K260773 · Nakanishi, Inc. · ELC · Mar 10, 2026 · Dental
Device Facts
Record ID
K260773
Device Name
Varios Combi Pro2
Applicant
Nakanishi, Inc.
Product Code
ELC · Dental
Decision Date
Mar 10, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.4850
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The dental device is intended for the following application(s): [Tooth Polishing System] Removal of supragingival and subgingival accretions and deposits, and accretions and deposits on prostheses (including implants). [Ultrasonic Scalers] Removal of accretions and deposits from the surface of teeth and dental restorations (fillings and prostheses), finishing (preparation and adjustment/polishing), removal and cleaning of infected dental tissues, and root canal preparation.
Device Story
Varios Combi Pro2 is a dental unit integrating ultrasonic scaling and air-polishing; used in dental clinics by professionals. Input: electrical power, compressed air, water, and powder media. Operation: control unit drives piezoelectric transducer in ultrasonic handpiece to generate mechanical vibrations for scaling; powder handpiece directs air-powder-water mixture for cleaning. Features: automatic chamber detection (Prophy/Perio) for mode selection; Bluetooth-connected wireless foot control for hands-free operation of spray, power, and on/off. Output: mechanical vibration and abrasive spray for tooth surface cleaning/preparation. Benefits: efficient oral hygiene management and periodontal therapy. Healthcare providers use the device to remove hard deposits and stains, facilitating clinical procedures like root canal preparation and surface finishing.
Clinical Evidence
Bench testing only. Verification and validation performed for ultrasonic/air-polishing performance, reprocessing, software, electrical safety, EMC, and cybersecurity. Biocompatibility testing conducted per ISO 10993-1 for externally communicating devices with limited contact (<24 hours). No clinical data presented.
Technological Characteristics
Combined oral hygiene system; AC powered. Ultrasonic handpiece (VA2-LUX-HP Ti) uses piezoelectric transducer. Powder handpiece for air-polishing. Features touch panel display, Bluetooth wireless foot control. Materials biocompatible per ISO 10993. Sterilization: handpieces/tips/nozzles cleaned and sterilized via moist heat (ISO 17665-1). Software: IEC 62304 compliant.
Indications for Use
Indicated for dental patients requiring removal of supragingival/subgingival accretions, deposits, and stains from teeth, dental restorations, prostheses, and implants; also indicated for finishing, cleaning of infected dental tissues, and root canal preparation.
Regulatory Classification
Identification
An ultrasonic scaler is a device intended for use during dental cleaning and periodontal (gum) therapy to remove calculus deposits from teeth by application of an ultrasonic vibrating scaler tip to the teeth.
Predicate Devices
EMS AIRFLOW Prophylaxis Master, EMS AIRFLOW One (K190124)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
March 10, 2026
Nakanishi, Inc.
% Prithul Bom
Most Responsible Person
Regulatory Technology Services, LLC
1000 Westgate Dr. Suite #510k
Saint Paul, Minnesota 55114
Re: K260773
Trade/Device Name: Varios Combi Pro2
Regulation Number: 21 CFR 872.4850
Regulation Name: Ultrasonic scaler
Regulatory Class: Class II
Product Code: ELC, KOJ
Dated: March 9, 2026
Received: March 9, 2026
Dear Prithul Bom:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260773 - Prithul Bom
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K260773 - Prithul Bom
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
MICHAEL E. ADJODHA -S
Michael E. Adjodha, MChE, RAC, CQIA
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K260773 | ? |
| Please provide the device trade name(s). | | ? |
| Varios Combi Pro2 | | |
| Please provide your Indications for Use below. | | ? |
| The dental device is intended for the following application(s):
[Tooth Polishing System]
Removal of supragingival and subgingival accretions and deposits, and accretions and deposits on
prostheses (including
implants).
[Ultrasonic Scalers]
Removal of accretions and deposits from the surface of teeth and dental restorations (fillings and
prostheses), finishing
(preparation and adjustment/polishing), removal and cleaning of infected dental tissues, and root canal
preparation. | | |
| Please select the types of uses (select one or both, as
applicable). | ☑ Prescription Use (21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
| 510(k) #: K260773 | 510(k) Summary | Prepared on: 2026-03-05 |
| --- | --- | --- |
| Contact Details | | 21 CFR 807.92(a)(1) |
| Applicant Name | NAKANISHIINC. | |
| Applicant Address | 700 Shimohinata Kanuma Tochigi 322-8666 Japan | |
| Applicant Contact Telephone | +81-289-64-7277 | |
| Applicant Contact | Mr. Kikuchi Masaaki | |
| Applicant Contact Email | m-kikuchi@nsk-nakanishi.co.jp | |
| Correspondent Name | NAKANISHI INC. | |
| Correspondent Address | 700 Shimohinata Kanuma Tochigi 322-8666 Japan | |
| Correspondent Contact Telephone | +81-289-64-4325 | |
| Correspondent Contact | Ms. Kim Junghee | |
| Correspondent Contact Email | j-kim@nsk-nakanishi.co.jp | |
| Device Name | | 21 CFR 807.92(a)(2) |
| Device Trade Name | Varios Combi Pro2 (Varios Combi Pro2 Basic Set) | |
| Common Name | Ultrasonic scaler | |
| Classification Name | Scaler, Ultrasonic | |
| Regulation Number | 872.4850 | |
| Product Code(s) | ELC, KOJ, EJR | |
| Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| K190124 | EMS AIRFLOW Prophylaxis Master, EMS AIRFLOW One | ELC |
| Device Description Summary | | 21 CFR 807.92(a)(4) |
| Varios Combi Pro2 consists of the control unit, the foot control, the ultrasonic handpiece, the powder handpiece, and accessories. The control unit integrates two operational systems: an ultrasonic generator for scaling and a powder supply system for air polishing. It is used to control functions such as power output, irrigation flow, powder/air mixture, and operation mode. The foot control allows for "hands-free" control of the spray, power output, and turning the device on/off. The ultrasonic handpiece (VA2-LUX-HP Ti) incorporates a piezoelectric transducer that converts the electrical signal into mechanical vibrations and is also equipped with LED illumination. The powder handpiece directs a mixture of powder, air, and water to the tooth surface for cleaning. | | |
| Varios Combi Pro2 is intended for use in dental periodontal therapy and tooth surface cleaning for procedures such as scaling, root canal preparation, and the removal of stains and plaque. The device is designed to be used with the ultrasonic handpiece for removing hard deposits and with the powder handpiece for removing supragingival and subgingival accretions, deposits, and stains from teeth and prostheses. | | |
{5}
A feature of this product is the integration of ultrasonic scaling and powder polishing capabilities into a single unit, allowing for efficient workflow in oral hygiene management. The system automatically detects the attached chamber (Prophy or Perio) to select the appropriate powder operating mode. It also features Bluetooth connectivity with the foot control for wireless operation.
The product is supplied non-sterile. The handpieces, tips, and nozzles are to be cleaned and sterilized at a medical facility before use and are intended for repeated use.
## Intended Use/Indications for Use
**21 CFR 807.92(a)(5)**
The dental device is intended for the following application(s):
[Tooth Polishing System]
Removal of supragingival and subgingival accretions and deposits, and accretions and deposits on prostheses (including implants).
[Ultrasonic Scalers]
Removal of accretions and deposits from the surface of teeth and dental restorations (fillings and prostheses), finishing (preparation and adjustment/polishing), removal and cleaning of infected dental tissues, and root canal preparation.
## Indications for Use Comparison
**21 CFR 807.92(a)(5)**
The intended use of the Tooth Polishing System falls under the intended uses of the predicate device, which include the removal of accretions, deposits, and stains; cleaning of implant fixtures and prior to bonding of orthodontic appliances; and surface preparation.
The intended use of the Ultrasonic Scalers falls under the intended uses of the predicate device, which include the removal of supragingival and subgingival accretions, deposits, and stains; scaling; root planing; root canal preparation, cleaning, and irrigation; retrograde preparation; and cavity preparation.
Therefore, this does not constitute a new intended use.
## Technological Comparison
**21 CFR 807.92(a)(6)**
The Varios Combi Pro2 is a combined oral hygiene system driven by an AC power source, which supplies power to and controls the functions of compatible ultrasonic and powder handpieces. Its operating principle is identical to that of the Predicate device (K190124), wherein an electrical signal from the generator is converted into mechanical vibrations by a piezo transducer for scaling, and compressed air is used to spray a powder mixture for tooth polishing. This fundamental technology, as well as functions such as controlling ultrasonic output, powder spray, and irrigation volume, are shared with the Predicate device.
On the other hand, the subject device differs from the Predicate device in several aspects, including the output specifications, specific patient contacting materials, the adoption of a touch panel display for flow control, and the connection method for the wireless foot control. These modifications reflect user convenience and market strategy and do not alter the intended use or the fundamental scientific technology. It has been confirmed through compliance with relevant standards (such as IEC 60601-1, ISO 18397, and ISO 10993) and comparison with the Predicate device that these changes do not raise new questions of safety or effectiveness. Therefore, these minor differences do not affect the substantial equivalence of the device.
## Non-Clinical and/or Clinical Tests Summary & Conclusions
**21 CFR 807.92(b)**
Non-Clinical Tests:
The following non-clinical tests were performed to support substantial equivalence of the subject device.
Performance Test:
The subject device, Varios Combi Pro2, was subjected to verification and validation testing for ultrasonic scaler and air-polishing performance, reprocessing, software, electrical safety, EMC, and cybersecurity to support substantial equivalence. The results of these tests demonstrate compliance with the requirements of the following standards and guidance.
- ISO 18397 First edition 2016-01, Dentistry - Powered scaler
- ISO 20608 First edition 2018-04, Dentistry - Powder jet handpieces and powders
- ISO 17664-1:2021, Processing of health care products - Information to be provided by the medical device manufacturer for the processing of medical devices - Part 1: Critical and semi-critical medical devices
- ISO 17664-2 First edition 2021-02, Processing of health care products - Information to be provided by the medical device manufacturer for the processing of medical devices - Part 2: Non-critical medical devices.
- ISO 17665-1 First edition 2006-08-15, Sterilization of health care products - Moist heat - Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices
- IEC 62304 Edition 1.1 2015-06 CONSOLIDATED VERSION, Medical device software - Software life cycle processes
{6}
- IEC 60601-1 Edition 3.2 2020-08 CONSOLIDATED VERSION, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance - Note: This standard is recognized with relevant US national differences applied, see references #1 and #2 in the Relevant FDA Guidance and/or Supportive Publication section below.
- IEC 60601-1-2 Edition 4.1 2020-09 CONSOLIDATED VERSION, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- IEC 80601-2-60 Edition 2.0 2019-06, Medical electrical equipment - Part 2-60: Particular requirements for the basic safety and essential performance of dental equipment
- IEC 81001-5-1 Edition 1.0 2021-12, Health software and health IT systems safety, effectiveness and security - Part 5-1: Security - Activities in the product life cycle
- FDA guidance document "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling"
- FDA guidance document "Content of Premarket Submissions for Device Software Functions"
- FDA guidance document "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions"
## Biocompatibility Test:
Based on its Indications for Use, the patient-contacting components of the subject device, specifically the Ultrasonic Handpiece and Powder Handpiece, are classified as externally communicating devices with limited contact (less than 24 hours) with tissue, bone, and dentin. To support substantial equivalence, biocompatibility testing was conducted using the final finished device (or representative samples). The test results demonstrate compliance with the requirements of the following standards and guidance:
- ISO 10993-1 Fifth edition 2018-08, Biological evaluation of medical devices - Part 9: Framework for identification and quantification of potential degradation products
- FDA guidance document "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process"
N/A
Nonclinical tests concluded that the device is as safe, as effective, and performs as well as its predicates.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.