BioMesh® Biodegradable GTR Barrier is intended for use in the surgical management and treatment of periodontal defects to aid in the regeneration and integration of tissue components in guided tissue regeneration procedures.
Device Story
BioMesh® Biodegradable GTR Barrier is a microporous, biodegradable membrane used in guided tissue regeneration (GTR) procedures. It is provided in various precut shapes and sizes, though surgeons may trim it to fit specific operative sites. The device is placed by a clinician during periodontal surgery to facilitate tissue regeneration and integration. It acts as a physical barrier to manage periodontal defects. The device is supplied sterile and is intended for single use only; it must not be resterilized. It is designed to be biocompatible and stable for 5 years.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing, including biocompatibility, toxicity, and tissue integration testing (BioMesh score 115 vs. Resolut score 110).
Technological Characteristics
Microporous membrane composed of polyglycolic acid coated with polylactic acid + polylactic glycolic acid copolymer. Features interconnective pores and an embossed surface. Supplied sterile via terminal EtO sterilization. Complies with various voluntary standards including USP, ASTM, and ISO 9000 series.
Indications for Use
Indicated for treatment of two or three wall vertical defects, Class II furcations, circumferential defects, recession type defects, and dehiscence defects associated with dental implants.
Regulatory Classification
Identification
Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.
Special Controls
*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
4/9/99
K990363
Regulatory Authority: Safe Medical Devices Act of 1990, 21 CFR 807.92
- Trade Name: 1.
BioMesh® Biodegradable GTR Barrier
Common Name(s): GTR Barrier
Classification Name(s): Unclassified device
- 2. Establishment Name & Registration Number: Name: SAMYANG CORPORATION Number: Pendina
- 3. Classification:
Device Class: ll Classification Panel: Dental Product Code(s): LYC
#### 4. Equivalent Predicate Device:
SAMYANG CORPORATION believes that BioMesh® Biodegradable GTR Barrier is substantially equivalent to Resolut Regenerative Material as characterized in K932866, K962624 and K973594.
- 1. Gore Resolut Regenerative Material, K932866, K962624 & K973594
Comparison of BioMesh® Biodegradable GTR Barrier brand GTR barrier to the predicate device(s) identified above based on available labeling, promotional materials, comparison testing, demonstrates the substantial equivalence of Biodegradable GTR Barrier.
#### 5. General Device Description:
BioMesh® Biodegradable GTR Barrier is provided as a single membrane type, but in a variety of precut shapes and sizes. Shapes and sizes. As noted on the graphic representation at the left, five different shapes are offered. The BioMesh® Biodegradable GTR Barrier is shape optimized to provide a ready made pattern well suited to the selected operative site needing a minimum of surgeon modification. If desired, the surgeon can trim any of the available shapes to fit unique or unusual circumstances. The BioMesh® Biodegradable GTR Barrier is made from polyglycolic acid with a coating of polylactic acid + polylactic glycolic acid copolymer. The membrane is constructed with interconnective pores with an embossed surface.
#### 6. Applicant Name & Address:
### SAMYANG CORPORATION
263 Yeonji-dong, Chongno-gu Seoul 110-725, Korea
{1}------------------------------------------------
#### 7. Company Contact:
Mr. Dong-Kee Yoo SAMYANG CORPORATION 263 Yeonji-dong, Chongno-gu Seoul 110-725. Korea Tel. 011.82.2.740.7296 Fax 011.82.2.743.6626
#### 8. Submission Correspondent:
Mr. David W. Schlerf Buckman Company, Inc. 200 Gregory Lane, Suite C-100 Pleasant Hill, CA 94523-3389 925.356.2640 - 925.356.2654 - fax
#### Performance Standards: ರು
United States Food and Drug Administration mandated performance standards for this device do not exist. Various voluntary performance standards are utilized. Voluntary standards utilized include U.S.P., ASTM, Standard Operating Procedures, vendor & process certification and qualification procedures, Quality Systems Regulations, ISO materials standards and ISO 9000 series quality regulations.
#### Storage, Packaging & Sterilization Information: 10.
Packaging. Packaging should be inspected on arrival for evidence of shipping damage. Damaged packaging may indicate the presence of unsafe product and it should not be used until carefully inspected. If the package or product is damaged, the product should not be used and should be returned. Once opened, the product should never be resterilized or reused. Storage. Product must be handled, stored and opened in such a way that it is protected from inadvertent damage or contamination. Rotate stock and observe shelf life dates. Discard when outdated or damaged. When used, the product must be placed into use following accepted surgical sterile technique. Sterilization. BioMesh® Biodegradable GTR Barrier is supplied "STERILE" and is terminally EtO sterilized in accordance with accepted standards.
| FEATURE | BioMesh | Resolut | SE? |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------|------------|
| Indications for Use: | Two or three wall vertical defects, Class II<br>furcations, Circumferrential defects, Recession<br>type defects, Dehisence defects associated with<br>dental implants | SAME | YES |
| Material: | polyglycolic acid with a coating of polylactic acid +<br>polylactic glycolic acid copolymer | Equivalent | YES |
| Structure: | microporous membrane made of biodegradable<br>polyglycolide, polylactide and d,l-lactide/glycolide<br>copolymer. The membrane is made up of<br>interconnective pores with an embossed surface. | Non-woven membrane yes, no<br>interconnecting pores, no<br>embossing | Equivalent |
| Tissue Integration: | Test result score 115 | Test result score 110 | YES |
| Biocompatability: | Good | Good | YES |
| Toxicity: | Good | Good | YES |
| Stability: | 5 Years | 5 Years | YES |
| Clinical Data: | Good | Good | YES |
| Manufacturer: | SAMYANG | GORE | YES |
| Product Code: | LYC | SAME | YES |
| K - Number: | Pending | K932866, K962624 & K973594 | YES |
#### 11. Summary Comparison Table:
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing to the right, stacked on top of each other. The profiles are connected and form a single, abstract shape. The profiles are surrounded by a circular border containing the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA".
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville, Maryland 20850
OCT 1 0 2007
Samyang Corporation C/O Mr. David W. Schlerf Buckman Company, Incorporated 200 Gregory Lane, Suite C-100 Pleasant Hill, California 94523-3389
Re: K990363
Trade Name: BioMesh® Biodegradable GTR Barrier Regulation Number: 872.3930 Regulation Name: Bone Grafting Material Regulatory Class: 2 Product Code: NPK Dated: December 28, 1998 Received: February 5, 1999
Dear Mr. Schlerf:
This letter corrects our substantially equivalent letter of April 9, 1999.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the require approval or on or a general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
### Page 2 -
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to continue marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Sayite y. Michailous.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Image /page/3/Picture/7 description: The image is a circular seal with the text "CALIFORNIA OLIVE ASSOCIATION" around the border. In the center of the seal, the text "1906-2006" is at the top, followed by the letters "COA" in a stylized font. Below the letters, the word "Centennial" is written, and there are three stars at the bottom of the seal.
ting Public Ho
{4}------------------------------------------------
## K990363 510(k) Number :_
## Device Name(s): BioMesh® Biodegradable GTR Barrier
# Intended Use(s) of the Device:
BioMesh® Biodegradable GTR Barrier is intended for use in the surgical management and treatment of periodontal defects to aid in the regeneration and integration of tissue components in guided tissue regeneration procedures.
## Indications for Use:
BioMesh® Biodegradable GTR Barrier is indicated for use in the treatment of the following specific conditions:
- 1. Two or three wall vertical defects.
- 2. Class II furcations.
- 3. Circumferrential defects.
- 4. Recession type defects.
- 5. Dehisence defects associated with dental implants
Concurrence of CDRH, Office of Device Evaluation (ODE)
*Susen Runse*
(Division Sign-Off)
Division of Dental. Infection Control,
OR
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY
and General Hospital Devices 510(k) Number
Prescription Use (Per 21 CFR 801.109)
Over-The-Counter Use
(Optional format 1-2-96)
SAMYANG/MESH.DOC
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.