K241202 · Hetaida Technology Co., Ltd. · BTA · Jul 1, 2024 · General, Plastic Surgery
Device Facts
Record ID
K241202
Device Name
Electric nasal aspirator (HTD2601US)
Applicant
Hetaida Technology Co., Ltd.
Product Code
BTA · General, Plastic Surgery
Decision Date
Jul 1, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4780
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Electronic Nasal Aspirator is intended for intermittent removal of nasal secretions and mucus from children (age 2-12 years old). This device is used in a home environment.
Device Story
Electric nasal aspirator (models HTD2601US, H2609AU) is a portable, battery-operated device for home use by caregivers to remove nasal secretions from children. The device consists of a main unit with a motor pump and a suction portion with silicone tips. The user operates the device via buttons and an LED display to control vacuum pressure. The motor pump generates negative pressure to aspirate mucus through the nasal tip. The device is powered by a rechargeable Li-ion battery. The output is the physical removal of nasal secretions, which assists in clearing nasal passages for the patient.
Clinical Evidence
No clinical data was used to support the decision of substantial equivalence. The device relies on bench testing, including biocompatibility, electrical safety, EMC, usability, and performance testing (negative pressure, flow rate, noise, and service life).
Indicated for intermittent removal of nasal secretions and mucus from children aged 2-12 years in a home environment.
Regulatory Classification
Identification
A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized design. To the right is the FDA logo, with the acronym "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in a larger, bolder font, and "ADMINISTRATION" in a smaller font below.
July 1, 2024
Hetaida Technology Co., Ltd. Tom Chen General Manager Rm 801, 802, 803, 804, 901, 2# Bldg Scientific Research Ctr Songhu Intelligent Valley, No.6 Minfu Road, Liaobu Town Dongguan. Guangdong 523423 China
Re: K241202
Trade/Device Name: Electric nasal aspirator (HTD2601US, H2609AU) Regulation Number: 21 CFR 878.4780 Regulation Name: Powered Suction Pump Regulatory Class: Class II Product Code: BTA Dated: April 19, 2024 Received: April 30, 2024
Dear Tom Chen:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shuchen Peng -S
Shu-Chen Peng, Ph.D. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia. Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
Submission Number (if known)
K241202
Device Name
Electric nasal aspirator (HTD2601US, H2609AU)
Indications for Use (Describe)
The Electronic Nasal Aspirator is intended for intermittent removal of nasal secretions and mucus from children (age 2-12 years old). This device is used in a home environment.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
[As required by 21 CFR 807.92]
| 1. Submission Information<br>510(k) Number:<br>Date:<br>Type of 510(k) Submission:<br>Basis for 510(k) Submission:<br>Submitter/Manufacturer:<br>Contactor: | K241202<br>April 19th, 2024<br>Traditional 510(k)<br>New device<br>Hetaida Technology Co., Ltd.<br>Room 801, 802, 803, 804, 901, 2# Building Scientific Research Center,<br>Songhu Intelligent Valley, No.6 Minfu Road, Liaobu Town, Dongguan<br>City, Guangdong Province, P.R.China<br>Tel: +86-0769-83326886<br>E-mail: tomchen@hetaida.com.cn<br>Tom Chen<br>Hetaida Technology Co., Ltd.<br>Room 801, 802, 803, 804, 901, 2# Building Scientific Research Center,<br>Songhu Intelligent Valley, No.6 Minfu Road, Liaobu Town, Dongguan<br>City, Guangdong Province, P.R.China<br>E-mail: tomchen@hetaida.com.cn<br>Tel: +86-0769-83326886 |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 2. Device Description | |
| Proprietary Name: | Electric nasal aspirator |
| Model Name:<br>Classification Name:<br>Product Code:<br>Device Class:<br>Regulation Number:<br>Review Panel:<br>Indications for use: | HTD2601US, H2609AU<br>Pump, Portable, Aspiration (Manual Or Powered)<br>BTA<br>2<br>21 CFR § 878.4780<br>General & Plastic Surgery<br>The Electronic Nasal Aspirator is intended for intermittent removal of<br>nasal secretions and mucus from children (age 2-12 years old). This device<br>is used in a home environment. |
| Device Description: | Electric nasal aspirator consists of main unit, and the suction portion<br>working together as one unit. The Electronic Nasal Aspirator is a portable<br>device which is intended for suction of nasal passages in children 2-12<br>years of age. The motor pump provides a negative pressure which removes<br>nasal secretions. The motor pump operates on a rechargeable battery. The<br>rechargeable battery can be charged from the external power adapter (not<br>included in this device) through the provided charging line. The user<br>interface consists of buttons and LED display, and the user can control the<br>vacuum pressure through the button. |
## 3. Predicate Device Identification
| 510(k) Number: | K222547 |
|----------------|-----------------------------------------|
| Product Name: | Electric nasal aspirator |
| Submitter: | Shenzhen XinLianFeng Technology CO.,LTD |
{4}------------------------------------------------
## 4. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows:
## 4.1 Non-clinical testing
A series of safety and performance tests were conducted on the subject device.
- Product service life
- Software validation
- Electromagnetic compatibility and electrical safety
- Usability
- Function test
- Biocompatibility test
All the test results demonstrate Electric nasal aspirator meets the requirements of its predefined acceptance criteria and intended use, and it is substantially equivalent to the predicate devices.
## 4.2 Clinical Testing
No clinical test data was used to support the decision of substantial equivalence.
#### 5. Substantially Equivalent Comparison
| Parameters | New Device | Predicate Device | Remark | |
|------------|---------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------|
| 1. | 510(k) Number: | K241202 | K222547 | -- |
| 2. | Marketing<br>clearance date: | No | October 20, 2022 | -- |
| 3. | Device Name | Electric nasal aspirator | Electric nasal aspirator | -- |
| 4. | Model | HTD2601US, H2609AU | BC026 (Predicate device<br>Labeling is BC023) | -- |
| 5. | 510(k) Owner | Hetaida Technology Co., Ltd. | Shenzhen XinLianFeng<br>Technology CO.,LTD | -- |
| 6. | Classification | Class II Device, BTA (21 CFR §<br>878.4780) | Class II Device, BTA (21 CFR §<br>878.4780) | Same |
| 7. | Classification<br>Panel | General & Plastic Surgery | General & Plastic Surgery | Same |
| 8. | Type of use | Over-The-Counter Use | Over-The-Counter Use | Same |
| 9. | Indications<br>for<br>Use | The Electronic Nasal Aspirator is<br>intended for intermittent removal<br>of nasal secretions and mucus<br>from children (age 2-12 years<br>old). This device is used in a<br>home environment. | The Electronic Nasal Aspirator is<br>intended for intermittent removal<br>of nasal secretions and mucus<br>from children (age 2-12 years<br>old). This device is used in a<br>home environment. | Same |
| 10. | Patient<br>Population | Age 2-12 years old | Age 2-12 years old | Same |
| 11. | Intended<br>Environment | Home use | Home use | Same |
| 12. | Device<br>Description | The HTD2601US and H2609AU<br>Electronic Nasal Aspirator is a<br>portable device which is intended<br>for suction of nasal passages in<br>children 2-12 years of age. The<br>motor pump provides a negative<br>pressure which removes nasal<br>secretions. | The BC026 Electronic Nasal<br>Aspirator is a portable device<br>which is intended for suction of<br>nasal passages in children 2-12<br>years of age. The motor pump<br>provides a negative pressure<br>which removes nasal secretions. | Same |
| | | | | |
| 13. | Vacuum pressure | <68kPa | 52-60Kpa | Different<br>Note 1 |
| 14. | Maximum flow<br>value | ≥1.6L/min | No | |
| 15. | Noise Level | <60dB | <80dBA | Similar |
| 16. | Power<br>consumption | 1.5W | 2.2W | |
| 17. | Motor Type | 3.7V DC | 3.7V DC | Different<br>Note 2 |
| 18. | Power Source | DC 3.7 V / 500mAh<br>Rechargeable Li-ion battery | DC 3.7 V / 700mAh<br>Rechargeable Li-ion battery | |
| 19. | Device<br>Dimension | HTD2601US: 185 mm x 42 mm<br>x 42 mm (Lx WxH)<br>H2609AU: 191 mm x 42 mm x<br>42 mm (Lx WxH) | 160 (H) x 41 (L) x 41 (W)mm | Different<br>Note 3 |
| 20. | Weight | Approx.167g | 320±5g | |
| 21. | Tips Dimension<br>(ψ) | Type1: OD4.1mm/ID2.3mm<br>Type2: OD7.9mm/ID3.0mm | OD4.3/ID2.4 | Different<br>Note 4 |
| 22. | Main Materials | ABS, PCTG, Silicone | ABS, PC, Silicone | Different<br>Note 5 |
| 23. | Operating<br>condition | 5°C~40°C(41°F~104°F);<br><85% R.H.; | 5°C(41°F) to 40°C(104°F);<br>15% to 93% R.H. | Similar |
| 24. | Storage condition | -20°C~55°C(-4°F~131°F);<br><93% R.H.; | -10°C(-23°F) to 70°C<br>(158°F) ; 10% to 95% R.H. | Different<br>Note 6 |
| 25. | Expected service<br>life | 5 years | 2 years | Different<br>Note 7 |
| 26. | Type BF applied<br>part | Type BF applied part | Type BF applied part | Same |
| 27. | Safety | ANSI AAMI ES60601-1<br>ANSI AAMI HA60601-1-11<br>IEC 60601-1-6<br>IEC 62133-2 | IEC 60601-1<br>IEC 60601-1-11<br>IEC 62133-2<br>IEC 62471 | Note 8 |
| 28. | EMC | ANSI AAMI IEC 60601-1-2 | IEC 60601-1-2 | |
| 29. | Water-resistance | IP22 | IP22 | Same |
| 30. | Biocompatibility | HTD2601US and H2609AU<br>operates in conjunction with<br>silicone nasal aspiration tips,<br>which come into the contact with<br>nasal skin and mucosa for less<br>than 24 hours. | BC026 operates in conjunction<br>with silicone nasal aspiration tips,<br>which come into the contact with<br>nasal skin and mucosa for less<br>than 24 hours. | Note 9 |
| 31. | Standard of<br>Biocompatibility | ISO 10993-1<br>ISO 10993-5<br>ISO 10993-10<br>ISO 10993-23 | ISO 10993-1<br>ISO 10993-5<br>ISO 10993-10<br>ISO 10993-12 | Note 9 |
| 32. | Contacted Parts | Silicone Tip (funnel nozzle) | Silicone Tip (funnel nozzle) | Same |
| 33. | Material of<br>contacted parts | Silicone | Silicone | Same |
{5}------------------------------------------------
The discussion of differences exist between the proposed and predicate devices is listed as follows:
Note 1: Although the pressures of the proposed device and the predicate device are slightly different, and the proposed device adds the flow rate to performance, both can evaluate that the nasal aspirator provides sufficient suction to clean the nasal cavity. Moreover, the performance test results have shown that the
{6}------------------------------------------------
proposed device meets the necessary requirements in terms of suction level properties and daily application. Therefore, the differences do not affect the safety and effectiveness.
Note 2: Although the Power consumption and Power Source between the proposed device and the predicate device are different, the battery of the proposed device is widely available on the market and has been tested to the IEC 62133-2. In addition, the proposed device has been tested to AAMI/ANSI ES60601-1 and IEC 60601-1-2. Therefore, the difference does not raise any new safety and effectiveness issues.
Note 3: Although the Device Dimension and the Weight between the proposed device and the predicate device are different, they are all complied with IEC 60601-1-2, so the differences do not affect the safety and effectiveness.
Note 4: Although the Tips Dimension of proposed device is different from that of predicate device, the Tip's OD Dimension (Type 2) of proposed device is larger than that of predicate device, but the Tip's OD Dimension (Type 1) of proposed device is slightly smaller than that of predicate device, in general, the OD size is with the similar range of the predicate device;
The Tip's ID Dimension (Type 2) of the proposed device is slightly larger than that of predicate device, but the Tip's ID Dimension (Type 1) of proposed device is slightly smaller than that of predicate device, in general, the ID size is with the similar range of predicate device. Besides, both the predicate device and proposed device are complied with the ANSI AAMI ES60601-1 and ANSI AAMI HA60601-1-11, so the differences do not affect the safety and effectiveness.
Note 5: Although the materials of the proposed device and the predicate device are slightly different, the PCTG is BPA-free and safer than PC. Moreover, the biocompatibility test results have shown that the proposed device and the predicate device meet the necessary requirements in terms of biocompatibility.
Note 6: The Storage environment of proposed device is different from that of predicate devices, but they all complied with the IEC 60601-1-11 and IEC 60601-1, so the difference will not raise any new safety and effectiveness issues.
Note 7: Although the Expected service life of the proposed device and the predicate device are slightly different, the service life of the proposed device is longer than that of predicate device. Moreover, the life test results have shown that the service life of proposed device meets the requirements in terms of 5 years. Therefore, the difference does not raise any new safety and effectiveness issues.
Note 8: Although the Safety and EMC standards of the proposed device and the predicate device are slightly different, the proposed device does not have the Music and Light function, so there is no need to meet the IEC 62471, and complied with the IEC 60601-1-6 for usability is also important. Therefore, the difference does not raise any new safety and effectiveness issues.
Note 9: Although the Biocompatibility standards of the proposed device and the predicate device are slightly different, the standard ISO 10993-10 has undergone a new revision that has resulted in the addition of standard ISO 10993-23, so should also comply with the new revision and the new standard for biocompatibility.
The proposed device has the same indication for use as the predicate device as well as comparable technical and biocompatibility properties and characteristics, and the minor differences don't raise any additional questions for safety and effectiveness. Therefore, the proposed device is substantially equivalent to the predicate device.
#### 6. Non-clinical Testing Summary
{7}------------------------------------------------
Bench tests were conducted to verify that the proposed device met all requirements. The test results demonstrated that the proposed device complies with the following standards:
- ANSI AAMI ES60601-1:2005/(R)2012 And A1:2012, C1:2009/(R)2012 And A2:2010/(R)2012+A2:2021 (Cons. Text) [Incl. AMD2:2021] Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD) [Including Amendment 2 (2021)];
- ANSI AAMI IEC 60601-1-2:2014 [Including AMD 1:2021] Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests [Including Amendment 1 (2021)];
- ANSI AAMI HA60601-1-11:2015 [Including AMD1:2021] Medical Electrical Equipment -- Part 1-11: General requirements for basic safety and essential performance -- Collateral Standard: Requirements for medical electrical equipment and medical electrical equipment and medical electrical systems used in the home healthcare environment (IEC 60601-1-11:2015 MOD) [Including Amendment1 (2021)];
- IEC 60601-1-6 Edition 3.2 2020-07 CONSOLIDATED VERSION Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability;
- IEC 62304 Edition 1.1 2015-06 CONSOLIDATED VERSION Medical device software - Software life cycle processes;
- IEC 62133-2 Edition1.0 2017-02 Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems;
- ISO 10993-5 Third Edition 2009-06-01, Biological Evaluation Of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity. (Biocompatibility)
- ISO 10993-10 Fourth edition 2021-11, Biological evaluation of medical devices - Part 10: Tests for skin sensitization.
- ISO 10993-23 First edition 2021-01, Biological evaluation of medical devices - Part 23: Tests for irritation.
- Performance tests: Negative pressure measuring, Maximum flow rate, Automatic power-off, and Noise test.
#### 7. Conclusion
The conclusion drawn from the non-clinical testing, comparison, and analysis demonstrates that the subject device in 510(k) submission, the subject device is substantially equivalent to the legally marketed predicate device K222547.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.