K212828 · Brius Technologies, Inc. · PNN · Oct 1, 2021 · Dental
Device Facts
Record ID
K212828
Device Name
BRIUS Planner Software
Applicant
Brius Technologies, Inc.
Product Code
PNN · Dental
Decision Date
Oct 1, 2021
Decision
SESE
Submission Type
Special
Regulation
21 CFR 872.5470
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
BRIUS Planner is intended for use as a medical front-end device providing tools for management of orthodontic models, systematic inspection, treatment simulation and virtual appliance design options (Patient-Specific Nitinol Wires, Sequential Aligners, Export of Models, Indirect Bonding Transfer Media) based on 3D models of the patient's dentition before the start of an orthodontic treatment. It can also be applied during the treatment to inspect and analyze the progress of the treatment. It can be used at the treatment to evaluate if the outcome is consistent with the planned/desired treatment objectives. The use of the BRIUS Planner requires the user to have the necessary training and domain knowledge in the practice of orthodontics, as well as to have received a dedicated training in the use of the software.
Device Story
BRIUS Planner Software is a browser-based orthodontic planning tool for clinicians. It ingests 3D dental scans (STL) and DICOM CT data to create digital models of patient dentition. Clinicians use the software to perform systematic inspections, analyze occlusion, simulate tooth movement, and design custom orthodontic appliances, including patient-specific Nitinol wires and sequential aligners. The software facilitates virtual appliance preparation and exports design files for manufacturing. Used in clinical settings by orthodontists or trained technicians to plan treatment, monitor progress, and verify outcomes against objectives. The output assists clinicians in visualizing treatment goals and generating patient-specific orthodontic solutions.
Clinical Evidence
Bench testing only. No clinical data provided. Substantial equivalence established through software integration, verification, and validation testing per FDA guidance.
Technological Characteristics
Standalone software module; browser-based (Windows/Mac). Imports STL and DICOM data. Supports 3D visualization, virtual tooth movement, and CAD design for orthodontic appliances. Minimum requirements: Windows 10, 8GB RAM, 1GB video memory. Login via username/password.
Indications for Use
Indicated for orthodontic patients requiring management of dental models, treatment simulation, and virtual design of orthodontic appliances (Nitinol wires, sequential aligners, indirect bonding media). Requires trained orthodontic professionals.
Regulatory Classification
Identification
An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.
K191720 — BRUIS Software Suite · Mechanodontics, Inc. · Apr 17, 2020
K253565 — SureSmile Software · Dentsply Sirona, Inc. · Mar 4, 2026
K201036 — PlaniMax Orthodontic Software · Choice Biotech, Inc. · Dec 22, 2020
K223518 — iOrtho · Shanghai EA Medical Instruments Co., Ltd. · Jun 13, 2023
K181112 — Orchestrate 3D · Orchestrate 3D · Nov 26, 2018
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The Department of Health & Human Services logo features a stylized human figure, while the FDA part includes the acronym "FDA" in a blue square, followed by "U.S. FOOD & DRUG" and "ADMINISTRATION" in blue text.
October 1, 2021
BRIUS Technologies Inc % Breanne Butler Regulatory Affairs Consultant Prime Path Medtech 1321 Upland Dr. Suite 6792 Houston, Texas 77043
Re: K212828
Trade/Device Name: BRIUS Planner Software Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: Class II Product Code: PNN Dated: September 2, 2021 Received: September 3, 2021
Dear Breanne Butler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
#### K212828
Device Name BRIUS Planner Software
#### Indications for Use (Describe)
BRIUS Planner is intended for use as a medical front-end device providing tools for management of orthodontic models, systematic inspection, treatment simulation and virtual appliance design options (Patient-Specific Nitinol Wires, Sequential Aligners, Export of Models, Indirect Bonding Transfer Media) based on 3D models of the patient's dentition before the start of an orthodontic treatment. It can also be applied during the treatment to inspect and analyze the progress of the treatment. It can be used at the treatment to evaluate if the outcome is consistent with the planned/desired treatment objectives.
The use of the BRIUS Planner requires the user to have the necessary training and domain knowledge in the practice of orthodontics, as well as to have received a dedicated training in the use of the software.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Section 5. 510(k) Summary
# 510(k) SUMMARY
A summary of 510(k) substantial equivalence information in accordance with the requirements of 21 CFR 807.92.
| Submitter: | BRIUS Technologies<br>2611 Westgrove Drive<br>Carrollton TX 75006 |
|--------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| Company Contact<br>Person:<br>Phone:<br>Email: | Mehdi Roein-Peikar<br>(312) 479-1449<br>mehdi@brius.com |
| Submission<br>Correspondent:<br>Address:<br>Phone:<br>Email: | Breanne Butler, Regulatory Affairs Consultant<br>1321 Upland Dr. Suite 6792 Houston, TX 77043<br>(860) 810-5594<br>bbutler@medavice.com |
| Date Prepared: | October 1, 2021 |
| Proprietary Name: | BRIUS Planner Software |
| Common Name: | Orthodontic Software |
| Product Code: | PNN - Orthodontic Software |
| Device Classification: | Class II, 21 CFR 872.5470 |
| Predicate Devices: | BRIUS Software Suite (K191720) (Primary Predicate) |
### Device Description:
The BRIUS Planner Software is an orthodontic appliance design and treatment simulation software. This software is used as a manufacturing software for orthodontic appliances. It is also used by professional technicians or physicians to design solutions for patients. Digital scans (3D) of a patient denture can be loaded into the software and a technician or physician can then create treatment plans for each individual patient and their needs. After approval by the patient's physician, the system can be used to fabricate dental appliances using standard stereolithographic (STL) for the design of custom shape-set nitinol appliances or clear aligners.
{4}------------------------------------------------
### Indications for Use:
BRIUS Planner is intended for use as a medical front-end device providing tools for management of orthodontic models, systematic inspection, treatment simulation and virtual appliance design options (Patient-Specific Nitinol Wires, Sequential Aliqners, Export of Models, Indirect Bonding Transfer Media) based on 3D models of the patient's dentition before the start of an orthodontic treatment. It can also be applied during the treatment to inspect and analyze the progress of the treatment. It can be used at the end of the treatment to evaluate if the outcome is consistent with the planned/desired treatment objectives.
The use of the BRIUS Planner requires the user to have the necessary training and domain knowledge in the practice of orthodontics, as well as to have received a dedicated training in the use of the software.
# Comparison to Predicate Devices:
BRIUS Planner Software is functionally equivalent to the following predicate device: BRIUS Software Suite (K191720 cleared April 2020).
The following table demonstrates the functional specifications of BRIUS Planner Software are substantially equivalent to the predicate devices.
| Specification | BRIUS Planner Software | BRIUS Software Suite (K191720) | Comparison Result |
|--------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|
| Indication<br>for Use | BRIUS Planner is intended for<br>use as a medical front-end<br>device providing tools for<br>management of orthodontic<br>models, systematic inspection,<br>treatment simulation and<br>virtual appliance design<br>options (Patient-Specific<br>Nitinol Wires, Sequential<br>Aligners, Export of Models,<br>Indirect Bonding Transfer<br>Media) based on 3D models of<br>the patient's dentition before<br>the start of an orthodontic<br>treatment. It can also be<br>applied during the treatment to<br>inspect and analyze the<br>progress of the treatment. It<br>can be used at the end of the<br>treatment to evaluate if the<br>outcome is consistent with the | BRIUS Software Suite is<br>intended for use as a medical<br>front-end device providing tools<br>for management of orthodontic<br>models, systematic inspection,<br>treatment simulation and virtual<br>appliance design options<br>(Patient-Specific Nitinol Wires,<br>Export of Models, Indirect<br>Bonding Transfer Media) based<br>on 3D models of the patient's<br>dentition before the start of an<br>orthodontic treatment. It can<br>also be applied during the<br>treatment to inspect and<br>analyze the progress of the<br>treatment. It can be used at the<br>end of the treatment to<br>evaluate if the outcome is<br>consistent with the | Similar |
| | | | |
| | planned/desired treatment<br>objectives. | planned/desired treatment<br>objectives. | |
| | The use of the BRIUS Planner<br>requires the user to have the<br>necessary training and domain<br>knowledge in the practice of<br>orthodontics, as well as to<br>have received a dedicated<br>training in the use of the<br>software. | The use of BRIUS Software<br>Suite requires the user to have<br>the necessary training and<br>domain knowledge in the<br>practice of orthodontics, as well<br>as to have received a<br>dedicated training in the use of<br>the software. | |
| Technology<br>Features | Stand Alone Software Module Imports Digital Patient Scans Can be used to design Dental Casts Useful for Diagnosis, treatment planning, and CAD design Virtual Planning of tooth movement Supports STL Files | Stand Alone Software Module Imports Digital Patient Scans Can be used to design Dental Casts Useful for Diagnosis, treatment planning, and CAD design Virtual Planning of tooth movement Supports STL Files | Same |
| Minimum<br>Hardware/Soft<br>ware<br>Requirements | OS: Windows 10 64-bit RAM: 8 GB Monitor Resolution: 1280 X 800 Video Card Memory: 1 GB Hard Drive Space: 10 GB CPU: Intel compatible 2.6 GHz/Dual or Quad core 2.6 GHz Mouse: Any Mouse with scrolling wheel or button | OS: Windows 10 64-bit RAM: 8 GB Monitor Resolution: 1280 X 800 Video Card Memory: 1 GB Hard Drive Space: 10 GB CPU: Intel compatible 2.6 GHz/Dual or Quad core 2.6 GHz Mouse: Any Mouse with scrolling wheel or button | Same |
| Login Method | Username and Password | Username and Password | Same |
#### Table 1: Functional Specification Comparison
{5}------------------------------------------------
# Table 2. Feature Comparison Table for BRIUS Planner and K191720
| Feature Comparison | BRIUS Planner | BRIUS Software<br>(K191720) |
|------------------------------------------------------|-------------------------------------|-------------------------------------|
| Supported anatomic<br>areas | Maxilla/Mandible | Maxilla/Mandible |
| Intended Use | | |
| Managing Patient and<br>case base data | Yes | Yes |
| | | |
| Collection of study material | Yes | Yes |
| Alignment of study material | Yes | Yes |
| Measuring study material | Yes | Yes |
| Analyzing Study Material | Yes | Yes |
| Treatment Simulation | Yes | Yes |
| Virtual Appliance Design | Yes | Yes |
| Supported PC formats | Windows/Mac: Internet Browser-based | Windows/Mac: Internet Browser-based |
| Managing patient and case base data | | |
| Creating, editing, deleting and copying patient data | Yes | Yes |
| Creating, editing, deleting and copying case data | Yes | Yes |
| Collection of study material | | |
| Surface scan for intraoral scanner | Yes | Yes |
| Surface scan from STL file | Yes | Yes |
| CT image data (DICOM) | Yes | Yes |
| Analyzing study material | | |
| Arch shape | Yes | Yes |
| Wire length | No | No |
| Tooth width | No | No |
| Bolton | No | No |
| Space Analysis | No | No |
| Overjet/Overbite | Yes | Yes |
| Occlusion Map | Yes | Yes |
| Treatment Simulation | | |
| 2D | Yes | Yes |
| 3D | Yes | Yes |
| Virtual Appliance Design | | |
| Orthodontic Appliance Search | Yes | Yes |
| Orthodontic Appliance Virtual Preparation | Yes | Yes |
| Orthodontic Appliance Design | Yes | Yes |
| Orthodontic appliance Export | Yes | Yes |
{6}------------------------------------------------
### Comparison of Indications for Use to Predicate Devices:
Based on the above comparison, the indications for use of BRIUS Planner Software is similar to that of the BRIUS Software Suite (K191720). The only change is the addition of Sequential Aligners to the appliance design options. This difference uses the same
{7}------------------------------------------------
technology to produce similar files for the production of orthodontic appliances. Therefore, BRIUS Planner Software can be considered substantially equivalent to its predicate device.
# Comparison of Technological Features to Predicate Devices:
Based on the above comparison, the design, construction, and performance characteristics of BRIUS Planner Software is the same as that of BRIUS Software Suite Therefore. BRIUS Planner Software can be considered (K191720). substantially equivalent to its predicate device.
### Summary of Performance Data and Substantial Equivalence:
Utilizing FDA Guidance document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" (issued May 11, 2015), BRIUS Planner Software underwent appropriate integration, verification, and validation testing.
### Conclusion:
Based on comparison of indications for use, technological features, performance testing, and software validation testing, BRIUS Planner Software have been shown to be appropriate for its indications for use and is substantially equivalent to the legally marketed predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.