eCligner

K143499 · Eclear International Co., Ltd. · NXC · Oct 9, 2015 · Dental

Device Facts

Record IDK143499
Device NameeCligner
ApplicantEclear International Co., Ltd.
Product CodeNXC · Dental
Decision DateOct 9, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.5470
Device ClassClass 2
AttributesTherapeutic

Indications for Use

eCligner® is a series of clear, lightweight, plastic appliances indicated for the correction of dental malocclusion in patients with permanent dentition (i.e. all second molars). The eCligner® is intended for minor anterior tooth movement by way of continuous gentle force.

Device Story

eCligner is a transparent, removable orthodontic appliance system; used for minor anterior tooth movement. Process: dentist takes patient tooth molds; 3D CAD/CAM system designs treatment plan; plastic models printed; aligners produced via vacuum forming. Treatment: patient wears three sequential aligners of varying thickness (soft, medium, hard) per step; 17 hours/day wear required; 3-week intervals per step. Ultra-hard version available for bruxism patients. Dentist monitors treatment from delivery of first to final aligner. Output: continuous gentle force on teeth to achieve progressive realignment. Benefits: gradual, aesthetic tooth correction.

Clinical Evidence

Bench testing only. Biocompatibility testing (cytotoxicity, irritation, sensitization) performed per ISO 10993. Software validation performed for the CAD/CAM system used in fabrication per FDA guidance.

Technological Characteristics

Thermoplastic material; vacuum-formed clear aligners; 0.5mm (soft), 0.62mm (medium), 0.75mm (hard), and 1.00mm (ultra-hard) thicknesses. 3D CAD/CAM system for treatment planning and model fabrication. Standalone workflow involving dentist-approved digital model schemes.

Indications for Use

Indicated for correction of dental malocclusion in patients with permanent dentition (all second molars) requiring minor anterior tooth movement.

Regulatory Classification

Identification

An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, with three figures or lines representing the staff and intertwined snakes. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 October 9, 2015 eClear International Co., Ltd. c/o Ms. April Lee Consultant WithUS Consulting 2531 Pepperdale Drive Rowland Heights, California 91748 Re: K143499 Trade/Device Name: eCligner® Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: II Product Code: NXC Dated: September 3, 2015 Received: September 9, 2015 Dear Ms. Lee: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Tina Kiang -S for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital. Respiratory. Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Clear Internat 30 Teheranro 27 gil, Gangnam-gu, Seoul, Rep Tel. 82-2-515-5945 / Fax. 82-2-3443-9092 ## Indication for Use Statement #### Indication for Use K143499 510(K) Number (if known): Device Name: eCligner® #### Indication for Use: eCligner® is a series of clear, lightweight, plastic appliances indicated for the correction of dental malocclusion in patients with permanent dentition (i.e. all second molars). The eCligner® is intended for minor anterior tooth movement by way of continuous gentle force. Prescription Use _____________________________________________________________________________________________________________________________________________________________ AND/OR Over-The-Counter _____________ (Part 21 CFR 801 Subpart D) (Per 21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) For FDA Use Only Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) Page 1 of 1 {3}------------------------------------------------ # eClearInternational Co.,L 30Teheranro27 gil,Gangnam-gu, Seoul,Republic ofKorea Tel.82-2-515-5945/Fax.82-2-3443-9092 #### 510(k) Summary # 510(k) Summary Submitter eClear International Co., Ltd. Dr. Tae Weon Kim 30 Teheranro 27 gilGangnam-gu Seoul, 135-915 Republic of Korea Email: info@ecligner.com Tel. +82-2-515-5945 Fax. +82-2-3443-9092 #### Official Correspondent With US Group, Inc. April Lee 2531 Pepperdale Drive, Rowland Heights, CA 91748 USA Email: withus6664@gmail.com Phone: 1-909-274-9971 Fax: 1-909-460-8122 #### Device Information - . Trade (Proprietary)Name: eCligner® - . Common Name: Aligners, Sequential - Classification Name: Orthodontic Plastic Bracket - Product Code: NXC - Panel: Dental - Regulation Number: 21 CFR872.5470 - Device Class: ClassII - Date prepared:10/08/2015 ● ## Predicate Device Invisalign® System (K081960) ## General Description eCligner® is a transparent and removable orthodontic appliance. eCligner® is a digitally- made Clear Aligner by the 3DCAD/CAM system. By printing the completed setup data into a series of plastic models, eCligne® is produced by a simple stamping procedure (Vacuum forming). eCligner® is designed for orthodontic tooth movement. Each step for using orthodontic appliance, eCligner® consists of 3 weeks (1 week for soft, 1 week for medium and 1 week for hard) and wearing the appliance 17 hours a day for each sequence is required. * More than 3 weeks of wearing may be required per step depending on the direction of orthodontic treatment and dentist's instruction. {4}------------------------------------------------ * eCligner® -H is used for 3 weeks during Final process. During one individual step of treatment, three aligners of different layer thickness are worn in weekly intervals. The applied layer thicknesses are 0.5 mm (medium) and 0.75 mm (hard). Each individual step is consisting of three different thickness appliances as follows: - 1. eCligner® -S (Soft) - 2. eCligner® -M (Medium) - 3. eCligner® -H (Hard) *eCligner® -UH (Ultra Hard, 1.00mm) is used only for the patient with Bruxism (Teeth Grinding) since the aligner can be broken while the patient grinds their teeth unintentionally. #### Indication for Use eCligner® is a series of clear, lightweight, plastic appliances indicated for the correction of dental malocclusion in patients with permanent dentition (i.e.all second molars). The eCligner® is intended for minor anterior tooth movement by way of continuous gentle force. #### Mechanism of Action Each eCligner® exerts gentle force to achieve progressive realignmen to the teeth until the final correction has been attained. ## Summary of Technological Characteristics A dental health care professional (e.g. orthodontist or dentist), prescribes the eCligner system based on an assessment of the patient's teeth, determines a course of treatment with the system, takes molds of the patient's teeth, and completes a prescription form, eCligner designs a series of plastic trays intended to gradually realign the patient's teeth in accordance with the physician's prescribing physician reviews and approves the model scheme before the molds are produced. Once approved, eCligner produces the trays, which are formed of clear, thin, thermoformed plastic. The trays are sent back to the dentist who then provides them to the patient, confirming fit and design. Over a period, additional trays are provided sequentially to the patient by the physician to gradually move the target teeth to the designed position. The dental care professional monitors treatment from the moment the first aligner is delivered to when the final aligner is delivered. The trays are held in place by pressure and can be removed by the patient at anytime. The technology is essentially identical to that used by a number of sequential alignment systems including the predicate device. {5}------------------------------------------------ #### Performance Testing Biocompatibility tests such as cytotoxicity, irritation, and sensitization were performed in accordance with ISO 10993. In addition, software validation test for CAD/CAM system used to fabricate this subject device was performed according to Guidance for Industry and FDA Staff, Guidance for the content of premarket submissions for software contained in medical devices. A performance qualification of the available software utilized in the processing steps of the subject device has been included to support substantial equivalence. ## Comparison to the Predicate Devices The subject device is substantially equivalent to the following predicate devices: - K081960, Invisalign system manufactured by Align Technology. {6}------------------------------------------------ # Comparison to the Predicate Devices | | Subject Device | Predicate Device | |--------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------| | Company Name | eClear International Co., Ltd. | ALIGNTECHNOLOGY | | Device Name | eCligner® | INVISALIGNSYSTEM | | 510(k)Number | NA | K081960 | | Device Classification<br>Name | Aligner, Sequential | Aligner, Sequential | | Classification Product<br>Code | NXC | NXC | | Regulation Number | 21 CFR872.5470 | 21 CFR872.5470 | | Indications for Use | eCligner® is a series of<br>clear, lightweight, plastic<br>appliances indicated for the<br>correction of dental<br>malocclusion in patients<br>with permanent dentition<br>(i.e.all second molars). The<br>eCligner® is intended for<br>minor anterior tooth<br>movement by way of<br>continuous gentle force. | The Invisalign system is<br>indicated for the alignment of<br>teeth during orthodontic<br>treatment of malocclusion. | | Material | Thermoplastic | Thermoplastic | | Design | Image: eCligner design | Image: Invisalign design | | Mechanism of Action | Gentle continuous pressure on<br>teeth | Gentle continuous pressure on<br>teeth | {7}------------------------------------------------ #### Substantial Equivalence Discussion The eCligner® has a substantially equivalent intended use as the identified predicate. The subject device is similar in fundamental scientific technology to the predicate device in that they all have been designed, manufactured and tested in compliance with 21 CFR 872.5470. The subject and predicate devices are similar in indications, design, mechanism of action and materials. The difference between the subject device and the predicate device is the aligners with various layer thicknesses. The subject device needs three steps of treatment with 0.5 mm (soft), 0.62 mm (medium) and 0.75 mm (hard) thickness. The verbiage of the Indications for Use of the subject device is slightly different than that of the declared predicates; however, these slight differences in wording does not change the intended use of the subject device has compared to the declared predicate. Any differences in technology characteristics are accompanied by information that demonstrated the device is as safe and as effective as the predicate and do not raise different questions of safety and effectiveness than the predicate. #### Conclusion Testing to demonstrate the acceptability of the software utilized in the treatment planning steps has been included to support substantial equivalence. Based on the similarities of the two devices, it is concluded that the eCligner® is substantially equivalent to the predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...