IS3

K180527 · Hiossen, Inc. · EKX · May 16, 2018 · Dental

Device Facts

Record IDK180527
Device NameIS3
ApplicantHiossen, Inc.
Product CodeEKX · Dental
Decision DateMay 16, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.4200
Device ClassClass 1

Intended Use

IS3 is indicated for use in measuring the stability of dental implants in the oral cavity and the maxillofacial region.

Device Story

IS3 is a hand-held, battery-operated dental implant stability analyzer used by dentists in clinics or hospitals. It measures implant stability via a 'Multi Peg' pin attached to the implant. The device emits short magnetic pulses to induce vibration in the pin; it then samples the resulting alternating magnetic field to determine the resonance frequency. This frequency is converted into an 'ISQ value' (1-100), where higher values indicate greater stability. The measurement process takes approximately one second. The device features a simple interface with one operating key and two LED displays for output. It lacks data memory, timers, or computer connectivity to minimize user error during patient sessions. The device provides clinicians with objective stability data to assist in clinical decision-making regarding implant treatment.

Clinical Evidence

Bench testing only. The device was validated for EMC (IEC 60601-1-2), software (FDA guidance), sterilization (ISO 17665-1/2), and biocompatibility (ISO 10993-1). No clinical data was required or provided.

Technological Characteristics

Hand-held, battery-driven (NiMh). Materials: PC/ABS (body), PEEK (tip, USP VI), Silicone (seal/key, USP VI), Stainless steel (driver, ASTM F899), Titanium grade 5 (peg, ASTM F136). Sensing: Magnetic pulse induction and resonance frequency analysis. Output: 2-digit LED display. Connectivity: None. Sterilization: Not specified (implied reusable/disposable components). Software: Microcontroller-based, non-networked.

Indications for Use

Indicated for measuring dental implant stability in the oral cavity and maxillofacial region for patients with one or more dental implants.

Regulatory Classification

Identification

A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym followed by the full name of the agency, "U.S. Food & Drug Administration," on the right. The FDA part of the logo is in blue, with the acronym in a solid blue square and the agency's name in a lighter shade of blue. Hiossen, Inc. David Kim RA Manager 85 Ben Fairless Dr. Fairless Hills, Pennsylvania 19030 May 16, 2018 Re: K180527 Trade/Device Name: IS3 Regulation Number: 21 CFR 872.4200 Regulation Name: Dental Handpiece And Accessories Regulatory Class: Class I Product Code: EKX Dated: February 28, 2018 Received: February 28, 2018 Dear David Kim: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); {1}------------------------------------------------ and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Mary S. Runner -S For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K180527 Device Name IS3 Indications for Use (Describe) IS3 is indicated for use in measuring the stability of dental implants in the oral cavity and the maxillofacial region. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for HIOSSEN. The logo is in two colors, red and gray. The letters "HI" are in red, and the rest of the word "OSSEN" is in gray. There is a red swoosh above the "O" in "OSSEN". 85 Ben Fairless Dr. Fairless Hills, PA 19030 Tel : 1-888-678-0001 / Fax : 1-267-759-7004 www.hiossen.com # 510(k) Summarv This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92. Date : Feb 26, 2018 ### 1. Company and Correspondent making the submission | 1) Submitter's Name : | HIOSSEN, Inc. | |-----------------------|--------------------------------------------------| | 2) Address : | 85 Ben Fairless Dr. Fairless Hills, PA 19030 USA | | 3) Telephone No. | 888 678 0001 | | 4) Contact : | Mr. David Kim | ## 2. Device | Trade or (Proprietary) Name : | IS3 | |-------------------------------|-----------------------------------| | Common or usual name : | Dental implant stability analyzer | | Accessories name : | Mult-Peg | | Regulation number : | 872.4200 | | Classification code : | EKX | | Device class : | Class I | # 3. Predicate Device Primary Predicate Device : Tellos ISQ Buddy ( K143445 ) Reference Predicate Device : Osstell ISQ ( K082523 ) # 4. Device Description : - 4.1. Description : IS3 measurement system consists of: IS3 Instrument Hand-held instrument | | Multi Peg Driver | Driver to attach the ISQ peg to the implant | |--|------------------|--------------------------------------------------------------------------------------------------------| | | IS3 Charger | 100-240 VAC to 5VDC charger for the instrument batteries | | | Multi Peg | Measurement pin to attach to the implant. Different pins are available to fit different implant types. | SECTION 008 {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for Hiossen Implant. The logo is composed of the word "Hiossen" in a stylized font. The "Hi" portion of the word is in red, while the "ossen" portion is in gray. The "H" in "Hi" is connected to a curved line above the "i", adding a unique design element to the logo. 85 Ben Fairless Dr. Fairless Hills. PA 19030 Tel : 1-888-678-0001 / Fax : 1-267-759-7004 www.hiossen.com #### 4.2. Basic principle and functionality The IS3 instrument is a hand-held, battery-driven device for measuring the relative stability of a dental implant. A small pin "Multi peg" is attached to the dental implant by a screw-connection, with the Multi Peg Driver. The pin has a small magnet incorporated into its top. The instrument is held towards the Multi Peg, and sends short magnetic pulses that bring the pin into vibration. After a pulse has been sent, the instrument measures the vibration by sampling the signal from the alternating magnetic field that follows from the vibrating pin. The frequency of the signal is determined by the instrument and is presented as an "ISQvalue", 1 to 100, where a higher number means higher stability. The measurement takes about 1 second. #### 4.3. Technical detail The instrument consists of a microcontroller and circuits to send the magnetic pulses, to receive the measurement signal, and to present the measurement value. A circuit for battery-charging is also included. Two 2-digit LED displays, one on each side of the instrument, show the measurement value and also communicate possible error codes and software id number at start-up. One operating key is used to turn the instrument on and off. The electronics are contained in a plastic body, which is sealed except for the back of the instrument which has a charging connector. The plastic body is made from PC/ABS plastic except for the tip, which is made from PEEK. No part of the instrument is intended to contact the patient; however, there can be unintentional contact with the tip of the instrument. The instrument is battery-driven and contains re-chargeable NiMh-batteries. The batteries can be charged during use, but it should not be attached while measuring due to the risk of power line interference making it impossible to measure. For safety, a charger complying with IEC 60601-1 is used.The charger connector is of a type that does not allow other chargers to be connected, thereby eliminating the risk of the wrong charger being used. ### 4.4. Material used Instrument body: PC/ABS Instrument tip: PEEK (USP VI) Instrument seal : Silicone (USP VI) Instrument key : Silicone (USP VI) Multi Peg Driver: Stainless steel, ASTM F899 Multi Peg: Titanium grade 5, ASTM F136 ### 5. Indication for Use IS3 is indicated for use in measuring the stability of dental implants in the oral cavity and the maxillofacial region. ### 6. Comparison and differences to the Predicate Device The instrument functions the same way as the predicate device regarding the measurement principle and technology used. There is no difference in between the subject device and primary predicate device. The differences with reference device are in the IS3(subject device) and Tellos(primary predicate device) are hand-held instruments while the Osstell ISO(reference {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for Hiossen Implant. The logo is composed of the word "Hiossen" in a stylized font. The "H" and "I" are in red, with a red swoosh above the "H". The rest of the word, "ossen", is in gray. SECTION 008 85 Ben Fairless Dr. Fairless Hills, PA 19030 Tel : 1-888-678-0001 / Fax : 1-267-759-7004 www.hiossen.com predicate device) consists of two parts; the instrument and the measurement probe. The IS3(subject device) and Tellos(primary predicate device) have no memory to store the measurement values; the reference predicate device has a memory that stores the measurements and also has an inbuilt clock timer, and a computer connection for transferring measurement values. IS3 and Primary predicate device has no timer or computer connection. The lack of these features (memory, timer, data transfer) is not considered to be an inconvenience for the user. Instead, it decreases the risk of mixing up values with each other. | | Proposed device | Primary<br>Predicate devices | Reference<br>Predicate device | |--------------------------------|--------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | IS3 | Tellos ISQ Buddy<br>(k143445) | Osstell ISQ<br>(k082523) | | Manufacturer | HIOSSEN, INC. | Tellos Medical AB | Osstell AB | | Product Code<br>/Class | EKX / Class I | EKX / Class I | EKX / Class I | | Regulation<br>Number | 872.4200 | 872.4200 | 872.4200 | | Indication<br>for Use | Same as primary<br>predicate device. | The device is indicated for use in<br>measuring the stability of<br>implants in the oral cavity and<br>the maxillofacial region. | The device is indicated for use in<br>measuring the stability of<br>implants in the oral cavity and<br>craniofacial region. | | Target<br>Population | Same as primary<br>predicate device. | Patients with one or more<br>dental implants | Patients with one or more<br>dental implants | | Anatomical<br>site | Same as primary<br>predicate device. | Oral cavity or<br>craniofacial region | Oral cavity or<br>craniofacial region | | Usage<br>location | Same as primary<br>predicate device. | Dentists office or hospital | Dentists office or hospital | | Energy<br>used/delivered | Same as primary<br>predicate device. | No energy is intended to be<br>delivered. A small amount of<br>energy could reach the dental<br>implant from the self-vibrating<br>measurement pin. | No energy is intended to be<br>delivered. A small amount of<br>energy could reach the dental<br>implant from the self-vibrating<br>measurement pin. | | Technology | Same as primary<br>predicate device. | A microcontroller sends electric<br>pulses to a coil in the instrument<br>tip. As a consequence, magnetic<br>pulses are emitted that affect the<br>pin connected to the implant.<br>The pin then starts to vibrate<br>with its resonance frequency.<br>Vibration creates an alternating | A microcontroller sends electric<br>pulses to a coil in the instrument<br>tip. As a consequence, magnetic<br>pulses are emitted that affect the<br>pin connected to the implant.<br>The pin then starts to vibrate<br>with its resonance frequency.<br>Vibration creates an alternating | | | | magnetic field which is being | magnetic field which is being | | | | picked up by another coil in the | picked up by another coil in the | | | | instrument tip. The electrical | instrument tip. The electrical | | | | signal from the receiving coil is | signal from the receiving coil is | | | | analyzed and the frequency is | analyzed and the frequency is | | | | determined. The Frequency is | determined. The Frequency is | | | | presented on the display as an | presented on the display as an | | | | "ISQvalue" | "ISQvalue" | | Measurement<br>output | Same as primary<br>predicate device. | It presents the resonance<br>frequency of the ISQ peg as an<br>ISQ number, 1-100. The ISQ<br>number is calculated from the<br>resonance frequency. | It presents the resonance<br>frequency of the ISQ peg as an<br>ISQ number, 1-100. The ISQ<br>number is calculated from the<br>resonance frequency. | | | | The operating key is pushed and<br>the instrument is held towards<br>the attached ISQ peg. The<br>instrument then presents the ISQ<br>value. | Any key is pushed and the<br>measurement probe is held<br>towards the attached SmartPeg.<br>The instrument then presents the<br>ISQ value. | | Operation | Same as primary<br>predicate device. | | | | System<br>components | Same as primary<br>predicate device. | Tellos ISQ Buddy instrument,<br>ISQ Peg, ISQ Peg Driver and<br>instrument charger. | Osstell ISQ instrument,<br>measurement probe, SmartPeg<br>and instrument charger. | | Power, weight<br>& size | Same as primary<br>predicate device. | 3VA, 0.1 Kg, 200 x 30 mm | 8 VA, 0.4 Kg, 195 x 20 x 45 mm | | Materials<br>used | Same as primary<br>predicate device. | Instrument; PC/ABS and PEEK.<br>Gasket and operating key:<br>Silicone<br>ISQ Peg Driver: Stainless steel<br>ISQ Peg: Titanium | Instrument: Not available<br>Measurement probe: Not<br>available<br>SmartPeg driver: Not Available<br>SmartPeg: Aluminum | | Instrument<br>memory | Same as primary<br>predicate device. | Tellos ISQ Buddy has no data<br>memory.The instrument displays<br>the measurement value after the<br>measurement. The memory is<br>not a necessary feature and the<br>absence of a memory decreases<br>the risk of mixing up values<br>between different implants<br>during a patient session. | Osstell ISQ has a memory for<br>storing measurement values. | | Instrument<br>setup-menu | Same as primary<br>predicate device. | No setup-menu is needed for<br>Tellos ISQ Buddy. The<br>instrument has one operating key<br>and no parameters to change. | Instrument has a setup-menu for<br>various instrument parameters<br>such as display contrast and<br>computer connection. | | Data transfer | Same as primary<br>predicate device | Since Tellos ISQ Buddy has no<br>memory, there are no data to<br>transfer. This is not a necessary<br>feature, and the absence of | Osstell ISQ has a cable<br>attachment to a PC computer,<br>for transferring data. | | Clock Timer | Same as primary<br>predicate device | the risk of mixing up values<br>between different implants<br>during a patient session.<br>Since Tellos ISQ Buddy has no<br>memory, there is no need for a<br>clock timer. | Osstell ISQ has a built-in clock<br>timer. | | Number of<br>operating<br>keys | Same as primary<br>predicate device | Tellos ISQ Buddy has one<br>operating key, used for turning<br>on and off the instrument. Since<br>the instrument has no memory or<br>setup-menu, not more than one<br>key is needed. With one key,the<br>instrument is more intuitive and<br>easier to use, which leaves less<br>room for user errors compared to<br>the predicate device. | Osstell ISQ has 5 operating keys,<br>for working the setup-menu and<br>for navigating the instrument<br>memory. | | Display | Same as primary<br>predicate device | Tellos ISQ Buddy has two LED-<br>displays; one on each side of the<br>instrument for easy reading. | Osstell ISQ has one LCD display<br>on the instrument. | | Measurement<br>pins | Same as primary<br>predicate device | Tellos ISQ Buddy uses the<br>Osstell SmartPegs, or<br>corresponding Tellos pins, “ISQ<br>Pegs” from titanium. | Osstell ISQ uses disposable<br>aluminum pins, “SmartPegs”. | | Electrical<br>safety | Same as primary<br>predicate device | It is designed to the standard<br>IEC 60601-1 Medical Electrical | Osstell ISQ is designed to the<br>standard IEC 60601-1 Medical | {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the logo for Hiossen Implant. The logo features the word "Hiossen" in a stylized font. The "H" and "I" are in red, while the rest of the word is in gray. There is a red swoosh above the "H" and "I". Hiossen Inc. 85 Ben Fairless Dr. Fairless Hills, PA 19030 Tel : 1-888-678-0001 / Fax : 1-267-759-7004 www.hiossen.com {7}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the Hiossen logo. The "HI" part of the logo is in red, while the "OSSEN" part is in gray. There is a red swoosh above the "HI" part of the logo. **Hiossen Inc.** 85 Ben Fairless Dr. Fairless Hills, PA 19030 Tel : 1-888-678-0001 / Fax : 1-267-759-7004 www.hiossen.com ### 7. Performance testing The IS3 instrument brings a pin attached to a dental implant into vibration and then measures the resonance frequency of the vibrating pin. The resonance frequency is then presented as an "ISQ-value" 1-100. The primary predicate device does the same technological principle and the electronic, software since same circuits are used. Equipment The IS3 instrument has been tested according to, and found to comply with the following: - · EMC standard IEC 60601-1-2 - · Software validation according to FDA guidance for Software Contained in a Medical Device - · Sterilization validation according to ISO 17665-1 and ISO 17665-2 - · Biocompatibility standard ISO 10993-1 SECTION 008 Electrical Equipment {8}------------------------------------------------ Image /page/8/Picture/0 description: The image shows the Hiossen logo. The "HI" is in red, and the "OSSEN" is in gray. The "H" is capitalized, and the "I" is lowercase. The "OSSEN" is all capitalized. There is a red swoosh above the "HI". SECTION 008 85 Ben Fairless Dr. Fairless Hills, PA 19030 Tel : 1-888-678-0001 / Fax : 1-267-759-7004 www.hiossen.com #### 8. Substantial Equivalence Conclusion The purpose of using the instrument is to measure the ISQ value (the relative stability) of the implant. The IS3 has the same indication for use, measurement output and technology as the primary predicate device. The device description, software validation, performing testing, sterilization and biocompatibility assessment demonstrate the substantial equivalence of the submission device to the identified primary predicate device. Therefore, the IS3 can be found substantially equivalent to the Tellos ISQ Buddy.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...