The ClearPath Aligner is a series of clear, lightweight, plastic appliances indicated for the correction of dental malocclusion in patients with permanent dentition (i.e. all second molars). The ClearPath Aligner is intended to position teeth by way of continuous gentle force.
Device Story
ClearPath Aligner is a series of clear, thermoformed plastic appliances; used for correction of dental malocclusion. Clinician takes patient impressions/molds; sends to manufacturer for custom fabrication. Patient wears aligners for 14-day intervals, removing only for meals. Device exerts continuous gentle force to achieve progressive tooth realignment. Used in dental clinic/home settings; prescribed by dentists or orthodontists. Output is physical tooth movement; clinical benefit is correction of malocclusion. Device is identical in design and material to predicate; clinical case reports demonstrate effectiveness for full permanent dentition realignment.
Clinical Evidence
Clinical case reports were submitted to support the expanded indication for full permanent dentition realignment. No bench or animal testing was required due to identical design and materials as the predicate device.
Technological Characteristics
Customized, thermoformed plastic aligners. Non-sterile. Mechanical force application via appliance geometry. No energy source. No software or connectivity.
Indications for Use
Indicated for correction of dental malocclusion in patients with permanent dentition (all second molars).
Regulatory Classification
Identification
An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.
K173785 — Custom Clear Aligner System · Derby Dental Laboratory · Mar 13, 2018
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a series of interconnected human profiles.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
ClearPath Orthodontics Ltd c/o Patsy J Trisler, JD, RAC Qserve Group US, Inc. 5600 Wisconsin Avenue #509 Chevy Chase, Maryland 20815
Re: K162609
Trade/Device Name: ClearPath Aligner Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: Class II Product Code: NXC Dated: May 30, 2017 Received: May 30, 2017
Dear Patsy J Trisler, JD, RAC:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Isting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
July 6, 2017
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Mary S. Runner -A
Lori A. Wiggins, MPT, CLT Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Change Control Table, Change History
## Change Control Table
| Version | Document Author | Document Approver | Date Approved |
|---------|---------------------|---------------------|---------------|
| 1.00 | Name, Title, Office | Name, Title, Office | MM/DD/YYYY |
Complete Change Control Table (all versions) retained in SWIFT Docs.
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K162609
Device Name
ClearPath Aligner
Indications for Use (Describe)
The ClearPath Aligner is a series of clear, lightweight, plastic appliances indicated for the correction of dental malocclusion in patients with permanent dentition (i.e. all second molars). The ClearPath Aligner is intended to position teeth by way of continuous gentle force.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary – K162609
| Submitter Name: | ClearPath Orthodontics Ltd |
|-------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter Address: | 6-N, Main Boulevard, Johar Town, Lahore-Pakistan |
| Phone Number: | +92 42 111 333 276 |
| Contact Person: | Dr. Waqas Wahab, Chief Executive Officer |
| Date Prepared: | 19 September 2016; updated 5 July 2017 |
| Device Trade Name: | ClearPath Aligner |
| Common Name | Clear Braces |
| Classification Name<br>Number<br>Product Code<br>Requlatory Class | Orthodontic Plastic Bracket<br>21 CFR 872.5470<br>NXC<br>2 |
| Predicate Device:<br>Reference Device: | K123514 ClearPath Aligner, ClearPath Orthodontics<br>No reference devices were included |
| Statement of<br>Intended Use | The ClearPath Aligner is a series of clear, lightweight, plastic<br>appliances indicated for the correction of dental malocclusion in<br>patients with permanent dentition (i.e. all second molars). The<br>ClearPath Aligner is intended to position teeth by way of<br>continuous gentle force. |
| Device Description | ClearPath Orthodontics uses clear, thin, thermoformed plastic<br>material for the manufacture of its ClearPath Aligner. The aligners<br>are provided non-sterile and are customized for each patient<br>according to the molds provided by the treating dentist or<br>orthodontist. |
| | Depending on the patient need and the treatment plan, a series of<br>aligners may be used. The duration of use for each aligner is 14<br>days and it is to be worn except during meals. |
| Mechanism of Action | Each ClearPath Aligner exerts gentle force to achieve progressive<br>realignment of the teeth until the final correction has been<br>attained. |
| Summary of<br>Technological<br>Characteristics | ClearPath Orthodontics manufactures the customized aligners<br>based on standard molds sent to the company by the prescribing<br>dentist or orthodontist. The molds are made after the clinician has<br>assessed the patient's teeth, designed a treatment plan, and<br>taken the impressions. |
| | The plastic used for fabrication of the aligners is a commonly used<br>thermoformed plastic, used in many dental appliances. It has a<br>long history of safe and biocompatible use for this purpose. |
{5}------------------------------------------------
#### Performance Testing Bench/Animal Testing
No in vitro or animal testing were required for this product because the identical design and materials compared to the predicate device assure that the preformed aligners are capable of performing similarly to the predicate device for the specified intended use.
### Clinical Evaluations
The reason for this 510(k) was to expand the indication to include realignment of all permanent teeth. The predicate ClearPath Aligner (K123514) is intended "for minor anterior tooth movement".
Clinical case reports were submitted in the 510(k) to demonstrate that the aligners can effectively realign all permanent teeth as indicated.
Comparison to The ClearPath Aligner is identical to the predicate device. The Predicate Device: difference is the original ClearPath Aligner (predicate) had the indication, as noted, that was limited to anterior tooth movement. The data submitted in the 510(k) support the broader indications for use.
### Substantial Equivalence Conclusion
- Thus, based on: the same intended use, ●
urvivalce
onclusion
- . the same design,
- identical materials,
- identical methods used for fabricating the aligners, and
- . submission of clinical reports that demonstrate the device performs as intended,
it can be concluded that the ClearPath Aligner is substantially equivalent to the predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.