K143445 · Tellos Medical AB · EKX · Jan 7, 2016 · Dental
Device Facts
Record ID
K143445
Device Name
Tellos ISQ Buddy
Applicant
Tellos Medical AB
Product Code
EKX · Dental
Decision Date
Jan 7, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.4200
Device Class
Class 1
Indications for Use
Tellos ISQ Buddy is indicated for use in measuring the stability of dental implants in the oral cavity and the maxillofacial region.
Device Story
Hand-held, battery-operated dental implant stability analyzer; uses magnetic pulse technology to induce vibration in a pin (ISQ Peg) attached to an implant; measures resonance frequency of vibrating pin via alternating magnetic field; calculates and displays stability value (ISQ 1-100) on LED screens; used by dentists in clinics/hospitals; provides objective stability metric to assist clinical decision-making regarding implant loading or treatment progression; benefits patient by ensuring implant stability before restoration.
Clinical Evidence
Bench testing only. Comparison performed using artificial frequency signals (1,000–10,200 Hz) and implants in artificial bone (3–85 ISQ range). Results showed 0 ISQ deviation compared to predicate device. Compliance with IEC 60601-1-2 (EMC), ISO 17665-1/2 (sterilization), and ISO 10993-1 (biocompatibility) confirmed.
Technological Characteristics
Hand-held device; PC/ABS body, PEEK tip, silicone seals. Microcontroller-based electronics; magnetic pulse excitation; inductive signal sensing. Battery-powered (NiMh); 3VA power. Dimensions 200x30mm; weight 0.1kg. Two-digit LED displays. No memory, no data transfer. Sterilization per ISO 17665. Biocompatibility per ISO 10993.
Indications for Use
Indicated for patients with one or more dental implants requiring stability measurement in the oral cavity and maxillofacial region.
Regulatory Classification
Identification
A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name encircling a symbol. The symbol is a stylized representation of three human profiles facing right, with flowing lines suggesting movement or connection. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 7, 2016
Tellos Medical AB Mr. Anders Petersson CEO Nedergardsgatan 5 Goteborg, 41654 SWEDEN
Re: K143445
Trade/Device Name: Tellos ISQ Buddy Regulation Number: 21 CFR 872.4200 Regulation Name: Dental Handpiece and Accessories Regulatory Class: I Product Code: EKX Dated: November 30, 2015 Received: December 2, 2015
Dear Mr. Petersson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina
Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K143445
Device Name
Tellos ISO Buddy
Indications for Use (Describe)
Tellos ISO Buddy is indicated for use in measuring the stability of dental implants in the maxillofacial region.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
# 1. 510(k) Summary
## 1.1 Submitter
Tellos Medical AB Nedergårdsgatan 5 416 54 Göteborg Sweden Ph: +46 709 741269 Email: info@tellosmedical.com Contact person: Anders Petersson This Summary is made on January 2, 2016
# 1.2 Device name
| Proprietary name: | Tellos ISQ Buddy |
|-----------------------------|-----------------------------------|
| Common name: | Dental implant stability analyzer |
| Accessories name: | ISQ Peg |
| Regulation Number | 872.4200 |
| Classification Product Code | EKX |
# 1.3 Identification of Predicate Device
| The predicate device trade name: | Osstell ISQ |
|----------------------------------|-------------|
| Accessories name: | SmartPeg |
| Predicate Device 510(k) information: | |
|--------------------------------------|-------------------------------------|
| Device Classification Name | Handpiece, Direct Drive, AC-Powered |
| 510(k) Number | K082523 |
| Device Name | OSSTELL ISQ IMPLANT STABILITY METER |
| Regulation Number | 872.4200 |
| Classification Product Code | EKX |
# 1.4 Device description
## 1.4.1 Description
The Tellos ISQ Buddy measurement system consists of:
| Tellos ISQ Buddy Instrument | Hand-held instrument |
|-----------------------------|--------------------------------------------------------------------------------------------------------|
| ISQ Peg Driver | Driver to attach the ISQ peg to the implant |
| Tellos ISQ Buddy Charger | 100-240 VAC to 5VDC charger for the instrument batteries |
| ISQ Peg | Measurement pin to attach to the implant. Different pins are available to fit different implant types. |
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image shows a digital thermometer with the words "Tellos Medical" printed on the side. The thermometer is white and has a small digital display that reads "82". The thermometer has a probe at the end that is used to take a temperature reading. The background is a solid gray color.
Tellos ISQ Buddy Instrument
Image /page/4/Picture/4 description: The image shows a cylindrical object with a ribbed pattern on the top half and a smooth surface on the bottom half. The object is light gray and appears to be made of a solid material. The ribbed pattern consists of vertical grooves that run along the circumference of the cylinder. The object is standing upright on a gray surface.
Image /page/4/Picture/5 description: The image shows a dental implant. The implant is made of metal and has a cylindrical shape. The top of the implant has a few grooves, and the bottom has screw threads. The implant is set against a gray background.
ISQ Peg Driver
ISQ Peg
Image /page/4/Picture/8 description: The image shows a collection of black power adapters with different plug types. There are four different plug types visible in the image. The adapters are arranged in a way that highlights the variety of plug configurations. The adapters appear to be interchangeable, allowing the user to use the same power supply in different countries.
Tellos ISQ Buddy Charger
## 1.4.2 Basic principle and functionality
The Tellos ISQ Buddy instrument is a hand-held, battery-driven device for measuring the relative stability of a dental implant.
A small pin "ISQ peg" is attached to the dental implant by a screw-connection, with the Peg Driver. The pin has a small magnet incorporated into its top. The instrument is held towards the ISQ Peg, and sends short magnetic pulses that bring the pin into vibration. After a pulse has been sent, the instrument measures the vibration by sampling the signal from the alternating magnetic field that follows from the vibrating pin. The frequency of the signal is determined by the instrument and is presented as an "ISQvalue", 1 to 100, where a higher number means higher stability. The measurement takes about 1 second.
{5}------------------------------------------------
Image /page/5/Picture/2 description: In the image, there is a gray background with two gray objects. The object on the right is a cylinder with a rounded end and a black ring around the base. The object on the left is a smaller cylinder with a screw thread at the bottom and a notched top.
## Instrument tip held towards an ISQ Peg
## 1.4.3 Technical details
The instrument consists of a microcontroller and circuits to send the magnetic pulses, to receive the measurement signal, and to present the measurement value. A circuit for battery-charging is also included. Two 2-digit LED displays, one on each side of the instrument, show the measurement value and also communicate possible error codes and software id number at start-up. One operating key is used to turn the instrument on and off.
The electronics are contained in a plastic body, which is sealed except for the instrument which has a charging connector. The plastic body is made from PC/ABS plastic except for the tip, which is made from PEEK. No part of the instrument is intended to contact the patient; however there can be unintentional contact with the tip of the instrument.
The instrument is battery-driven and contains re-chargeable NiMh-batteries can be charged during use, but it should not be attached while measuring due to the risk of power line interference making it impossible to measure. For safety, a charger complying with IEC 60601-1 is used. The charger connector is of a type that does not allow other chargers to be connected, thereby eliminating the risk of the wrong charger being used.
#### 1.4.4 Materials used
Instrument body: PC/ABS Instrument tip: PEEK (USP VI) Instrument seal: Silicone (USP VI) Instrument key: Silicone (USP VI) ISQ Peg Driver: Stainless steel, ASTM F899 ISQ Peg: Titanium grade 5, ASTM F136
## 1.5 Indications for Use
Tellos ISQ Buddy is indicated for use in measuring the stability of dental implants in the oral cavity and the maxillofacial region.
{6}------------------------------------------------
# 1.6 Comparison and differences to the Predicate Device
The instrument functions the same way as the predicate device regarding the measurement principle and technology used. The differences are in the handling; Tellos ISQ Buddy is a hand-held instrument while the predicate device consists of two parts; the instrument and the measurement probe.
Tellos ISQ Buddy has no memory to store the measurement values; the predicate device has a memory that stores the measurements. The predicate device also has an inbuilt clock timer, and a computer connection for transferring measurement values. Tellos ISQ Buddy has no timer or computer connection. The lack of these features (memory, timer, data transfer) is not considered to be an inconvenience for the user. Instead, it decreases the risk of mixing up values with each other.
| | 510(k) submission | Predicate Device |
|--------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device name | Tellos ISQ Buddy | Osstell ISQ |
| Company name | Tellos Medical AB | Osstell AB |
| Product Code/Class | EKX / Class I | EKX / Class I |
| Regulation number | 872.4200 | 872.4200 |
| Indications for use | The device is indicated for use in<br>measuring the stability of implants in<br>the oral cavity and the maxillofacial<br>region.<br>The differences in Indications for use<br>are based on clarification of the use of<br>the device with dental implants and the<br>anatomical region appropriate for<br>dental implants. This clarification is<br>within the indications for use of the<br>primary predicate and does not change<br>the intended use of the device. | The device is indicated for use in<br>measuring the stability of implants in<br>the oral cavity and craniofacial region. |
| Target population | Same as predicate device | Patients with one or more dental<br>implants |
| Anatomical site | Same as predicate device | Oral cavity or craniofacial region |
| Usage location | Same as predicate device | Dentists office or hospital |
| Energy used/delivered | Same as predicate device | No energy is intended to be delivered.<br>A small amount of energy could reach<br>the dental implant from the self-<br>vibrating measurement pin. |
| Technology | Same as predicate device | A microcontroller sends electric pulses<br>to a coil in the instrument tip. As a<br>consequence, magnetic pulses are<br>emitted that affect the pin connected<br>to the implant. The pin then starts to<br>vibrate with its resonance frequency.<br>Vibration creates an alternating<br>magnetic field which is being picked up<br>by another coil in the instrument tip.<br>The electrical signal from the receiving<br>coil is analyzed and the frequency is<br>determined. The Frequency is<br>presented on the display as an "ISQ-<br>value" |
| Measurement output | Same as predicate device | Osstell ISQ presents the resonance<br>frequency of the ISQ peg as an ISQ<br>number, 1-100. The ISQ number is |
| | | calculated from the resonance |
| | | frequency. |
| Operation | The operating key is pushed and the | Any key is pushed and the |
| | instrument is held towards the | measurement probe is held towards |
| | attached ISQ peg. The instrument then | the attached SmartPeg. The instrument |
| | presents the ISQ value. | then presents the ISQ value. |
| System components | Tellos ISQ Buddy instrument, ISQ Peg, | Osstell ISQ instrument, measurement |
| | ISQ Peg Driver and instrument charger. | probe, SmartPeg and instrument |
| | | charger. |
| Power, weight & size | 3VA, 0.1 Kg, 200 x 30 mm | 8 VA, 0.4 Kg, 195 x 20 x 45 mm |
| Materials used | Instrument; PC/ABS and PEEK. | Instrument: Not available |
| | Gasket and operating key: Silicone | Measurement probe: Not available |
| | ISQ Peg Driver: Stainless steel | SmartPeg driver: Not Available |
| | ISQ Peg: Titanium | SmartPeg: Aluminum |
| Instrument memory | Tellos ISQ Buddy has no data memory. | Osstell ISQ has a memory for storing |
| | The instrument displays the | measurement values. |
| | measurement value after the | |
| | measurement. The memory is not a | |
| | necessary feature and the absence of a | |
| | memory decreases the risk of mixing up | |
| | values between different implants | |
| | during a patient session. | |
| Instrument setup-menu | No setup-menu is needed for Tellos ISQ | Instrument has a setup-menu for |
| | Buddy. The instrument has one | various instrument parameters such as |
| | operating key and no parameters to | display contrast and computer |
| | change. | connection. |
| Data transfer | Since Tellos ISQ Buddy has no memory, | Osstell ISQ has a cable attachment to a |
| | there are no data to transfer. This is not | PC computer, for transferring data. |
| | a necessary feature, and the absence of | |
| | a data transfer function decreases the | |
| | risk of mixing up values between | |
| | different implants during a patient | |
| | session. | |
| Clock Timer | Since Tellos ISQ Buddy has no memory, | Osstell ISQ has a built-in clock timer. |
| | there is no need for a clock timer. | |
| Number of operating keys | Tellos ISQ Buddy has one operating key, | Osstell ISQ has 5 operating keys, for |
| | used for turning on and off the | working the setup-menu and for |
| | instrument. Since the instrument has | navigating the instrument memory. |
| | no memory or setup-menu, not more | |
| | than one key is needed. With one key, | |
| | the instrument is more intuitive and | |
| | easier to use, which leaves less room | |
| | for user errors compared to the | |
| | predicate device. | |
| Display | Tellos ISQ Buddy has two LED-displays; | Osstell ISQ has one LCD display on the |
| | one on each side of the instrument for | instrument. |
| | easy reading. | |
| Measurement pins | Tellos ISQ Buddy uses the Osstell | Osstell ISQ uses disposable aluminum |
| | SmartPegs, or corresponding Tellos | pins, "SmartPegs". |
| | pins, "ISQ Pegs" from titanium. | |
| Electrical safety | Same as predicate device | Osstell ISQ is designed to the standard |
| | | IEC 60601-1 Medical Electrical |
| | | Fauinment…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.