ProSomnus EVO [PH] Sleep and Snore Device, ProSomnus EVO [PH] Sleep and Snore Device with Patient Monitoring
Applicant
Prosomnus Sleep Technologies
Product Code
LRK · Dental
Decision Date
Oct 6, 2022
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 872.5570
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ProSomnus® EVO™ [PH] Sleep and Snore Device is intended to reduce night time snoring and mild to moderate obstructive sleep apnea (OSA) in adults. Optionally, if the DentiTrac micro-recorder is completely embedded in the ProSomnus® EVO™ [PH] Sleep and Snore Device, the micro-recorder is intended to measure patient compliance therapy in combination with the DentiTrac System.
Device Story
ProSomnus EVO [PH] is a custom-fabricated, CAD/CAM-milled mandibular advancement device (MAD) consisting of upper and lower arches connected by a Herbst-style telescopic advancement mechanism. Used by patients at home or in sleep labs to treat snoring and mild-to-moderate OSA by holding the mandible forward, increasing pharyngeal space. The device allows for jaw movement (side-to-side, opening/closing). Physicians prescribe the device; advancements are adjusted by the patient via a hex nut or pinhole mechanism. Optionally, a DentiTrac micro-recorder can be embedded to monitor patient compliance with therapy. The device is non-sterile and made of medical-grade polymer and stainless steel.
Clinical Evidence
Bench testing only. Performance testing included mechanical strength (button hook, shear, pull tests), biocompatibility (ISO 10993-1, -5, -10, -11, -12, -23), and physical property testing (ASTM D638, D790, D695, D256, D785, D648, D570). Packaging distribution testing followed ASTM D4169.
Indicated for adults with mild to moderate obstructive sleep apnea (OSA) and snoring. Optional use of embedded DentiTrac micro-recorder for measuring patient compliance with oral appliance therapy.
Regulatory Classification
Identification
Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea are devices that are worn during sleep to reduce the incidence of snoring and to treat obstructive sleep apnea. The devices are designed to increase the patency of the airway and to decrease air turbulence and airway obstruction. The classification includes palatal lifting devices, tongue retaining devices, and mandibular repositioning devices.
Special Controls
*Classification.* Class II (special controls). The special control for these devices is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA.”
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 6, 2022
ProSomnus Sleep Technologies Divya Mavalli Director of Regulatory & Quality 5860 W Las Positas Blvd. Suite 25 Pleasanton, California 94588
Re: K221889
Trade/Device Name: ProSomnus EVO [PH] Sleep and Snore Device, ProSomnus EVO [PH] Sleep and Snore Device with Patient Monitoring Regulation Number: 21 CFR 21 CFR 872.5570 Regulation Name: Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea Regulatory Class: Class II Product Code: LRK, LQZ, PLC Dated: August 26, 2022 Received: August 29, 2022
Dear Divya Mavalli:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For Michael E. Adjodha, M. ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K221889
#### Device Name
ProSomnus® EVO™ [PH] Sleep and Snore Device ProSomnus® EVO™ [PH] Sleep and Snore Device with Patient Monitoring
#### Indications for Use (Describe)
The ProSomnus® EVO™ [PH] Sleep and Snore Device is intended to reduce night time snoring and mild to moderate obstructive sleep apnea (OSA) in adults.
Optionally, if the DentiTrac micro-recorder is completely embedded in the ProSomnus® EVO™ [PH] Sleep and Snore Device, the micro-recorder is intended to measure patient compliance therapy in combination with the DentiTrac System.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for ProSomnus Sleep Technologies. The logo consists of a blue circular graphic on the left, followed by the text "ProSomnus" in a larger, bolder blue font. Below "ProSomnus" is the text "SLEEP TECHNOLOGIES" in a smaller, lighter blue font. The overall design is clean and professional, suggesting a focus on sleep-related technology and solutions.
# 510(K) SUMMARY - K221889
### 1. Manufacturer and Submitter
| Company Name: | ProSomnus Sleep Technology, Inc. |
|------------------|-----------------------------------------------------------------|
| Company Address: | 5860 West Las Positas Blvd.<br>Suite 25<br>Pleasanton, CA 94588 |
| Contact Person: | Divya Mavalli |
- Email: dmavalli@prosomnus.com Phone: 925.307.5337
- Date Prepared: August 30, 2022
#### 2. Device Name and Classification
| Trade/Proprietary Name: | ProSomnus® EVO™ [PH] Sleep and Snore Device |
|-----------------------------|--------------------------------------------------------------------------------------------------------------------|
| | ProSomnus® EVO™ [PH] Sleep and Snore Device with Patient<br>Monitoring |
| Common Name: | Mandibular Advancement Device |
| Classification Name:<br>and | Intraoral devices for snoring and intraoral devices for snoring<br>obstructive sleep apnea. |
| Regulatory Class: | II (special controls) |
| Regulatory Number: | 21 CFR 872.5570 |
| Product Code: | LRK (Device, Anti-snoring), LQZ (Device, Jaw Repositioning), and<br>PLC (Sleep Appliances with Patient Monitoring) |
| Panel: | Dental |
### 3. Predicate and Reference Devices
| Device Names: | Kava and Kava with Herbst<br>Respire Pink Series with DentiTrac |
|-----------------|-----------------------------------------------------------------|
| 510(k) Numbers: | K182661<br>K170692 |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for ProSomnus Sleep Technologies. The logo consists of a blue circular graphic on the left, followed by the company name "ProSomnus" in a larger, bolder blue font. Below "ProSomnus", the words "SLEEP TECHNOLOGIES" are written in a smaller, lighter blue font. The overall design is clean and professional, suggesting a focus on technology and sleep-related products or services.
#### 4. Device Description
The ProSomnus® EVO™ [PH] Sleep and Snore Device consists of maxillary and mandibular device arches that when interfaced together reduce snoring and mild to moderate obstructive sleep apnea by holding the mandible forward during sleep, providing increased pharyngeal space. These separate upper and lower device arches are connected with the herbst advancer. The device is designed and manufactured using Computer Aided Design/Computer Aided Manufacturing (CAD/CAM) dental technologies, which allows the delivery of a well-fitting, aesthetic, and durable intraoral device for the patient. These devices are digitally milled to be patient-specific according to physician prescription. Prescribed advancements can be achieved by simply rotating the telescopic advancer hex nut or pinhole mechanism in an upward direction to increase the forward positioning of the mandible by smaller increments and according to the physician schedule. Advancements up to 12 mm past the initial bite position may be achieved with additional arches as prescribed by the physician The patient will be able to move the lower jaw forward, side-to-side, and open and close the mouth while wearing the device. The proposed device materials include medical grade polymer, stainless steel herbst arms, and biocompatible adhesive. The device is supplied nonsterile.
#### 5. Intended Use/Indication for Use
The ProSomnus® EVO™ [PH] Sleep and Snore Device is intended to reduce night time snoring and mild to moderate obstructive sleep apnea (OSA) in adults.
Optionally, if the DentiTrac micro-recorder is completely embedded in the ProSomnus® EVO™ [PH] Sleep and Snore Device, the micro-recorder is intended to measure patient compliance to oral device/appliance therapy in combination with the DentiTrac System.
#### 6. Non-Clinical Performance Data
Product performance testing was performed to support the above indications for use as well as the claim of Substantial Equivalence.
Bench testing of the ProSomnus® EVO™ [PH] Sleep and Snore Device was conducted in accordance with ProSomnus Sleep Technologies' risk analysis and all applicable FDA guidance documents and international standards, including:
ISO 10993-1, Biological evaluation of medical device – Part 1 Evaluation and testing within a risk management process
ISO 10993-5 - Biological evaluation of medical device - Part 5 Tests for in vitro cytoxicity
ISO 10993-10 – Biological evaluation of medical device – Part 10 Tests irritation and skin sensitization
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for ProSomnus Sleep Technologies. The logo consists of a blue circular design on the left and the company name on the right. The circular design is made up of curved lines that create a sense of movement. The text "ProSomnus" is in a larger, bolder font, while "Sleep Technologies" is in a smaller font below it.
ISO 10993-11 – Biological evaluation of medical devices - Part 11: Tests for systemic toxicity
ISO 10993-12 – Biological evaluation of medical devices - Part 12: Sample preparation and reference materials
ISO 10993- 23 Biological evaluation of medical devices – Part 23: Tests for irritation
ISO 14971 Medical devices - Application of risk management to medical devices
Guidance for industry and FDA Staff: Class II Special Controls Intraoral Devices for Snoring and/or Obstructive Sleep Apnea
Bench testing was conducted on finished devices, unless otherwise specified. The following testing was completed:
- . Metal-Free Button Hook Strength Testing
- . Shear Testing
- . Loctite 3922 Cure Time Test
- . Polymer Adhesive Pull Test
### Packaging
Packaging distribution testing was performed following ASTM D4169 procedure, following Distribution Cycle 13 at Assurance Level II.
#### Biocompatibility
Biocompatibility testing, per ISO 10993-1 was performed, which included the following tests: Cytotoxicity, Sensitization, and Irritation
#### Physical Properties Testing
| Modulus of elasticity (tensile test) | ASTM D 638 |
|---------------------------------------|------------|
| Tensile strength at yield | ASTM D 638 |
| Elongation at break (tensile test) | ASTM D 638 |
| Flexural strength | ASTM D 790 |
| Modulus of elasticity (flexural test) | ASTM D 790 |
| Compression strength | ASTM D 695 |
| Compression modulus | ASTM D 790 |
| Notched impact strength (Izod) | ASTM D 256 |
| Rockwell hardness | ASTM D 785 |
| Deflection temperature | ASTM D 648 |
| Moisture absorption | ASTM D 570 |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for ProSomnus Sleep Technologies. The logo consists of a blue circular graphic on the left and the company name on the right. The words "SLEEP TECHNOLOGIES" are written in a smaller font size below the company name.
## 7. Comparison of Technological Characteristics
The ProSomnus® EVO™ [PH] Sleep and Snore Device has the same intended use as the predicate and reference devices. The materials, design, and performance characteristics are similar to the predicate and reference devices.
The below table provides a comparison of technological characteristics to predicate and reference devices.
| Characteristics | ProSomnus®<br>EVOTM [PH] Sleep<br>and Snore Device<br>(Subject Device) | Kava and Kava with<br>Herbst<br>(Predicate Device) | Respire Pink Series<br>with DentiTrac<br>(Reference Device) |
|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Body Material | Medical Grade<br>Polymer | Polymethyl<br>methacrylate<br>(PMMA) | Polymethyl<br>methacrylate<br>(PMMA) |
| Usability | Single-patient,<br>multiple use | Single-patient,<br>multiple use | Single-patient,<br>multiple use |
| Splint | Comprised of two<br>customizable<br>splints (upper and<br>lower) | Comprised of two<br>splints (upper and<br>lower) | Comprised of two<br>splints (upper and<br>lower) |
| Splints Jointure | Herbst style<br>adjustable medical<br>grade stainless<br>steel advancement<br>mechanism | Adjustable metal<br>screws or lugs | Surgical grade<br>stainless steel<br>herbst hardware |
| Technical Method | Advances the<br>patient's mandible<br>in a forward<br>position to increase<br>airway space.<br>Optionally,<br>monitors patient's<br>compliance to the<br>oral appliance<br>therapy. | Advances the<br>patient's mandible<br>in a forward position<br>to maintain an open<br>airway. | Advances the<br>patient's mandible<br>in a forward<br>position to<br>maintain an open<br>airway. Optionally,<br>monitors the<br>patient compliance<br>to oral appliance<br>therapy. |
| Embed Micro-<br>Recorder | Optional | No | Optional |
- 8. Substantial Equivalence
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for ProSomnus Sleep Technologies. The logo consists of a blue circular design on the left and the words "ProSomnus" and "SLEEP TECHNOLOGIES" on the right. The word "ProSomnus" is in a larger font than the words "SLEEP TECHNOLOGIES". The logo is simple and modern.
| Substantially | ProSomnus® EVO™<br>[PH] Sleep and Snore<br>Device (Subject Device) | Kava and Kava with<br>Herbst<br>(Predicate Device) | Respire Pink Series<br>with DentiTrac<br>(Reference Device) | Indications for<br>Use | The ProSomnus®<br>EVOTM [PH] Sleep and<br>Snore Device is<br>intended to reduce<br>night time snoring and<br>mild to moderate<br>obstructive sleep apnea<br>(OSA) in adults.<br><br>Optionally, if the<br>DentiTrac micro-<br>recorder is completely<br>embedded in the<br>ProSomnus® EVOTM<br>[PH] Sleep and Snore<br>Device, the micro-<br>recorder is intended to<br>measure patient<br>compliance to oral<br>device/appliance<br>therapy in combination<br>with the DentiTrac<br>System. | The Kava and Kava<br>with Herbst device(s)<br>are intended to reduce<br>nighttime snoring and<br>mild to moderate<br>obstructive sleep<br>apnea (OSA) in adults. | The Respire Pink<br>Series intraoral<br>appliances are<br>intended to treat<br>snoring and mild to<br>moderate<br>Obstructive Sleep<br>Apnea (OSA) in<br>adult patients 18<br>years of age or<br>older.<br><br>Optionally, the<br>DentiTrac micro-<br>recorder may be<br>incorporated into a<br>Respire Pink Series<br>device. The micro-<br>recorder is<br>intended to<br>measure patient<br>compliance to oral<br>appliance therapy<br>in combination<br>with the DentiTrac<br>system. |
|--------------------------|--------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------|-----------------------------------------------------------------------------|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Equivalent<br>Connection | | | | Usability | Single patient, multiple<br>use | Single patient, multiple<br>use | Single patient,<br>multiple use |
| 510(k) Number | K221889 | K182661 | K170692 | Target<br>Population | Adult patients | Adult patients | Adult patients |
| Manufacturer | ProSomnus Sleep<br>Technologies | Sketchpad Innovations<br>LLC | Respire Medical<br>Holding | Where Used | For personal nighttime<br>use at home or in sleep<br>laboratories | For personal use | For personal use |
| Relation | Subject Device | Predicate Device | Reference Device<br>with Patient<br>Monitoring | Biocompatible | Yes | Yes | Yes |
| Device | • Device, Anti-Snoring<br>• Device, Jaw<br>Repositioning<br>• Sleep Appliances<br>with Patient<br>Monitoring | • Device, Anti-<br>Snoring<br>• Device, Jaw<br>Repositioning | Sleep Appliances<br>with Patient<br>Monitoring | Anatomical Site | Oral Cavity | Oral Cavity | Oral Cavity |
| Product Code | LRK<br>LQZ<br>PLC | LRK<br>LQZ | PLC | Prescription | Yes | Yes | Yes |
| Device<br>Classification | Class II | Class II | Class II | Principle of<br>Operation | Mandibular<br>Repositioners | Mandibular<br>Repositioners | Mandibular<br>Repositioners |
| Regulation<br>Number | 21 CFR 872.5570 | 21 CFR 872.5570 | 21 CFR 872.5570 | Non-Sterile | Yes | Yes | Yes |
| Design | Allows mandibular<br>movement and to open<br>and close mouth during<br>wear | Allows mandibular<br>movement and to open<br>and close mouth<br>during wear | Allows mandibular<br>movement and to<br>open and close<br>mouth during wear | | | | |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for ProSomnus Sleep Technologies. The logo consists of a blue circular design on the left and the words "ProSomnus" and "SLEEP TECHNOLOGIES" on the right. The word "ProSomnus" is in a larger font than "SLEEP TECHNOLOGIES". The color of the text is also blue.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for ProSomnus Sleep Technologies. The logo consists of a blue circular design on the left and the company name on the right. The words "ProSomnus" are in a larger font than the words "Sleep Technologies", which are stacked below the first word.
| Adjustment<br>Mechanism | Herbst style<br>advancement | Herbst style<br>advancement | Herbst style<br>advancement |
|-------------------------|-----------------------------|-----------------------------|-----------------------------|
| Clean | Daily | Daily | Daily |
### 9. Conclusion
The ProSomnus® EVO™ [PH] Sleep and Snore Device is substantially equivalent to the predicate device with respect to safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.