspondent Submitter's information: Company Name: Shenzhen Kingboom Technology Co., Ltd Street Address: #201, 2nd floor, No.5, Tongxin Road, Pingdong Community, Pingdi Street, Longgang District City: Shenzhen State/Province: Guangdong Country: China Telephone: +86-13590244726 Fax: / Contact Person: Ma Qiang Contact Title: General Manager Contact Email: 156673750@qq.com Date Prepared: Jan. 15, 2026 Submission correspondent's information: Shenzhen Reanny Medical Devices Management Consulting Co., Ltd Address: Room 1509, Jingting Building, Dongzhou Community, Guangming Street, Guangming District, Shenzhen 518107, China Contact Person: Reanny Wang E-mail: reanny@reanny.com Phone: +86(755) 27391220 2. Device Information Trade Name: Electric nasal aspirator {5} Model: BC026, BC025, BC023, BC028, BC029, BC036 Common Name: Powered suction pump Classification Name: Pump, portable, aspiration (powered) Regulation: 21 CFR § 878.4780 Device Class: Class 2 Product Code: BTA ## 3. Identification of Predicate Device(s) | Manufacturer | Shenzhen Kingboom Technology Co., Ltd | | --- | --- | | Legally Marketed Device | Electric nasal aspirator, model: BC026, BC023, BC025 | | 510(K) Number | K250697 | ## 4. Description of Device Electric nasal aspirator (model: BC026, BC025, BC023, BC028, BC029, BC036) consists of main unit, and suction portion working together as one unit. The Electric nasal aspirator is a portable device which is intended for suction of nasal passages in children 2-12 years of age. The motor pump provides a negative pressure which removes nasal secretions. The motor pump is powered by a rechargeable battery. The rechargeable battery can be charged via an external power adapter (not included in this device) using the provided charging line. The user interface consists of buttons and LED display, and the user can control the air flow level through the button. The differences between BC026, BC025, BC023, BC028, BC029 and BC036 are as follows: 1) The flow rate range of BC036 at different flow levels is different from other models. 2) BC025 and BC036 do not have "Light" function (used to distract and pacify the child), while other models do. 3) BC026, BC028, BC029 can pause suction during the suction process, while other models do not. 4) BC026, BC028, BC029 and BC023 have 4 buttons, BC025 and BC036 have 2 buttons. 5) In addition, BC026, BC028, BC029, BC025, BC036 and BC023 have different Display, Size, Weight, and Appearance. ## 5. Indications for Use The Electric nasal aspirator is intended for intermittent removal of nasal secretions and mucus from children (age 2-12 years old). This device is used in a home environment. {6} 6. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows: 6.1 Non-clinical testing A series of safety and performance tests were conducted on the subject device. - Product service life - Software validation - Electromagnetic compatibility and electrical safety - Function test All the test results demonstrate Electric nasal aspirator meets the requirements of its predefined acceptance criteria and intended use, and it is substantially equivalent to the predicate devices. 6.2 Clinical Testing No clinical test data was used to support the decision of substantial equivalence. 7. Performance Summary The devices conform to applicable standards as follow table: | Test Type | Standard Designation Number | FDA Recogniti on Status | Outcome for Device | | --- | --- | --- | --- | | Safety | ES60601-1:2005/(R)2012 & A1:2012, C1:2009/(R)2012 & A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021] | Yes | Conforms | | EMC | IEC 60601-1-2:2020 IEC TS 60601-4-2: 2024 | Yes | Conforms | | Home healthcare environment | IEC 60601-1-11:2020 | Yes | Conforms | | Performance | Enterprise standard | Yes | Conforms | | Biocompatibility | ISO 10993-1:2018 | Yes | Conforms | | Software | ANSI AAMI IEC 62304:2006/A1:2016 | Yes | Conforms | | Safety of Lithium battery | IEC 62133-2:2017 | Yes | Conforms | {7} Biocompatibility: The patient contact part (suction nozzles) of subject device uses the same materials, material suppliers, and production processes as the FDA-cleared Electric nasal aspirator (K250697). The subject device's patient contact part is identical to that of the predicate device in terms of formation, processing, and surface characteristics, and no additional chemicals (e.g., plasticizers, fillers, color additives, cleaning agents, mold release agents, etc.) have been added. # 8. Discussion of Comparison to Predicate Devices. The Electric nasal aspirator submitted in this 510(k) submission is substantially equivalent in intended use, technological characteristics, materials, and performance to the cleared Electric nasal aspirator K250697. Differences between the subject and predicate device do not affect the substantial equivalence comparison. | Device | Subject device | Primary Predicate device | Comparison | | --- | --- | --- | --- | | Manufacturer | Shenzhen Kingboom Technology Co., Ltd | Shenzhen Kingboom Technology Co., Ltd | / | | 510(K) number | TBD | K250697 | / | | Product name | Electric nasal aspirator, model: BC026, BC025, BC023, BC028, BC029, BC036 | Electric nasal aspirator, model: BC026, BC025, BC023 | / | | Classification | Class II Device, BTA (21 CFR § 878.4780) | Class II Device, BTA (21 CFR § 878.4780) | Identical | | Classification Panel | General & Plastic Surgery | General & Plastic Surgery | Identical | | Indication for Use | The Electric nasal aspirator is intended for intermittent removal of nasal secretions and mucus from children (age 2-12 years old). This device is used in a home environment. | The Electric nasal aspirator is intended for intermittent removal of nasal secretions and mucus from children (age 2-12 years old). This device is used in a home environment. | Identical | | Patient Population | Age 2-12 years old | Age 2-12 years old | Identical | | Intended Environment | Home use | Home use | Identical | | Device Description | Electric nasal aspirator is a portable device which is intended for suction of nasal passages in children 2-12 years of age. The motor pump provides a negative | The BC026, BC025 and BC023 Electric nasal aspirators are portable device which are intended for suction of nasal passages in children 2-12 years | Identical | {8} | Device | Subject device | Primary Predicate device | Comparison | | --- | --- | --- | --- | | | pressure which removes nasal secretions. | of age. The motor pump provides a negative pressure which removes nasal secretions. | | | Vacuum pressure | 50 to 62kPa | 52-60kPa | Similar Note 1 | | Air Flow rate | BC026, BC029, BC028, BC023: ● Level1: 1.5-2.7L/min; ● Level 2: 2.8-3.7L/min; ● Level 3: 3.8~5L/min. BC025: ● Gentle flow level: 1.5-2.7L/min; ● Normal flow level: 2.8-3.7L/min; ● Strong flow level: 3.8~5L/min. BC036: ● Gentle flow level: 1.5-2.5L/min; ● Normal flow level: 2.6-3.5L/min; ● Strong flow level: 3.6~5L/min | No | | | Music function | Yes | Yes | Identical | | Light function | BC026 and BC023, BC028, BC029: yes BC025 and BC036: No | BC026 and BC023: yes BC025: No | Identical | | Noise Level | <80dBA | <80dBA | Identical | | Power consumption | 2.55W | 2.2W | Similar note 2 | | Motor Type | DC 3.0V | 3.7V DC | Similar note 2 | | Power Source | DC 3.7 V / 700mAh Rechargeable Li-ion battery | DC 3.7 V / 700mAh Rechargeable Li-ion battery | Identical | | Device Dimension | BC026 and BC025: 160 (H) x 41 (L) x 41 (W) mm; BC023: 160 (H) x 41(L) x 39(W) mm BC028: 156 (H) x41.5(L) x 38.5(W) mm BC029: 158 (H) x40.5(L) x 39.5(W) mm BC036: 156 (H) x41.5(L) x 40.5(W) mm | BC026 and BC025: 160 (H) x 41 (L) x 41 (W) mm; BC023: 160 (H) x 41(L) x 39(W) mm | Similar Note 3 | | Weight | BC026: 320±5g BC025, BC036: 305±5g BC023: 317±5g BC028, BC029: 300±5g | BC026: 320±5g BC025: 305±5g BC023: 317±5g | | | Tips Dimension | Gourd Nozzle: OD5/ID3 Flat nozzle: OD3.8/ID2 | OD4.3/ID2.4 | Similar Note 4 | {9} | Device | Subject device | Primary Predicate device | Comparison | | --- | --- | --- | --- | | (ψ) | Angle nozzle: OD4/ID2.3 | | | | Main Materials | ABS, PC, Silicone | ABS, PC, Silicone | Identical | | Operating condition | 5°C(41°F) to 40°C(104°F); 15% to 93% R.H | 5°C(41°F) to 40°C(104°F); 15% to 93% R.H. | Identical | | Storage condition | -10°C(-23°F) to 70°C (158°F); 10% to 95% R.H. | -10°C(-23°F) to 70°C (158°F); 10% to 95% R.H. | Identical | | Expected service life | 2 years | 2 years | Identical | | Type BF applied part | Type BF applied part | Type BF applied part | Identical | | Safety | IEC 60601-1 | IEC 60601-1 | Identical | | | IEC 60601-1-11 | IEC 60601-1-11 | | | | IEC 62133-2 | IEC 62133-2 | | | | IEC 62471 | IEC 62471 | | | EMC | IEC 60601-1-2 | IEC 60601-1-2 | Identical | | Water - resistance | IP22 | IP22 | Identical | | Biocompatibility | BC026, BC025, BC023, BC028, BC029 and BC036 operate in conjunction with silicone nasal aspiration tips, which come into contact with nasal skin and mucosa for less than 24 hours. | BC026, BC025 and BC023 operates in conjunction with silicone nasal aspiration tips, which come into contact with nasal skin and mucosa for less than 24 hours. | Identical | | Contacted Parts | Silicone nozzle | Silicone Tip (funnel nozzle) | Identical Only the name is different | | Material of contacted parts | Silicone | Silicone | Identical | The discussion of differences between the subject and predicate device is listed as follows: Note 1 Although the pressures of the subject device and the predicate device are slightly different, and the subject device adds the air flow rate to its performance, it can be evaluated that both electrical nasal aspirator provide sufficient suction to clean the nasal cavity. Moreover, the performance test results have shown that the subject device meets the necessary requirements in terms of suction properties. Therefore, the differences do not affect the substantial equivalence comparison. {10} Note 2: Although the "Motor Type" and "Power consumption" between the predicate device and subject device are different, both the subject device and the predicate device comply with ANSI AAMI ES60601-1. So the differences do not affect the substantial equivalence comparison. Note 3 Although there are some differences in the "Device Dimension" and "Weight" between the predicate device and BC028, BC029, and BC036, they all comply with ANSI AAMI ES60601-1 and IEC 60601-1-2, so the differences do not affect the substantial equivalence comparison. Note 4: Although the Tips Dimension of the subject device is different from that of the predicate device, the Tip's OD Dimension (Gourd Nozzle) of the subject device is slightly larger than that of the predicate device, while the Tip's OD Dimension (Flat nozzle and Angle nozzle) of the subject device is slightly smaller than that of the predicate device, in general, the OD size is with the similar range of the predicate device; The Tip's ID Dimension (Gourd Nozzle) of the subject device is slightly larger than that of the predicate device, while the Tip's ID Dimension (Flat nozzle and Angle nozzle) of the subject device is slightly smaller than that of the predicate device, in general, the ID size is with the similar range of predicate device. Besides, both the predicate device and the subject device are comply with the ANSI AAMI ES60601-1 and IEC 60601-1-11, so the differences do not affect the substantial equivalence comparison. 9. Conclusions Based on performance testing, comparison and analysis, the subject device Electric nasal aspirator (model: BC026, BC025, BC023, BC028, BC029, BC036) is substantially equivalent to the predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...