K181888 · Osstell AB · EKX · Aug 22, 2019 · Dental
Device Facts
Record ID
K181888
Device Name
Osstell Beacon
Applicant
Osstell AB
Product Code
EKX · Dental
Decision Date
Aug 22, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.4200
Device Class
Class 1
Indications for Use
Osstell Beacon is intended for use as a Dental Implant Stability Analyzer
Device Story
Osstell Beacon is a portable, handheld dental implant stability analyzer. It utilizes Resonance Frequency Analysis (RFA) to measure implant stability. A Smartpeg (aluminum rod) is screwed onto the implant; the device emits a magnetic pulse from its integrated measurement tip to excite the Smartpeg. The device detects the resulting resonance frequency via a second coil in the tip. A microcontroller calculates the Implant Stability Quotient (ISQ) value (0-100) from the frequency response. The ISQ value is displayed on the device's OLED screens. Used by professional clinicians in dental practices or operating rooms. The device is battery-operated and non-sterile, requiring a transparent barrier sleeve for use. The output assists clinicians in assessing implant stability over time; higher ISQ values generally indicate increased stability, while lower values may suggest stability loss or potential failure.
Clinical Evidence
No clinical data. Bench testing only.
Technological Characteristics
Portable, handheld instrument; battery-powered (Lithium-ion). Materials: ABS and PC plastic (food grade compliant). Sensing: Resonance Frequency Analysis (RFA) using magnetic excitation and detection coils. Connectivity: Bluetooth 4.0 LE (for production purposes only). Dimensions: 210 x 35 x 25 mm. Weight: 0.07 kg. Sterilization: Not autoclavable; requires transparent barrier sleeve. Software: Embedded firmware for signal processing and ISQ calculation.
Indications for Use
Indicated for use in measuring the stability of implants in the oral cavity and maxillofacial region.
Regulatory Classification
Identification
A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 22, 2019
Osstell AB % Cherita James Regulatory Consultant M Squared Associates, Inc 575 8th Ave Suite 1212 New York, NY 10018
Re: K181888
Trade/Device Name: Osstell Beacon Regulation Number: 21 CFR 872.4200 Regulation Name: Dental Handpiece and Accessories Regulatory Class: Class I Product Code: EKX Dated: July 23, 2019 Received: July 24, 2019
Dear Cherita James:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-devicereporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Srinivas Nandkumar, Ph.D. Acting Assistant Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K181888
Device Name Osstell Beacon
Indications for Use (Describe)
Osstell Beacon is indicated for use in measuring the stability of implants in the oral cavity and maxillofacial region.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
{3}------------------------------------------------
# 510(k) Summary
The following information is provided as required by 21 CFR § 807.87 for the Osstell Beacon 510(k) premarket notification. In response to the Safe Medical Devices Act of 1990, the following is a summary of the information upon which the substantial equivalence determination is based.
| <b>Sponsor:</b> | <b>Osstell AB</b><br>Stampgatan 14<br>411 01 Goteborg<br>Sweden |
|-----------------|-----------------------------------------------------------------|
|-----------------|-----------------------------------------------------------------|
Contact: Cherita James M Squared Associates, Inc. 575 8th Ave, Suite 1212 New York, New York 10018 Ph. 347-954-0624 Fax. 703-562-9797 CJames@MSquaredAssociates.com
| Revision Date: | August 19, 2019 |
|------------------------------------------------------------------------|-------------------------------------------|
| Proprietary Name: | Osstell Beacon |
| Common Name: | Dental implant stability analyzer |
| Regulatory Class: | Class 1 |
| Regulation: | 872.4200 |
| Product Codes: | EKX - handpiece, direct drive, ac-powered |
| Predicate Device: | Osstell ISQ (K082523)-Primary |
| Tellos ISO Buddy (K143445) and Implantmed SI-1015 (K161957)- Reference | |
#### Device Description:
The Osstell Beacon is a modification of the Osstell ISQ (K082523). The system is designed to measure implant stability in the oral cavity and maxillofacial region. Similar to K082523, the Osstell Beacon is a portable, handheld instrument that involves the use of the noninvasive technique, Resonance Frequency Analysis. The system involves the use of a Smartpeg (aluminum rod) attached to the implant by means of a screw. The Smartpeg is excited by a magnetic pulse from the measurement tip on the handheld instrument. The resonance frequency, which is the measure of implant stability, is calculated from the response signal. Results are displayed as the Implant Stability Quotient (ISQ). The ISQ is a measurement of the implant and is
{4}------------------------------------------------
#### K181888
derived from the resonance frequency value obtained from the Smartpeg.
#### Intended Use:
Osstell Beacon is intended for use as a Dental Implant Stability Analyzer
#### Indications for Use:
Osstell Beacon is indicated for use in measuring the stability of implants in the oral cavity and maxillofacial region.
#### Summary of the Technological Characteristics:
The modifications to the Osstell ISQ since its previous clearance in K082523 include the following changes:
- . Enclosure design is smaller to be fully handheld. The measurement probe with tip is integrated into the enclosure and referred to as the measurement tip
- . Not possible to sterilize (autoclave). The device must be used with a transparent barrier sleeve. Commercially available transparent, barrier sleeves are recommended in the Instructions for Use.
- Different plastic material used in the enclosure (all being food grade compliant, 21 CFR §181.32 and 21 CFR §177.1580)
- . Updated user interface to make the measurement procedure even more easy and intuitive
- The device cannot measure while charging
- Bluetooth data communication .
These differences do not affect the substantial equivalence or performance of the device and do not change the intended use of the Osstell Beacon.
{5}------------------------------------------------
#### K181888
#### Summary of the Nonclinical Testing:
Based on the Risk Analysis, the verification and validation testing confirms the Beacon performs as intended. The Osstell Beacon was subjected to the same preclinical requirements and testing as the primary predicate device.
Though the recommendation require use of a transparent barrier sleeve, due to material changes, the device was evaluated in accordance with ISO 10993-1.
Additionally, cleaning and disinfection recommendations were validated in the event the Osstell Beacon should become contaminated due to a damaged barrier sleeve.
Performance testing was conducted to confirm compliance to the design specifications.
Osstell has provided information to support compliance of Beacon with applicable portions of the following standards and FDA Guidance documents:
- FDA Dental Handpieces Premarket Notification [510(k)] Submissions, 2007 ●
- . FDA Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling, 2015
- . FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, 2005
- . ISO 10993-1
- . ISO 10993-5
- . AAMI TIR12:2010
- AAMI TIR30:2011
- . ASTM E1837-96 (2014)
- ISO 17665-1
- ISO 17664
- IEC 60601-1, Ed 3 2005 ●
- . IEC 60601-1-2:2015
- . IEC 62133
#### Substantial Equivalence Discussion:
The changes to the enclosure, no longer capable to withstand sterilization and instead use of barrier sleeves, electronics, user interface and communication do not change the intended use, nor do they affect the substantial equivalence as compared to the Osstell ISQ previously cleared in
{6}------------------------------------------------
# K082523.
| | Osstell<br>Beacon | Predicate<br>Device: Osstell<br>ISQ | Reference<br>Device:<br>Tellos ISQ<br>Buddy | Reference Device:<br>Implantmed SI-<br>1015 incl.<br>Accessories | Substantial<br>Equivalence |
|-----------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K# | To be assigned | K082523 | K143445 | K161957 | |
| Device name | Osstell Beacon | Osstell® ISQ | Tellos ISQ<br>Buddy<br>(marketed as<br>Penquin RFD) | Implantmed SI-1015<br>incl. Accessories | - |
| Company name | Osstell AB | Osstell AB | Tellos Medical<br>AB | W & H<br>DENTALWERK<br>BÜRMOOS GMBH | |
| Classification | Class I | Same | Same | Same | - |
| Regulation<br>Number | 872.4200 | Same | Same | Same | - |
| Classification<br>name | Handpiece,<br>Direct Drive,<br>AC-powered | Same | Same | Same | - |
| Intended Use | Dental implant<br>stability<br>analyzer | Same | Same | Mechanical drive for<br>dental instruments<br>with ISQ module<br>used to measure<br>implant stability. | Same |
| Indication for<br>use | Indicated for use<br>in measuring the<br>stability of<br>implants in the<br>oral cavity and<br>maxillofacial<br>region. | Indicated for use<br>in measuring the<br>stability of<br>implants in the<br>oral cavity and<br>craniofacial<br>region. | Indicated for use<br>in measuring the<br>stability of<br>dental implants<br>in the oral<br>cavity and the<br>maxillofacial<br>region. | Mechanical drive unit<br>with coolant supply<br>for transmission<br>instruments<br>with ISO 3964<br>(DIN13940)<br>compatible coupling<br>system, for use in<br>dental surgery,<br>implantology and<br>maxillofacial surgery<br>(CMF) for treatment<br>of dental hard tissue.<br>Includes ISQ module<br>used to measure<br>implant stability. | Subject, primary<br>predicate, and<br>K143445<br>reference device<br>are identical.<br>K161957<br>reference device<br>used for<br>Bluetooth<br>technology has a<br>different<br>indication,<br>however the<br>comparison is<br>only for<br>Bluetooth<br>technology and<br>does not have<br>specific<br>indications for<br>this technology |
| Description | Portable,<br>handheld<br>battery driven<br>instrument<br>indicated for use<br>in measuring the<br>stability of<br>implants in the<br>oral cavity and<br>maxillofacial<br>region. | Portable,<br>handheld, or<br>freestanding<br>instrument<br>indicated for use<br>in measuring the<br>stability of<br>implants in the<br>oral cavity and<br>craniofacial<br>region. | Hand-held.<br>battery-driven<br>device for<br>measuring the<br>relative stability<br>of a dental<br>implant. | - the control unit.<br>- a motor with cable<br>with or without light<br>(EM-19 LC/EM-19),<br>- a wireless or wired<br>foot control (S-NW<br>or S-N2).<br>- the Osstell Module<br>(SI-SQ)<br>- and as an attachment<br>the surgical<br>handpieces | All are portable<br>handheld<br>devices, the<br>change in<br>configuration<br>compared to ISQ<br>which<br>incorporates the<br>device controls<br>and displays into<br>a single handheld<br>does not impact<br>the substantial |
| | Osstell<br>Beacon | Predicate<br>Device: Osstell<br>ISQ | Reference<br>Device:<br>Tellos ISQ<br>Buddy | Reference Device:<br>Implantmed SI-<br>1015 incl.<br>Accessories | Substantial<br>Equivalence |
| | | | | - (WS-56 L, WS-75<br>L, WS-91 L, WS-92<br>L and S-11 L).<br>The user can select<br>five different<br>programs. Switching<br>between<br>these programs is<br>performed by foot<br>control or via touch<br>display. | the device.<br><br>ISQ Buddy is<br>also portable,<br>handheld and<br>Implantmed uses<br>a wireless foot<br>pedal via<br>Bluetooth<br>technology.<br><br>Verification of<br>the Beacon<br>confirms it<br>performs as<br>intended. |
| Operation of<br>System | Measures the<br>frequency<br>response from<br>Smartpeg that is<br>directly attached<br>to the implant or<br>abutment. The<br>system includes<br>the following<br>components:<br>Instrument,<br>Osstell<br>Smartpeg,<br>Osstell<br>SmartPeg<br>Mount, Osstell<br>USB cable,<br>Osstell TestPeg<br><br>The technique<br>involves a<br>SmartPeg (10<br>mm x 3 mm)<br>that is attached<br>to the implant or<br>abutment. The<br>SmartPeg is<br>excited over a<br>range of<br>frequencies (1<br>kHz to 10 kHz)<br>and the<br>resonance<br>frequency is<br>measured with<br>the Osstell<br>Beacon<br>instrument and<br>software. The<br>resonance<br>frequency is | Measures the<br>frequency<br>response from<br>Smartpeg that is<br>directly attached<br>to the implant or<br>abutment. The<br>system includes<br>the following<br>components:<br>Instrument,<br>Osstell SmartPeg,<br>Osstell<br>Measurement<br>Probe, Osstell<br>SmartPeg Mount,<br>Osstell USB<br>Cable, Osstell<br>TestPeg<br><br>Same technique is<br>used in operating<br>the system as the<br>Osstell Beacon<br>device. | A<br>microcontroller<br>sends electric<br>pulses to a coil<br>in the<br>instrument tip.<br>As a<br>consequence,<br>magnetic pulses<br>are emitted that<br>affect the pin<br>connected to the<br>implant. The pin<br>then starts to<br>vibrate with its<br>resonance<br>frequency.<br>Vibration<br>creates an<br>alternating<br>magnetic field<br>which is being<br>picked up by<br>another coil in<br>the instrument<br>tip. The<br>electrical signal<br>from the<br>receiving coil is<br>analyzed and<br>the frequency is<br>determined. The<br>Frequency is<br>presented on the<br>display as an<br>"ISQ- value"<br><br>Resonance<br>frequency of the<br>ISQ peg as an<br>ISQ number, 1- | Referenced specific<br>to wireless foot pedal;<br>however, devices<br>include a ISQ module<br>used to measure<br>implant stability. | There is no<br>difference in the<br>system operation<br>between the<br>subject device<br>and K082523 or<br>K143445. |
| | Osstell<br>Beacon | Predicate Device: Osstell<br>ISQ | Reference<br>Device:<br>Tellos ISQ<br>Buddy | Reference Device:<br>Implantmed SI-<br>1015 incl.<br>Accessories | Substantial<br>Equivalence |
| | the stiffness of<br>the implant<br>system. The<br>Osstell Beacon<br>presents the<br>resonance<br>frequency as an<br>Implant<br>Stability<br>Quotient (ISQ)<br>value (scaled 0-<br>100). The ISQ<br>value is<br>proportional to<br>the stability of<br>the implant.<br>(In general, an<br>increase in ISQ<br>value from one<br>measurement<br>time to the next<br>indicates a<br>progression<br>towards higher<br>stability and a<br>lower ISQ value<br>may indicate a<br>loss in stability<br>and perhaps,<br>implant failure.)<br>Bluetooth<br>functionality for<br>device<br>production | number is<br>calculated from<br>the resonance<br>frequency. | | | |
| | purposes only | | | | |
| System<br>Components | Instrument<br>A portable,<br>handheld<br>instrument with<br>2 built-in<br>graphical<br>displays. The<br>unit operates<br>from a<br>rechargeable<br>power source<br>offering more<br>than 400 ISQ<br>measurements<br>between<br>charges.<br>The size of the<br>displays is 14 x<br>11 mm. | Instrument<br>A portable,<br>handheld, or<br>freestanding<br>instrument with<br>built-in graphical<br>display. The unit<br>operates from a<br>rechargeable<br>power source<br>offering over 6<br>hours of<br>continuous use<br>between charges.<br>The size of the<br>LCD display is 69<br>x 37 mm<br>The instrument<br>can be connected<br>to a PC via the<br>USB cable and the | Tellos ISQ<br>Buddy<br>instrument, ISQ<br>Peg, ISQ Peg<br>Driver and<br>instrument<br>charger.<br>Tellos ISQ<br>Buddy has two<br>LED-displays;<br>one on each side<br>of the<br>instrument for<br>easy reading.<br>Tellos ISQ<br>Buddy uses the<br>Osstell<br>SmartPegs, or<br>corresponding<br>Tellos pins, | - the control unit,<br>- a motor with cable<br>with or without light<br>(EM-19 LC/EM-19),<br>- a wireless or wired<br>foot control (S-NW<br>or S-N2),<br>- the Osstell Module<br>(SI-SQ)<br>- and as an<br>attachment the<br>surgical handpieces<br>- (WS-56 L, WS-75<br>L, WS-91 L, WS-92<br>L and S-11 L). | The change in<br>configuration and<br>components from<br>the ISQ which<br>incorporates the<br>device controls<br>and displays into<br>a single handheld<br>for the Beacon<br>does not impact<br>the substantial<br>equivalence of<br>the device.<br>Tellos ISQ<br>Buddy also is<br>portable,<br>handheld device<br>with LED<br>displays, and<br>employs<br>"SmartPegs or |
| Osstell<br>Beacon | Predicate<br>Device: Osstell<br>ISQ | Reference<br>Device:<br>Tellos ISQ<br>Buddy | Reference Device:<br>Implantmed SI-<br>1015 incl.<br>Accessories | Substantial<br>Equivalence | |
| The<br>measurement tip | can be transferred<br>to the optional | "ISQ Pegs"<br>from titanium. | | ISQ Pegs. The<br>Implantmed SI | |
| is held close to<br>the Smartpeg.<br>The | ISQ Data Manager<br>Software. | | | device, is<br>referenced<br>specific to | |
| measurement<br>electronics<br>sends the<br>excitation signal | Measurement<br>Probe with tip<br>The Measurement<br>Probe is connected | | | wireless foot<br>pedal, which also<br>utilizes Bluetooth<br>software | |
| to the coils in<br>the tip, and also | to the instrument<br>via the probe | | | functionality.. | |
| detects the<br>response signal<br>from the<br>detection coil in | cable and is held<br>close to the<br>Smartpeg. The<br>measurement | | | Verification of<br>the Beacon<br>confirms it | |
| the tip. The<br>microcontroller<br>in the | electronics sends<br>the excitation<br>signal to the coil | | | performs as<br>intended. | |
| instrument<br>calculates the<br>frequency of the | in the probe with<br>tip, and also<br>detects the | | | | |
| response signal,<br>and presents it<br>on the display | response signal<br>from the detection<br>coil in the probe | | | | |
| as a number, the<br>Implant<br>Stability | with tip. The<br>microcontroller in<br>the instrument…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.