The 3Shape Ortho System™ is intended for use as a medical front-end device providing tools for management of orthodontic models, systematic inspection, detailed analysis, treatment simulation and virtual applians (Custom Metal Bands, Export of Models, Indirect Bonding Transfer Media) based on 3D models of the patient's dentition before the start of an orthodontic treatment. It can also be applied during the treatment to inspect and analyze the progress of the treatment. It can be used at the treatment to evaluate if the outcome is consistent with the planned/desired treatment objectives.
Device Story
Software system for orthodontic case management; inputs include 3D dental scans (STL), DICOM images, and 2D images (JPG, BMP, PNG). Operates by visualizing 3D models, measuring dentition, simulating tooth movements, and designing virtual orthodontic appliances (e.g., indirect bonding trays, metal bands). Used in dental clinics by trained orthodontic professionals. Output consists of digital files for fabrication of appliances by third parties. Assists clinicians in diagnosis, treatment planning, and progress monitoring. Benefits include improved visualization of treatment objectives and customized appliance design.
Clinical Evidence
No clinical data. Bench testing only; software, hardware, and integration verification and validation performed per FDA guidance.
Technological Characteristics
Software-only device programmed in Delphi. Runs on Windows 7/8/10 (64-bit) with minimum 8GB RAM, Intel Core i5, and 250GB HDD. Supports DICOM, STL, JPG, BMP, PNG formats. No patient contact. Features include 2D/3D measurement, arch shape analysis, occlusion mapping, and virtual appliance design.
Indications for Use
Indicated for orthodontic patients requiring management of dental models, inspection, analysis, treatment simulation, and virtual appliance design (custom metal bands, indirect bonding transfer media) prior to, during, and at the end of orthodontic treatment.
Regulatory Classification
Identification
An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.
{0}------------------------------------------------
Janmuary 17, 2018
Image /page/0/Picture/1 description: The image shows the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. The logos are simple and professional in appearance.
3Shape A/S Antje Marquardsen Director Regulatory Affairs Holmens Kanal 7 Copenhagen, 1060 Dk
Re: K171634
Trade/Device Name: Ortho System Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: Class II Product Code: PNN Dated: November 21, 2017 Received: November 24, 2017
Dear Antje Marquardsen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
# Page 2 - Antje Marquardsen
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Mary S. Runner -S
ForTina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K171634
Device Name Ortho System™
#### Indications for Use (Describe)
The 3Shape Ortho System™ is intended for use as a medical front-end device providing tools for management of orthodontic models, systematic inspection, detailed analysis, treatment simulation and virtual applians (Custom Metal Bands, Export of Models, Indirect Bonding Transfer Media) based on 3D models of the patient's dentition before the start of an orthodontic treatment. It can also be applied during the treatment to inspect and analyze the progress of the treatment. It can be used at the treatment to evaluate if the outcome is consistent with the planned/desired treatment objectives.
The use of the Ortho System™ requires the user to have the necessary training and domain knowledge in the practice of orthodontics, as well as to have received a dedicated training in the use of the software.
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY – Special 510(K) – K171634
# Submitter Information
| Company Name: | 3Shape A/S |
|--------------------------------|---------------------------------------------------|
| Company Address: | Holmens Kanal 7<br>DK-1060 Copenhagen K |
| Company Phone:<br>Company Fax: | +45 7027 2620<br>+45 7027 2621 |
| Contact Person: | Antje Marquardsen<br>Director, Regulatory Affairs |
| Date Summary Prepared: | January 16, 2018 |
| Device Identification | |
| 510(k) number: | K171634 |
| Trade/proprietary Name: | Ortho System™ |
| Regulation Number: | 872.5470 |
| Regulation Name: | Orthodontic Plastic Bracket |
| Classification: | Class II |
# Primary Predicate Device
The Ortho System™ intended for dental retainers has the same intended use, scientific concept, performance specification and technical characteristics as the primary predicate device 3Shape Ortho System™ (K152086) from 3Shape A/S.
Based on the information and supporting documentation provided, the Ortho System™ and the primary predicate device (K152086) have same intended use.
Both software devices are used by Dental Professionals in orthodontic treatment planning (before, during, after treatment) covering management of patients and models, inspection, 2D and 3D measurement and orthodontic analysis of models, 2D & 3D treatment simulation, as well as virtual appliance preparation, handling and export, and they are both providing device output.
Additionally, the indirect bonding functionality of both systems is intended for use with commercially available brackets and wires.
The only modifications compared to the primary predicate device, besides look and feel of the user interface, are:
- 1. A workflow improvement to view present and ideal position of teeth at
{4}------------------------------------------------
the same time (dual view) with the introduction of arch wire in Indirect Bonding.
sshar
- 2. The ScanIt Manager is used for all scanners instead of ScanIt Orthodontics and ScanIt Ortho Impression that both are retracted.
Therefore, the Ortho System™ and the predicate (K152086) are found to be similar in their intended use, supported anatomic areas and the majority of the available features and functionalities.
# Reference Predicate devices
The Ortho System have two reference predicate devices K161884 and K163677, which uses the functions and tools of Ortho System™ to produce different orthodontic dental treatments than the primary predicate device. Both of these devices have the same fundamental scientific technology as the primary predicate device, as they are consolidated in the single medical software.
Based on the on the information and supporting documentation provided, the reference predicate devices do not alter the performance specification nor technical characteristics of the primary predicate device.
# Indications for Use
Ortho System™ is intended for use as a medical front-end device providing tools for management of orthodontic models, systematic inspection, detailed analysis, treatment simulation and virtual appliance design options (Custom Metal Bands, Export of Models, Indirect Bonding Transfer Media) based on 3D models of the patient's dentition before the start of an orthodontic treatment. It can also be applied during the treatment to inspect and analyze the progress of the treatment. It can be used at the end of the treatment to evaluate if the outcome is consistent with the planned/desired treatment objectives.
The use of the Ortho System™ requires the user to have the necessary training and domain knowledge in the practice of orthodontics, as well as to have received a dedicated training in the use of the software.
# Device Description
3Shape's Ortho System™ is a software system used for the management of 3D scanned orthodontic models of the patients, orthodontic diagnosis by measuring, analysing, inspecting and visualize 3D scanned orthodontic models, virtual planning of orthodontic treatments by simulating tooth movements, virtual placement of orthodontic brackets on the 3D models and design of orthodontic appliances based on 3D scanned orthodontic models, including transfer methods for indirect bonding of brackets. Output includes only Export Model (also called dental casts), Custom Metal Bands (also called metal bands), and Indirect Bonding Transfer Trays (also called orthodontic bracket placement trays). All devices are to be fabricated from FDA cleared materials by third party under the responsibility of the user.
The device has no patient contact.
{5}------------------------------------------------
## Scientific Concept
The underlying scientific concept of the Ortho System™ is to apply digital imaging tools for in orthodontic case archiving, diagnosis, treatment planning and CAD design of customized appliances.
Virtual positioning of brackets is possible with the use of encrypted libraries of the bracket geometry provided by the manufacturers and available through a dedicated download center in the software.
The system supports the following types of digital data: DICOM, STL, JPG, BMP, PNG.
### Summary of the technological characteristics
Ortho System™ is a software only device programmed in Delphi and has the following PC/laptop hardware requirements:
| Item | Minimum Requirements<br>Ortho System<br>(K171634) | Minimum Requirements<br>3Shape Ortho System<br>(K152086) |
|-------------------------|---------------------------------------------------|----------------------------------------------------------|
| OS: | Windows 7, 8 or 10 64-bit | Windows 7, 8 or 10, 64-bit |
| RAM: | 8 GB | 8 GB |
| Monitor<br>Resolution: | 1280x800 or similar | 1280x800 or similar |
| Video Card<br>Memory: | 1 GB | 1 GB |
| Available<br>HDD Space: | 250 GB | 250 GB |
| CPU: | Intel Core i5 or equivalent | Intel Core i5 or equivalent |
| Network: | Network Internet connection | Network Internet connection |
| Mouse: | With wheel button | With wheel button |
The Ortho System™ Software has the same intended uses and technical characteristics as the 3Shape Ortho System™ (K152086) from 3Shape A/S:
| Feature name | Ortho<br>System™<br>(K171634) | 3Shape<br>Ortho<br>System™<br>(K152086) |
|------------------------------------------------------|-------------------------------|-----------------------------------------|
| Supported anatomic areas | Maxilla | Maxilla |
| | Mandible | Mandible |
| Intended use | | |
| Managing patient and case base data | Yes | Yes |
| Collection of study material | Yes | Yes |
| Alignment of study material | Yes | Yes |
| Measuring study material | Yes | Yes |
| Analyzing study material | Yes | Yes |
| Treatment simulation | Yes | Yes |
| Virtual appliance design | Yes | Yes |
| Supported PC formats | Windows | Windows |
| Managing patient and case base data | | |
| Creating, editing, deleting and copying patient data | Yes | Yes |
| Creating, editing, deleting and copying case data | Yes | Yes |
| Feature name | Ortho<br>System™<br>(K171634) | 3Shape<br>Ortho<br>System™<br>(K152086) |
| Collection of study material | | |
| Surface scan for intra-oral scanner | Yes | Yes |
| Surface scan from STL file | Yes | Yes |
| CT image data | DICOM | DICOM |
| 2D overlay | PNG, JPG,<br>BMP | PNG, JPG, BMP |
| Alignment of study material | | |
| Aligning surface scan and CT image | Yes | Yes |
| Aligning cephalometric images | Yes | Yes |
| Alignment of 2D overlays (e.g. ideal arch) | Yes | Yes |
| Ability to check/adjust DICOM visibility | Yes | Yes |
| DICOM scan segmentation | No | No |
| Measuring study material | | |
| 2D measurement toolbox | Yes | Yes |
| 3D measurement toolbox | Yes | Yes |
| Analyzing study material | | |
| Arch shape | Yes | Yes |
| Wire length | Yes | Yes |
| Tooth width | Yes | Yes |
| Bolton | Yes | Yes |
| Space analysis | Yes | Yes |
| Overjet/overbite | Yes | Yes |
| Occlusion map | Yes | Yes |
| Treatment simulation | | |
| 2D & 3D simulation | Yes | Yes |
{6}------------------------------------------------
# ORTHO SYSTEM™ SOFTWARE - SPECIAL 510(K) SUBMISSION
зshap
{7}------------------------------------------------
## ORTHO SYSTEM™ SOFTWARE - SPECIAL 510(K) SUBMISSION
| Feature name | Ortho<br>System™<br>(K171634) | 3Shape<br>Ortho<br>System™<br>(K152086) |
|-------------------------------------------|-------------------------------|-----------------------------------------|
| Virtual appliance design | | |
| Orthodontic appliance search | Yes | Yes |
| Orthodontic appliance virtual preparation | Yes | Yes |
| Orthodontic appliance design | Yes | Yes |
| Orthodontic appliance export | Yes | Yes |
# Nonclinical Testing
Software, hardware, and integration verification and validation testing was performed in accordance with the FDA Guidance Document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" (Issued on May 11, 2005).
The validation suite includes validation of implemented mitigations related to device hazards identified in the risk management procedures.
All test results have been reviewed and approved, showing the Ortho System™ to be substantially equivalent to the primary predicate device.
# Clinical Testing
Clinical testing is not a requirement and has not been performed.
# Conclusion
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for 3Shape. The text "3shape" is written in a dark blue sans-serif font. To the right of the text is a red triangle made up of three smaller triangles.
Based on a comparison of intended use, indications, principle of operations, features and technical data, and the test results, the Ortho System™ is found to be as safe and as effective as the primary predicate device.
Intended use and performance is found to be substantially equivalent to the primary predicate Device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.