K241852 · Avita Corporation · BTA · Dec 31, 2024 · General, Plastic Surgery
Device Facts
Record ID
K241852
Device Name
Nasal Aspirator (NS 13)
Applicant
Avita Corporation
Product Code
BTA · General, Plastic Surgery
Decision Date
Dec 31, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4780
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The device is designed for using suction to remove nasal secretion and mucus in Children (age 2~12 years old) at home environment. The additional spray/irrigation function helps moisten and loosen/soften nasal secretions in Children (over 3 years old), making them easier to aspirate.
Device Story
Battery-powered (2x AAA) nasal aspirator; generates negative pressure suction to remove nasal mucus; includes integrated sprayer module to moisten/soften secretions. Operated by user/caregiver in home environment. Device features silicone tips (gourd, long, short) for different mucus consistencies. User activates suction and spray via push buttons. Output is physical removal of nasal secretions. Benefits include relief of nasal congestion in pediatric patients.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including biocompatibility (ISO 10993-5, -10, -11, -23), cleaning/disinfection verification, and electrical safety (IEC 60601-1, IEC 60601-1-11).
Indicated for children aged 2-12 years for removal of nasal secretions and mucus via suction in home environments; spray/irrigation function indicated for children over 3 years to moisten and soften secretions.
Regulatory Classification
Identification
A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
December 31, 2024
AViTA Corporation Maggie Chao Adm. Manager 9F, No.78, Sec.1, Kwang Fu Road, Sanchong Dist. New Taipei City, 24158 Taiwan
Re: K241852
Trade/Device Name: Nasal Aspirator (NS13) Regulation Number: 21 CFR 878.4780 Regulation Name: Powered Suction Pump Regulatory Class: Class II Product Code: BTA, KMA Dated: December 2, 2024 Received: December 2, 2024
Dear Maggie Chao:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely,
Joyce C. Lin -S
for Shu-Chen Peng, Ph.D. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Submission Number (if known)
K241852
Device Name
Nasal Aspirator (NS13)
Indications for Use (Describe)
The device is designed for using suction to remove nasal secretion and mucus in Children (age 2~12 years old) at home environment.
The additional spray/irrigation function helps moisten and loosen/soften nasal secretions in Children (over 3 years old), making them easier to aspirate.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for AViTA. The logo consists of the word "AViTA" in blue, with a stylized heart shape to the right. The heart is colored in three sections: the top is yellow, the bottom is red, and the right side is green. There is a registered trademark symbol to the bottom right of the heart.
# AVITA CORPORATION
510(k) Submission for AViTA Nasal Aspirator
510(K) Summary
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for AViTA. The logo features the word "AViTA" in blue, with the "i" dotted with a yellow semi-circle. The word is set to the left of a heart shape, which is formed by a red curve at the bottom and a green curve at the top and right. A registered trademark symbol is located at the bottom right of the heart.
| Applicant: | AVITA Corporation |
|---------------------------------|--------------------------------------------------------------------------------------|
| Address: | 9F, No.78, Sec.1, Kwang-Fu Rd., San-Chung District, New Taipei<br>City 24158, Taiwan |
| Applicant Establishment Number: | 9617543 |
| Applicant Contact: | Maggie Chao |
| Email: | Maggie_chao@avita.com.tw |
| Phone Number: | +886-2-8512-1568 |
| Fax Number: | +886-2-8512-1347 |
| Date of Submission: | November 28, 2024 |
## DEVICE
| Trade/Proprietary Name: | AViTA Nasal Aspirator |
|------------------------------|--------------------------------|
| Common Name: | Nasal Aspirator |
| Model: | NS13 |
| Review Panel: | General & Plastic Surgery |
| Classification Product Code: | BTA KMA |
| Regulation Number: | 870.4780- Powered suction pump |
| Device Class: | II |
## DEVICE DESCRIPTION
This AViTA nasal aspirator (Model NS13) consists of a pump that can generate negative pressure suction, a silicone tip for contacting the nostrils, a collecting nasal mucus, a sprayer head module for moistening the nasal cavity, and two buttons that activate suction and sprayer functions. This nasal aspirator is powered by two AAA batteries. The accessories are the batteries, silicone tips.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for AViTA. The logo consists of the word "AViTA" in blue, with a heart shape to the right. The heart shape is made up of three different colors: yellow at the top, red at the bottom, and green on the right side. There is also a registered trademark symbol to the bottom right of the heart.
## INDICATIONS FOR USE
The device is designed for using suction to remove nasal secretion and mucus in children (age 2~12 years old) at home environment.
The additional spray function helps moisturize and soften nasal secretions in children (over 3 years old), making them easier to aspirate.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for AVITA. The logo has the word "AVITA" in blue font on the left side of the image. To the right of the word is a heart shape that is made up of three different colors: yellow, red, and green.
# COMPARISON WITH PREDICATE DEVICE
| Item | Subject<br>(AVITA Nasal Aspirator NS13) | Predicate<br>(AVITA Nasal Aspirator NS1 Series) | Predicate<br>(Geonic Nasal Cleaner) |
|--------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K number | - | K090379 | NA |
| Classification | 21CFR 878.4780 | 21CFR 878.4780 | 21 CFR 874.5550 |
| Product code | BTA | BTA | KMA |
| FDA class | II | II | I |
| Intended use | The device is designed for using suction to remove nasal<br>secretion and mucus in Children (age 2~12 years old) at home<br>environment. The additional spray/irrigation function helps<br>moisten and loosen/soften nasal secretions in Children (over 3<br>years old), making them easier to aspirate. | The device is designed for using intermittent suction to remove<br>nasal secretion and mucus in Children (age 2-12 years old) at<br>home environment. | Geonic Nasal Cleaner (Model Name:<br>H2) additionally provide<br>positive<br>pressure to moisturize the nasal cavities<br>by using saline solution in the form of<br>spray for adults and pediatrics (3 years<br>old). The nasal cleaner is intended for<br>over-the-counter (OTC) use at home<br>environment. |
| Effective<br>Vacuum Pressure | Vacuum Pressure: 100-120 mmHg | Max.120mmHg | N/A |
| Optional Levels | N/A | N/A | N/A |
| Sound<br>Pressure Level | ≤60dB | 65±3dB | N/A |
| Spray Capacity | 0.4~0.5 ml/10sec<br>(2.4~3 ml/min) | N/A | 5 ml/min |
| Operating<br>Temperature Range | 16°C ~ 35°C<br>(60.8°F ~ 95°F)<br>15%~85% RH | 15°C ~ 40°C<br><95% RH | 5°C ~ 40°C (41°F~104°F)<br>RH15%~90% 700-1060pa |
| Storage Temperature<br>Range | - 25°C ~ 55°C<br>(-13°F ~ 131°F) | - 20°C ~ 50°C<br><95% RH | -25°C ~ 70°C<br>(-13°F~158°F) |
{8}------------------------------------------------
| | | 15%~85% RH<br>2 x 1.5V AAA batteries<br>ABS | | | | 2 x 1.5V AA batteries<br>ABS | | | | | | RH up to 90% | | |
|--|----------------------|------------------------------------------------------------------------------------------------------------------------|------------------------------------|--------------------------------------------|---------|------------------------------------|-----------|------------------|--------------------------------------------|--------|-----------------------------------|--------------|-----|--|
| | A Power Requirements | | | | | | | | | | 3.7 V Li-ion rechargeable battery | | | |
| | Housing ® | | | | | | | | | | ABS | | | |
| | Storage pouch | No | | | | No Traditional 510(k) | | | | | No | | | |
| | Unit Weight | Approx.158g<br>(Exclude batteries) | | | | approx.250g<br>(Exclude batteries) | | | | | N/A | | | |
| | | Tip type | Picture | Suggestions for use: | Depth: | Size: OD x D x L (mm) | Tip type | Picture | Suggestions for use: | Depth: | Size: OD x D x L (mm) | | | |
| | | Gourd shape tip | Image: Gourd shape tip | With a stopper that does not go deep into. | Shallow | $Ф5хФ2.6x35.68$ | Long tip | Image: Long tip | Recommended for use when mucus is thinner. | Deep | $Ф4.02 x Ф2.6 x 28.0$ | | | |
| | | Long tip | Image: Long tip | Recommended for use when mucus is thinner. | Deep | $Ф5хФ3х35.68$ | Short tip | Image: Short tip | Recommended for use when mucus is thick. | Medium | $Ф6.23 x Ф2.0 x 22.2$ | | | |
| | Tips Information | Short tip | Image: Short tip | Recommended for use when mucus is thick. | Medium | $Ф5.68хФ3х27.68$ | | | | | | | N/A | |
| | | | OD: outside diameter / D: diameter | | | | | | | | | | | |
| | | OD: outside diameter / D: diameter | | | | | | | | | | | | |
| | | NS13 tip type based on the anticipated use. These three types of tips are suitable<br>for children aged 2 to 12 years. | | | | | | | | | | | | |
## Comparison:
The proposed device and the predicated device have the same principle. There is no additional question of safety and effectiveness as compared to the predicate device.
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the logo for AViTA. The logo has the word "AViTA" in blue, with a heart shape behind it. The top of the heart is yellow, the right side is green, and the bottom left is red. There is a registered trademark symbol in the bottom right corner of the heart.
There is a device currently on the US market that has the same clinical and biological characteristic as the subject device AVITA Nasal Aspirator NS 13.
In Clinical terms, Subject Device and Predicate device have the same intended uses for moistening and clearing nasal congestion and are applicable to the same patient groups, users, and usage environments. In Technical terms, the two have exactly the same working principle and performance specifications. In Biological terms, the parts that contact the human body are exactly the same in terms of materials, process, contact time and contact location.
The following technological differences exist between the subject and predicate devices:
- About clinical difference (differences in intended use) ।
Regarding the intended use there no difference.
- About technical difference (differences in Power Supply, Dimensions, Weight, Spray Capacity and Number of Push Button)
Regarding the difference in power requirement, the subject device has passed the relevant tests of IEC 60601-1.
The differences in dimensions, weight, number of push button and spray capacity do not affect the device's safety and effectiveness, as the subject device has passed the relevant tests in accordance with IEC 60601-1-11, and these aspects have been verified through the relevant performance tests.
In conclusion, the difference in power requirement, dimensions, weight, spray capacity and number of push button do not affect product performance.
Performance testing has been completed to demonstrate that the Subject device, AVITA Nasal Aspirator NS 13 is substantially equivalent to the Predicate device. AVITA Nasal Aspirator NS 1 Series and Geonic Nasal Cleaner.
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image shows the logo for AViTA. The logo consists of the word "AViTA" in blue, with a stylized heart shape to the right. The heart is colored with three different colors: yellow at the top, green in the middle, and red at the bottom. The registered trademark symbol is located at the bottom right of the heart.
## PERFORMANCE DATA
Performance testing has been carried out to demonstrate that this device meets the performance specifications for its intended use. The following tests were performed on the device. The results of those testing show that the required limits for mean difference and standard deviation are fulfilled by the subject device.
# Biocompatibility testing
The biocompatibility evaluation and testing of the Product name was conducted in accordance with the following standards and guidance, as recognized by the FDA:
- FDA Draft Guidance Use of International Standard ISO- 10993, "Biological Evaluation of - Medical Devices, Part 1: Evaluation and Testing".
- -ISO 10993-5, Biological evaluation of medical devices- Part 5: Tests for in vitro cytotoxicity
- 10993-10, Biological evaluation of medical devices- Part 10: Tests for irritation and skin sensitization
- ISO 10993-11. Biological evaluation of medical devices- Part 11: Tests for systemic toxicity. -
- -ISO 10993-23, Biological evaluation of medical devices- Part 23: Test for irritation.
# Cleaning and Disinfecting Test
The Nasal Aspirator is a medical device that does not supply sterilized. This product has passed the cleaning verification of 1000 times with mild mom-abrasive soap and 70℃ water and soft dry cloth wiping, confirming that the product can be used continuously without any functional abnormalities.
# Safety Test:
- -IEC 60601-1:2005+A1:2012, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- -IEC 60601-1-11:2010, Medical electrical equipment-Part 1-11: General Requirement for basic safety and essential performance- Collateral Standard: Requirements for medical electrical systems used in the home healthcare environment
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for AViTA. The logo consists of the word "AViTA" in blue, with a stylized heart shape surrounding it. The heart shape is made up of three different colors: yellow at the top, red at the bottom left, and green on the right side. There is also a registered trademark symbol to the right of the heart.
## Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "Basic documentation" level of concern, since a failure or latent flaw in the software could not directly result in serious injury or death to the patient or operator.
Clinical Studies None
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the logo for AViTA. The logo has the text "AViTA" in blue, with a stylized heart shape around it. The heart shape is made up of three different colored sections: yellow at the top, red at the bottom, and green on the right side. There is a registered trademark symbol to the right of the heart.
# CONCLUSION
The substantial equivalence is supported by the results of non-clinical performance testing, labeling and other provided data. The non-clinical data supports the safety of the device and the hardware and software verification and validation demonstrate that the AViTA Nasal Aspirator NS 13 should perform as intended in the specified use conditions and performs comparably to the predicate device that is currently marketed for the same intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.