The Custom-made Invisible Aligners is indicated for the treatment of tooth malocclusion in patients with permanent dentition (i.e., all second molars). The Custom-made Invisible Aligners positions teeth by way of continuous gentle force.
Device Story
Custom-made Invisible Aligners are a series of clear, thermoformed plastic dental aligners used to progressively reposition teeth. The device is fabricated from a composite multilayer material (TPU+PETG). The system operates by applying continuous gentle force to the dentition, with minor positional changes programmed into each subsequent aligner in the sequence. The device is used in a clinical setting under the direction of a dental practitioner; the patient wears the aligners for approximately 22 hours per day, replacing them every two weeks as prescribed. The output is the physical movement of teeth toward a target position, which benefits the patient by correcting malocclusion. The treatment plan is facilitated by software that processes dental data to design the sequence of aligners.
Clinical Evidence
No clinical data. Bench testing only. Performance testing included thickness, appearance, odor, density, water absorption, dissolution, color stability, tear resistance, and tensile modulus. Manufacturing accuracy was validated across 12 patient cases. Biocompatibility testing (cytotoxicity, irritation, sensitization, sub-chronic systemic toxicity, genotoxicity) confirmed the material is non-toxic and non-sensitizing. Shelf life of 3 years established via accelerated aging.
Technological Characteristics
Material: Multilayer thermoplastic polyurethane (TPU) and polyethylene terephthalate-1, 4-cyclohexanedimethanol ester (PETG). Manufacturing: Thermoforming. Form factor: Clear, thin, removable dental aligners. Biocompatibility: ISO 10993 compliant. Software: Moderate level of concern.
Indications for Use
Indicated for treatment of tooth malocclusion in patients with permanent dentition (all second molars).
Regulatory Classification
Identification
An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.
Invisalign System with Mandibular Advancement Feature (K181739)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. To the right of it is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. The logos are placed side by side.
April 17, 2023
Zhejiang Yinchili Medical Technology Co., Ltd. % Ivy Wang Technical Manager Shanghai Sungo Management Consulting Company Limited 14th Floor, 1500# Central Avenue Shanghai, Shanghai 200122 China
Re: K223454
Trade/Device Name: Custom-made Invisible Aligners Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: Class II Product Code: NXC Dated: January 16, 2023 Received: January 17, 2023
Dear Ivy Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Bobak Shirmohammadi -S
For Michael E. Adjodha, M. ChE., CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K223454
Device Name Custom-made Invisible Aligner
Indications for Use (Describe)
The Custom-made Invisible Aligners is indicated for the treatment of tooth malocclusion in patients with permanent dentition (i.e., all second molars). The Custom-made Invisible Aligners positions teeth by way of continuous gentle force.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) Summary
(As requirement by 21 CFR 807.92)
#### A. Applicant:
Zhejiang Yinchili Medical Technology Co., Ltd. Address: North 4F, No.239 Yatai Road, Nanhu District, Jiaxing, Zhejiang, CHINA Contact Person: Ms. Xinyan Zhang Tel: +86- 15800780940 Email: zhangxinyan@smartee.cn Date of summary prepared: 2023-03-30
Submission Correspondent:
Primary contact: Ms. Ivy Wang Shanghai SUNGO Management Consulting Co., Ltd. Room 1401, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-58817802 Email: haiyu.wang@sungoglobal.com Secondary contact: Mr. Raymond Luo Room 1401, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-68828050 Email: fda.sungo@gmail.com
#### B. Device:
Trade Name: Custom-made Invisible Aligners Common Name: Sequential Aligners Model: YCL-C Regulatory Information Classification Name: Aligner, Sequential Classification: Class II Product code: NXC Regulation Number: 21 CFR 872.5470 Review Panel: Dental
## C. Predicate device:
K203624 (Primary Predicate) Custom-made Invisible Aligners Zhejiang Yinchili Medical Technology Co., Ltd.
Reference device K181739 Invisalign System with Mandibular Advancement Feature Align Technology, Inc.
{4}------------------------------------------------
# K223454
# D. Indications for use of the device:
The Custom-made Invisible Aligners is indicated for the treatment of tooth malocclusion in patients with permanent dentition (i.e., all second molars). The Custom-made Invisible Aligner positions teeth by way of continuous gentle force.
# E. Device Description:
The Custom-made Invisible Aligners System is a series of dental aligners that fabricated of clear, thin thermoformed plastic to progressively reposition the teeth. The product is made from a composite multilayer material with thermoplastic polyurethane and polyethylene terephthalate-1, 4-cyclohexanedimethanol ester (TPU+PETG). Corrective force to reposition the teeth is delivered via minor changes into a position in each subsequent aligner. They are designed to move the teeth to the target position and deliver desired clinical effect.
# F. Comparison to predicate device
The Custom-made Invisible Aligners is substantially equivalent in intended use, indications for use, mode of action, mode of use, design, and manufacturing to the predicate device. Only minor differences exist between the subject product and the predicate, which do not affect the safety or effectiveness of the subject device.
| Table 1: Comparison to Predicate Device | | | | |
|-----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------|
| Device | Subject Device | Predicate Device | Reference Device | Result |
| 510K number | K223454 | K203624 | K181739 | - |
| Model Name | Custom-made Invisible<br>Aligners | Custom-made<br>Invisible Aligners | Invisalign System<br>with Mandibular<br>Advancement<br>Feature | - |
| Classification | Class II Device, NXC (21<br>CFR 872.5470) | Class II Device, NXC<br>(21 CFR 872.5470) | Class II Device, NXC<br>(21 CFR 872.5470) | Same |
| Classification<br>Name | Aligner, Sequential | Aligner, Sequential | Aligner, Sequential | Same |
| Indications for<br>use | The Custom-made<br>Invisible Aligners is<br>indicated for the<br>treatment of tooth<br>malocclusion in<br>patients with<br>permanent dentition<br>(i.e., all second molars).<br>The Custom-made<br>Invisible Aligner<br>positions teeth by way<br>of continuous gentle | The Custom-made<br>Invisible Aligners is<br>indicated for the<br>treatment of tooth<br>malocclusion in<br>patients with<br>permanent dentition<br>(i.e., all second<br>molars). The<br>Custom-made<br>Invisible Aligner<br>positions teeth by way | The Invisalign<br>System is intended<br>for the orthodontic<br>treatment of<br>malocclusion. | Same |
| | force. | of continuous gentle force. | | |
| Mode of<br>Action | Orthodontic tooth<br>movement occurs<br>through forces applied<br>by the device to the<br>dentition as each tooth<br>follows the<br>programmed<br>displacement based on<br>a doctor's prescription. | Orthodontic tooth<br>movement occurs<br>through forces<br>applied by the device<br>to the dentition as<br>each tooth follows the<br>programmed<br>displacement based<br>on a doctor's<br>prescription. | Orthodontic tooth<br>movement occurs<br>through forces<br>applied by the<br>device to the<br>dentition as each<br>tooth follows the<br>programmed<br>displacement<br>based on a doctor's<br>prescription. | Same |
| Anatomical<br>Site of Use | Oral cavity | Oral cavity | Oral cavity | Same |
| Mode of Use | Each aligner is worn by<br>the patient as<br>determined by the<br>treating dental<br>practitioner, generally<br>for 22 hrs/day (or full<br>time except for eating<br>and hygiene) for 2<br>weeks prior to being<br>replaced by the next<br>aligner in sequence.<br>This is repeated for a<br>duration as prescribed<br>by a Dental<br>Professional. | Each aligner is worn by<br>the patient as<br>determined by the<br>treating dental<br>practitioner, generally<br>for 22 hrs/day (or full<br>time except for eating<br>and hygiene) for 2<br>weeks prior to being<br>replaced by the next<br>aligner in sequence.<br>This is repeated for a<br>duration as prescribed<br>by a Dental<br>Professional. | Aligners are worn<br>for approximately<br>1-2 weeks of 20-22<br>hours of wear per<br>day, after which it<br>is replaced by the<br>next stage aligners.<br>This is repeated for<br>duration as<br>prescribed by the<br>Dental Practitioner | Same |
| Application | Removable | Removable | Removable | Same |
| Raw Material<br>Used | Multilayer<br>(Thermoplastic<br>polyurethane) + PETG<br>(Polyethylene<br>terephthalate-1,<br>4-cyclohexanedimethan<br>ol ester) | TPU Thermoplastic<br>copolyester<br>(polyethylene<br>terephthalate-ethylen<br>e glycol copolyester) | The Invisalign<br>System uses either:<br>1. Multilayer<br>aromatic<br>thermoplastic<br>polyurethane<br>/copolyester.<br>or<br>2. thermoformed<br>polyurethane | Same<br>with the<br>reference<br>device. |
| Method of<br>Manufacturing | Thermoforming | Thermoforming | Thermoforming | Same |
| OTC or Rx | Rx | Rx | Rx | Same |
| Sterile | No | No | No | Same |
| Biocompatibility | In compliance with ISO 10993, tests including<br>Cytotoxicity<br>Oral Mucosa Irritation<br>Sensitization<br>Sub chronic systemic toxicity<br>Genotoxicity | In compliance with ISO 10993, tests including<br>Cytotoxicity<br>Oral Mucosa Irritation<br>Sensitization | In compliance with ISO 10993 | Similar.<br>Both are in compliance with ISO 10993 |
| Design | Image: Clear aligner | Image: Clear aligner | Not available | Similar.<br>Both are transparent plastic films. |
Table 1 provides a comparison of the subject and predicate device.
{5}------------------------------------------------
{6}------------------------------------------------
K223454
#### G. Non-clinical Test
#### 1) Performance Testing
Bench testing has demonstrated that the device is in compliance with pertinent standards and specifications, the expectations of the dental community and the product labeling. Performance testing was performed to the subject device including thickness, appearance, odor, density, water absorption, dissolution, color stability, tear resistance, Tensile modulus of elasticity to demonstrate its effectiveness.
## Manufacturing validation accuracy testing
Manufacturing accuracy validation were conducted to the Custom-made Invisible Aligners. Aligners from 12 different patient case were evaluated at the beginning, middle and end throughout the sequence. The accuracy of 3D molding and aligner molding are checked and meet the pre-established specification. The suitability, function and form of the aligner were checked and comparing it to the treatment design in the software, and the results were complied with the pre-established specifications and acceptance criteria.
#### Shelf life – 3 years
A 3-year shelf life was determined by accelerated aging testing. Performance testing were conducted after 81 days of accelerated aging under 60°C. The test results showed conformity with the pre-established specifications and acceptance criteria.
## 2) Biocompatibility Testing
The biocompatibility evaluation for the device was conducted in accordance with "Use of International Standard ISO 10993-1, Biological evaluation and testing within a risk management process -- Guidance for Industry and Food and Drug Administration Staff" as recognized by FDA. The aligner is considered mucosal membrane contacting for a duration of greater than 30 days. The testing included the following tests:
- Cytotoxicity ●
- Irritation
{7}------------------------------------------------
# K223454
- Sensitization
- Sub chronic systemic toxicity
- Genotoxicity
The results of the testing met the requirements of the study protocols and the material is considered non-toxic, non-sensitizing and is not an intracutaneous irritant. The results of the studies further support a determination of substantial equivalence.
3) Software Verification and Validation Testing
Software verification and validation testing were conducted on the software that facilitates ordering and processing of the Custom-made Invisible Aligner to support that the device is as safe and effective as the predicates. Documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level of concern.
#### H. Clinical Test Conclusion
No clinical study is included in this submission.
#### l. Conclusion
The conclusions drawn from the nonclinical tests demonstrate that the subject device is as safe, as effective, and performs as well as or better than the legally marketed predicate device K203624.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.