K170692 · Respire Medical Holding · LRK · May 26, 2017 · Dental
Device Facts
Record ID
K170692
Device Name
Respire Pink Series with DentiTrac
Applicant
Respire Medical Holding
Product Code
LRK · Dental
Decision Date
May 26, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.5570
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Respire Pink Series intraoral appliances are intended to treat snoring and mild to moderate Obstructive Sleep Apnea (OSA) in adult patients 18 years of age or older. Optionally, the DentiTrac® micro-recorder may be incorporated into a Respire Pink Series device. The micro-recorder is intended to measure patient compliance to oral appliance therapy in combination with the DentiTrac® system.
Device Story
Mandibular advancement splint; holds jaw in forward position to maintain open airway; treats snoring and mild to moderate OSA. Device consists of upper and lower hard acrylic trays; Herbst hardware allows forward/lateral movement while restricting backward movement; includes expansion screw for mandibular adjustment. Optional DentiTrac micro-recorder embedded in device via Loctite epoxy; tracks compliance via oral temperature, movement, and head position. Used in home or sleep labs; patient-operated. Data uploaded to cloud-based web application via DentiTrac system for clinician review. Compliance data informs clinical decision-making regarding therapy effectiveness.
Clinical Evidence
Bench testing only. Risk assessment, EMC, electrical safety, biocompatibility, and software elements evaluated. Process validation for micro-recorder incorporation certified by Braebon. 100% visual and functional inspections performed.
Technological Characteristics
Mandibular advancement splint; hard acrylic fitting surfaces; chrome (BEGO Wironit®) plates (EF model); medical grade stainless steel hardware. DentiTrac micro-recorder embedded in Loctite epoxy. Connectivity: cloud-based web application. Sterilization: N/A (patient-cleaned).
Indications for Use
Indicated for adult patients 18+ with snoring and mild to moderate Obstructive Sleep Apnea (OSA).
Regulatory Classification
Identification
Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea are devices that are worn during sleep to reduce the incidence of snoring and to treat obstructive sleep apnea. The devices are designed to increase the patency of the airway and to decrease air turbulence and airway obstruction. The classification includes palatal lifting devices, tongue retaining devices, and mandibular repositioning devices.
Special Controls
*Classification.* Class II (special controls). The special control for these devices is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA.”
{1}------------------------------------------------
Image /page/0/Picture/2 description: The image is a circular seal for the Department of Health & Human Services - USA. The seal features the department's logo, which is a stylized representation of three human profiles facing right, stacked on top of each other. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper half of the circle, following its curvature.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 26, 2017
Respire Medical Holding Mr. Jonathan Sandler Ouality Assurance Manager 18 Bridge Street. Suite 4i Brooklyn. New York 11201
Re: K170692
Trade/Device Name: Respire Pink Series With DentiTrac® Regulation Number: 21 CFR 872.5570 Regulation Name: Intraoral Devices For Snoring And Intraoral Devices For Snoring And Obstructive Sleep Apnea Regulatory Class: Class II Product Code: PLC Dated: March 7, 2017 Received: March 7, 2017
Dear Jonathan Sandler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{2}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
# Lori A. Wiggins -S6
for
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K170692
Device Name Respire Pink Series with DentiTrac®
Indications for Use (Describe)
The Respire Pink Series intraoral appliances are intended to treat snoring and mild to moderate Obstructive Sleep Apnea (OSA) in adult patients 18 years of age or older.
Optionally, the DentiTrac® micro-recorder may be incorporated into a Respire Pink Series device. The micro-recorder is intended to measure patient compliance to oral appliance therapy in combination with the DentiTrac® system.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Respire Medical. The logo features a stylized figure with outstretched arms above a green and white structure, all enclosed within a circle. Below the circle, the words "Respire Medical" are written in an elegant, cursive font.
# 510(k) Summary
510(k) Submission Number: K170692
# Submitter
Respire Medical 18 Bridge Street, Suite 4J Brooklyn, NY 11201 Phone: 718-643-7326
# Official Contact:
Jonathan Sandler Quality Assurance Manager Phone: 718-360-9209 Email: Jonathan@respiremedical.com
# Date Prepared:
May 16, 2017
#### Device Identification
| Proprietary Name: | Respire Pink Series with DentiTrac® |
|------------------------|------------------------------------------------------------------------------------------------|
| Common Name: | Device, Anti-Snoring |
| Classification Name: | Intraoral devices for snoring and intraoral devices for snoring<br>and obstructive sleep apnea |
| Device Classification: | Class II |
| Product Code: | PLC |
| Regulation Number: | 21 CFR 872.5570 |
# Predicate Devices
| Candidate | Predicate | Manufacturer | 510(k) Number |
|----------------------------------------|--------------------------------|----------------------|---------------|
| Respire Pink Series<br>with DentiTrac® | SomnoDent® with Micro-Recorder | SomnoMed, Inc. | K150369 |
| | Respire Pink Series-Herbst | Respire Medical, LLC | K131138 |
| | Respire Pink Series-Herbst-EF | Respire Medical, LLC | K150572 |
The primary predicate device used for substantial equivalence is the SomnoMed SomnoDent® with Micro-Recorder. The reference predicate devices are the Respire Pink Series-Herbst devices which includes both the standard Respire Pink Series-Herbst device and the Respire Pink Series-Herbst-EF.
# Device Description
The Respire Pink Series with DentiTrac® refers to two devices: Respire Pink Series-Herbst and the Respire Pink Series-Herbst-EF, herein referred to collectively as Respire Pink Series. These devices function as mandibular advancement splints which hold the jaw in a forward
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Respire Medical. The logo features a person with their arms outstretched above a green structure, all enclosed in a circle. Below the circle, the words "Respire Medical" are written in an elegant, cursive font.
position. The forward jaw placement moves the tongue and pharyngeal tissue into a position to help maintain an open airway, which allows the passage of more air per breath and helps in the treatment of snoring and mild to moderate Obstructive Sleep Apnea (OSA). Respire Pink Series appliances are customizable devices which are produced according to the individual patient anatomy and prescription.
The Respire Pink Series-Herbst consists of full hard acrylic upper and lower fitting surfaces, and the Respire Pink Series-Herbst-EF consists of hard acrylic fitting side plates and chrome (BEGO Wironit®) upper palatal and lower lingual plates. The Herbst hardware on the side of the device allows for forward and lateral jaw movement while restricting backward jaw movement, and allows the patient to open and close the mouth. The Herbst hardware contains an expansion screw for mandibular position adjustment, and is attached to the appliance using medical grade stainless steel fixing elements and screws.
This submission is to provide clinicians the option to incorporate a Braebon DentiTrac® micro-recorder into a Respire Pink Series device. The micro-recorder is encapsulated in Loctite epoxy and fully encased into the device acrylic, and is not patient contacting. The micro-recorder is utilized to track patient compliance to prescribed oral appliance therapy by collecting data for wear-time through oral temperature, movement tracking, and head position. The data can be uploaded to a cloud-based web application using the DeniTrac® system which allows clinicians to review the patient data. The inclusion of the microrecorder provides additional treatment information to clinicians without effecting the operating principles of the Respire Pink Series devices.
Image /page/4/Picture/4 description: The image shows two figures of the Respire Pink Series with DentiTrac. Figure 1 shows a front view of the dental device, which is pink and has metal components. Figure 2 shows a side view of the same device, highlighting the DentiTrac component.
#### Intended Use
The Respire Pink Series intraoral appliances are intended to treat snoring and mild to moderate Obstructive Sleep Apnea (OSA) in adult patients 18 years of age or older.
Optionally, the DentiTrac® micro-recorder may be incorporated into a Respire Pink Series device. The micro-recorder is intended to measure patient compliance to oral appliance therapy in combination with the DentiTrac® system.
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Respire Medical. The logo features a circular emblem with a stylized figure with outstretched arms above a green structure, possibly representing a medical facility or a person in a relaxed pose. Below the emblem, the words "Respire Medical" are written in an elegant, cursive font, emphasizing the brand's name and suggesting a focus on respiratory health or well-being.
# Comparison to Predicate Devices:
| Substantial | Proposed: | Primary Predicate: | Reference Predicate: |
|-------------------------------------------------------------|-----------------------------------------------------------------------|------------------------------------------------------------------------|--------------------------------------------------------------------|
| Equivalence Topic | Respire Pink Series with | SomnoMed | Respire Pink Series- |
| | DentiTrac® | SomnoDent® with | Herbst / Pink Series- |
| | | Micro-Recorder | Herbst-EF |
| 510(k) | N/A | K150369 | K131138 / |
| | | | K150572 (EF) |
| Company Name | Respire Medical | SomnoMed | Respire Medical |
| Device Name | Device, Anti-Snoring | Device, Anti-Snoring | Device, Anti-Snoring |
| Indications for Use | The Respire Pink Series intraoral<br>appliances are indented to treat | The SomnoDent Intraoral devices<br>are intended for the treatment of | The Respire Pink Series-Herbst is<br>indicated to treat mild to |
| Statement | snoring and mild to moderate | night time snoring and mild to | moderate OSA. |
| | Obstructive Sleep Apnea (OSA) in | moderate obstructive sleep | |
| | adult patients 18 years of age or<br>older. | apnea in patients 18 years of age<br>or older | The Respire Pink Series-Herbst-EF<br>is indicated to treat mild to |
| | | | moderate OSA. |
| | Optionally, the DentiTrac micro-<br>recorder may be incorporated | Optionally, if the DentiTrac micro-<br>recorder is completely embedded | |
| | into a Respire Pink Series device. | into the SomnoDent device, the | |
| | The micro-recorder is intended to | micro-recorder is intended to | |
| | measure patient compliance to<br>oral appliance therapy in | measure patient compliance to<br>oral device/appliance therapy in | |
| | combination with the DentiTrac | combination with the DentiTrac | |
| Intended Use | system. | System. | |
| Intended as an | Yes | Yes | Yes |
| intraoral device | | | |
| Intended to reduce | Yes | Yes | Yes |
| snoring or help | | | |
| alleviate snoring | | | |
| Intended for night | Yes | Yes | Yes |
| time use | | | |
| Indicated for single | Yes | Yes | Yes |
| patient multiuse | | | |
| Indicated for use at | Yes | Yes | Yes |
| home or sleep | | | |
| laboratories | | | |
| Design | | | |
| Separate upper and | Yes | Yes | Yes |
| lower tray pieces | | | |
| Works by mandibular | Yes | Yes | Yes |
| advancement | | | |
| Can be adjusted or | Yes | Yes | Yes |
| refit | | | |
| Lower jaw adjustment | Yes | Yes | Yes |
| using supplied | | | |
| components | | | |
| Permits patient to | Yes | Yes | Yes |
| breathe through | | | |
| mouth | | | |
| Upper and lower trays<br>disengage for easy<br>removal | Yes | Yes | Yes |
| Cleaned and inspected<br>daily by patient | Yes | Yes | Yes |
| Materials | | | |
| Advancement<br>mechanism: surgical<br>grade stainless steel | Yes | Yes | Yes |
| Acrylic fitting surface | Yes | Yes | Yes |
| Acrylic/Wironit®<br>fitting surfaces | Yes (EF) | No | Yes (EF) |
| Micro-Recorder<br>embedded into<br>device | Yes | Yes | No |
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Respire Medical. The logo features a circle with a person with their arms raised above a green bridge. Below the circle is the text "Respire Medical" in a cursive font.
# Performance Testing
Non-clinical performance testing:
The addition of the DentiTrac® micro-recorder does not adversely affect the original design, therefore, not introducing any new issues of concern. A risk assessment shows that risks are mitigated to acceptable levels, and performance testing related to the DentiTrac® microrecorder confirms that EMC and electrical safety, biocompatibility, and software elements were evaluated. DentiTrac® performance testing is located in the Braebon DentiTrac® Master File MAF 2557. Process validation concerning the incorporation and functionality of the DentiTrac® micro-recorder is certified by Braebon. Respire Medical performs 100% visual and functional inspections throughout the production process to ensure that the device meets manufacturing specifications for the Respire Pink Series device and specifications for the incorporation of the DentiTrac® micro-recorder.
#### Discussion of Similarities and Differences Between the Subject and Predicate Devices
The Respire Pink Series predicate devices (K131138 and K150572) are identical to the Respire Pink Series with DentiTrac® candidate device apart from the addition of the DentiTrac® micro-recorder. The general characteristics and specifications are unchanged in that each devices utilizes the same materials, manufacturing procedures, principles of operation, and overall design. The DentiTrac® micro-recorder used in the Respire Pink Series with DentiTrac® is identical to that which is used in the SomnoDent® (K150369) predicate device, and the indications for use are similar between the two devices. However, the age population has been added which is not present on K131138 and K150572 predicates.
#### Conclusion
Based on the similarities in the indications for use, technology, biocompatibility assessment, and non-clinical performance testing, Respire Pink Series with DentiTrac® is substantially equivalent to the SomnoMed SomnoDent® with Micro-Recorder and the Respire Pink Series predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.