EMS AIR-FLOW MASTER PIEZON

K110173 · E.M.S Electro Medical Systems S.A · ELC · Apr 7, 2011 · Dental

Device Facts

Record IDK110173
Device NameEMS AIR-FLOW MASTER PIEZON
ApplicantE.M.S Electro Medical Systems S.A
Product CodeELC · Dental
Decision DateApr 7, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.4850
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The AIR-FLOW MASTER PIEZON combines the functions of an ultrasonic scaler and air-polishing unit within a single chassis. The AIR-FLOW MASTER PIEZON is intended for use in the following dental and periodontal applications: - Removing supra and subgingival calculus deposits and stains from teeth . - Periodontal pocket lavage with simultaneous ultrasonic tip movement . - Scaling and root planing . - Releasing crowns, bridges, inlays, and posts as well as condensing gutta percha . - Plugging for amalgam condensation . - Amalgam burnishing ● - Preparing, cleaning and irrigating root canals ● - Cavity preparation . - . Cementing inlays and onlays - Retrograde preparation of root canals . The AIR-FLOW MASTER PIEZON is intended for use in the cleaning and polishing of teeth by the projection of water, air, and dental powders onto the tooth surface. The device removes dental plaque, soft deposits, and surface stains from pits, grooves, interproximal spaces, or smooth surfaces of teeth. The AIR-FLOW MASTER PIEZON can be used for the following cleaning procedures: - · Plaque removal for placement of sealants - Surface preparation prior to bonding/cementation of inlays, crowns and . . veneers - Surface preparation prior to placing composite restorations . - Effective plaque and stain removal for orthodontic patients . - Cleaning prior to bonding ortho brackets ● - Cleaning implant fixture prior to loading ● - Stain removal for shade determination . - Plaque removal for fluoride treatment . - Plaque and stain removal prior to whitening procedure . The AIR-FLOW MASTER PIEZON is also intended for use as an air-polisher in patients suffering from periodontal disease. The AIR-FLOW MASTER PIEZON is indicated for the non-surgical removal of subgingival plaque in pockets up to 5 mm after initial periodontal treatment.

Device Story

Dental prophylaxis device combining ultrasonic scaler and air-polishing unit in single chassis. Inputs: electrical power, water, air, and prophylaxis powders (sodium bicarbonate or glycine). Operation: ultrasonic generator produces piezo-electric vibrations for scaling handpiece; air-polishing handpiece projects air/water/powder mixture. Used in dental clinics by practitioners. Practitioner selects function by removing specific handpiece from holder; footswitch controls activation/flow. Touch panels regulate air/powder/water flow rates. Output: mechanical vibration for scaling; abrasive spray for cleaning/polishing. Benefits: efficient plaque/stain removal, periodontal pocket lavage, and surface preparation for restorative procedures.

Clinical Evidence

No clinical data. Bench testing only.

Technological Characteristics

Ultrasonic scaler and air-polishing unit. Piezo-electric vibration generation. Handpieces: Piezon LED, AIR-FLOW, PERIO-FLOW. Compatible with sodium bicarbonate and glycine powders. Connectivity: footswitch, touch panel controls. Standards: IEC 60601-1, IEC 60601-1-2, ISO 17664, ISO 17665-1, ANSI/AAMI ST79, ISTA 2A.

Indications for Use

Indicated for dental/periodontal patients requiring calculus/stain removal, scaling, root planing, endodontic preparation, or surface cleaning/polishing. Includes patients with periodontal disease for non-surgical subgingival plaque removal in pockets up to 5 mm post-initial treatment.

Regulatory Classification

Identification

An ultrasonic scaler is a device intended for use during dental cleaning and periodontal (gum) therapy to remove calculus deposits from teeth by application of an ultrasonic vibrating scaler tip to the teeth.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K110173 APR - 7 2011 # 510(k) Summary for the E.M.S. Electro Medical Systems SA EMS AIR-FLOW MASTER PIEZON #### 1. SUBMITTER/510(K) HOLDER E.M.S. Electro Medical Systems SA Ch. de la Vuarpillière 31 CH - 1260 Nyon Switzerland Contact Person: Suzanne Fassio-Hardy 022 994 47 00 Telephone: January 17, 2011 Date Prepared: #### 2. DEVICE NAME EMS AIR-FLOW MASTER PIEZON Proprietary Name: Common/Usual Name: Ultrasonic Scaler/Air Polishing Unit Ultrasonic Scaler (21 CFR 872.4850, Product Code ELC)* Classification Name: *As will be discussed in Section 4, the EMS AIR-FLOW MASTER PIEZON combines of an ultrasonic scaler and air-polishing unit. The predicate AIR-FLOW S2, which was cleared with dual ultrasonic scaler and air-polishing functions, was classified by FDA as an ultrasonic scaler. #### 3. PREDICATE DEVICES The proposed EMS AIR-FLOW MASTER PIEZON is substantially equivalent to the following legally marketed medical devices: - E.M.S. Electro Medical Systems S.A., PIEZON MASTER 700 (K093000) . - E.M.S. Electro Medical Systems S.A.,AIR-FLOW HANDY PERIO (K092289) . - E.M.S. Electro Medical Systems S.A., AIR-FLOW MASTER (K082791) . - E.M.S. Electro Medical Systems S.A., AIR-FLOW MASTER STANDARD . (K073284) - E.M.S. Electro Medical Systems S.A., AIR-FLOW S2 (K900709) . - E.M.S. Electro Medical Systems S.A., miniMaster LED (K093723) . The instruments used with the AIR-FLOW MASTER PIEZON were previously cleared by FDA: {1}------------------------------------------------ - E.M.S. Electro Medical Systems S.A., miniMaster Ultrasonic Scaler (K050710) . - E.M.S. Electro Medical Systems S.A., EMS Kermit (K992504) . - E.M.S. Electro Medical Systems S.A., miniPiezon (K953026) . - E.M.S. Electro Medical Systems S.A., Piezon Master 400 (K896749) . - E.M.S. Electro Medical Systems S.A., PIEZON MASTER 600 (K022328) . ### 4. DEVICE DESCRIPTION The EMS AIR-FLOW MASTER PIEZON is a dental device that combines the functions of an ultrasonic scaler and air-polishing unit within a single chassis. The proposed device consists of the base control unit, hoses and connectors for two handpieces, and a foot control containing a four-position switch. There are two mounts on the top of the base unit, one for an irrigation liquid container and the other for one of the two air-polishing chambers supplied with the device. The AIR-FLOW MASTER PIEZON is supplied with the Piezon Handpiece LED for performing ultrasonic scaling functions. The proposed device is compatible with EMS instruments legally marketed for ultrasonic scaling procedures. The AIR-FLOW MASTER PIEZON is also supplied with the AIR-FLOW Handpiece and PERIO-FLOW Handpiece for performing air-polishing procedures. The proposed device is compatible with the AIR-FLOW CLASSIC (sodium bicarbonate), AIR-FLOW SOFT (glycine), and AIR-FLOW PERIO (glycine) prophylaxis powders. The AIR-FLOW MASTER PIEZON is supplied with accessories for attaching and removing instruments and nozzles from the handpieces and containers for storage and sterilization of the reusable components and accessories. ## 5. INTENDED USE The AIR-FLOW MASTER PIEZON combines the functions of an ultrasonic scaler and air-polishing unit within a single chassis. The AIR-FLOW MASTER PIEZON is intended for use in the following dental and periodontal applications: - Removing supra and subgingival calculus deposits and stains from teeth . - Periodontal pocket lavage with simultaneous ultrasonic tip movement . - Scaling and root planing . - Releasing crowns, bridges, inlays, and posts as well as condensing gutta percha . - Plugging for amalgam condensation . - Amalgam burnishing ● - Preparing, cleaning and irrigating root canals ● {2}------------------------------------------------ - Cavity preparation . - . Cementing inlays and onlays - Retrograde preparation of root canals . The AIR-FLOW MASTER PIEZON is intended for use in the cleaning and polishing of teeth by the projection of water, air, and dental powders onto the tooth surface. The device removes dental plaque, soft deposits, and surface stains from pits, grooves, interproximal spaces, or smooth surfaces of teeth. The AIR-FLOW MASTER PIEZON can be used for the following cleaning procedures: - · Plaque removal for placement of sealants - Surface preparation prior to bonding/cementation of inlays, crowns and . . veneers - Surface preparation prior to placing composite restorations . - Effective plaque and stain removal for orthodontic patients . - Cleaning prior to bonding ortho brackets ● - Cleaning implant fixture prior to loading ● - Stain removal for shade determination . - Plaque removal prior to fluoride treatment . - Plaque and stain removal prior to whitening procedure . The AIR-FLOW MASTER PIEZON is also intended for use as an air-polisher in patients suffering from periodontal disease. The AIR-FLOW MASTER PIEZON is indicated for the non-surgical removal of subgingival plaque in pockets up to 5 mm after initial periodontal treatment. #### 6. PRINCIPLES OF OPERATION The proposed AIR-FLOW MASTER PIEZON combines the functions of an ultrasonic scaler and air-polishing unit within a single chassis to provide a complete prophylaxis solution. The working instrument for both the scaling and air-polishing functions is the handpiece, which is connected to the control unit via a power cord. The AIR-FLOW MASTER PIEZON is configured with the scaling handpiece mounted in a holder on the left side of the control unit, and the air-polishing handpiece mounted on the right side of the control unit. The treatment method (Piezon ultrasonic scaling or Air-Flow air-polishing) is selected by removing the handpiece associated with the treatment method of interest from the holder. {3}------------------------------------------------ As with other ultrasonic scalers, the ultrasonic generator housed in the AIR-FLOW MASTER PIEZON control unit generates piezo-electric vibrations. Instruments mounted in the Piezon Handpiece LED vibrate with a controlled oscillatory movement when activated. Air-polishing procedures are carried out by placing the nozzle at the end of the handpiece at the treatment site and activating the handpiece. As with the ultrasonic scaling functions, the footswitch activates the flow of the air/water/powder mixture. When the footswitch is released, the air/water/powder flow stops immediately. The practitioner controls the air-polishing treatment using the touch panels on the control panel and the footswitch to regulate the air/powder flow rate and the flow rate of the water into the handpiece. #### 7. SUMMARY OF TECHNOLOGICAL CHARACTERISTICS COMPARED TO THE PREDICATE DEVICES The proposed EMS AIR-FLOW MASTER PIEZON is substantially equivalent to the previously cleared AIR-FLOW S2, which also combines ultrasonic scaling and airpolishing functions in a single chassis. The design and function of the ultrasonic scaling and air-polishing functional units of the proposed AIR-FLOW MASTER PIEZON are substantially equivalent to the PIEZON MASTER 700 ultrasonic scaler and the AIR-FLOW MASTER and AIR-FLOW MASTER STANDARD airpolishing units. Differences between the proposed AIR-FLOW MASTER PIEZON and the predicate PIEZON MASTER 700 ultrasonic scaler and AIR-FLOW MASTER and AIR-FLOW MASTER STANDARD air-polishing units are limited to those hardware and firmware modifications necessary to integrate these functionalities into a single chassis. Testing demonstrated that the AIR-FLOW MASTER PIEZON fulfills the prospectively defined performance specifications. The similarities in intended use, operational characteristics, and functional technological characteristics between the proposed AIR-FLOW MASTER PIEZON and the predicate AIR-FLOW S2, PIEZON MASTER 700, AIR-FLOW MASTER, and AIR-FLOW MASTER STANDARD lead to a conclusion of substantial equivalence between the proposed and predicate devices. A side-by-side comparison of the predicate devices and the proposed device is provided in Table 5-1 at the end of this section. #### 8. SUMMARY OF NON-CLINICAL PERFORMANCE TESTING AS BASIS FOR SUBSTANTIAL EQUIVALENCE Testing was performed to verify compliance of the AIR-FLOW MASTER PIEZON with the following standards: {4}------------------------------------------------ - ANSI/AAMI ST79:2006, "Comprehensive guide to steam sterilization and . sterility assurance in health care facilities" ' - ISO 17664 (2004), "Sterilization of Medical Devices Information to be . provided by the Manufacturer for the Processing of Resterilizable Medical Devices" - ANSI/AAMI/ISO 17665-1:2006 "Sterilization of health care products -- Moist . heat -- Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices" - IEC 60601-1:1988 +A1: 1991 + A2: 1995, "Medical electrical equipment, Part1: . General requirements for safety" - . IEC 60601-1-2, "Medical electrical equipment. Part 1 General requirements for safety - Part 2 Collateral standard: Electromagnetic compatibility: Requirements and Tests" (2007). - . . ISTA 2A (2008), "ISTA Preshipment Testing Procedures- Combination Tests for Packaged-Products 150 lb (68 kg) or Less" The results of this testing confirm that the AIR-FLOW MASTER PIEZON is safe and effective for the intended use described in Section 5. #### 9. SUMMARY OF CLINICAL TESTING AS BASIS FOR SUBSTANTIAL EQUIVALENCE No clinical testing was conducted for this submission. #### 10. SUMMARY OF OTHER INFORMATION No other information is available. #### 11. Conclusions Drawn from Non-Clinical and Clinical Tests Based on the information and supporting documentation provided in the premarket notification, the EMS AIR-FLOW MASTER PIEZON is substantially equivalent to the cited predicate devices. Testing demonstrates that the EMS AIR-FLOW MASTER PIEZON fulfills prospectively defined design and performance specifications. {5}------------------------------------------------ | Item for Comparison | Indications for Use | | | Treatment Site | Function | Mechanism of action | Components | Prophylaxis Powders for Use with System | |--------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|-------------------------------|--------------------------------------|----------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------| | EMS AIR-FLOW MASTER<br>PIEZON<br>Proposed | Intended for use in dental and periodontal applications performed by an ultrasonic scaler. | Intended for use in the cleaning and polishing of teeth by the projection of water, air, and dental powders onto the tooth surface. | Intended for the non-surgical removal of subgingival plaque in pockets up to 5 mm after initial periodontal treatment. | Subgingival and supragingival | Ultrasonic scaling and air-polishing | Ultrasonic energy Projection of water/air/powder mixture | Control Unit Powder chambers Irrigation liquid bottle Foot pedal Piezon Handpiece LED AIR-FLOW Handpiece PERIO-FLOW Handpiece PERIO-FLOW Slim Nozzle* Instruments | PERIO (Glycine) SOFT (Glycine) CLASSIC (Sodium Bicarbonate) | | EMS PIEZON MASTER 700<br>K093000 | Intended for use in dental and periodontal applications performed by an ultrasonic scaler. | | | Subgingival and supragingival | Ultrasonic scaling | Ultrasonic energy | Control Unit Irrigation liquid bottle Foot pedal Piezon Handpiece LED Instruments | Not applicable | | EMS AIR-FLOW MASTER<br>K082791 | Intended for use in the cleaning and polishing of teeth by the projection of water, air, and dental powders onto the tooth surface. | Intended for the non-surgical removal of subgingival plaque in pockets up to 5 mm after initial periodontal treatment. | | Subgingival and supragingival | Air-polishing | Projection of water/air/powder mixture | Control Unit Powder chambers Foot pedal AIR-FLOW Handpiece PERIO-FLOW Handpiece PERIO-FLOW Nozzle | PERIO (Glycine) SOFT (Glycine) CLASSIC (Sodium Bicarbonate) | | EMS AIR-FLOW MASTER<br>STANDARD<br>K073284 | Intended for use in the cleaning and polishing of teeth by the projection of water, air, and dental powders onto the tooth surface. | | | Supragingival | Air-polishing | Projection of water/air/powder mixture | Control Unit Powder chamber Foot pedal AIR-FLOW Handpiece | AIR-FLOW (Sodium bicarbonate) | | AIR-FLOW S2<br>K900709 | Intended for use in dental and periodontal applications performed by an ultrasonic scaler. | Intended for use in the cleaning and polishing of teeth by the projection of water, air, and dental powders onto the tooth surface. | | Supragingival | Ultrasonic scaling and air-polishing | Ultrasonic energy Projection of water/air/powder mixture | Control Unit Powder chamber Foot pedal Piezon Handpiece AIR-FLOW Handpiece Instruments | AIR-FLOW (Sodium bicarbonate) | able 5-1. Comparison Table for Determination of Substantial Equivalen : Cleared for marketing in K092289 (AIR-FLOW Handy Per anuary 17, 201 Page 5-6 Electro Medical Systems SA 510(k) EMS AIR-FLOW MASTER PIEZON {6}------------------------------------------------ #### DEPARTMENT OF HEALTH & HUMAN SERVICES Food and Drug Administration 10903 New Hampshire Avenue Document Control Room ~WO66-G609 Silver Spring, MD 20993-0002 EMS Electro Medical Systems, S.A. C/O Cynthia J. M. Nolte, PhD, RAC Principal Consultant Medical Device Consultants, Incorporated 49 Plain Street North Attleboro, Massachusetts 02760 APR - 7 2011 Re: K110173 Trade/Device Name: EMS AIR-FLOW MASTER PIEZON Regulation Number: 21 CFR 872.4850 Regulation Name: Ultrasonic Scaler Regulatory Class: II Product Code: ELC, EFB, and EJR Dated: January 17, 2011 Received: January 20, 2011 Dear Dr. Nolte: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {7}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, hn for Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {8}------------------------------------------------ # Indications for Use 510(k) Number (if known): #### Device Name: EMS AIR-FLOW MASTER PIEZON #### Indications for Use: The AIR-FLOW MASTER PIEZON combines the functions of an ultrasonic scaler and airpolishing unit within a single chassis. The AIR-FLOW MASTER PIEZON is intended for use in the following dental and periodontal applications: - Removing supra and subgingival calculus deposits and stains from teeth . - Periodontal pocket lavage with simultaneous ultrasonic tip movement . - Scaling and root planing . - Releasing crowns, bridges, inlays, and posts as well as condensing gutta percha . - Plugging for amalgam condensation . - Amalgam burnishing . - Preparing, cleaning and irrigating root canals . - Cavity preparation . - Cementing inlays and onlays . - · Retrograde preparation of root canals The AIR-FLOW MASTER PIEZON is intended for use in the cleaning and polishing of teeth by the projection of water, air, and dental powders onto the tooth surface. The device removes dental plaque, soft deposits, and surface stains from pits, grooves, interproximal spaces, or smooth surfaces of teeth. The AIR-FLOW MASTER PIEZON can be used for the following cleaning procedures: - plaque removal for placement of sealants . - surface preparation prior to bonding/cementation of inlays, onlays, crowns and veneers . - surface preparation prior to placing composite restorations . - effective plaque and stain removal for orthodontic patients . - cleaning prior to bonding ortho brackets . - cleaning implant fixture prior to loading . - · stain removal for shade determination - plaque removal prior to fluoride treatment . - plaque and stain removal prior to whitening procedure . The AIR-FLOW MASTER PIEZON is also intended for use as an air-polisher in patients suffering from periodontal disease. The AIR-FLOW MASTER PIEZON is indicated for the non-surgical removal of subgingival plaque in pockets up to 5 mm after initial periodontal treatment. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)Concurrence of CDRH, Office of Device Evaluation (ODE Division Sign-Off) Tivision of Anesthesiology, General Hospital nfection Control. Dental Devices 510(k) Number: K110173
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...