K171050 · Geistlich Pharma AG · NPL · Nov 9, 2017 · Dental
Device Facts
Record ID
K171050
Device Name
Geistlich Fibro-Gide
Applicant
Geistlich Pharma AG
Product Code
NPL · Dental
Decision Date
Nov 9, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3930
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Geistlich Fibro-Gide® is intended for soft tissue augmentation. It is indicated for: - Localized gingival augmentation to increase keratinized tissue (KT) around teeth and implants - Alveolar ridge reconstruction for prosthetic treatment - Recession defects for root coverage
Device Story
Geistlich Fibro-Gide is a fully resorbable, porous, porcine-derived collagen matrix (spongious consistency). Implanted by licensed dentists in clinical settings for soft tissue augmentation. Device acts as a scaffold; porous structure enables cell invasion and soft tissue ingrowth; provides volume stability while withstanding mechanical stresses post-wound closure. Degrades over time via natural resorption, eliminating need for removal. Wettability and suturability facilitate integration into surrounding tissue. Benefits include increased keratinized tissue, ridge reconstruction, and root coverage, utilizing patient's own healing capacity.
Clinical Evidence
Evidence includes biocompatibility testing (ISO 10993), bench testing, animal studies (rat/dog models), and a controlled, parallel, randomized clinical study. Animal studies confirmed tissue integration and comparable degradation rates to the predicate. Clinical study in patients with insufficient soft tissue volume demonstrated safety and effectiveness for the indicated uses. All biocompatibility endpoints (cytotoxicity, irritation, sensitization, systemic toxicity, genotoxicity, pyrogenicity) passed.
Technological Characteristics
Resorbable porcine collagen matrix (Type I/III collagen, elastin). Chemically cross-linked with EDC and NHS. Porosity >90%. Dimensions: 15x20mm, 20x40mm (approx. 6mm thick). Sterile (Gamma-irradiation, SAL 10^-6). Standards: ISO 10993 (biocompatibility), ISO 11137 (sterilization), ISO 11607 (packaging), ISO 22442 (animal tissues), ASTM F1980 (aging), ASTM F2450-10 (tissue engineering).
Indications for Use
Indicated for soft tissue augmentation in patients requiring localized gingival augmentation to increase keratinized tissue around teeth and implants, alveolar ridge reconstruction for prosthetic treatment, or treatment of recession defects for root coverage.
Regulatory Classification
Identification
Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.
Special Controls
*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
November 9, 2017
Image /page/0/Picture/1 description: The image shows the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the FDA acronym in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.
Geistlich Pharma AG % Janice M. Hogan Regulatory Counsel Hogan Lovells US LLP 1835 Market Street, 29th Floor Philadelphia, Pennsylvania 19103
Re: K171050
Trade/Device Name: Geistlich Fibro-Gide Regulation Number: 21 CFR 872.3930 Regulation Name: Bone Grafting Material Regulatory Class: Class II Product Code: NPL Dated: October 12, 2017 Received: October 12, 2017
Dear Janice Hogan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
#### Page 2 - Janice Hogan
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Mary S. Runner -S
for
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known)
#### K171050
Device Name
Indications for Use (Describe)
Geistlich Fibro-Gide® is intended for soft tissue augmentation. It is indicated for:
- Localized gingival augmentation to increase keratinized tissue (KT) around teeth and implants A
- Alveolar ridge reconstruction for prosthetic treatment A
- Recession defects for root coverage A
Type of Use (Select one or both, as applicable)
Z Prescription Use (Part 21 CFR 801 Subpart D)
□ Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
#### GEISTLICH FIBRO-GIDE®
#### SPONSOR
Geistlich Pharma AG Bahnhofstrasse 40 CH-6110 Wolhusen Switzerland
| Contact Person: | Marco Steiner |
|-----------------|----------------------------|
| Company: | Geistlich Pharma AG |
| Phone Number: | Direct 011 41 41 492 67 64 |
| | Company 011 41 492 55 55 |
| E-mail: | marco.steiner@geistlich.ch |
| Date Prepared: | November 9, 2017 |
#### DEVICE NAME
| Proprietary Name: | Geistlich Fibro-Gide® |
|----------------------|-----------------------------------------|
| Common/Usual Names: | Porcine Collagen Matrix |
| Classification Name: | Barrier, animal source, intraoral (NPL) |
#### PREDICATE DEVICES
Geistlich Mucograft® (K102531) (Primary predicate device)
Geistlich Mucograft® and Geistlich Mucograft® Seal (K140518) (Reference device)
Geistlich Mucograft® (K073711, K012423) (Reference devices)
GENOSS Collagen Membrane (K102307) (Reference device)
#### DEVICE DESCRIPTION
Geistlich Fibro-Gide® is a fully resorbable, porous, collagen matrix of porcine origin of a spongious consistency. The collagen is extracted from veterinary certified pigs and is carefully purified to avoid antigenic reactions. The collagen scaffold is weakly crosslinked. Geistlich Fibro-Gide® is sterilized in double packaging by Gamma-irradiation.
Geistlich Fibro-Gide® is an implantable device intended for use in soft tissue augmentation procedures. As described in more detail below, the device is indicated
{4}------------------------------------------------
specifically for insufficient tissue volume at the alveolar ridge and for soft tissue recession. It has mechanical properties appropriate to withstand the mechanical stresses that occur after wound closure in soft tissue augmentation procedures, i.e., it has good volume stability and it withstands early resorption to allow the formation of new soft tissue and degrades over time. In addition, the matrix is designed with an appropriate thickness to provide sufficient space for the ingrowth of new soft tissue. Due to its good wettability, suturability and biological properties, the device becomes well integrated into the surrounding soft tissue.
## INTENDED USE AND INDICATIONS FOR USE
Geistlich Fibro-Gide® is intended for soft tissue augmentation. It is indicated for:
- > Localized gingival augmentation to increase keratinized tissue (KT) around teeth and implants
- Alveolar ridge reconstruction for prosthetic treatment A
- > Recession defects for root coverage.
## COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
Geistlich Fibro-Gide® and its primary predicate device Geistlich Mucograft® are implantable, collagen matrices (scaffolds) used in dental soft tissue augmentation procedures. Both devices consist of collagen fibers in a loose, porous arrangement to enable cell invasion and settlement, providing an environment to allow for soft tissue growth by using the patient's own healing capacities. Both devices are fully resorbable and do not require a second intervention for removal. Like its reference device, Geistlich Fibro-Gide® is cross-linked. The device dimensions are similar and both devices require adaptation to fit the defect size.
Bench testing was conducted to compare Geistlich Fibro-Gide® against its predicate device in terms of its appearance, porosity, amino acid and protein composition, capillarity and wettability, as well as trimming, suturing and degradation capabilities.
| Characteristic | Geistlich Fibro-Gide® | Geistlich Mucograft<br>(K102531 – Predicate<br>device; K140518, K073711,<br>K012423 – Reference<br>devices) | |
|------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Intended Use | | | |
| Intended Use | Soft Tissue Augmentation | Soft Tissue Augmentation | |
| Indications for Use | Geistlich Fibro-Gide® is | | |
| Characteristic | Geistlich Fibro-Gide® | Geistlich Mucograft<br>(K102531 - Predicate<br>device; K140518, K073711,<br>K012423 - Reference<br>devices) | |
| | intended for soft tissue<br>augmentation. It is indicated<br>for:<br>Localized gingival<br>augmentation to<br>increase keratinized<br>tissue (KT) around<br>teeth and implants Alveolar ridge<br>reconstruction for<br>prosthetic treatment Recession defects<br>for root coverage | Covering of implants<br>placed in immediate<br>or delayed extraction<br>sockets Localized gingival<br>augmentation to<br>increase keratinized<br>tissue (KT) around<br>teeth and implants Alveolar ridge<br>reconstruction for<br>prosthetic treatment Recession defects<br>for root coverage | |
| Physical Characteristics | | | |
| Dimensions | 15x20 mm, 20x40mm<br>(thickness app. 6mm) | 15x20 mm, 20x30 mm<br>(K102531)<br>30x40 mm (K073711,<br>K012423)<br>(thickness app. 2.5 - 5 mm) | |
| Form | Sponge-like matrix | Sponge-like matrix | |
| Color | White to almost white | Almost white | |
| Porosity | Very porous materials with<br>average pore values of 90%<br>or more. The average<br>surface area and the bulk<br>density of Geistlich Fibro-<br>Gide® is almost identical to<br>that of Geistlich Mucograft. | Very porous materials with<br>average pore values of 90%<br>or more. The average<br>surface area and the bulk<br>density of Geistlich<br>Mucograft is almost identical<br>to that of Geistlich Fibro-<br>Gide. | |
| Characteristic | Geistlich Fibro-Gide® | Geistlich Mucograft<br>(K102531 - Predicate<br>device; K140518, K073711,<br>K012423 - Reference<br>devices) | |
| Capillarity and wettability | Spontaneously wettable and<br>completely soaked in<br>aqueous solution in less<br>than one minute | Spontaneously wettable and<br>completely soaked in<br>aqueous solution in less<br>than one minute | |
| Cross-linking | Chemically cross-linked with<br>EDC and NHS | Not chemically cross-linked | |
| Easy to trim with surgical<br>instruments (to fit a<br>template) | Can be cut with surgical<br>instruments | Can be cut with surgical<br>instruments | |
| Can be sutured | Can be fixed with sutures,<br>as demonstrated by<br>sutureability testing | Can be fixed with sutures,<br>as demonstrated by<br>sutureability testing | |
| Composition Materials | | | |
| Raw Material | Porcine connective tissue, | Porcine connective tissue, | |
| | Porcine skin tissue | Porcine skin tissue | |
| Composition | Porcine collagen: Product is<br>produced from the same<br>intermediate collagenous<br>products | Porcine collagen: Product is<br>produced from the same<br>intermediate collagenous<br>products | |
| Collagen / Elastin | Major protein components<br>Collagen I and Collagen III<br>(no Collagen II) and Elastin | Major protein components<br>Collagen I and Collagen III<br>(no Collagen II) and Elastin | |
| Amino Acid Composition | Very similar with equal<br>amounts of amino acids | Very similar with equal<br>amounts of amino acids | |
| Fat | Trace (less than 0.5%) | Trace (less than 0.5%) | |
| Glycosaminoglycans | Trace (less than 0.5%) | Trace (less than 0.5%) | |
| Other proteins > 0.5% | None | None | |
| pH | 3-7 | 3-7 | |
| Characteristic | Geistlich Fibro-Gide® | Geistlich Mucograft<br>(K102531 - Predicate<br>device; K140518, K073711,<br>K012423 - Reference<br>devices) | |
| Other Characteristics | | | |
| Source of raw material | Identical porcine tissue | Identical porcine tissue | |
| Manufacture | Multistage validated, SOP<br>controlled purification<br>process | Multistage validated, SOP<br>controlled purification<br>process | |
| Manufacturing conditions | Quality systems regulation<br>(CFR Part 820) | Quality systems regulation<br>(CFR Part 820) | |
| Packaging | Conforms to ISO 11607,<br>Parts 1 and 2.<br>Sterile double layer<br>packaging, including<br>aluminum layer to protect<br>against vapor penetration. | Conforms to ISO 11607,<br>Parts 1 and 2.<br>Sterile double layer<br>packaging, in aluminum<br>pouch to protect against<br>vapor penetration. | |
| Manufacture / Packaging<br>location | Geistlich Pharma AG,<br>Wolhusen,<br>Switzerland | Geistlich Pharma AG,<br>Wolhusen,<br>Switzerland | |
| Sterilization location | Synergy Health, Daeniken,<br>Switzerland | Synergy Health, Daeniken,<br>Switzerland | |
| Sterility | Gamma irradiation<br>SAL 10-6; Device provided<br>sterile, for single use only | Gamma irradiation<br>SAL 10-6; Device provided<br>sterile, for single use only | |
| User | Restricted to licensed<br>dentists | Restricted to licensed<br>dentists | |
| Principles of Operation | | | |
| Principles of Operation | Implantable resorbable<br>collagen matrix (scaffold)<br>consisting of collagen fibers<br>in a loose, porous | Implantable resorbable<br>collagen matrix (scaffold)<br>consisting of collagen fibers<br>in a loose, porous | |
| Characteristic | Geistlich Fibro-Gide® | Geistlich Mucograft<br>(K102531 - Predicate<br>device; K140518, K073711,<br>K012423 - Reference<br>devices)…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.