Mini Screws

K231502 · Proimtech Saglik Urunleri Anonim Sirketi · OAT · Sep 4, 2024 · Dental

Device Facts

Record IDK231502
Device NameMini Screws
ApplicantProimtech Saglik Urunleri Anonim Sirketi
Product CodeOAT · Dental
Decision DateSep 4, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3640
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Orthodontic Screw is indicated for use as a fixed anchorage point for attachment of orthodontic appliances to facilitate the orthodontic movement of teeth. It is used temporarily and is removed after orthodontic treatment has been completed. Screws are intended for single use only. The orthodontic screw is intended in patients aged 12 years and older.

Device Story

Mini Screws provide fixed anchorage in maxillary/mandibular bone for orthodontic appliances. Device consists of head (hexagonal socket for rotational force/attachment of chains/springs), neck (smooth surface), and body (threaded, cylindrical). Applied by orthodontists/dentists via hand tool or low-torque rotary instrument; typically self-drilling. SLA surface treatment (sandblasted, large-grid, acid-etched) on apical half increases stability. Used temporarily during treatment; removed upon completion. Benefits include firm anchorage to facilitate tooth movement. Single-use, sterile, prescription-only.

Clinical Evidence

Bench testing only. Mechanical strength testing performed per TS EN ISO 19023 and ASTM F543. Biocompatibility evaluated per EN 10993-1. Sterilization validated per EN 11137-1/2.

Technological Characteristics

Material: Pure Titanium Grade 4 (ISO 5832-2, ASTM F67-13). Surface: SLA (sandblasted, large-grid, acid-etched) on apical 1/2. Dimensions: 1.4-2.0mm diameter, 5-16mm length. Connection: Hexagon socket. Sterilization: Gamma radiation. Packaging: Class 100,000 clean room (ISO 11607).

Indications for Use

Indicated for patients aged 12 years and older requiring fixed anchorage for orthodontic tooth movement. Contraindications: Not evaluated for MRI safety.

Regulatory Classification

Identification

An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.

Special Controls

*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2) *Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use; (ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads; (iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system; (iv) The device must be demonstrated to be biocompatible; (v) Sterility testing must demonstrate the sterility of the device; (vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment; (vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians; (viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and (ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue rectangle with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue. September 04, 2024 Proimtech Saglik Urunleri Anonim Sirketi Hakan Cevik General Manager Imes Sitesi, No:3 Dudullu Osb Mahallesi Imes 305.Sokak Istanbul. Istanbul 34773 TURKEY Re: K231502 Trade/Device Name: Proimtech Mini Screw Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: OAT Dated: May 12, 2024 Received: August 8, 2024 Dear Hakan Cevik: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" {1}------------------------------------------------ (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). {2}------------------------------------------------ Sincerely, Sherrill Lathrop Blitzer for Andrew Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K231502 Device Name Mini Screws Indications for Use (Describe) The Orthodontic Screw is indicated for use as a fixed anchorage point for attachment of orthodontic appliances to facilitate the orthodontic movement of teeth. It is used temporarily and is removed after orthodontic treatment has been completed. Screws are intended for single use only. The orthodontic screw is intended in patients aged 12 years and older. Type of Use (Select one or both, as applicable) | <span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | |----------------------------------------------------------------------------------------------------------| | <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ ## 510(k) Summary 510k number: K231502 Mini Screws Proimtech Saglik Urunleri Anonim Sirketi | Company Name | : | Proimtech Saglik Urunleri A.S. | |---------------------------|---|-----------------------------------------------------------------------------------| | Authorised Person / Title | : | Hakan Çevik / General Manager | | Address-1 | : | Dudullu OSB Mahallesi Imes 305 Sokak C Blok No:3 PK:34775<br>Umraniye / İstanbul | | Address-2 | : | Dudullu OSB Mahallesi Imes 501 Sokak E Blok No:25 PK:3477.<br>Umraniye / İstanbul | | Phone | : | +90 (216) 999 5041-44 | | Web/Mail | : | www.bilimplant.com / info@bilimplant.com | | Official Contact | : Hakan Cevik, General Manager of Proimtech Saglik Urunleri Anonim Sirket | |------------------|---------------------------------------------------------------------------| | Phone Number | : +90 535 355 51 00 | | E-Mail | : hakancevik@bilimplant.com | ### DEVICE INFORMATION | Preparation Date | : September 04, 2024 | |---------------------|-------------------------------------------| | Device Trade Name | : Mini Screws | | Common Name | : Mini Screw | | Classification Name | : screw, fixation, bone (21 CFR 872.3640) | | Device Class | : II | | Product Code | : OAT | | Regulation Number | : 872.3640 | | Review Panel | : Dental | ## INDICATION FOR USE The Orthodontic Screw is indicated for use as a fixed anchorage point for attachment of orthodontic appliances to facilitate the orthodontic movement of teeth. It is used temporarily and is removed after orthodontic treatment has been completed. Screws are intended for single use only. The orthodontic screw is intended in patients aged 12 years and older. {5}------------------------------------------------ ### DEVICE DESCRIPTION The Mini Screws system has been designed to provide a firm anchorage point in the oral cavity for orthodontic treatment. Mini screws are manufactured with CP Ti Gr 4 and are SLA treated. The Mini Screw system is designed to be applied with different orthodontic solutions and is versatile enough to satisfy different clinical cases. Titanium orthodontic mini screws are intended to provide fixed anchorage for orthodontic treatment used by implanting in the maxillary and mandibular bone. These mini screws used for anchoring purposes can be divided into 3 parts as head, neck and body. ## ► Head Part: Head part, it is the part where both the rotational force necessary for the screw to move in the bone is applied and where it is used as a fulcrum for the force to occur. The heads of the mini screws are designed in the shape of a hexagon in which the screwdriver bit is fully inserted. The hole in the head is designed for the attachment of chains, elastic wires or springs to initiate the necessary tooth movements. ## ► Neck Part: Neck part It is the part between the gingiva and the cortical bone, has no grooves and a shiny surface. The maximum compression of the screw occurs in this area and during insertion into the cortical bone. ## ► Body Part: Body part is the threaded part of the screw that remains in the bone. The inner part of the body (core) provides the resistance of the screw against fracture and the groove depth provides the resistance of the screw against pulling. The head part of the screws is cross-shaped or button-shaped and this part is the part where the necessary rotational force is applied to ensure that the screw advances in the bone. It is also the part where mechanics such as wire, spring or rubber are connected to the teeth to be moved. The neck part is designed to be smooth to make the use more comfortable, especially in patients with thin gingival biotype. The body part is designed as cylindrical and grooved in order to keep the screw fixed in the bone. The tip is angled conical for ease of application. ### > Length of Screws: Screw lengths vary between 5 mm and 16 mm. The length of the screw is determined by some factors such as the angle of application, thickness of the mucosa, bone quality and proximity of neighboring structures. ### > Screw Diameters: Screw diameters vary between 1.4 and 2 mm. The stability of the screws is affected by the screw surface area in contact with the cortical bone. As the diameter increases, the contact surface area will increase and the stability will also increase. {6}------------------------------------------------ ## > Surface Characteristics of Screws: The tips of the mini screws with 0 1.4 mm, Ø 1.65 mm and Ø 2 mm diameters used for orthodontic anchorage have a sandblasted and acid etched (SLA) surface structure and have a surface roughness of 1.5-2 Ra. Since similar screws on the market are not expected to be osseointegrated, the surfaces of most of them are not treated. However, in order to apply high anchorage force and increase screw stability, the ends of these products are treated with "Sandblasted, large-grid, acid etched" (SLA) process, which is sandblasting followed etching. ## > Groove Structure of Screws: Screws can be classified as self-tapping (ST) and self-drilling (SD) according to their thread design. Our mini screws, which are used for tooth mobilization are placed directly as SD without predrilling. However, in areas with dense cortical bone, the need for pilot slot preparation with drills may rarely be encountered. Our mini screws are placed with a hand tool or a low-torque rotary instrument (contraangle handpiece) at the recipient site. ## > Connection Type of Products: The connection of the mini screws features a hexagon socket. Image /page/6/Picture/6 description: The image contains three different views of a medical device. The first view shows a screw with a hexagonal head and a threaded shaft. The second view shows a spherical component with a hexagonal indentation. The third view shows a tool with a blade and a mechanism for gripping the screw. ## Hexagon Connection Structure Image /page/6/Picture/8 description: The image shows three different views of a screw. The first view is a side view of the screw, showing the threads and the head. The second view is a top-down view of the screw head, showing a cross-shaped indentation. The third view shows the screw being inserted into a piece of material using a screwdriver. ## Length of SLA Surface Treatment Area: All our mini screws have 1/2 apical part with SLA surface. {7}------------------------------------------------ #### MATERIALS: Only materials approved for medical purposes are used in the manufacture of products: Pure Titanium Grade 4 according to standard ISO 5832-2 and ASTM F67-13. ## PERFORMANCE AND SAFETY SPECIFICATION ### Surface Characterization: SLA is used to roughen the surface by applying a strong acid to the surface sandblasted with coarse grainy particles. The SLA technique has been used as an alternative to TPS coatings to achieve better surface chemistry and topography. This process is applied in combination after osseointegration to achieve macroroughness and micropits, as well as grit-sandblasting and acid etching to increase surface roughness. ### MRI Compatibility: The Mini Screw has not been evaluated for safety in the MR environment. It has not been tested for heating or unwanted movement in the MR environment. The safety of Mini Screws in the MR environment is unknown. Performing an MR exam on a person who has this medical device may result in injury or device malfunction. The product design has been made to include all kinds of safety and performance requirements considering the intended use and location of the product. In this direction, especially; - √ The product is manufactured from pure titanium (Grade 4) raw material in accordance with EN 1642, TS EN ISO 19023, ISO 5832-2 and ASTM F67-13 Standard. - > Our products are packaged in a Class 100,000 Clean Room according to ISO 11607 standard to ensure microbiological safety. They are then sterilized with Gamma radiation in accordance with the parameters specified in our user manual. - > The gamma sterilization method declared by us has been validated in accordance with EN 11137-1/2 Standard. - > The biocompatibility of our finished products has been carried out in accordance with EN 10993-1 Standard. - > Tests for the mechanical strength of our products have been carried out in accordance with TS EN ISO 19023 and ASTM F543 Standard. ### Primary Predicate Device and Reference Device For primary device Mini Screws: K161197; Orthodontic screw, OSSTEM Implant Co., Ltd. For reference device Mini Screws: K152297; OBS Anchorage Screw, BOMEI For reference device Mini Screws: K191041; Orthodontic Fixation Screw (Smart Anchor Miniscrew), GNI Co., LTD For reference device Mini Screw: K122069; Storm Mini Screw, Lancer Orthodontics {8}------------------------------------------------ For reference device Mini Screw: K161335; Dual Top Screw System, Jeil Medical Corporation # SUMMARIES OF TECHNOLOGICAL CHARACTERISTICS & SUBSTANTIAL EQUIVALENCE DISCUSSION {9}------------------------------------------------ | Technological<br>Characteristics | Subject Device | Primary Predicate Device | Reference Device | Reference Device | Reference Device | Reference Device | | | | | | | removed after<br>orthodontic<br>treatment has<br>been<br>completed. The<br>Storm Mini<br>Screw is<br>provided sterile<br>and is intended<br>for single use<br>only. | | |----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|---------------------------------|------------------------------------------|---------------------------|-----------------------------------------------------------------------------------------------------------|---------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------| | | Mini Screws | Orthodontic screw | OBS Anchorage Screw | Orthodontic Fixation<br>Screw (Smart Anchor<br>Miniscrew) | Storm Mini<br>Screw | Dual Top<br>Screw<br>System | Design | Image: screw | Image: screw | Image: screw | Image: screw | Image: screw | Image: screw | Image: screw | | 510(k)<br>Number | K231502 | K161197 | K152297 | K191041 | K122069 | K161335 | Materi<br>al | Titanium (Ti Grade 4, ASTM F67) | | Ti-6Al-4V ELI, ASTM F 136 | Stainless Steel (ISO 5832-<br>1:2007/ ASTM F138-13)<br>Titanium Alloy (ISO 5832-<br>3:1996/ ASTM F136-13) | Ti-6Al-4V ELI, ASTM F 136 | Ti-6Al-4V ELI,<br>ASTM F 136 | Titanium<br>Alloy (ASTM<br>F136) | | Manufacturer | Proimtech Saglik Urunleri A.S | OSSTEM Implant Co., Ltd. | BOMEI | GNI Co., LTD | Lancer<br>Orthodontics | Jeil Medical<br>Corporation | Body<br>Diamet<br>er | Ø1.4 mm<br>Ø1.65 mm<br>Ø2 mm | Ø1.2 mm<br>Ø1.4 mm<br>Ø1.6 mm<br>Ø1.8 mm | | Ø1.5<br>Ø2 | Ø1.2 mm<br>Ø1.4 mm<br>Ø1.6 mm<br>Ø1.8 mm<br>Ø2 mm | Ø1.5 mm<br>Ø2 mm | Ø1.3 – Ø2.5 | | Product Code | OAT | OAT | OAT | OAT | OAT | OAT | | | | | | | | | | Indications for<br>Use | The Orthodontic Screw is indicated<br>for use as a fixed anchorage point for<br>attachment of orthodontic<br>appliances to facilitate the<br>orthodontic movement of teeth. It is<br>used temporarily and is removed<br>after orthodontic treatment has been<br>completed. Screws are intended for<br>single use only. The orthodontic<br>screw is intended in patients aged 12<br>years and older. | The Orthodontic Screw is<br>indicated for use as a fixed<br>anchorage point for<br>attachment of orthodontic<br>appliances to facilitate the<br>orthodontic movement of<br>teeth. It is used temporarily<br>and is removed after<br>orthodontic treatment has<br>been completed. Screws are<br>intended for single use only | Temporary anchorage<br>screws are intended to<br>provide a fixed anchorage<br>point for attachment of<br>orthodontic appliances to<br>facilitate the orthodontic<br>movement of the teeth. It<br>is used temporarily and is<br>removed after orthodontic<br>treatment has been<br>complete. | The Orthodontic Fixation<br>Screw[Smart Anchor<br>Miniscrew] is intended<br>for use as a temporary<br>anchor for orthodontic<br>treatment for use in<br>patients aged 12 and<br>older. | The Storm Mini<br>Screw is a<br>threaded<br>titanium dental<br>implant screw<br>intended to<br>provide a fixed<br>anchorage<br>point for !he<br>attachment .of<br>orthodontic<br>appliances and<br>facilltate the<br>orthodontic<br>movement of,<br>teeth. it is used<br>temporarily<br>and must be | The Dual<br>Top Screw<br>System is<br>intended<br>for use as a<br>temporary<br>anchor for<br>orthodontic<br>treatment<br>for use in<br>patients<br>aged 12<br>and older. | | | | | | | | | {10}------------------------------------------------ {11}------------------------------------------------ | Length | Ø1.4 x 6 mm<br>Ø1.4 x 7 mm<br>Ø1.4 x 8 mm<br>Ø1.4 x 9 mm<br>Ø1.4 x 10 mm<br>Ø1.65 x 6 mm<br>Ø1.65 x 7 mm<br>Ø1.65 x 8 mm<br>Ø1.65 x 9 mm<br>Ø1.65 x 10 mm<br>Ø1.65 x 11 mm<br>Ø1.65 x 12 mm<br>Ø2 x 6 mm<br>Ø2 x 7 mm<br>Ø2 x 8 mm<br>Ø2 x 9 mm<br>Ø2 x 10 mm<br>Ø2 x 11 mm<br>Ø2 x 12 mm<br>Ø2 x 14 mm | 6 mm<br>8 mm<br>10 mm | Ø1.5 x 8 mm<br>Ø1.5 x 10 mm<br>Ø1.5 x 12 mm<br>Ø2 x 8 mm<br>Ø2 x 10 mm<br>Ø2 x 12 mm<br>Ø2 x 14 mm | 6 mm<br>8 mm<br>10 mm | 8 mm<br>10 mm | 5 mm – 16<br>mm | |------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|-----------------------------------------------------|----------------------------------------------------------------------------------------------------------------------| | Single<br>Use | Yes | Yes | Yes | Yes | Yes | Yes | | Surface | Sandblasted Large grit Acid etched (SLA) | Acid etching | --- | Machined (not acid<br>etched) | --- | No surface<br>treatment | | Steriliz<br>ation | Radiation Sterile | Radiation Sterile | Non-Sterile (Steam<br>sterilized by user) | Non-Sterile (Steam<br>sterilized by user) | Sterile via beta<br>irradiation | Non-Sterile<br>(Steam<br>sterilized by<br>user) or<br>Gamma-<br>Sterilized | | Shelf<br>life | 5 years | 8 years | 5 years | 5 years | 5 years | - | | Target<br>population | Professional use only – qualified<br>dentists, orthodontist or oral<br>surgeon. Strictly reserved to<br>specialized and trained users. | Professional use only –<br>qualified dentists. Strictly<br>reserved to specialized and<br>trained users. | Professional use only -<br>qualified dentists. Strictly<br>reserved to specialized and<br>trained users. | Professional use only –<br>qualified dentists. Strictly<br>reserved to specialized<br>and trained users. | People having<br>orthodontia<br>work | Professional use only –<br>qualified<br>dentists.<br>Strictly<br>reserved to<br>specialized<br>and trained<br>users. | | Principl<br>e of<br>Operati<br>on | Anchorage mini screws are used to<br>move the teeth. | Orthodontic screw is inserted<br>into either jaw to help the<br>orthodontist move the correct<br>teeth and stop the wrong<br>teeth from moving in the<br>wrong direction. | Anchorage mini screws are<br>used to move the teeth. | Anchorage mini screws<br>are used to move the<br>teeth. | Dental and<br>orthodontic<br>specialists<br>offices | Anchorage<br>mini screws<br>are used to<br>move the<br>teeth. | | Substantial Equivalence Discussion | | | | | | | {12}------------------------------------------------ #### Similarities The device has similar characteristics compared to the following respects. Indication for use, product code, design, diameter, disposable, intended population and operating principle. #### Difference The differences between the subject device is the surface treatment. However, testing data such as mechanical testing provided in the submission show that these differences do not raise issues in performance. Additionally, the subject device are available in slightly different however the such dimensional changes do not raise any performance concerns. #### Conclusion On the basis of the above discussion, it is concluded that the device in question is substantially equivale device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...