K223624 · Island Dental Lab, Inc Dba Emerald Dental · LRK · May 10, 2023 · Dental
Device Facts
Record ID
K223624
Device Name
Emerald Herbst
Applicant
Island Dental Lab, Inc Dba Emerald Dental
Product Code
LRK · Dental
Decision Date
May 10, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.5570
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Emerald Herbst appliance is intended to treat snoring and mild to moderate Obstructive Sleep Apnea (OSA) in adults 18 years of age or older.
Device Story
Customized intraoral mandibular advancement device; upper and lower acrylic trays connected by adjustable Herbst telescopic hardware; patient-worn during sleep; guides mandible forward to prevent tongue/soft tissue airway collapse; increases pharyngeal space; improves air exchange; non-sterile; provided with adjustment screw; patient-operated; used at home or in sleep labs; provides freedom of lateral/forward movement for comfort; hard or hard/soft laminate options; no electronic components or tracking.
Clinical Evidence
No clinical data. Bench testing only. Biocompatibility and physical properties assessment based on predicate material composition. Accelerated aging testing performed to measure deflection strength; results showed no decrease in strength post-aging.
Technological Characteristics
Customized upper/lower interlocking trays; materials include standard orthodontic acrylics, stainless steel Herbst telescopic hinges, and cobalt chrome mesh; hard or hard/soft laminate options; retention via standard orthodontic ball clasps; non-sterile; mechanical operation; no software or connectivity.
Indications for Use
Indicated for snoring and mild to moderate Obstructive Sleep Apnea (OSA) in adults 18 years of age or older.
Regulatory Classification
Identification
Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea are devices that are worn during sleep to reduce the incidence of snoring and to treat obstructive sleep apnea. The devices are designed to increase the patency of the airway and to decrease air turbulence and airway obstruction. The classification includes palatal lifting devices, tongue retaining devices, and mandibular repositioning devices.
Special Controls
*Classification.* Class II (special controls). The special control for these devices is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA.”
K131138 — RESPIRE PINK SERIES - HERBST · Respire Medical · Sep 9, 2013
K181416 — Sapphire Dorsal Comfort Fit · Crane Dental Laboratory, Inc. · Sep 17, 2019
K210694 — Silent Nite Sleep Appliance with the Glidewell Hinge · Prismatik Dentalcraft, Inc. · Jun 16, 2021
K192581 — The NightBlocks Appliance · Advanced Facialdontics, LLC · Jan 28, 2020
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue. Underneath the FDA logo is the word "ADMINISTRATION".
May 10, 2023
Island Dental Lab, Inc dba Emerald Dental % Colette Cozean, Phd Regulatory Consultant The EyeDeas Company 21581 Midcrest Dr. Lake Forest, California 92630
Re: K223624
Trade/Device Name: Emerald Herbst Regulation Number: 21 CFR 872.5570 Regulation Name: Intraoral Devices For Snoring And Intraoral Devices For Snoring And Obstructive Sleep Apnea Regulatory Class: Class II Product Code: LRK, LQZ Dated: November 30, 2022 Received: December 5, 2022
Dear Colette Cozean, Phd:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Michael E. Adjodha -2
Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K223624
Device Name Emerald Herbst
Indications for Use (Describe)
The Emerald Herbst appliance is intended to treat snoring and mild to moderate Obstructive Sleep Apnea (OSA) in adults 18 years of age or older.
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 5. 510(k) Summary
| Applicant: | Island Dental Lab, Inc dba Emerald Dental Lab<br>76 S. Central Ave., # 1D<br>Valley Stream, NY, 11580<br>516-327-0810<br>347-522-9800 |
|--------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Colette Cozean, PhD<br>21581 Midcrest Drive<br>Lake Forest, CA 92630<br>(949) 855-2885<br>colettecozean@gmail.com |
| Date Prepared | August 15, 2022 |
| Proprietary Name | Emerald Herbst Appliance (Hard and Hard/Soft) |
| Common Name | Device, Snoring and Mild to Moderate Obstructive Sleep Apnea |
| Classification Name | Intraoral devices for snoring; intraoral devices for snoring and obstructive sleep apnea<br>(Class II, 21 CFR872.5570, Product Code LRK and LQZ) |
| Primary Predicate Device | Respire Pink Series with DentiTrac(K170692) refers to Herbst ( (K131138),<br>Respire Pink Series – Herbst EF (K150572) |
| Reference Device | Dent-O-Cryl Orthodontic Acrylic (K942667) is used to justify the<br>use of a different color. Emerald Dental certifies that the Herbst<br>Emerald Device uses the following identical materials found in<br>the Respire Pink Series: |
Description: The Emerald Herbst appliance is a customized oral device featuring both lower and upper trays and interlocking system. The product is non-sterile, biocompatible, and provided in a sealed box with instructions for use and an adjustment screw. The upper and lower trays are connected by an adjustable hinge allowing the patient to open and close the mouth while wearing the appliance. The Herbst Telescopic Hardware on the side of the device allows the patient to move forward and left and right but not backwards. This freedom of movement is important for comfort and overall success of the device.
The Emerald Herbst appliance is intended to treat snoring and mild to moderate sleep apnea (OSA) in adults 18 years of age or older by guiding the mandible forward during sleep, preventing the tongue and soft tissues of the throat from collapsing into the airway.
The Emerald Herbst is offered in two options:
- Hard devices which are all acrylic and retained with ball clasps that allow the device to be tightened if 1. it becomes loose.
- 2. Hard / Soft which has a laminate layer that provides a soft layer on the tooth surface.
These two options are also available in the predicate, Respire Pink Series.
Indications for Use: The Emerald Herbst appliance is intended to treat snoring and mild to moderate obstructive sleep apnea (OSA) in adult patients 18 years of age or older. This indication for use is identical to that of the predicate device(s).
Technological Characteristics: The Emerald Herbst appliance consists of upper and lower interlocking, customized trays. The Emerald Herbst appliance is customized on models of the patient's teeth, using standard orthodontic acrylics and standard orthodontic ball clasps for retention. The Emerald Herbst appliance allows for inter1ocking of the upper and lower trays to adjust the mandibular position of the user. The technical characteristics are identical to the predicate devices.
{4}------------------------------------------------
Mechanism of Action: All predicate devices (Respire Pink Series) also function as mandibular advancement devices to increase the patient's pharyngeal space and improve the ability to exchange air during sleep. They each have customized upper and lower trays that interlock to advance the mandible. The mechanism of action of the subject device is substantially equivalent to the predicate devices.
# Substantial Equivalence Table:
| | | Subject<br>Device | Primary<br>Predicate<br>Device | Included in<br>Primary<br>Predicate<br>Device | Included in<br>Primary<br>Predicate<br>Device | |
|--------------|----------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------|-------------------------------------------------|--|
| | | Emerald<br>Dental | Respire Pink<br>Series | Respire Pink<br>Series -<br>Herbst | Respire Pink<br>Series -<br>Herbst EF | |
| | | | K170692 | K131138 | K150572 | |
| Intended Use | | | | | | |
| | Indication for<br>Use | Intended to<br>treat snoring<br>and mild to<br>moderate<br>obstructive<br>sleep apnea<br>(OSA) in adult<br>patients 18<br>years of age<br>or older | Intended to<br>treat snoring<br>and mild to<br>moderate<br>Obstructive<br>Sleep Apnea<br>(OSA) in adult<br>patients 18<br>years of age<br>or older | Intended to<br>treat mild to<br>moderate<br>OSA | Intended to<br>treat mild to<br>moderate<br>OSA | |
| | Class | II | II | II | II | |
| | Intended as<br>Intraoral<br>Device | Yes | Yes | Yes | Yes | |
| | Intended to<br>reduce snoring | Yes | Yes | Yes | Yes | |
| | Intended to<br>treat mild to<br>moderate OSA | Yes | Yes | Yes | Yes | |
| | Intended for<br>nighttime use | Yes | Yes | Yes | Yes | |
| | Intended for<br>single patient<br>multiuse | Yes | Yes | Yes | Yes | |
| | Intended for<br>use at home<br>or sleep labs | Yes | Yes | Yes | Yes | |
| | Intended for<br>adults, 18<br>years or older | Yes | Yes | Yes | Yes | |
| Design | | | | | | |
| | Customized fit<br>for each<br>patient | Yes | Yes | Yes | Yes | |
| | Mechanism of<br>action:<br>Mandibular | Yes | Yes | Yes | Yes | |
| | Can be<br>adjusted or<br>refit | Yes | Yes | Yes | Yes | |
| | Range of<br>forward<br>movement of<br>mandible | Yes | Yes | Yes | Yes | |
| | Upper & lower<br>trays<br>disengage for<br>easy removal | Yes | Yes | Yes | Yes | |
| | Permits<br>patient to<br>breathe<br>through the<br>mouth | Yes | Yes | Yes | Yes | |
| | Has tracking<br>device | No | Yes | No | No | |
| | Cleaned and<br>inspected daily<br>by patient | Yes | Yes | Yes | Yes | |
| Materials | | | | | | |
| | Hard surface<br>material | Yes | Yes | Yes | Yes | |
| | Soft lining<br>material | No Hard, Yes<br>Hard/Soft | No Herbst,<br>Yes Herbst EF | No | Yes | |
| | Advancement<br>mechanism | Yes | Yes | Yes | Yes | |
| | Advancement<br>mechanism -<br>Herbst<br>telescoping<br>hinge,<br>stainless steel | Yes | Yes | Yes | Yes | |
| | Cobalt Chrome<br>Mesh inside<br>Acrylic | Yes | Yes | Yes | Yes | |
| Mechanism | | | | | | |
| | Mechanism of<br>Action | Mandibular<br>Advancement | Mandibular<br>Advancement | Mandibular<br>Advancement | Mandibular<br>Advancement | |
| | Method of<br>Positioning the<br>Mandible | Herbst<br>Telescoping<br>Hinge | Herbst<br>Telescoping<br>Hinge | Herbst<br>Telescoping<br>Hinge | Herbst<br>Telescoping<br>Hinge | |
| | Maximum<br>Amount of<br>Mandible<br>Protrusion | 6 mm | 6 mm | 6 mm | 6 mm | |
| Labeling | | | | | | |
| | Prescription<br>Only | Yes | Yes | Yes | Yes | |
{5}------------------------------------------------
{6}------------------------------------------------
Clinical and Non-Clinical Data: A biocompatibility and physical properties assessment was completed based on the material composition of the primary predicate, which concluded that the subject device was substantially equivalent to the primary predicate device. A risk assessment has also been conducted with the subject device, which concluded there are no additional risks as compared to the predicate device(s).
Performance Testing: The Emerald Herbst is equivalent to the predicate device, Pink Series, as it uses all the same materials, same manufacturing procedure and same manufacturer. Therefore, no mechanical testing was conducted to establish equivalence. The Emerald Herbst was subjected to pre and post accelerated aging test measured by deflection strength. Aging did not decrease the strength of the device.
Differences: The Emerald Herbst is green unlike the Pink Series, which is pink. Both colors have received premarket approval in K942667 for the reference device, Dent-O-Cryl. This reference device justifies the use of all colors submitted by Dentauraum for Dent-O-Cryl, including the Emerald Herbst green and the Pink Series pink color additives. This 510(k) contains all the information considered by the FDA to determine substantial equivalence, including physical parameters and biocompatibility.
Summary: Based on the intended use, technical characteristics, biocompatibility assessment, labeling and other data provided in this submission, the Emerald Herbst appliance demonstrates substantial equivalence to the predicate devices in both safety and efficacy.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.