The LOCATOR® F-Tx Attachment System is designed to support fixed, partial or full arch restorations on endosseous dental implants in the mandible or maxilla for the purpose of restoring masticatory function. It is used in fixed hybrid restorations that can be attached with a snap-in system. The LOCATOR® F-Tx Attachment System is compatible with the following implants: Implant Manufacturer Connection Type / Platform Astra Tech OsseoSpeed Plus EV: 3.6, 4.2, 4.8 mm BioHorizons Internal Hex: 3.0, 3.5, 4.5, 5.7 mm Biomet 3i Osseotite® Certain®: 3.4, 4.1, 5.0, 6.0 mm Camlog Camlog, Conelog: 3.3, 3.8, 4.3, 5.0 mm iSy: 3.8, 4.4, 5.0 mm Dentsply Ankylos®: 3.5, 4.5, 5.5, 7.0 mm MIS Implants Internal Hex: 3.75, 4.5 mm Nobel Biocare NobelActive™: 3.0, 3.5, 3.9 mm Replace™ Select: 3.5, 4.3, 5.0, 6.0 mm Straumann Brånemark: 3.5, 4.1, 5.1 mm Bone Level: 3.3, 4.1, 4.8 mm Standard: 4.8, 6.5 mm Zimmer Tapered Screw-Vent®: 3.5, 4.5, 5.7 mm
Device Story
LOCATOR F-Tx Attachment System provides rigid connection for fixed, partial, or full arch dental restorations; utilizes snap-in attachment mechanism. System components include titanium alloy abutments (various gingival cuff heights) and PEEK healing caps. Abutments interface with specific endosseous dental implants; optional TiN or TiCN coatings applied to abutments. Clinician installs abutments onto implants; hybrid restoration snaps onto abutment. Unlike predicate devices removable by patients, F-Tx restorations are fixed and require clinician-only removal using specialized tool. System benefits include secure retention for hybrid restorations, facilitating masticatory function restoration.
Clinical Evidence
No clinical data. Bench testing included sterilization validation (ISO 17665-1/2), biocompatibility (ISO 10993-1/5/12), and mechanical denture retention testing. Retention strength of F-Tx system with High Retention Balls was statistically greater than tensile force of masticating worst-case sticky food (p<0.05).
Technological Characteristics
Abutments: Ti-6Al-4V ELI (ASTM F136); optional TiN or TiCN coating. Healing caps: PEEK. Interface: Conical, External Hex, Internal Hex, Internal Multi Lobe. Divergence allowance: 20°. Sterilization: Moist heat.
Indications for Use
Indicated for patients requiring fixed, partial, or full arch restorations supported by endosseous dental implants in the mandible or maxilla to restore masticatory function.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
Predicate Devices
Locator Implant Anchor Abutment for Endosseous Dental Implant (K072878)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol is a stylized representation of a human figure, with three profiles overlapping to create a sense of depth and connection.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 4, 2016
Zest Anchors, LLC c/o Kevin A. Thomas, Ph.D. Regulatory Affairs PaxMed International, LLC 12264 El Camino Real, Suite 400 San Diego, CA 92130
Re: K151789
Trade/Device Name: LOCATOR® F-Tx Attachment System Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: February 2, 2016 Received: February, 03, 2106
Dear Dr. Thomas:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Kevin A. Thomas, Ph.D.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and Part 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours, Tina Kiang
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
S
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known)
K151789
#### Device Name
LOCATOR® F-Tx Attachment System
#### Indications for Use (Describe)
The LOCATOR® F-Tx Attachment System is designed to support fixed, partial or full arch restorations on endosseous dental implants in the mandible or maxilla for the purpose of restory function. It is used in fixed hybrid restorations that can be attached with a snap-in system.
The LOCATOR® F-Tx Attachment System is compatible with the following implants: Implant Manufacturer Connection Type / Platform
| Astra Tech | OsseoSpeed Plus EV: 3.6, 4.2, 4.8 mm |
|---------------|---------------------------------------------------------------------|
| BioHorizons | Internal Hex: 3.0, 3.5, 4.5, 5.7 mm |
| Biomet 3i | Osseotite <sup>®</sup> Certain <sup>®</sup> : 3.4, 4.1, 5.0, 6.0 mm |
| Camlog | Camlog, Conelog: 3.3, 3.8, 4.3, 5.0 mm |
| | iSy: 3.8, 4.4, 5.0 mm |
| Dentsply | Ankylos <sup>®</sup> : 3.5, 4.5, 5.5, 7.0 mm |
| MIS Implants | Internal Hex: 3.75, 4.5 mm |
| Nobel Biocare | NobelActive <sup>™</sup> : 3.0, 3.5, 3.9 mm |
| | Replace <sup>™</sup> Select: 3.5, 4.3, 5.0, 6.0 mm |
| | Brånemark: 3.5, 4.1, 5.1 mm |
| Straumann | Bone Level: 3.3, 4.1, 4.8 mm |
| | Standard: 4.8, 6.5 mm |
| Zimmer | Tapered Screw-Vent <sup>®</sup> : 3.5, 4.5, 5.7 mm |
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW! *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary ZEST Anchors, LLC LOCATOR® F-Tx Attachment System
March 1, 2016
### ADMINISTRATIVE INFORMATION
| Manufacturer Name | Zest Anchors, LLC<br>2061 Wineridge Place<br>Escondido, CA 92029 |
|---------------------------|------------------------------------------------------------------|
| | Telephone: +1 (760) 743-7744 ext. 140<br>Fax: +1 (760) 743-7975 |
| Official Contact | Annie Wright<br>Regulatory Affairs Manager |
| Representative/Consultant | Kevin A. Thomas, PhD<br>Linda K. Schulz, RSDH, RDH |
| ntative/Consultant | Kevin A. Thomas, PhD<br>Linda K. Schulz, BSDH, RDH |
|---------------------------------|----------------------------------------------------|
| PaxMed International, LLC | |
| 12264 El Camino Real, Suite 400 | |
| San Diego, CA 92130 | |
| Telephone: | +1 (858) 792-1235 |
| Fax: | +1 (858) 792-1236 |
| Email: | KThomas@paxmed.com<br>LSchulz@paxmed.com |
# DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name | LOCATOR® F-Tx Attachment System |
|---------------------------|------------------------------------|
| Common Name | Dental implant abutment |
| Classification Name | Endosseous dental implant abutment |
| Classification Regulation | Class II, 21 CFR 872.3630 |
| Product Code | NHA |
| Classification Panel | Dental Products Panel |
| Reviewing Branch | Dental Devices Branch |
# PREDICATE DEVICE INFORMATION
The primary predicate is K072878. The reference predicate device for the TiCN coating is K150295. The reference predicate devices for the abutment-implant interface compatibilities are K120414, K071638, K093321, K143022, K063341, K000100, K083496, K113779, K133991, K041509, K040807, K071370, K102436, K020646, K022562, K062129, K130222, and K061410.
{4}------------------------------------------------
# INTENDED USE
The LOCATOR® F-Tx Attachment System is designed to support fixed, partial or full arch restorations on endosseous dental implants in the mandible or maxilla for the purpose of restoring masticatory function. It is used in fixed hybrid restorations that can be attached with a snap-in system.
The LOCATOR® F-Tx Attachment System is compatible with the following implants: Implant Manufacturer Connection Type / Platform
| Astra Tech | OsseoSpeed Plus EV: 3.6, 4.2, 4.8 mm |
|---------------|--------------------------------------------------------------------------------------|
| BioHorizons | Internal Hex: 3.0, 3.5, 4.5, 5.7 mm |
| Biomet 3i | Osseotite® Certain®: 3.4, 4.1, 5.0, 6.0 mm |
| Camlog | Camlog, Conelog: 3.3, 3.8, 4.3, 5.0 mm<br>iSy: 3.8, 4.4, 5.0 mm |
| Dentsply | Ankylos®: 3.5, 4.5, 5.5, 7.0 mm |
| MIS Implants | Internal Hex: 3.75, 4.5 mm |
| Nobel Biocare | NobelActive™: 3.0, 3.5, 3.9 mm<br>Replace™ Select: 3.5, 4.3, 5.0, 6.0 mm |
| Straumann | Brånemark: 3.5, 4.1, 5.1 mm<br>Bone Level: 3.3, 4.1, 4.8 mm<br>Standard: 4.8, 6.5 mm |
| Zimmer | Tapered Screw-Vent®: 3.5, 4.5, 5.7 mm |
# DEVICE DESCRIPTION
The LOCATOR® F-Tx Attachment System is for rigid connection of fixed, partial and full arch restorations on endosseous dental implants using a snap-in or screw-retained attachment system. The system includes abutments and healing caps. LOCATOR F-Tx abutments are compatible with the implant systems, connection types, and platform sizes listed above, and are provided in various gingival cuff heights ranging from 1 to 6 mm. LOCATOR F-Tx System abutments are made of titanium alloy conforming to ASTM F136 Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications (UNS R56401). LOCATOR F-Tx abutments may be provided with an optional coating of either TiCN (titanium carbon nitride) or TiN (titanium nitride). LOCATOR F-Tx System Healing Caps are made of PEEK.
# PERFORMANCE DATA
Performance testing to demonstrate substantial equivalence included methods described in the following standards: ISO 17665-1 Sterilization of health care products – Moist heat – Part 1: Requirements for the development, validation, and routine control of a sterilization process for medical devices; ISO 17665-2 Sterilization of health care products - Moist heat - Part 2: Guidance on the application of ISO 17665-1; ISO 10993-1 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process; ISO 10993-5 Biological Evaluation of Medical Devices – Part 5: Tests for In Vitro Cytotoxicity; and
{5}------------------------------------------------
ISO 10993-12 Biological Evaluation of Medical Devices - Part 12: Sample Preparation and Reference Materials. Non-clinical testing data submitted, referenced, or relied upon to demonstrate substantial equivalence included: sterilization testing, characterization and biocompatibility testing of the TiCN coating, biocompatibility testing of the PEEK Healing Caps, and mechanical denture retention testing. The mechanical testing demonstrated the retention strength of the LOCATOR F-Tx Attachment System when using the High Retention Balls was statistically greater than the tensile force created when masticating worst case sticky food (p<0.05). No clinical data were included in this submission.
# EQUIVALENCE TO MARKETED DEVICES
LOCATOR F-Tx Attachment System is substantially equivalent in indications and design principles to the legally marketed predicate devices shown in the following table.
| K072878 | Locator Implant Anchor Abutment for Endosseous Dental Implant | Zest Anchors, Inc. |
|-------------------------------------------|---------------------------------------------------------------|-----------------------------------|
| K150295 | LOCATOR RTx | Zest Anchors, LLC |
| Predicates for Compatible Implant Systems | | |
| K120414 | OsseoSpeed™ Plus | Astra Tech AB |
| K071638 | BioHorizons Tapered Internal Implant System | BioHorizons Implant Systems, Inc. |
| K093321 | BioHorizons Laser-Lok 3.0 Implant System | BioHorizons Implant Systems, Inc. |
| K143022 | BioHorizons Tapered Internal Implants | BioHorizons Implant Systems, Inc. |
| K063341 | 3i OSSEOTITE® Certain® Dental Implants | Implant Innovations, Inc. |
| K000100 | CAMLOG Rootform Implant System | Altatec Biotechnologies |
| K083496 | CAMLOG Implant System Modified Implants and Abutments | Altatec GmbH |
| K113779 | CONELOG® Implant System | Altatec GmbH |
| K133991 | iSy® Implant System | Altatec GmbH |
| K041509 | ANKYLOS® Dental Implant System | Friadent GmbH |
| K040807 | MIS Dental Implant System | MIS - Implant Technologies Ltd. |
| K071370 | NobelActive Internal Connection Implant | Nobel Biocare AB |
| K102436 | NobelActive 3.0 | Nobel Biocare AB |
| K020646 | Replace™ HA Coated Implant | Nobel Biocare USA Inc. |
| K022562 | Various Brånemark System Implants | Nobel Biocare AB |
| K062129 | P.004 Implants | Institut Straumann AG |
| K130222 | Straumann® Dental Implant System | Straumann USA, LLC |
| K061410 | Zimmer Dental Implant System | Zimmer Dental Inc. |
# Table of Predicate Devices
The primary predicate is K072878.
The reference predicate device for the TiCN coating is K150295.
The reference predicate devices for the abutment-implant interface compatibilities are: K120414, K071638, K093321, K143022, K063341, K000100, K083496, K113779, K133991, K041509, K040807, K071370, K102436, K020646, K022562, K062129, K130222, and K061410.
{6}------------------------------------------------
| Comparison | Subject Device | Primary Predicate Device |
|--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Zest Anchors, LLC<br>LOCATOR® F-Tx Attachment System<br>K151789 | Zest Anchors, Inc.<br>Locator Implant Anchor Abutment<br>for Endosseous Dental Implant<br>K072878 |
| Indications for Use | The LOCATOR® F-Tx Attachment System is<br>designed to support fixed, partial or full arch<br>restorations on endosseous dental implants in<br>the mandible or maxilla for the purpose of<br>restoring masticatory function. It is used in<br>fixed hybrid restorations that can be attached<br>with a snap-in system.<br>See page 2 for complete list of compatible<br>dental implants. | The LOCATOR Implant Anchor Abutment<br>for Endosseous Dental Implants is appropriate<br>for use with overdentures or partial dentures<br>retained in whole or in part by endosseous<br>implants in the mandible or maxilla. |
| Design | | |
| Abutment Platform Diameter | 3.0 mm to 7.0 mm; Multiple Systems | 2.5 mm to 6.5 mm, Multiple Systems |
| Abutment Angle | Straight | Straight |
| Abutment/Implant Interface | Conical, External Hex, Internal Hex,<br>Internal Multi Lobe | Conical, External Hex, Internal Hex,<br>Internal Multi Lobe |
| Divergence Allowance | 20° | 20° |
| Prosthesis Attachment Type | PEEK Retention Ball attached to Denture<br>Attachment Housing | Nylon Insert engaged to Denture Attachment<br>Housing |
| Material | | |
| Abutment | Ti-6Al-4V ELI | Ti-6Al-4V ELI |
| Abutment Coating | TiN, TiCN | TiN |
| Prosthetic Retention Component | PEEK | Nylon |
### Comparison of Subject Device and Primary Predicate Device
LOCATOR F-Tx is substantially equivalent to Locator (K072878) and the reference predicate LOCATOR RTx (K150295) in that it uses the identical interface for compatible implant systems. The previous interfaces have been demonstrated through contractual agreements with the OEM implant manufacturers. The compatible dental implants with the corresponding interfaces are listed above in the Table of Predicate Devices. LOCATOR F-Tx and Locator (K072878) are each provided with varving cuff heights. The abutment/implant interfaces of all LOCATOR F-Tx abutments are identical to those of the corresponding Locator (K072878) abutments.
The subject device and primary predicate have slightly different Indications for Use language. However, the difference in language does not change the intended use of abutments. The subject device and primary predicate device abutments are intended to be used in conjunction with dental implants for the purpose of supporting partial or full arch restorations. The subject device is for use with fixed hybrid restorations removed only by the clinician, whereas the primary predicate device is for use with restorations that may be removed by the patient.
The subject device is indicated for fixed hybrid implant restorations using a snap-in connection. The LOCATOR F-Tx snap-in connection is similar to the Locator (K072878), but has a PEEK retention ball attached to the center of the denture attachment housing instead of a nylon insert
{7}------------------------------------------------
on the abutment. Removal of the hybrid denture restoration by the clinician requires the use of the LOCATOR F-Tx denture removal tool.
The TiN coating for the subject device is the same as that used on the predicate Locator device (K072728). The TiCN coating for the subject device is the same as that used on the reference predicate LOCATOR RTx (K150295).
# CONCLUSION
The data included in this submission demonstrate substantial equivalence to the predicate devices listed above.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.