SomnoDent Classic with DentiTrac Micro-recorder, SomnoDent Flex with DentiTrac Micro-recorder, SomnoDent Herbst Advance (Classic or Flex Retention) with DentiTrac Micro-recorder, SomnoDent G2 (Classic or Flex Retention) with DentiTrac Micro-recorder, SomnoDent Fusion (Classic or Flex Retention) with DentiTrac Micro-recorder
K150369 · Somnomed, Inc. · PLC · Jun 18, 2015 · Dental
Device Facts
Record ID
K150369
Device Name
SomnoDent Classic with DentiTrac Micro-recorder, SomnoDent Flex with DentiTrac Micro-recorder, SomnoDent Herbst Advance (Classic or Flex Retention) with DentiTrac Micro-recorder, SomnoDent G2 (Classic or Flex Retention) with DentiTrac Micro-recorder, SomnoDent Fusion (Classic or Flex Retention) with DentiTrac Micro-recorder
Applicant
Somnomed, Inc.
Product Code
PLC · Dental
Decision Date
Jun 18, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.5570
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SomnoDent® intraoral devices are intended for the treatment of night time snoring and mild to moderate obstructive sleep apnea in patients 18 years of age or older. Optionally, if the DentiTrac® micro-recorder is completely embedded into the SomnoDent® appliance, the micro-recorder is intended to measure patient compliance to oral appliance therapy in combination with the DentiTrac® System.
Device Story
SomnoDent is a custom-fitted mandibular repositioning device; increases pharyngeal space during sleep to improve airway exchange. Device consists of upper and lower trays; adjustable coupling mechanism allows dentist-controlled mandibular advancement. This submission adds an optional, fully embedded DentiTrac micro-recorder. Recorder monitors patient compliance by tracking oral temperature, movement, and head position. Data is uploaded via a base station at the clinician's office to a web application for cloud-based reporting. Device is used at home; operated by patient. Clinicians use output to track therapy adherence. Embedding the recorder does not alter the primary mandibular advancement function or patient safety; device remains sealed under acrylic.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by non-clinical bench testing, including EMC, electrical safety, biocompatibility, and process validation for the embedding procedure, as documented in MAF2557.
Technological Characteristics
Custom-fitted intraoral mandibular repositioner. Materials include surgical grade stainless steel for advancement mechanisms and acrylic/soft lining for trays. Features an embedded DentiTrac micro-recorder for compliance monitoring. Connectivity via base station to cloud-based reporting application. Biocompatible, sealed construction.
Indications for Use
Indicated for patients 18 years or older with night time snoring and mild to moderate obstructive sleep apnea.
Regulatory Classification
Identification
Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea are devices that are worn during sleep to reduce the incidence of snoring and to treat obstructive sleep apnea. The devices are designed to increase the patency of the airway and to decrease air turbulence and airway obstruction. The classification includes palatal lifting devices, tongue retaining devices, and mandibular repositioning devices.
Special Controls
*Classification.* Class II (special controls). The special control for these devices is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA.”
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other. The profiles are simple and abstract, with only the outlines of the faces visible.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 18, 2015
SomnoMed, Inc. c/o Kien T. Nguyen, Ph.D., MBA President - North America 7460 Warren Parkway, Suite 190 Frisco, Texas, 75034
Re: K150369
Trade/Device Name: SomnoDent® with Micro-Recorder Regulation Number: 21 CFR 872.5570 Regulation Name: Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea Regulatory Class: II Product Code: PLC Dated: May 19, 2015 Received: May 19, 2015
### Dear Dr. Nguyen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina
Kianq-S
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
### K150369
Device Name
# SomnoDent® with Micro-Recorder
Indications for Use (Describe)
The SonmoDent® intraoral devices are intended for the treatment of night time snoring and mild to moderate obstructive sleep apnea in patients 18 years of age or older.
Optionally, if the DentiTrac® micro-recorder is completely embedded into the SomnoDent® device, the microrecorder is intended to measure patient compliance to oral device/appliance therapy in combination with the DentiTrac® System.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
■ Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# (As required by 21 CFR 807.92) SomnoDent® with Micro-Recorder
#### 1.0 Submitter
SomnoMed, Inc. 7460 Warren Parkway, #190 Frisco, TX 75034 Telephone: 972-377-3400 Fax: 972-377-3404
### Official Contact
Kien Nguyen, Ph.D., MBA President - North America Telephone: 972-377-3400, Ext. 101 Email: knguyen@somnomed.com
### 2.0 Date Prepared
June 16, 2015
### 3.0 Device Identification
| Proprietary Names: | SomnoDent® with Micro-Recorder |
|------------------------|----------------------------------------------------|
| Common Name: | Device, Anti-Snoring |
| Classification Name: | Intraoral device for snoring and Intraoral devices |
| | for snoring and obstructive sleep apnea |
| Device Classification: | Class II |
| Product Code: | PLC |
| Regulation Number: | 21 CFR 872.5570 |
### 4.0 Legally Marketed Predicate Device
| Candidate(s) | Predicate | Manufacturer | Document<br>Number |
|-----------------------------------|------------------------------------------------------|----------------|--------------------|
| SomnoDent® with<br>Micro-Recorder | SomnoDent® Fusion Classic,<br>SomnoDent® Fusion Flex | SomnoMed, Inc. | K140278 |
The SomnoDent with Micro-Recorder is substantially equivalent to the SomnoDent® Fusion Classic and Flex products listed above, currently in commercial distribution.
### Device Description 5.0
The SomnoDent family of intraoral devices consists of the following models: Classic, Flex, G2, Herbst Advance, and Fusion. The devices functions as a mandibular repositioner, which acts to increase the patient's pharyngeal space during sleep. The increase in the patient's pharyngeal space improves their ability to exchange air during sleep. The devices are patient specific (they are customized for each patient) and have an adjustable coupling mechanism enabling the amount of mandibular advancement to be seep dentist at the time of fitting the device. The devices can then be adjusted by the sleep dentist to control the treatment to find the best possible adjustment.
This submission adds the option for any clinician to decide to incorporate a DentiTrac® microrecorder embedded into a SomnoDent® to record a patient's compliance to the prescribed oral appliance therapy in combination with the DentiTrac® System. This option can be incorporated into the family member models listed above. During scheduled visits, the data within the DentiTrac® can be uploaded to a web application for cloud-based reporting and tracking using a DentiTrac® Base Station at the clinician's office. The DentiTrac® micro-recorder monitors the wear time through the oral temperature, as well as tracks movements and head position. The inclusion of the embedded DentiTrac® micro-recorder provides additional information when used in combination with the DentiTrac® System, but does not impact the operating principles or safety of the SomnoDent®
{4}------------------------------------------------
### 6.0 Intended Use
The SomnoDent® intraoral devices are intended for the treatment of night time snoring and mild to moderate obstructive sleep apnea in patients 18 years of age or older.
Optionally, if the DentiTrac® micro-recorder is completely embedded into the SomnoDent® appliance, the micro-recorder is intended to measure patient compliance to oral appliance therapy in combination with the DentiTrac® System.
### 7.0 Comparison to the Predicate
| Technological Characteristics | Predicate<br>SomnoDent® Fusion<br>K140278 | Proposed<br>SomnoDent® Family with<br>DentiTrac® Micro-<br>Recorder | | | |
|------------------------------------------------------------------------------------|-------------------------------------------|---------------------------------------------------------------------|--|--|--|
| Intended Use | | | | | |
| Intended as an intraoral device | Yes | Yes | | | |
| Intended to reduce snoring or help<br>alleviate snoring | Yes | Yes | | | |
| Treatment of mild to moderate<br>obstructive sleep apnea | Yes | Yes | | | |
| Intended for nighttime use | Yes | Yes | | | |
| Indicated for single patient multiuse | Yes | Yes | | | |
| Indicated for use at home or sleep<br>laboratories | Yes | Yes | | | |
| Target population: adults | Yes | Yes | | | |
| Prescription device | Yes | Yes | | | |
| Design | | | | | |
| Customized fit for each patient | Yes | Yes | | | |
| Separate upper and lower tray pieces | Yes | Yes | | | |
| Works by mandibular advancement | Yes | Yes | | | |
| Can be adjusted or refit | Yes | Yes | | | |
| Lower jaw adjustment using supplied<br>components | Yes | Yes | | | |
| Permits patient to breathe through mouth | Yes | Yes | | | |
| Upper and lower trays disengage for easy<br>removal | Yes | Yes | | | |
| Cleaned and inspected daily by patient | Yes | Yes | | | |
| Material | | | | | |
| Trays constructed from a soft lining<br>material adhered to a hard surface acrylic | Yes (Flex retention) | Yes (Flex retention) | | | |
| | No (Classic retention) | No (Classic retention) | | | |
| Advancement mechanism constructed of<br>surgical grade stainless steel | Yes | Yes | | | |
| DentiTrac® Micro-recorder embedded<br>into SomnoDent® | No | Yes | | | |
{5}------------------------------------------------
The addition of the DentiTrac Micro-Recorder to the SomnoDent device does not change the intended use of the SomnoDent device for the treatment of night time snoring and mild to moderate obstructive sleep apnea in patients 18 years of age or older. Based on how the DentiTrac Micro-Recorder is embedded inside the SomnoDent device, the patient will not be physically exposed to the micro-recorder. The DentiTrac Micro-Recorder is completely sealed under a layer of acrylic within the SomnoDent device. Furthermore, the size, position and location of the embedded DentiTrac Micro-Recorder in the SomnoDent device does not increase any risk to patient safety as there is ample space in the patient's oral cavity to accommodate the additional volume.
#### 8.0 Performance Testing
The SomnoDent® family of devices with DentiTrac® micro-recorder are identical to the predicate devices with the exception of the addition of the DentiTrac® micro-recorder. Since the characteristics of the SomnoDent® did not change, no new clinical performance data was necessary to substantiate this change. Risk assessments and non-clinical testing demonstrated that EMC and electrical safety, biocompatibility, and software elements were assessed and The data is reported in master file, MAF2557. Process validation was generated from tested. SomnoMed to ensure that the SomnoDent® devices can be repeatedly and reliably embedded and retain the Quality Control functionality of SomnoDent® and the DentiTrac®.
#### 9.0 Conclusion
Based on the similarities in the primary intended use, principles of operation, functional design, established materials and medical use, and the information presented in the summary and in Masterfile MAF2557, the conclusions demonstrate that the SomnoDent® family members with an embedded DentiTrac® micro-recorder are as safe, as effective, and performs as well as the predicate device. The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.