Silent Nite Sleep Appliance with the Glidewell Hinge
K210694 · Prismatik Dentalcraft, Inc. · LRK · Jun 16, 2021 · Dental
Device Facts
Record ID
K210694
Device Name
Silent Nite Sleep Appliance with the Glidewell Hinge
Applicant
Prismatik Dentalcraft, Inc.
Product Code
LRK · Dental
Decision Date
Jun 16, 2021
Decision
SESE
Submission Type
Special
Regulation
21 CFR 872.5570
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Silent Nite Sleep Appliance with the Glidewell Hinge is indicated to reduce snoring and mild to moderate obstructive sleep apnea (OSA) in patients 18 years of age or older. It is worn while sleeping to support the lower jaw in a forward position prescribed by the dentist. The appliance is removable by the patient.
Device Story
Mandibular advancement device; holds mandible in protrusive position to increase airway space and decrease air turbulence during sleep. Consists of custom-fabricated upper and lower trays (soft polyurethane inner layer, hard polyester outer layer) connected by Glidewell Hinge (stainless steel). Clinician prescribes mandibular position; patient adjusts hinge via hex driver in 0.25 mm increments up to 10 mm. Optional orthodontic elastic bands attach to hooks/lugs to assist in keeping mouth closed. Used at home or in sleep lab; patient-removable. Device custom-fitted to patient dentition via thermoforming on dental models. Forward jaw position prevents tongue from falling back into airway, reducing OSA symptoms.
Clinical Evidence
Bench testing only. Verification of mandibular advancement function, hinge displacement, and hinge positioning. Validation of device fit, retention, and finish. Biocompatibility testing per ISO 10993 (cytotoxicity, sensitization, oral mucosal irritation) performed on stainless steel hinge component.
Technological Characteristics
Mandibular advancement device. Materials: dual-layered thermoplastic (soft polyurethane inner, hard polyester outer), medical-grade stainless steel hinge. Dimensions: adjustable up to 10 mm in 0.25 mm increments. Non-sterile. Custom-fabricated per patient dentition. No software or electronic components.
Indications for Use
Indicated for patients 18 years of age or older with snoring and mild to moderate obstructive sleep apnea (OSA).
Regulatory Classification
Identification
Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea are devices that are worn during sleep to reduce the incidence of snoring and to treat obstructive sleep apnea. The devices are designed to increase the patency of the airway and to decrease air turbulence and airway obstruction. The classification includes palatal lifting devices, tongue retaining devices, and mandibular repositioning devices.
Special Controls
*Classification.* Class II (special controls). The special control for these devices is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA.”
Predicate Devices
Silent Nite sl, Prismatik Dentalcraft, Inc. (K183270)
Reference Devices
Thermoforming Sheet Materials, Erkodent Erich Kopp GmbH (K200125)
Acrylic Herbst Splint Appliance, Specialty Appliances Works (K083209)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of the square.
June 16, 2021
Prismatik Dentalcraft, Inc. % Patsy Trisler Regulatory Consultant Qserve Group US, Inc. 7949 Beaumont Green East Drive Indianapolis, Indiana 46250
Re: K210694
Trade/Device Name: Silent Nite Sleep Appliance with the Glidewell Hinge Regulation Number: 21 CFR 872.5570 Regulation Name: Intraoral Devices For Snoring And Intraoral Devices For Snoring And Obstructive Sleep Apnea Regulatory Class: Class II Product Code: LRK Dated: May 13, 2021 Received: May 17, 2021
Dear Patsy Trisler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K210694
Device Name
Silent Nite Sleep Appliance with the Glidewell Hinge
#### Indications for Use (Describe)
Silent Nite Sleep Appliance with the Glidewell Hinge is indicated to reduce snoring and mild to moderate obstructive sleep apnea (OSA) in patients 18 years of age or older. It is worn while sleeping to support the lower jaw in a forward position prescribed by the dentist. The appliance is removable by the patient.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------|----------------------------------------------------------|
| <div><span style="font-family: Arial;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div>□ Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary – K210694
| Submitter Name: | Prismatik Dentalcraft, Inc. |
|---------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter Address: | 2144 Michelson Drive<br>Irvine, CA 96212 |
| Contact Person: | Herbert Schoenhoefer<br>Director of RA/QA |
| Telephone: | 949.440.2632 |
| Date Prepared: | May 13, 2021 |
| Device Trade Name: | Silent Nite Sleep Appliance with the Glidewell Hinge |
| Common Name | Anti-Snoring Device |
| Classification Name<br>and Number<br>Product Code<br>Regulatory Class | Intraoral devices for snoring and for obstructive sleep apnea<br>21 CFR 872.5570<br>LRK<br>II |
| Primary Predicate<br>Device: | K183270, Silent Nite sl, Prismatik Dentalcraft, Inc. (LRK) |
| Reference Devices<br>and Product Codes: | Reference #1: K200125, Thermoforming Sheet Materials,<br>Erkodent Erich Kopp GmbH (MQC, KMY)<br>Reference #2: K083209, Acrylic Herbst Splint Appliance,<br>Specialty Appliances Works (LRK) |
| Indications for Use<br>Statement: | Silent Nite Sleep Appliance with the Glidewell Hinge is indicated<br>to reduce snoring and mild to moderate obstructive sleep apnea<br>(OSA) in patients 18 years of age or older. It is worn while<br>sleeping to support the lower jaw in a forward position prescribed<br>by the dentist. The appliance is removable by the patient. |
| Device Description.<br>Mechanism of Action,<br>and Summary of<br>Technological<br>Characteristics | The Silent Nite Sleep Appliance with the Glidewell Hinge is a<br>mandibular advancement device. It holds the mandible in a<br>protrusive position to increase the patient's ability to exchange air<br>and decrease air turbulence, thus improving airflow during sleep.<br>The device consists of upper and lower trays, which are fabricated<br>using a biocompatible dual-layered thermoplastic material, and an<br>engaging mechanism, which is the Glidewell Hinge (biocompatible<br>medical grade stainless steel bars), and band lugs and hooks for<br>attaching the optional orthodontic (elastic) bands, which are<br>provided as accessories.<br>The trays consist of a soft polyurethane inner layer that provides<br>patient comfort and a hard polyester outer layer for durability. The<br>hinge is adjustable by turning an inner adjustment screw with a<br>hex driver, enabling the amount of mandibular advancement |
{4}------------------------------------------------
prescribed by the clinician. The design allows for the advancement of the mandible to be adjusted (increased or decreased) in 0.25 mm increments up to 10 mm.
The orthodontic bands accessories are used to aid in holding the mouth closed. The end result, a forward position of the iaw, is the same as the predicate, however, the method of action of the predicate device and subject device are slightly different. Silent Nite Sleep Appliance with the Glidewell Hinge's method of action is identical to the Reference device and both use orthobands. Without orthobands, the patient's mouth can open during sleep which allows the tongue to fall back into the airway. The orthobands help keep the mouth closed to avoid this. With just a slight protrusion, the orthobands can be placed on the provided hooks. More protrusion requires the orthoband to be placed on the forward hook and the band lug.
The device is customized to conform to the patient's upper and lower dentition based on the clinician's prescription. Upon receipt of the prescription, models for the trays are made and used during the thermoforming fabrication of the upper and lower trays.
The difference between the Silent Nite Sleep Appliance with the Glidewell Hinge and the Primary Predicate Silent Nite sl is the mechanism connecting the trays: The new device's stainless steel hinge allows adjustment within the mechanism, while the Predicate's adjustment requires changing the plastic connectors.
This is a non-sterile device, to be used only by the single patient for whom it is custom-designed and is to be worn during sleep.
#### Device Testing Laboratory Testing
Verification and validation of the manufacturing process was performed by Prismatik Dentalcraft, Inc.
Verification testing was conducted to assure the following functional attributes of the finished device were correct:
- Mandibular advancement function ●
- Maximum hinge displacement (mm) ●
- Hinge placement/positioning .
Validation showed that the following were acceptable:
- Device fits and retains to the model
- Hinge is attached well to the trays ●
- Device's final finish is smooth and free of roughness or burs
## Biocompatibility
ISO 10993 testing was performed to assess the safety and biocompatibility of the Thermoforming Sheet Materials and documentation was provided in the 510(k) for the Reference device. K200125. The testing showed the material is safe and biocompatible for the intended use "fabrication of orthodontic and dental appliances".
{5}------------------------------------------------
The following biocompatibility testing was conducted according to Good Laboratory Practices and per ISO 10993 on the stainless steel hinge component of the device:
- In vitro Cytotoxicity (Part 5)
- . Guinea Pig Maximization Sensitization (Part 10)
- . Oral Mucosal Irritation in Hamsters:14 days (Part 10)
The test article met the requirements of each test and is safe and biocompatible for the intended use.
## Animal | Human Testing
No animal or human testing are required for this product because it is composed of the same materials, is designed similarly, and is manufactured by method similar to the predicate device.
Comparison to Predicate Device:
Refer to the Comparison Table below for similarities and differences between Silent Nite Sleep Appliance with the Glidewell Hinge and the Primary Predicate device. In summary:
- The intended uses are the same.
- . The mechanisms of actions are the same.
- . The materials used for the trays are the same as the Predicate and Reference #1.
- I The methods used for manufacturing the appliances are similar.
- . The materials used for the connector mechanism are different and the design for adjusting the jaw position is different, as follows:
The technological design of the subject device allows for microadjustments of the mandibular position to achieve the needed improvement in airflow during sleep, which is similar to Reference #2, while the Predicate device allows for millimeter incremental adjustment. Reference #2 also includes the optional orthobands like the subject device, but unlike the Predicate.
The design and material differences in the connector mechanism do not raise different questions of safety and effectiveness, and were addressed with standard verification and validation testing submitted in the 510(k).
Based on the documentation presented in the 510(k), as Substantial summarized above and the following side-by-side comparison, it Equivalence can be concluded subject device is substantially equivalent to the Conclusion predicate device.
{6}------------------------------------------------
| Substantial Equivalence Comparison Table | | | |
|----------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------|
| Element | New Device | Primary Predicate | Comparison |
| Device name: | Silent Nite Sleep Appliance<br>with the Glidewell Hinge | Silent Nite® sl | N/A |
| 510(k) #: | K210694 | K183270 | |
| Regulatory<br>Classification:<br>Product Code:<br>Class: | 21 CFR 872.5570<br>LRK<br>II | 21 CFR 872.5570<br>LRK<br>II | Same |
| Submitter: | Prismatik Dentalcraft, Inc. | Prismatik Dentalcraft, Inc. | Same |
| Indications for Use: | To reduce snoring and mild<br>to moderate obstructive<br>sleep apnea (OSA) in<br>patients 18 years of age or<br>older. Silent Nite with the<br>Glidewell Hinge is worn while<br>sleeping to support the lower<br>jaw in a forward position<br>prescribed by the dentist.<br>The appliance is removable<br>by the patient. | To reduce snoring and mild<br>to moderate obstructive<br>sleep apnea (OSA) in<br>patients 18 years of age or<br>older. Silent Nite® sl is<br>worn while sleeping to<br>support the lower jaw in a<br>forward position prescribed<br>by the dentist. The<br>appliance is removable by<br>the patient. | Same |
| Method of Use: | Single patient use,<br>removable and reusable<br>appliance | Single patient use,<br>removable and reusable<br>appliance | Same |
| Prescription Use<br>Only: | Yes | Yes | Same |
| Environment of<br>Use: | In patient's home or in sleep<br>laboratory | In patient's home or in<br>sleep laboratory | Same |
| DESIGN: | | | |
| Rigid Trays | Yes | Yes | Same |
| Adjustable | Yes | Yes | Same |
| MATERIALS: | | | |
| Trays: | Heat-sensitive impression<br>dual-layer material | Heat-sensitive impression<br>dual-layer material | Same as<br>Primary<br>Predicate and<br>Reference #1 |
| Advancement<br>Mechanism: | Stainless steel rod and tube<br>mechanism to position the<br>mandible forward | Plastic connectors to<br>position the mandible<br>forward | Similar to<br>Primary<br>Predicate and<br>same as<br>Reference #2 |
| STERILITY: | Non-sterile | Non-sterile | Same |
| BIOCOMPATIBILITY | All materials meet<br>requirements of ISO 10993:<br>• Cytotoxicity<br>• Irritation | All materials meet<br>requirements of ISO 10993:<br>• Cytotoxicity<br>• Irritation | Same |
# Substantial Equivalence Comparison Table
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.