OssBuilder System is a metal device intended for use with a dental implant to stabilize and support of bone graft in dento-alveolar bony defect sites.
Device Story
OssBuilder System is a non-resorbable, 3D pre-formed titanium membrane system used in dental surgery to stabilize bone grafts at dento-alveolar defect sites. The system comprises an OssBuilder membrane, Healing Cap, Cover Cap, and OB Anchor. It is used by dental professionals in clinical settings. The device is placed over the bone graft site and secured using the anchor and caps to prevent foreign material intrusion and support bone augmentation during osseointegration. The system is not intended for permanent implantation. It provides mechanical support to the graft, facilitating bone regeneration, which benefits patients by improving the site for subsequent dental implant placement.
Clinical Evidence
No clinical studies were submitted. Substantial equivalence is supported by non-clinical bench testing, including mechanical properties testing (tensile strength, yield strength, elongation) and biocompatibility evaluation based on the use of established dental materials (titanium/titanium alloy). Sterilization was validated per ISO 11137.
Technological Characteristics
Materials: Pure Titanium Grade 4 (ASTM F67) and Ti-6Al-4V (ASTM F136). Components: 3D pre-formed membrane, healing cap, cover cap, and anchor. Sterilization: Gamma radiation. Shelf life: 8 years. Connectivity: None (mechanical device).
Indications for Use
Indicated for patients requiring stabilization and support of bone grafts in dento-alveolar bony defect sites in conjunction with dental implants.
Regulatory Classification
Identification
A bone plate is a metal device intended to stabilize fractured bone structures in the oral cavity. The bone segments are attached to the plate with screws to prevent movement of the segments.
K222451 — SAVE GBR · Dentis Co., Ltd. · Feb 7, 2023
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 26, 2018
Osstem Implant Co., Ltd. David Kim Manager Hiossen Inc. 85 Ben Fairless Dr. Fairless Hills, Pennsylvania 19030
Re: K172354
Trade/Device Name: OssBuilder System Regulation Number: 21 CFR 872.4760 Regulation Name: Bone Plate Regulatory Class: Class II Product Code: JEY, NHA Dated: February 20, 2018 Received: February 20, 2018
Dear David Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Michael J. Ryan -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo and contact information for OSSTEM Implant Co., Ltd. The logo includes the company name in bold, black font, with the word "OSSTEM" in orange to the left. Below the company name is the address: 66-16, Bansong-ro 513beon-gil, Haeundae-gu, Busan, Republic of Korea, as well as the telephone number: +82 51 850-2500, fax number: +82 51 850-4341, and website: www.osstem.com.
510(k) Number: K172354
Device Name: OssBuilder System
Indication for Use:
OssBuilder System is a metal device intended for use with a dental implant to stabilize and support of bone graft in dento-alveolar bony defect sites.
Image /page/2/Picture/5 description: The image contains a square with the letter 'X' inside. The 'X' is centered within the square and appears to be in a simple, sans-serif font. The square is a solid black outline with a white fill, providing a clear contrast for the black 'X'. The image is simple and straightforward, focusing on the single letter within the geometric shape.
Prescription Use (21 CFR 801 Subpart D)
Over-The Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo and contact information for OSSTEM Implant Co., Ltd. The address is 66-16, Bansong-ro 513beon-gil, Haeundae-gu, Busan, Republic of Korea. The telephone number is +82 51 850-2500, the fax number is +82 51 850-4341, and the website address is www.osstem.com.
### 510(k) Summary
Date: March 24, 2018
- 1. Company and Correspondent making the submission
| - Submitter's Name | : OSSTEM IMPLANT Co., Ltd. |
|------------------------|-----------------------------------------------------------------------------------|
| - Address | : 66-16, Bansong-ro 513beon-gil, Haeundae-gu, Busan<br>612-070, Republic of Korea |
| - Contact | : Ms. Jungmin Yoo |
| - Phone | : +82-51-850-2575 |
| - Correspondent's Name | : HIOSSEN Inc. |
| - Address | : 85 Ben Fairless Dr. Fairless Hills, PA 19030 |
| - Contact | : Mr. David Kim |
| - Phone | : 267-759-7031 |
- 2. Device
| Trade or (Proprietary) Name | : OssBuilder System |
|-----------------------------|---------------------|
| Classification Name | : Bone Plate |
| Regulation Number | : 21CFR872.4760 |
| Devce Classification | : Class II |
| Classification Product Code | : JEY |
| Subsequent Product Code | : NHA |
- Predicate Device 3.
- Primary Predicate -SMARTbuilder System, OSSTEM IMPLANT Co., Ltd. (K120951)
- -Reference Devices SMARTBUILDER SYSTEM (SB1) (K130840) SB Anchor, OSSTEM Implant Co., Ltd. (K140600)
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo and contact information for OSSTEM Implant Co., Ltd. The logo includes the company name in bold, black font, with the word "OSSTEM" also appearing in orange and gray on the left side of the image. The contact information includes the company's address, phone number, fax number, and website address. The address is 66-16, Bansong-ro 513beon-gil, Haeundae-gu, Busan, Republic of Korea, the phone number is +82 51 850-2500, the fax number is +82 51 850-4341, and the website address is www.osstem.com.
#### 4. Description
OssBuilder System is a metal device intended for use with a dental implant to stabilize and support of bone graft in dento-alveolar bony defect sites. It is consisted of OssBuilder, Healing Cap, Cover Cap, and OB Anchor.
The OssBuilder System is not indicated for permanent implantation.
| Item | Content | |
|-------------------|-----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| OssBuilder | Description | OssBuilder is 3D pre-formed titanium made non-resorbable<br>membrane that helps bone augmentation by performing bone graft<br>on the intraoral site with autogenous bone defect. |
| | Material | Pure Titanium Grade 4 (ASTM F67) |
| | Dimension<br>(mm) | OB2 Proximal<br>Buccal width<br>Buccal length<br>Buccal distance |
| | | OB3 Buccal width<br>Buccal length<br>Lingual length<br>Buccal distance<br>Lingual distance |
| | | OB Anchor |
| Material | Ti-6Al-4V (ASTM F136) | |
| Dimension<br>(mm) | D (Ø) | |
| | | L |
| Healing Cap | Description | Healing Cap is used by connecting with the OB Anchor to fix<br>OssBuilder during osseointegration and prevents intrusion of foreign<br>material. |
| | Material | Pure Titanium Grade 4 (ASTM F67) |
| | Dimension<br>(mm) | D (Ø) |
| | | H |
| Cover Cap | Description | Cover Cap is used by connecting with the OB Anchor to fix<br>OssBuilder during osseointegration and prevents intrusion of foreign<br>material. |
| | Material | Ti-6Al-4V (ASTMF 136) |
| | Dimension<br>(mm) | D (Ø) |
| | | H |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo and contact information for OSSTEM Implant Co., Ltd. The logo includes the company name in a bold, sans-serif font, with the word "OSSTEM" in a larger size. Below the company name is the address: 66-16, Bansong-ro 513beon-gil, Haeundae-gu, Busan, Republic of Korea, as well as the telephone number: +82 51 850-2500, fax number: +82 51 850-4341, and website: www.osstem.com.
The OssBuilder System is compatible with OSSTEM Implant System, K161604; US SA Implant System, K161103; and NT/IT System, K081078.
- 5. Substantial Equivalence Matrixes
| Change on the 'name' of the predicated devices | | | |
|------------------------------------------------|----------------|-----------------------------|--------------------------------------------------------------------------------|
| | Proposed | Predicated | Remark |
| Device Name | Oss builer OB2 | SMARTbuilder SB2<br>K120951 | No other<br>modifications<br>happened except<br>for the name of<br>the device. |
| | Healing Cap | Healing Abutment<br>K120951 | |
| | OB Anchor | SB Anchor<br>K140600 | |
| | Oss Builder (OB3) | SMARTbuilder (SB2) | Remark |
|--------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| 510(k) No. | Proposed | K120951 | - |
| Manufacturer | Osstem Implant Co., Ltd. | Osstem Implant Co., Ltd. | Identical |
| Design | Image: Oss Builder Design | Image: SMARTbuilder Design | Different |
| Indication for Use | Oss Builder Systemis a<br>metal device intended for<br>use with a dental implant<br>to stabilize and support of<br>bone graft in dento-alvolar<br>bony defect sites. | SMARTbuilder is a metal<br>device intended for use<br>with a dental implant to<br>stabilize and support of<br>bone graft in dento-<br>alveolar bony defect sites. | Identical |
| Material | Pure Titanium Grade 2<br>(ASTM F67) | Pure Titanium Grade 2<br>(ASTM F67) | Identical |
| Width (D, mm) | 10, 20 | 8, 9, 10, 12 | Different |
| Sterilization | Gamma Sterilization | Gamma Sterilization | Identical |
| Shelf Life | 8 years | 5 years | Different |
| S.E | Similarities | | |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Osstem Implant. The word "OSSTEM" is written in orange, with the word "IMPLANT" written in gray below it. There is a small orange circle with a white number 6 in the upper right corner of the logo.
OSSTEM Implant Co., Ltd.
66-16, Bansong-ro 513beon-gil, Haeundae-gu, Busan, Republic of Korea
Tel: +82 51 850-2500 Fax: +82 51 850-4341 www.osstem.com
| The proposed Os sBuilder (OB3) is made of same material and has same<br>indication for use and sterilization method; and is made by same<br>manufacturer compared to the predicated SMARTbuilder (SB2)<br>(K120951).<br><br>Differences<br>Application site is different between the proposed OssBuider (OB3) and the<br>predicated, and it causes shape differences as a result.<br><br>... While there are shape differences between the proposed Os sBuilder<br>(OB3) and the predicated, the proposed device is made with same material<br>and has same function and indication for use compared to that of the<br>predicated; therefore, the proposed OssBuilder (OB3) is substantially<br>equivalent to the predicated SMARTbuilder (SB2) (K120951) based<br>on the results of the mechanical properties testing. | | | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------|
| | Healing Cap | Healing Abutment | Remark |
| 510(k) No. | Proposed | Predicated (K120951) | |
| Manufacturer | Osstem Implant Co., Ltd. | Osstem Implant Co., Ltd. | Identical |
| Design | Image: Healing Cap | Image: Healing Abutment | Different |
| Indication for Use | Oss Builder System is a<br>metal device intended for<br>use with a detal implant to<br>stabilize and support of<br>bone graft in dento-<br>alveolar bony defect sites. | SMARTbuilder is a metal<br>device intended for use<br>with a dental implant to<br>stabilize and support of<br>bone graft in dento-<br>alveolar bony defect sites. | Identical |
| Material | Pure Titanium Grade 4<br>(ASTM F67) | Pure Titanium Grade 4<br>(ASTM F67) | Identical |
| Diameter (mm) | 4.0-5.0 | 4.0-7.0 | Within the range<br>of the predicated<br>device. |
| Height (mm) | 7.0-8.0 | 3.35-4.22 | Different |
| Sterilization | Gamma Sterilization | Gamma Sterilization | Identical |
| Shelf Life | 8 years | 5 years | Different |
| S.E. | Similarities<br>The proposed Healing Cap is made of same material and has same<br>indication for use and sterilization method; and is made by same<br>manufacturer compared to the predicated Healing Abutment (K120951). | | |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Osstem Implant. The word "OSSTEM" is in orange, with a small superscript "6" next to the "M". Below "OSSTEM" is the word "IMPLANT" in gray. The logo is simple and modern.
OSSTEM Implant Co., Ltd.
66-16, Bansong-ro 513beon-gil, Haeundae-gu, Busan, Republic of Korea
Tel: +82 51 850-2500 Fax: +82 51 850-4341 www.osstem.com
| Also, its diameter is within the range of that of the predicated device. |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <b>Differences</b><br>The proposed Healing Cap is used with internal type of anchor while the<br>predicated Healing Abutment is used with external type of anchor. That is<br>why the proposed device has longer height. |
| : While the proposed Healing Cap has different connection structure<br>because it is compatible with different type of anchor than the predicated,<br>its function and indication for use is same; therefore, the proposed Healing<br>Cap is substantially equivalent to the predicated Healing Abutment<br>(K120951) based on the results of the mechanical properties tests. |
| | Cover Cap | Cover Cap | Remark |
|--------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| 510(k) No. | Proposed | Predicated (K120951) | |
| Manufacturer | Osstem Implant Co., Ltd. | Osstem Implant Co., Ltd. | Identical |
| Identifier | SMCC415 | SMCC415 | Identical |
| Design | Image: Cover Cap Design | Image: Cover Cap Design | Identical |
| Indication for Use | Oss Builder Systemis a<br>metal device intended for<br>use with a detal implant to<br>stabilize and support of<br>bone graft in dento-<br>alveolar bony defect sites. | SMARTbuilder is a metal<br>device intended for use<br>with a dental implant to<br>stabilize and support of<br>bone graft in dento-<br>alveolar bony defect sites. | Identical |
| Material | Ti-6Al-4V<br>(ASTM F 136) | Pure Titanium Grade 4<br>(ASTM F67) | Different |
| Diameter (mm) | 4.0 | 4.0 | Identical |
| Height (mm) | 1.73 | 1.52 | Different |
| Sterilization | Gamma Sterilization | Gamma Sterilization | Identical |
| Shelf Life | 8 years | 5 years | Different |
| S.E. | Similarities<br>The proposed Cover Cap has same identifier, design, indication for use,<br>diameter, and sterilization method; and is made by same manufacturer<br>compared to the predicated Cover Cap (K120951).<br>Differences<br>The material for the proposed Cover Cap is Ti-6Al-4V while the predicated<br>is Ti Gr.4; and its height is different than that of the predicated.<br>While the material and the height of the proposed Cover Cap is different<br>from the predicated, its design function, and indication for use is same;<br>therefore, the proposed Cover Cap is substantially equivalent to the | | |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image contains the logo for Osstem Implant. The logo is comprised of the word "OSSTEM" in orange, with the word "IMPLANT" in gray underneath. The logo is simple and modern, and the colors are bright and eye-catching.
OSSTEM Implant Co., Ltd.
66-16, Bansong-ro 513beon-gil, Haeundae-gu, Busan, Republic of Korea
Tel: +82 51 850-2500 Fax: +82 51 850-4341 www.osstem.com
| predicated Cover Cap (K120951) based on the results of the biocompatibility<br>evaluations and mechanical properties testing. | | | |
|-------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| | Cover Cap | Cover Cap | Remark |
| 510(k) No. | Proposed | Predicated (K120951) | - |
| Manufacturer | Osstem Implant Co., Ltd. | Osstem Implant Co., Ltd. | Identical |
| Design | Image: Proposed Cover Cap | Image: Predicated Cover Cap | Different |
| Indication for Use | Oss Builder Systemis a<br>metal device intended for<br>use with a detal implant to<br>stabilize and support of<br>bone graft in dento-<br>alveolar bony defect sites. | SMARTbuilder is a metal<br>device intended for use<br>with a dental implant to<br>stabilize and support of<br>bone graft in dento-<br>alveolar bony defect sites. | Identical |
| Material | Ti-6Al-4V<br>(ASTM F 136) | Pure Titanium Grade 4<br>(ASTM F67) | Different |
| Diameter (mm) | 4.0 | 4.0 | Identical |
| Height (mm) | 4.3 | 1.52 | Different |
| Sterilization | Gamma Sterilization | Gamma Sterilization | Identical |
| Shelf Life | 8 years | 5 years | Different |
| S.E. | Similarities<br>The proposed additional shape of Cover Cap has same indication for use,<br>diameter, and sterilization method; and is made by same manufacturer<br>compared to the predicated Cover Cap (K120951).<br><br>Differences<br>The proposed Cover Cap is made of Ti-6Al-4V while the predicated is<br>made of Ti Gr. 4. The proposed Cover Cap is used with internal type of<br>anchor while the predicated Cover Cap is used with external type of anchor.<br>That is why the proposed device has longer height.<br><br>... While the proposed Cover Cap is made with different material and has<br>different connection structure because it is compatible with different type of<br>anchor than the predicated, its function and indication for use is same;<br>therefore, the proposed Cover Cap is substantially equivalent to the<br>predicated Cover Cap (K120951) based on the results of the<br>biocompatibility evaluations and mechanical properties testing. | | |
| | OB Anchor | SB Anchor | Remark |
|------------|-----------|----------------------|--------|
| 510(k) No. | Proposed | Predicated (K140600) | - |
IDD
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for Osstem Implant. The word "OSSTEM" is written in orange, with the superscript symbol of a degree symbol. Below the word "OSSTEM" is the word "IMPLANT" in gray. The logo is simple and modern.
## OSSTEM Implant Co., Ltd
66-16, Bansong-ro 513beon-gil, Haeundae-gu, Busan, Republic of Korea Tel: +82 51 850-2500 Fax: +82 51 850-4341 www.osstem.com
| Manufacturer | Osstem Implant Co., Ltd. | Osstem Implant Co.,<br>Ltd. | Identical |
|-----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------|
| Design | Image: [Osstem Implant Co., Ltd. Design] | Image: [Osstem Implant Co., Ltd. Design] | Different |
| Indication for<br>Use | OssBuilder Systemis a metal<br>device intended for use with a detal<br>implant to stabilize and support of<br>bone graft in dento-alveolar bony<br>defect sites. | SMARTbuilder is a<br>metal device intended<br>for use with a dental<br>implant to stabilize and<br>support of bone graft in<br>dento-alveolar bony<br>defect sites. | Identical |
| Material | Ti-6Al-4V<br>(ASTM F136) | Ti-6Al-4V<br>(ASTM F136) | Identical |
| Diameter<br>(mm) | 3.1~5.0 | 3.3~5.1 | Within the<br>range of the<br>predicated<br>device. |
| Height (mm) | 4.65~9.5 | 7.0~15 | Different |
| Sterilization | Gamma Sterilization | Gamma Sterilization | Identical |
| Shelf Life | 8 years | 8 years | Identical |
| S.E. | Similarities<br>The proposed OB Anchor is made of same material, has same indication for use<br>and sterilization method; and is made by same manufacturer compared to the<br>predicated SB Anchor (K140600). Also, its diameter is within the range of that<br>of the predicated device.<br>Differences<br>The proposed device is internal type of anchor while the predicated device is<br>external type of anchor. That is why the height of the proposed anchor is shorter<br>than that of the predicated anchor.<br>... The design of the proposed OB Anchor is different because it is internal type<br>of anchor while the predicated is external type of anchor, but it is made with<br>same material and has same function and indication for use; therefore, the<br>proposed OB Anchor is substantially equivalent to the predicated SB Anchor<br>(K140600) based on the results of the mechanical properties tests. | | |
#### 6. Indication for Use
OssBuilder System is a metal device intended for use with a dental implant to stabilize and support of bone graft in dento-alveolar bony defect sites.
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo and contact information for OSSTEM Implant Co., Ltd. The logo includes the company name in bold, black font, with the word "OSSTEM" also appearing in orange next to the company name. Below the company name is the address: 66-16, Bansong-ro 513beon-gil, Haeundae-gu, Busan, Republic of Korea, as well as the telephone number: +82 51 850-2500, fax number: +82 51 850-4341, and website: www.osstem.com.
- 7. Summary of Non-Clinical Performance Testing
Non-clinical testing data are submitted, referenced, or relied upon to demonstrate substantial equivalence.
#### Biocompatibility evaluation
Biocompatibility evaluation for OssBuilder System is not considered because the materials used for manufacturing OssBuilder System are titanium and titanium alloy which have been generally and widely used as a dental material such as implant for a long time. OssBuilder System is made of same material, chemical composition, and body contact with the predicated devices, SMARTbuilder System, K120951; SMARTbuilder System (SB1), K130840; and SB Anchor, K140600.
#### Sterilization Validation and Shelf-life
All subject devices in OssBuilder System are delivered in sterile. Therefore, the shelf-life and sterilization vas considered. For the sterilization validation, it was performed accordance with worst-case chosen among subject devices and following ISO 11137. For the shelf-life of subject devices, we considered their validation of packaging materials by leveraging the data that of our prior submissions.
#### Mechanical Properties
Bench tests evaluated for OssBuilder System included Tensile Strength, Yield Strength, and Elongation. Mechanical properties from bench tests of OssBuilder System found substantially equivalent to the predicate device.
- Summary of Clinical Testing 8.
No clinical studies are submitted.
- 9. Conclusion
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification. OSSTEM IMPLANT Co., Ltd. concludes that the OssBuilder System is substantially equivalent to the predicated devices as herein.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.