Mechanical drive unit with coolant supply for transmission instruments with ISO 3964 (DIN13940) compatible coupling system, for use in dental surgery, implantology and maxillofacial surgery (CMF) for treatment of dental hard tissue.
Device Story
Implantmed SI-1015 is a mechanical drive unit for dental surgery, implantology, and CMF procedures. System components include a control unit with a TFT capacitive touch display, an electric motor (EM-19/EM-19 LC), a wireless or wired foot control, an Osstell module (SI-SQ) for implant stability measurement, and various surgical handpieces. The device is operated by dental professionals in clinics. Users select from five programs to adjust speed and torque; switching is performed via foot control or touch display. The control unit provides irrigation (100ml/min) and drives the handpiece. The device facilitates precise control of surgical instruments, aiding in dental hard tissue treatment and implant placement. The wireless foot control improves ergonomics. The integrated Osstell module allows for implant stability assessment. The device benefits patients by providing a controlled, efficient, and stable surgical environment for dental and maxillofacial interventions.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), handpiece function and lifecycle testing (ISO 14457), software validation (IEC 62304), thermal safety (IEC 62471), and biocompatibility testing (EN ISO 10993-5, ISO 7405).
Technological Characteristics
Control unit with TFT capacitive touch display; electric motor with LED contacts; wireless/wired foot control; irrigation system (100ml/min). Materials: plastic housing, stainless steel motor, chromium-coated steel/brass handpieces. Standards: IEC 60601-1, IEC 60601-1-2, ISO 14457, IEC 62304, IEC 62471, EN ISO 10993-5. Sterilization: steam (dynamic-air-removal or gravity displacement).
Indications for Use
Indicated for use in dental surgery, implantology, and maxillofacial surgery (CMF) for the treatment of dental hard tissue in patients requiring such procedures.
Regulatory Classification
Identification
A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around it. Inside the circle is a stylized symbol featuring three human profiles facing to the right, layered on top of each other to create a sense of depth and connection.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 23, 2016
W&H Dentalwerk Bürmoos GmbH Anja Lindner Mag., Regulatory Affairs Ignaz-Glaser-Straße 53 Buermoos, Salzburg 5111 Austria
Re: K161957
Trade/Device Name: Implantmed SI-1015 incl. Accessories Regulation Number: 21 CFR 872.4200 Regulation Name: Dental Handpiece and Accessories Regulatory Class: Class I Product Code: EBW
Dated: October 24, 2016 Received: October 27, 2016
Dear Anja Lindner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
Michael J. Ryan -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows a logo with the letters "W&H" inside a gray hexagon with rounded corners. Above the logo is a green banner with the text "PEOPLE HAVE PRIORITY" in white letters. The background is white. The logo appears to be for a company or organization that values people.
510(k) Indications for Use Statement
### Indications for Use
510(k) number:
K161957
Device Name:
Implantmed SI-1015 incl. Accessories
Indication for Use:
Mechanical drive unit with coolant supply for transmission instruments with ISO 3964 (DIN13940) compatible coupling system, for use in dental surgery, implantology and maxillofacial surgery (CMF) for treatment of dental hard tissue.
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over The Counter Use (Part 21 CFR 807 Subpart C)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a logo with the text "W&H" inside a hexagon. Above the logo, there is a green banner with the text "PEOPLE HAVE PRIORITY" in white letters. The logo is simple and modern, with a focus on the company's initials and a statement about its values.
## 510(k) SUMMARY
| Submitter | W & H DENTALWERK BÜRMOOS GMBH<br>Ignaz-Glaser-Strasse 53<br>A - 5111 Bürmoos<br>Austria<br>Tel.: 0043 -6274 / 6236 -397<br>Fax: 0043 -6274 / 6236 -55 |
|------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|
| Registration<br>Number | 9681479 |
| Contact<br>Person | Mag. Anja LINDNER |
| Date Summary<br>was Prepared | 21st of November, 2016 |
| Device Name | Implantmed SI-1015 incl. Accessories |
| Classification<br>Name | Dental Handpiece and Accessories |
| Common/Usual<br>Name | Controller, Foot, Handpiece and Cord |
| Regulation<br>Number | 21 CFR 872.4200 |
| Regulatory<br>class | I |
| Product Code | EBW |
| Predicate<br>Devices | Primary Predicate:<br>"Implantmed SI-915/923", W&H Dentalwerk Bürmoos GmbH; cleared<br>under K052741 |
| | Reference devices for the accessories:<br>"Osstell ISQ Implant Stability Meter", Osstell®; cleared via K082523 |
| | „Surgical contra-angle handpieces”, W&H Dentalwerk, Bürmoos; cleared<br>under K011061 and K080939 |
| | The predicate has not been subject to a design-related recall. |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows a logo with the letters "W&H" inside a gray, rounded hexagon. Above the logo is a green banner with the text "PEOPLE HAVE PRIORITY" in white letters. The logo is simple and modern, with a focus on the company's initials and a statement about its values.
| Device<br>Description | The Implantmed SI-1015 is intended for use in dental surgery, implantology and<br>maxillofacial surgery (CMF) for treatment of dental hard tissue.<br>The new Implantmed SI-1015 is a redesigned version of the old one.<br>The submission consists of :<br>- the control unit,<br>- a motor with cable with or without light (EM-19 LC/EM-19),<br>- a wireless or wired foot control (S-NW or S-N2),<br>- the Osstell Module (SI-SQ)<br>- and as an attachment the surgical handpieces<br>- (WS-56 L, WS-75 L, WS-91 L, WS-92 L and S-11 L).<br>The user can select five different programs. Switching between<br>these programs is performed by foot control or via touch display.<br>Programs 1-3 are used for adjusting the speed and programs<br>4-5 are for adjusting the torque.<br>The control unit is intended to be used with the EM-19 or EM-19 L motor.<br>The Implantmed SI-1015 will be delivered with software on the control unit.<br>To run the Implantmed SI-1015 according to its intended use, W&H provides<br>five different surgical handpieces. | | |
|-------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|----------------|
| Type | Transmission ratio | Max. speed | Max. torque |
| WS-56 L | 1:1 | 50,000 | not applicable |
| WS-75 L | 20:1 | 50,000 | 70 Ncm |
| WS-91 L | 1:2.7 | 50,000 | not applicable |
| WS-92 L | 1:2.7 | 50,000 | not applicable |
| S-11 L | 1:1 | 50,000 | not applicable |
| Indications for<br>Use: | Implantmed SI-1015 (incl. EM-19/EM-19 L):<br>Mechanical drive unit with coolant supply for transmission instruments with ISO<br>3964<br>(DIN 13940) compatible coupling system, for use in dental surgery, implantology<br>and maxillofacial surgery (CMF) for treatment of dental hard tissue. | | |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a logo with the text "W&H" inside a hexagon. Above the logo, there is a green banner with the text "PEOPLE HAVE PRIORITY" in white letters. The logo and banner are set against a light gray background.
| | General specification SI-1015 incl. Accessories | | | | |
|---------------------------------------|----------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|--|
| Item | New device | predicate device | Difference | | |
| Picture | Image: Implantmed SI-1015 | Image: Implantmed SI-915/923 | - | | |
| Name | Implantmed SI-1015 | Implantmed SI-915/923 | - | | |
| Manufacturer | W&H Dentalwerk<br>Bürmoos GmbH | W&H Dentalwerk<br>Bürmoos GmbH | - | | |
| Where used | Dental practice, dental<br>clinic | Dental practice, dental<br>clinic | none | | |
| Technological<br>Characteristics | Indications<br>for use | Mechanical drive unit<br>with coolant supply for<br>transmission<br>instruments with ISO<br>3964 (DIN 13940)<br>compatible coupling<br>system, for use in<br>dental surgery,<br>implantology and<br>maxillofacial surgery<br>(CMF) for treatment of<br>dental hard tissue. | Mechanical drive unit<br>with coolant supply for<br>transmission<br>instruments with<br>coupling system<br>according to ISO 3964<br>(Din 13.940) The<br>equipment is a drive<br>unit for use in dental<br>surgery, implantology<br>and maxilla-facial<br>surgery for treatment of<br>dental hard tissue. | none | |
| | Control Unit | Main dimensions:<br>100x262x291mm<br>Front panel:<br>TFT display with<br>capacitive touch<br>Programs:<br>5 programs for various<br>stages of implantology<br>Irrigation:<br>100ml/min. | Main dimensions:<br>90x252x254mm<br>Front panel:<br>Graphical display<br>without backlighting<br>Programs:<br>5 programs for various<br>stages of implantology<br>Irrigation:<br>100ml/min. | none | |
| | | | | | |
| | Irrigation Tubing can<br>be inserted<br>ergonomically on the<br>unit's side face | inserted on the unit's<br>front face | | | |
| | Main dimension:<br>154x202x210 | Main dimension:<br>215x190x40 (without<br>cable) | | | |
| Foot Control | Features:<br>4 buttons for pump<br>on/off<br>Forward/reverse<br>Change programs | Features:<br>4 buttons for pump<br>on/off<br>Forward/reverse<br>Change programs | Yes,<br>the power supply of<br>the new Implantmed is<br>performed via wireless<br>foot control not via<br>cable.<br> | | |
| | Motor control (on/off<br>and variable) | Motor control (on/off<br>and variable) | | | |
| | Power supply:<br>wireless<br>via 3xAA batteries | Power supply:<br>wired<br>via cable | | | |
| Motor<br>with<br>cable | Length:<br>71,65 mm<br>With LED contacts | Length:<br>75 mm<br>Without LED contacts | Yes,<br>the new motor with<br>LED contacts (EM-19<br>LC) disposes of better<br>lightning at a lower<br>temperature as the<br>energy transfer is<br>provided via direct<br>voltage supply and not<br>via generator any<br>more. However, this<br>does not raise any<br>additional questions<br>regarding substantial<br>equivalence.<br> | | |
| SI-SQ | W&H SI-SQ:<br>Connection via USB-<br>cable | Osstell ISQ:<br>Stand-alone device | Yes, the Osstell ISQ is<br>already cleared for<br>market (K082523). The<br>W&H SI-SQ has the<br>same intended use,<br>consists of the same<br>materials and is<br>sterilized the same<br>way as the Osstell<br>ISQ. The only<br>differences are the<br>labeling (W&H) and<br>the connection - as the<br>SI-SQ is no stand-<br>alone device.<br>However, this does not<br>raise any additional<br>questions regarding<br>substantial | | |
| | | | | equivalence. | |
| | | Technical data/Functions<br>Implantmed SI-1015 | | | |
| New Device | | | Predicate Device | Difference | |
| Max.<br>mechanical<br>output<br>power | 80 W | | 70 W | Yes5 | |
| Torque at<br>the motor | 6.2 Ncm | | 5.5 Ncm | Yes5 | |
| Speed range<br>of motor | 200 – 40,000 rpm | | 200 – 40,000 rpm | None | |
| Supply<br>voltage | 100-130 VAC | | 100-130 VAC | None | |
| Rated<br>current | 0.3 - 1.6 A | | 0.2 - 1.6 A | None | |
| Frequency | 50-60 Hz | | 50-60 Hz | None | |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows a logo with the letters "W&H" inside a hexagon. Above the logo is a green banner with the text "PEOPLE HAVE PRIORITY" in white letters. The logo is simple and modern, with a focus on the company's initials and a statement about its values.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows a logo with the letters "W&H" inside of a hexagon. The hexagon is white with a gray outline. Above the logo is a green banner with the words "PEOPLE HAVE PRIORITY" in white letters. The background is white.
lt allows in its program 4 to adjust torque values up to 80 Ncm (which are needed for implant systems) - instead of previous maximum of 50 Ncm.
This does not lead to any negative effect regarding substantial equivalence because of the constructive power reserves of the motor (no big difference between 5.5 Ncm – 6.2 Ncm for the motor).
| Materials | | | | |
|-----------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------|-------------------------------------------------------------------------------|------------|
| Item | New device | Predicate device | Difference | |
| Control unit<br>housing | plastic material | plastic material | None | |
| Tubing outer<br>sheath | Customer specific | Customer specific | None | |
| Motor with cable | Stainless steel | Stainless steel | None | |
| Surgical<br>handpieces | Chromium coated steel and<br>chromium coated brass | Chromium coated steel and<br>chromium coated brass | None | |
| | SI-SQ | Stainless steel | Stainless steel | None |
| | Hygiene / Maintenance<br>Surgical Handpieces (WS-56 L, WS-75 L, S-11 L, WS-91 L, WS-92 L), EM-19, EM-19 LC<br>and SI-SQ | | | |
| | Item | New device | Predicate device | Difference |
| | Lubrication | after max. 30 minutes of use<br>EM-19 LC/ EM-19:<br>No lubrication is needed | after max. 30 minutes of use<br>Motor with cable:<br>No lubrication is needed | None |
| | Cleaning | rinse under demineralized<br>water (< 38°C/100°F) with<br>aid of brush | rinse under demineralized<br>water<br>(< 38°C/100°F) with aid of<br>brush | None |
| | Disinfection | wiping disinfection using<br>disinfectant cloths | wiping disinfection using<br>disinfectant cloths | None |
| | Sterilization | Dynamic-air-removal<br>sterilizers: 270°F (132°C)<br>for 4 minutes or | Dynamic-air-removal<br>sterilizers: 270°F (132°C) for<br>4 minutes or | None |
| | | Gravity displacement<br>sterilizers:<br>270°F (132°C) for 15<br>minutes | Gravity displacement<br>sterilizers:<br>270°F (132°C) for 15 minutes | |
| | W&H's Implantmed SI-1015 represents a redesigned and improved<br>version of the predicate device.<br>The technical principle is the same as within the predicate device.<br>The main technological characteristics are the same or, at least, quite similar to<br>those of the comparable product.<br>The Indications for Use statement for the Implantmed SI-1015 is<br>identical to the predicate device. | | | |
| Comparison of<br>the device to<br>the predicate<br>device | The target field of application, the intended use, functions and<br>technological features, performance parameter and material are the same or,<br>at least, quite similar to those of the predicate device.<br>The new device is substantially equivalent to the predicate device. | | | |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows a logo with the letters "W&H" inside a double-lined hexagon. Above the logo is a green banner with the text "PEOPLE HAVE PRIORITY" in white letters. The logo is simple and modern, with a focus on the company's initials and a statement about its values.
# Implantmed SI-1015 incl. Accessories
Section 5 Page 6 of 7
510(k) Summary
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows a logo with the letters "W&H" inside a gray, rounded hexagon. Above the logo, there is a green banner with the text "PEOPLE HAVE PRIORITY" written in white, sans-serif font. The logo appears to be for a company or organization that values its people.
| Performance<br>Testing | Electrical Safety Tests according to IEC 60601-1:2005, Medical electrical<br>equipment<br>– Part 1: General requirements for basic safety and essential performance. |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Electromagnetic Compatibility Test according to IEC 60601-1-2:2007:<br>General requirement for basic safety and essential performance –<br>Collateral standard: Electromagnetic compatibility – Requirements and tests |
| | Product testing of handpiece function and lifecycle testing were performed<br>per ISO 14457:2012: Dentistry – Handpieces and Motors.<br>The results demonstrate substantial equivalence in this regard. |
| | Software validation according to IEC 62304:2006: Medical device software –<br>Part 1:<br>Guidance on the application of ISO 14972 to medical device software |
| | Thermal safety according to the standard IEC 62471:2006:<br>Photobiological safety of lamps and lamp systems |
| | Evaluation of biocompatibility is based upon the fact that patient<br>contacting materials in the subject handpiece are identical to those in the<br>previously W&H surgical handpieces K011061 and K080939, which, as<br>handpieces,<br>present the same level and duration of contact.<br>In addition, Cytotoxicity Testing per EN ISO 10993-5:2009-06 was performed.<br>This evaluation meets the requirements of ISO 7405:2008 for preclinical<br>evaluation of biocompatibility of dental devices. |
| | For the new wireless foot control (S-NW) software verification/validation of the<br>functions of the foot control was conducted according to IEC 62304:2006.<br>Additionally, EMC testing was performed to evaluate the risk of communication<br>loss according to IEC 60601-1-2:2007 and Electrical Safety Tests have been<br>done according to IEC 60601-1-1:2005. |
| | Clinical<br>Testing |
| Conclusion | W&H considers the “Implantmed SI-1015 incl. Accessories” to be<br>substantially equivalent to the predicate devices listed above. This conclusion is<br>based on the similarities in intended use, principles of operation, functional<br>design, and establishment medical use. Differences between the devices shown<br>in the side-by-side comparison table above are minor and do not have any<br>negative effect on equivalence. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.