The OssBuilder System is used for space maintenance and stabilization of bone grafting material in an alveolar bone defect site.
Device Story
OssBuilder System comprises metal components: Tenting Screw, Healing Cap, Cover Cap, OB Anchor, and OssBuilder (titanium membrane). Tenting Screw acts as a temporary dental implant placed in maxilla or mandible bone defects to fixate bone graft material and support the membrane during osseointegration. Healing Cap connects to Tenting Screw to secure the membrane. System used in dental clinics by clinicians; requires second surgery for removal after bone regeneration. Device provides mechanical stabilization of graft material to facilitate bone growth; benefits patient by enabling successful bone augmentation in dento-alveolar defects.
Clinical Evidence
No clinical studies submitted. Substantial equivalence supported by non-clinical bench testing, including biocompatibility evaluation per ISO 10993-1 and mechanical property testing per ASTM F543 (fracture load, insertion/removal torque, axial pullout strength, torsional strength, breakage angle, self-tapping force).
Technological Characteristics
Materials: Ti-6Al-4V ELI (ASTM F136). Components: Tenting Screw (2.1mm diameter, 8.5-13.0mm length), Healing Cap (4-5mm diameter, 7-8mm length). Principle: Mechanical fixation of bone graft and membrane. Sterilization: Gamma radiation. Connectivity: None (mechanical device).
Indications for Use
Indicated for patients requiring space maintenance and stabilization of bone grafting material in alveolar bone defect sites for horizontal and vertical bone augmentation.
Regulatory Classification
Identification
An intraosseous fixation screw or wire is a metal device intended to be inserted into fractured jaw bone segments to prevent their movement.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of the square.
May 7, 2019
Osstem Implant Co., Ltd. % Peter Lee OA/RA Manger Hiossen Inc. 85 Ben Fairless Dr. Fairless Hills, Pennsylvania 19030
Re: K181854
Trade/Device Name: OssBuilder System Regulation Number: 21 CFR 872.4880 Regulation Name: Intraosseous Fixation Screw or Wire Regulatory Class: Class II Product Code: DZL, NHA Dated: April 4, 2019 Received: April 5, 2019
Dear Peter Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Malvina Eydelman, M.D. Director OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/1 description: This image contains the logo for OSSTEM Implant Co., Ltd. The logo includes the company name in bold, black text, with the word "OSSTEM" appearing in orange on the left side. Below the logo is the company's address: 66-16, Bansong-ro 513beon-gil, Haeundae-gu, Busan, Republic of Korea, along with their phone number: +82 51 850 2500, fax number: +82 51 861 4693, and website: www.osstem.com.
510(k) Number: K 181854
Device Name: OssBuilder System
## Indication for Use:
The OssBuilder System is used for space maintenance and stabilization of bone grafting material in an alveolar bone defect site.
X | Prescription Use (21 CFR 801 Subpart D) Over-The Counter Use (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
## Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
6-16, Bansong-ro 513beon-gil, Haeundae-gu, Busan, Republic of Korea Tel: +82 51 850 2500 Fax: +82 51 861 4693 www.osstem.com
STEM Implant Co., L
# 510(K) Summary K181854
# Date: May 07, 2019
## 1. Company and Correspondent making the submission
| Submitter's Name | OSSTEM IMPLANT Co., Ltd. |
|----------------------|---------------------------------------------------------------------------------|
| Address | 66-16, Bansong-ro 513beon-gil, Haeundae-gu, Busan<br>612-070, Republic of Korea |
| Contact | Ms. Jungmin Yoo |
| Phone | +82-51-850-2500 |
| Correspondent's Name | HIOSSEN Inc. |
| Address | 85 Ben Fairless Dr.<br>Fairless Hills, PA, 19030 |
| Contact | PETER LEE |
| Phone | 267-759-7031 |
- 2. Device
| - Trade or (Proprietary) Name | : OssBuilder System |
|-------------------------------|---------------------------------------|
| - Classification Name | : Intraosseous Fixation Screw or Wire |
| - Regulation Number | : 21CFR872.4880 |
| - Devce Classification | : Class II |
| - Classification Product Code | : DZL |
| - Subsequent Product Code | : NHA |
## 3. Predicate Device
- Primary Predicate -Neo GBR System, Neobiotech Co., Ltd. (K160991)
- -Reference Predicate OssBuilder System, OSSTEM Implant Co., Ltd. (K172354)
{4}------------------------------------------------
- 4. Description
OssBuilder System is consisted of metal devices, OssBuilder, Healing Cap, Cover Cap, OB Anchor, and Tenting Screw, and is intended for use to stabilize and support of bone graft in dento-alveolar bony defect sites.
Tenting Screw, a product in the OssBuilder System product group, is a temporary dental implant that can be implanted in the bone defect site of the alveolar bone to fix the guided tissue regeneration material.
Healing Cap is used by connecting with the Tenting Screw to bind OssBuilder during osseointegration.
| Item | Content | | |
|---------------|----------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------|
| Tenting Screw | Description | Tenting Screw is used for space maintenance of the bone graft material for horizontal and vertical bone augmentation in the alveolar bone defect site. | |
| | Material | Ti-6Al-4V (ASTM F 136) | |
| | Dimension (mm) | D (Ø, implanted) | 2.1 |
| | | Total Length | 8.5, 10.0, 11.5, 13.0 |
| Healing Cap | Description | Healing Cap is used by connecting with the Tenting Screw to bind OssBuilder during osseointegration. | |
| | Material | Ti-6Al-4V (ASTM F 136) | |
| | Dimension (mm) | D (Ø) | 4, 5 |
| | | Total Length | 7, 8 |
The OssBuilder System is compatible with OssBuilder, Healing Cap, and Cover Cap of OssBuilder System cleared in K172354.
- 5. Substantial Equivalence Matrixes
| | OssBuilder System | Neo GBR System | Remark |
|-------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(K) | Proposed | Predicated (K160991) | - |
| Manufacturer | Osstem Implant Co., Ltd. | Neobiotech. Co., Ltd. | Different |
| Device Name | Tenting Screw | Tent Screw | Different |
| Design | Image: Tenting Screw | Image: Tent Screw | Similar |
| Indication for<br>Use | The OssBuilder System is used<br>for space maintenance and<br>stabilization of bone grafting<br>material in an alveolar bone<br>defect site. | Neo GBR System is intended<br>to fixate and stabilize non-<br>resorbable barrier membranes<br>used for regeneration of tissue<br>in the oral cavity or in dental<br>situations that require<br>membrane use or fixation. | The propose device<br>and the predicated<br>device has different<br>indication for use in<br>language. The<br>indication for use<br>for the proposed<br>device is for<br>individual device<br>(Tenting Screw)<br>within the system,<br>while the indication<br>for use for the<br>predicated device is<br>for the overall<br>devices in the Neo<br>GBR System.<br>Therefore, it can be<br>said that the<br>indication for use<br>for both devices are<br>equivalent when<br>viewed as a whole<br>system. |
| Principle of<br>Operation | Tenting Screw is placed in<br>maxilla or mandible bone area<br>to be combined with<br>OssBuilder (non-resorbable<br>titanium membrane) and<br>Healing Cap or Cover Cap. It<br>helps to support and prevent<br>the mobility of the membrane<br>during the bone regeneration.<br>Since it is used for temporary<br>use, second surgery is required<br>to remove them after bone<br>regeneration. | The function of Bone screw<br>body is temporary dental<br>implant (Screw) that place in<br>maxilla or mandible bone area<br>to be combined with titanium<br>membrane and cover screw or<br>cover cap. It helps to support<br>the membrane. Second surgery<br>for removal is required. | The proposed<br>device and the<br>predicated device<br>has different<br>principle of<br>operation in<br>language, however<br>the difference in<br>language does not<br>change the intended<br>use or substantial<br>equivalence status. |
| Material | Ti-6A1-4V ELI (ASTM F136) | Ti-6Al-4V ELI (ASTM F136) | Identical |
| Head<br>Diameter<br>(mm) | 3.0 | 2.85 | Similar |
| Major<br>Thread<br>Diameter<br>(mm) | 2.1 | 2.0 | Similar |
| Length (mm) | 8.5, 10.0, 11.5, 13.0 | 7.2, 8.7, 10.2, 11.7, 13.2, 15.2,<br>18.2 | Within the range of<br>the predicated. |
| Sterilization | Gamma Sterilization | Gamma Sterilization | Identical |
| Shelf Life | 8 years | Unknown | Similar |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Osstem Implant. The word "OSSTEM" is in orange, with a stylized design. Below it, the word "IMPLANT" is written in gray, smaller font. The logo is simple and modern.
**OSSTEM Implant Co., Ltd.**
OSSTEM Implant Co., Ltd.
66-16, Bansong-ro 513beon-gil, Haeundae-gu, Busan, Republic of Korea
Tel: +82 51 850 2500 Fax: +82 51 861 4693 www.osstem.com
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Osstem Implant. The word "OSSTEM" is written in orange, with a small superscript "6" next to the "M". Below "OSSTEM", the word "IMPLANT" is written in gray. The logo is simple and modern.
OSSTEM Implant Co., Ltd.
66-16, Bansong-ro 513beon-gil, Haeundae-gu, Busan, Republic of Korea Tel: +82 51 850 2500 Fax: +82 51 861 4693 www.osstem.com
S.E. Similarities The proposed Tenting Screw has similar shape, is made of same material and has same sterilization method compared to the predicated Tent Screw of Neo GBR System (K160991). Differences The implanted diameter and the length of the proposed device and the predicated device are different. : . While its dimension is different from the predicated, the proposed Tenting Screw is made of same material and has same function and intended use; therefore, the proposed Tenting Screw is substantially equivalent to the predicated Tent Screw of Neo GBR System (K160991).
| | OssBuilder System | OssBuilder System | Remark |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
| 510(k) No. | Proposed | Predicated (K172354) | - |
| Manufacturer | Osstem Implant Co., Ltd. | Osstem Implant Co., Ltd. | Identical |
| Device Name | Healing Cap | Healing Cap Cover Cap | Identical |
| Design | Image: Healing Cap | Image: Healing Cap and Cover Cap | Identical |
| Intended Use | OssBuilder System is a metal device intended for use with a dental implant to stabilize and support of bone graft in dento-alveolar bony defect sites. | OssBuilder System is a metal device intended for use with a dental implant to stabilize and support of bone graft in dento-alveolar bony defect sites. | Identical |
| Principle of Operation | Where Tenting Screw is placed in maxilla or mandible bone area, it is combined with OssBuilder (non-resorbable titanium membrane) and Healing Cap or Cover Cap. Healing Cap is intended by connecting with Tenting Screw to fix OssBuilder during osseointegration. Second surgery is required to remove them after bone regeneration. | Healing Cap and Cover Cap is used by Connecting with the OB Anchor to fix OssBuilder during osseointegration. Second surgery is required to remove them after bone regeneration. | Healing Cap can be connected with Tenting Screw or OB Anchor to fix OssBuilder during osseointegration. |
| Material | Ti-6Al-4V<br>(ASTM F136) | Pure Titanium Grade 4<br>(ASTM F67)<br>Ti-6Al-4V<br>(ASTM F136) | Identical |
| Diameter (mm) | 4.0, 5.0 | 4.0, 5.0<br>4.0 | Identical |
| Height (mm) | 7.0, 8.0 | 7.0, 8.0<br>4.3 | Identical |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Osstem Implant. The word "OSSTEM" is in large, bold, orange letters. Below it, the word "IMPLANT" is in smaller, gray letters. There is a small orange circle with a white number 6 in the upper right corner of the logo.
CM Imr nt Co.. I t
undae-gu Busan Republic of Korea 51 850 2500 Fax: +82 51 861 4693 www.osstem.com
| Sterilization | Gamma Sterilization | Gamma Sterilization | Identical |
|---------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------|-----------|
| Shelf Life | 8 years | 8 years | Identical |
| S.E. | Similarities<br>The proposed Healing Cap has same dimension, function and indication for use; and is<br>treated with same sterilization method and made by same manufacturer compared to the<br>predicated Healing Cap (K172354).<br><br>Differences<br>The proposed Healing Cap is made of Ti-6Al-4V (ASTM F136) while the predicated<br>Healing Cap is made to Ti Grade 4 (ASTM F67), but has made with same material<br>compared to the predicated Cover Cap.<br><br>∴ The propose Healing Cap has different only in raw material compared to the<br>predicated Healing Cap (K172354), but it has made with same material compared to the<br>predicated Cover Cap (K172354). Except for this change/difference, its function,<br>dimension, indication for use, etc. is same compared to the predicated; therefore, the<br>proposed Healing Cap is substantially equivalent to the predicated Healing Cap<br>(K172354). | | |
- 6. Indication for Use
The OssBuilder System is used for space maintenance and stabilization of bone grafting material in an alveolar bone defect site.
- 7. Summary of Non-Clinical Performance Testing
Non-clinical testing data are submitted, referenced, or relied upon to demonstrate substantial equivalence.
#### Biocompatibility evaluation
Biocompatibility evaluation was conducted per the FDA Guidance Document "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process." No additional biocompatibility testing for OssBuilder System was deemed necessary as the OssBuilder System is made of same material, chemical composition, manufacturing process and body contact with the predicated devices, Neo GBR System, K160991; and OssBuilder System, K172354.
#### Sterilization Validation and Shelf-life
All subject devices in OssBuilder System are delivered in sterile. Therefore, the shelf-life and sterilization validation was considered. We considered them by leveraging the data that of our prior submissions.
#### Mechanical Properties
Bench tests evaluated for OssBuilder System according to ASTM F543. Comparative mechanical testing was conducted to evaluate the mechanical properties including fracture load, insertion torque, axial pullout strength, torsional strength, breakage angle, removal torque, and self-tapping force following per ASTM F543 for worstcase insertion in block material that represents hard/soft bone. Mechanical properties
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image contains the logo for Osstem Implant. The logo is in orange and gray. The word "OSSTEM" is in orange, and the word "IMPLANT" is in gray.
OSSTEM Implant Co., Lt
66-16, Bansong-ro 513beon-gil, Haeundae-gu, Busan, Republic of Korea Tel: +82 51 850 2500 Fax: +82 51 861 4693 www.osstem.com
from bench tests of OssBuilder System found substantially equivalent to the predicate device.
- 8. Summary of Clinical Testing
No clinical studies are submitted.
- 9. Conclusion
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification, OSSTEM IMPLANT Co., Ltd. concludes that the OssBuilder System is substantially equivalent to the predicated devices as herein.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.