Custom-made Invisible Aligners

K203624 · Zhejiang Yinchili Medical Technology Co., Ltd. · NXC · Jun 16, 2022 · Dental

Device Facts

Record IDK203624
Device NameCustom-made Invisible Aligners
ApplicantZhejiang Yinchili Medical Technology Co., Ltd.
Product CodeNXC · Dental
Decision DateJun 16, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.5470
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Custom-made Invisible Aligners is indicated for the treatment of tooth malocclusion in patients with permanent dentition (i.e., all second molars). The Custom-made Invisible Aligners positions teeth by way of continuous gentle force.

Device Story

Custom-made Invisible Aligners are a series of clear, thin thermoformed PETG plastic dental aligners. Used in a clinical setting under the direction of a dental practitioner, the device is worn by the patient (typically 22 hours/day) to progressively reposition teeth. Each aligner in the sequence applies continuous gentle force to move teeth toward a target position based on a doctor's prescription. The system is removable and replaced every two weeks. The aligners benefit patients by correcting malocclusion through programmed displacement.

Clinical Evidence

No clinical study included. Bench testing only. Performance testing included thickness, appearance, odor, density, water absorption, dissolution, color stability, tear resistance, wear resistance, and flexural modulus of elasticity, all meeting pre-established specifications. Biocompatibility testing (cytotoxicity, irritation, sensitization) per ISO 10993-1 confirmed non-cytotoxic and non-sensitizing properties. Manufacturing accuracy validated via 3D molding and aligner fit against treatment design.

Technological Characteristics

Material: Thermoplastic copolyester (PETG). Manufacturing: Thermoforming. Form factor: Clear, thin, removable dental aligners. Connectivity: Not specified. Sterilization: Non-sterile. Software: Moderate level of concern for ordering/processing software.

Indications for Use

Indicated for treatment of tooth malocclusion in patients with permanent dentition (all second molars).

Regulatory Classification

Identification

An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text. June 16, 2022 Zhejiang Yinchili Medical Technology Co., Ltd. % Ivy Wang Technical Manager Shanghai Sungo Management Consulting Company Limited 14th Floor, 1500# Central Avenue Shanghai, Shanghai 200122 CHINA Re: K203624 Trade/Device Name: Custom-made Invisible Aligners Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: Class II Product Code: NXC Dated: May 7, 2022 Received: May 12, 2022 Dear Ivy Wang: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K203624 Device Name Custom-made Invisible Aligner Indications for Use (Describe) The Custom-made Invisible Aligners is indicated for the treatment of tooth malocclusion in patients with permanent dentition (i.e., all second molars). The Custom-made Invisible Aligners positions teeth by way of continuous gentle force. Type of Use (Select one or both, as applicable) | Permitted Use (Part 21 CFR 601 Subpart E) | |-------------------------------------------| | Secondary Use (21 CFR 601 Subpart E) | > Prescription Use (Part 21 CFR 801 Subpart D) __ Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ### 510(K) Summary This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92. #### A. Applicant: Zhejiang Yinchili Medical Technology Co., Ltd. Address: North 4F, Building 1, No.239 Yatai Road, Daqiao Town, Nanhu District, Jiaxing, Zhejiang Province, China Contact Person: Ms. Hangli Chen Tel: +86-15024335314 Email: chenhangli@smartee.cn Date of summary prepared: 2022-05-24 Submission Correspondent: Primary contact: Ms. Ivy Wang Shanghai SUNGO Management Consulting Co., Ltd. Room 1309, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-58817802 Email: haiyu.wang@sungoglobal.com Secondary contact: Mr. Raymond Luo Room 1309, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-68828050 Email: fda.sungo@gmail.com #### B. Device: Trade Name: Custom-made Invisible Aligners Common Name: Sequential Aligners Regulatory Information Classification Name: Aligner, Sequential Classification: Class II Product code: NXC Regulation Number: 21 CFR 872.5470 Review Panel: Dental ### C. Primary Predicate device: K180346 Byte Aligner System Straight Smile, LLC ### D. Indications for use of the device: The Custom-made Invisible Aligners is indicated for the treatment of tooth malocclusion in patients with permanent dentition (i.e., all second molars). The Custom-made Invisible Aligner positions teeth {4}------------------------------------------------ by way of continuous gentle force. ### E. Device Description: The Custom-made Invisible Aligners System is a series of dental aligners that are fabricated of clear, thin thermoformed Polyethylene terephthalate (PETG) plastic to progressively reposition the teeth. Corrective force to reposition the teeth is delivered via minor changes into a position in each subsequent aligner. ### F. Comparison to predicate device The Custom-made Invisible Aligners is substantially equivalent in intended use, indications for use, mode of action, mode of use, design, and materials to the predicate Straight Smile Byte Aligner System (K180346). Only minor differences exist between the subject product and the predicate, which do not affect the safety or effectiveness of the subject device. | Table 1: Comparison to Predicate Device | | | | |-----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------| | Device | Subject Device | Primary Predicate Device | Result | | Manufacturer | Zhejiang Yinchili Medical<br>Technology Co., Ltd. | Straight Smile, LLC | - | | 510K number | K203624 | K180346 | - | | Model Name | Custom-made Invisible<br>Aligners | Byte Aligner System | - | | Classification | Class II Device, NXC (21 CFR 872.5470) | Class II Device, NXC (21 CFR 872.5470) | Same | | Classification Name | Aligner, Sequential | Aligner, Sequential | Same | | Indications for use | The Custom-made Invisible<br>Aligners is indicated for the<br>treatment of tooth<br>malocclusion in patients with<br>permanent dentition (i.e., all<br>second molars). The<br>Custom-made Invisible<br>Aligner positions teeth by<br>way of continuous gentle<br>force. | The Byte Aligner System is<br>indicated for the treatment<br>of tooth malocclusion in<br>patients with permanent<br>dentition (i.e., all second<br>molars). The Byte Aligner<br>System positions teeth by<br>way of continuous gentle<br>force. | Same | | Mode of Action | Orthodontic tooth movement<br>occurs through forces applied<br>by the device to the dentition<br>as each tooth follows the<br>programmed displacement<br>based on a doctor's<br>prescription. | Orthodontic tooth<br>movement occurs through<br>forces applied by the device<br>to the dentition as each<br>tooth follows the<br>programmed displacement<br>based on a doctor's<br>prescription | Same | | Anatomical Site of Use | Oral cavity | Oral cavity | Same | | Mode of Use | Each aligner is worn by the patient as determined by the treating dental practitioner, generally for 22 hrs/day (or full time except for eating and hygiene) for 2 weeks prior to being replaced by the next aligner in sequence. This is repeated for a duration as prescribed by a Dental Professional. | Each aligner is worn by the patient as determined by the treating dental practitioner, generally for 2 weeks prior to being replaced by the next aligner in sequence. This is repeated for a duration as prescribed by a Dental Professional. | Same | | Application | Removable | Removable | Same | | Raw Material Used | Thermoplastic copolyester (polyethylene terephthalate-ethylene glycol copolyester) | Thermoplastic polymers (polyethylene terephthalate glycol or PETG) | Similar. Both are thermoplastic forming materials that do not raise any additional questions of safety or efficacy. | | Method of Manufacturing | Thermoforming | Thermoforming | Same | | OTC or Rx | Rx | Rx | Same | | Sterile | No | No | Same | | Biocompatibility | In compliance with ISO 10993, tests including Cytotoxicity Oral Mucosa Irritation Sensitization | In compliance with ISO 10993, tests including Cytotoxicity Sensitization | Similar. Both are in compliance with ISO 10993 | | Design | Image: Clear aligner | Image: Clear aligner | Similar. Both are transparent plastic films. | Table 1 provides a comparison of the subject and predicate device. {5}------------------------------------------------ ### G. Non-clinical Test ### 1) Performance Testing Bench testing has demonstrated that the device is in compliance with pertinent standards and {6}------------------------------------------------ specifications, the expectations of the dental community and the product labeling. A comparison testing was performed in combination with the subject and predicate device including thickness, appearance, odor, density, water absorption, dissolution, color stability, tear resistance, wear resistance, flexural modulus of elasticity. Please see below table 2. | Test item | Test standard/method | Acceptance criteria | Result | |--------------------------------|----------------------|----------------------------------|--------| | Thickness | Internal standard | ≤1.2mm | Pass | | Appearance | "ZMT-FD-SS-052" | No crack or bubbling; No defects | Pass | | Odor | | Odorless | Pass | | Density | ASTM D792-2013 | ≤2.6g/cm3 | Pass | | Water absorption | ISO20795.1-2013 | ≤32 μg/mm3 | Pass | | Dissolution | ISO20795.1-2013 | ≤1.6 μg/mm3 | Pass | | Color stability | ISO20795.1-2013 | No change | Pass | | Tear resistance | ISO6383.1-2015 | >200N/cm | Pass | | Wear resistance | ISO9352-2012 | <0.25g/1000r | Pass | | Flexural modulus of elasticity | ISO20795.1-2013 | ≥600Mpa | Pass | The results showed that the subject device is as effective as the predicate device. ### Manufacturing validation accuracy testing Manufacturing accuracy validation were conducted to the Custom-made Invisible Aligners from 12 different patient case were evaluated at the beginning, middle and end throughout the sequence. The accuracy of 3D molding and aligner molding are checked and meet the pre-established specification. The suitability, function and form of the aligner were checked and comparing it to the treatment design in the software, and the results were comply with the pre-established specifications and acceptance criteria. ### Shelf life – 2 years A 2-year shelf life was determined by real-time aging testing. Performance testing were conducted after 30 months real-time aging under commercial storage condition. The test results showed conformity with the pre-established specifications and acceptance criteria. ### 2) Biocompatibility Testing The biocompatibility evaluation for the device was conducted in accordance with "Use of International Standard ISO 10993-1, Biological evaluation and testing within a risk management process –Guidance for Industry and Food and Drug Administration Staff" as recognized by FDA. The aligner is considered mucosal membrane contacting for a duration of greater than 30 days. The testing included the following tests: - Cytotoxicity - Irritation - Sensitization The results of the testing met the requirements of the study protocols and the material is considered non-cytotoxic, non-sensitizing and is not an intracutaneous irritant. The results of the studies further support a determination of substantial equivalence. {7}------------------------------------------------ ### 3) Software Verification and Validation Testing Software verification and validation testing were conducted on the software that facilitates ordering and processing of the Custom-made Invisible Aligner to support that the device is as safe and effective as the predicates. Documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level of concern. #### H. Clinical Test Conclusion No clinical study is included in this submission. #### l. Conclusion The conclusions drawn from the nonclinical tests demonstrate that the subject device is as safe, as effective, and performs as well as or better than the legally marketed predicate device K180346.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...