K233203 · Collagen Matrix, Inc. · NPL · May 1, 2024 · Dental
Device Facts
Record ID
K233203
Device Name
Soft Tissue Augmentation Resorbable Matrix
Applicant
Collagen Matrix, Inc.
Product Code
NPL · Dental
Decision Date
May 1, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3930
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Soft Tissue Augmentation Resorbable (STAR) Matrix is intended to support localized gingival augmentation to increase keratinized tissue. STAR Matrix is indicated for: • Covering of implants placed in immediate or delayed extraction sockets; • Localized gingival augmentation to increase keratinized tissue (KT) around teeth and implants.
Device Story
STAR Matrix is a cross-linked, resorbable membrane derived from purified porcine Achilles tendon Type I collagen. It features a dual-layer structure: a smooth outer barrier layer and a porous inner layer for tissue integration. Used in periodontal, oral, and maxillofacial surgery, the device is trimmed by a clinician and sutured over extraction sockets or sites requiring gingival augmentation. The porous side is placed against host bone or periosteum. The matrix acts as a scaffold for cell invasion and tissue ingrowth, eventually resorbing as native tissue regenerates. It provides a physical barrier to support healing and increase keratinized tissue, benefiting patients by facilitating soft tissue reconstruction around dental implants and teeth.
Clinical Evidence
No clinical studies were required. Evidence includes bench testing (collagen characterization, suture strength, thermal stability, density) and a canine model study. The animal study compared the subject device to the predicate in critical soft tissue periodontal defects, evaluating healing via gross pathology and histology at 4 and 12 weeks. Results showed positive healing, native structure return, and complete resorption without adverse effects, demonstrating substantial equivalence.
Technological Characteristics
Material: Purified porcine Achilles tendon Type I collagen. Structure: Dual-layer (smooth barrier/porous matrix). Cross-linking: Formaldehyde. Sterilization: Gamma irradiation (SAL 10^-6). Dimensions: Various sizes (1.5x2.0 cm to 3.0x4.0 cm; 8-12 mm diameter). Thickness: 1.5-5 mm. Biocompatibility: ISO 10993 compliant. Shelf-life standards: ISO 11607, ASTM D4169, ASTM F-1886-09, ASTM F-88/F88-M15, ISTA 3A.
Indications for Use
Indicated for patients requiring localized gingival augmentation to increase keratinized tissue around teeth and implants, and for covering implants placed in immediate or delayed extraction sockets.
Regulatory Classification
Identification
Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.
Special Controls
*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 01, 2024
Collagen Matrix, Inc. Daniel Fernandez Regulatory Affairs Specialist 15 Thornton Rd. Oakland. New Jersey 07436
Re: K233203
Trade/Device Name: Soft Tissue Augmentation Resorbable Matrix Regulation Number: 21 CFR 872.3930 Regulation Name: Bone grafting material Regulatory Class: Class II Product Code: NPL Dated: April 03, 2024 Received: April 04, 2024
## Dear Daniel Fernandez:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Sherrill Lathrop Blitzer
for Andrew Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality
{2}------------------------------------------------
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K233203
#### Device Name
Soft Tissue Augmentation Resorbable (STAR) Matrix
Indications for Use (Describe)
The Soft Tissue Augmentation Resorbable (STAR) Matrix is intended to support localized gingival augmentation to increase keratinized tissue.
STAR Matrix is indicated for:
- · Covering of implants placed in immediate or delayed extraction sockets;
· Localized gingival augmentation to increase keratinized tissue (KT) around teeth and implants.
Type of Use (Select one or both, as applicable)
| <div> <span>☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> |
|----------------------------------------------------------------------------|
| <div> <span>☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Collagen Matrix, Inc. 510(k) Summary 510(k) Submission No.:K233203 Soft Tissue Augmentation Resorbable (STAR) Matrix Page 1 of 4
#### 510(k) SUMMARY
| 1. Applicant Information | |
|-------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant Name: | Collagen Matrix, Inc. |
| Address: | 15 Thornton Road<br>Oakland, New Jersey 07436 |
| Telephone: | (201) 405-1477 |
| Fax: | (201) 405-1355 |
| Contact Person: | Daniel Fernandez<br>Regulatory Affairs, Specialist<br>dfernandez@collagenmatrix.com |
| Date Prepared: | May 01, 2024 |
| 2. Name of the Device | |
| Device Trade Name: | Soft Tissue Augmentation Resorbable (STAR) Matrix |
| Device Common Name(s): | Collagen Dental Membrane |
| Device Classification Name: | Barrier, animal source, intraoral<br>872.3930<br>NPL<br>Class II |
| 3. Legally Marketed Devices to Which Substantial Equivalence is Claimed | |
| Predicate Device: | Geistlich Mucograft®, Geistlich Mucograft Seal®<br>Geistlich Pharma AG<br>K210280 |
| Reference Device(s): | Collagen Dental Membrane - Porcine Type I<br>Collagen<br>Collagen Matrix, Inc.<br>K110600<br>Porcine Mineral Collagen Composite<br>Collagen Matrix, Inc.<br>K202183 |
| 4. Description of the Device | |
The Soft Tissue Augmentation Resorbable Matrix (STAR Matrix) is a cross – linked, resorbable membrane engineered from highly purified Type I collagen fibers derived from Porcine Achilles Tendon for use in periodontal, oral, and maxillofacial surgery. STAR Matrix is composed of two structures: a smooth outer layer that acts as a barrier membrane and a porous matrix layer to allow cell invasion and tissue ingrowth. The product is oriented so that the porous layer is in contact with the host tissue bone/bone graft or periosteum to facilitate tissue integration. The
{5}------------------------------------------------
product is provided sterile, non-pyrogenic, and for single use only. Product is provided in various sizes where it can be easily trimmed for appropriate fit and sutured into place during surgery.
#### 5. Indications for Use
STAR Matrix is intended to support localized gingival augmentation to increase keratinized tissue.
STAR Matrix is indicated for:
- Covering of implants placed in immediate or delayed extraction sockets; -
- -Localized gingival augmentation to increase keratinized tissue (KT) around teeth and implants.
#### 6. Summary/Comparison of Technical Characteristics
The subject device has equivalent technological characteristics as the predicate Geistlich Mucograft® and Geistlich Mucograft Seal® (510(k) K210280) device.
| Feature | Soft Tissue<br>Augmentation<br>Resorbable (STAR)<br>Matrix<br>(Subject Device) | Geistlich Mucograft®,<br>Geistlich Mucograft<br>Seal®<br>(K210280)<br>(Predicate Device) | Collagen Dental<br>Membrane - Porcine<br>Tendon Type I Collagen<br>(K110600)<br>(Reference Device) | Device<br>Comparison |
|------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|
| Indications for<br>Use | STAR Matrix is intended<br>to support localized<br>gingival augmentation to<br>increase keratinized<br>tissue.<br>STAR Matrix is indicated<br>for:<br>• Covering of<br>implants placed<br>in immediate or<br>delayed<br>extraction<br>sockets;<br>Localized<br>gingival<br>augmentation to<br>increase<br>keratinized tissue<br>(KT) around<br>teeth and<br>implants. | Geistlich Mucograft and<br>Mucograft seal are<br>indicated for<br>covering of implants<br>placed in immediate or<br>delayed extraction sockets,<br>localized gingival<br>augmentation to increase<br>keratinized tissue (KT)<br>around teeth and implants,<br>alveolar ridge<br>reconstruction for<br>prosthetic treatment,<br>and recession defects for<br>root coverage. | Collagen Dental<br>Membrane - Porcine<br>Tendon Type I Collagen is<br>intended for use in oral<br>surgical procedures as a<br>resorbable material for<br>placement in the area of<br>dental implants, bone<br>defect or ridge<br>reconstruction to aid in<br>wound healing. | Similar |
| Product Code<br>& Regulation | NPL<br>872.3930<br>Bone Grafting Material | NPL<br>872.3930<br>Bone Grafting Material | NPL<br>872.3930<br>Bone Grafting Material | Identical |
| Material Origin | Porcine | Porcine | Porcine | Similar |
| Material<br>Composition | Primarily Type I Porcine<br>collagen | Primarily Type I Porcine<br>collagen | Primarily Type I Porcine<br>collagen | Similar |
| Product Form | Smooth outer layer<br>Porous matrix layer | Smooth "outer" surface<br>Porous "inner" structure | Membrane | Identical |
{6}------------------------------------------------
| Feature | Soft Tissue<br>Augmentation<br>Resorbable (STAR)<br>Matrix<br>(Subject Device) | Geistlich Mucograft®,<br>Geistlich Mucograft<br>Seal®<br>(K210280)<br>(Predicate Device) | Collagen Dental<br>Membrane - Porcine<br>Tendon Type I Collagen<br>(K110600)<br>(Reference Device) | Device<br>Comparison |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Sizes | 1.5 x 2.0 cm<br>2.0 x 3.0 cm<br>3.0 x 4.0 cm<br>10 x 20 mm<br>15 x 15 mm<br>15 x 25 mm<br>10 x 40 mm<br>20 x 40 mm<br>15 x 45 mm<br>15 x 40 mm<br>25 x 45 mm<br>Diameter: 8 - 12 mm | Mucograft: 15 x 20 mm,<br>20 x30 mm, and 30 x 40<br>mm<br>Mucograft Seal: 8mm and<br>12 mm diameter | 1.5 x 2.0 cm<br>2.0 x 3.0 cm<br>3.0 x 4.0 cm | Similar; The<br>additional<br>potential<br>sizes of the<br>subject<br>device fall<br>within the<br>surface<br>area of the<br>largest<br>device. |
| Device<br>Thickness | Approx. 1.5 to 5 mm | Approx. 2.5 to 5 mm | 0.2 - 0.6 mm | Similar |
| Thermal<br>Stability | $53 \pm 8$ | $31.4 \pm 0.20$ | $57 \pm 7$ | Different |
| Suturability<br>(kg) | ≥ 0.065 | 0.166 ± 0.0816 | ≥ 0.065 | Similar |
| Density | 0.047 ± 0.018 | 0.054 ± 0.003 | N/A | Similar |
| Cross-linking | Crosslinked with<br>formaldehyde | N/A | Crosslinked with<br>formaldehyde | Different |
| Biocompatibility | Biocompatible<br>Meets ISO 10993<br>biocompatibility series of<br>testing | Biocompatible<br>Meets ISO 10993<br>biocompatibility series of<br>testing | Biocompatible<br>Meets ISO 10993<br>biocompatibility series of<br>testing | Identical |
| Sterility | Sterile using Gamma<br>Irradiation; SAL 10-6 | Sterile using Gamma<br>Irradiation | Sterile using Gamma<br>Irradiation; SAL 10-6 | Identical |
| Pyrogenicity | Non-pyrogenic | Non-pyrogenic | Non-pyrogenic | Identical |
| Single Use/<br>Reuse | Single use only | Single use only | Single use only | Identical |
## 7. Performance Data
In vivo and in vitro testing of the subject device was conducted to demonstrate substantial equivalence of the subject device to its predicate device. The following performance data are provided in support of the substantial equivalence determination.
Viral inactivation to ensure product safety was performed in accordance with ISO 22442-3.
## Biocompatibility Testing
Biocompatibility testing was conducted in accordance with the testing matrix of ISO 10993-1: Cytotoxicity, ISO 10993-5 Genotoxicity, ISO 10993-3 and ISO 10993-12 Muscle Implantation, ISO 10993-6 Implantation Assessment in Canine Study, ISO 10993-6 Intracutaneous Reactivity, ISO 10993-10 Acute System Toxicity, ISO 10993-11 Genotoxicity ISO 10993-12 Pyrogenicity ISO 10993-11
{7}------------------------------------------------
A Toxicology Risk Assessment was performed to evaluate formaldehyde residuals. The assessment included an evaluation of information presented in scientific literature and biocompatibility testing conducted on the subject device as well as similar devices. The results of the assessment concluded that the amount of formaldehyde residual for single product use has been addressed. The level of formaldehyde residual present in the subject device is within the acceptable limit as found in previous FDA approved Collagen Matrix, Inc devices.
## Bench Testing
- Purified Collagen Characterization Testing -
- -Hydroxyproline Content
- Amino Sugar Content -
- Lipid Content -
- -SDS page for purified porcine collagen, from reference device K110600, was leveraged in support of characterization testing.
- -Heavy Metals testing was leveraged from reference device K202183 in support of characterization testing.
#### Product Characterization Testing -
- Suture Pull-out Strength -
- Thermal Transition Temperature -
- Density -
#### Animal Testing
The performance of the device was compared to that of the predicate device. Geistlich Mucograft® in a canine model, and evaluated by gross pathology and histological assessments. Critical soft tissue periodontal defects were made at the alveolar ridge, where devices were then implanted. Measurements of the defects were taken at baseline, 4 weeks, and 12 weeks, which included anterior, central, and posterior measurements of the keratinized and soft tissue thickness. Measurement and assessment of the keratinized tissue length in the defect site was made at each termination time point. Conclusion of the study indicates the positive healing performance of the device as compared to the predicate, with noticeable return to native structures and complete resorption of the device with no unresolved adverse effects. The results demonstrate performance substantially equivalent to the predicate device with regards to the indications for use.
#### Clinical Studies
Clinical performance data was not required to determine substantial equivalence.
## Sterilization / Shelf Life
The device is provided sterile in following with ISO 11737. Sterilization for the subject device is based off of and leveraged from reference device K110600.
The subject device shelf-life program was developed taking into consideration ISO 11607, ASTM D4169, ASTM F-1886-09, ASTM F-88/F88-M15, and ISTA 3A. The shelf life of the subject device is 26 months.
#### 8. Conclusions Drawn from Non-clinical Studies
The conclusions drawn from the nonclinical tests demonstrate that the device is substantially equivalent to its predicate devices.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.