Medentika CAD/CAM Abutments

K150203 · Medentika GmbH · NHA · Oct 23, 2015 · Dental

Device Facts

Record IDK150203
Device NameMedentika CAD/CAM Abutments
ApplicantMedentika GmbH
Product CodeNHA · Dental
Decision DateOct 23, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3630
Device ClassClass 2

Indications for Use

Medentika TiBase CAD/CAM Abutments are intended for use with dental implants as a support for single or multiple tooth prostheses in the maxilla or mandible of a partially or fully edentulous patient. Medentika Preface CAD/CAM Abutments are intended for use with dental implants as a support for single or multiple tooth prostheses in the maxilla or mandible of a partially or fully edentulous patient.

Device Story

Medentika CAD/CAM Abutments (TiBase and PreFace) are dental implant abutments used to support prosthetic restorations (crowns/bridges). TiBase is a two-piece base for zirconia superstructures; PreFace is a blank for milling patient-specific titanium abutments. Devices are non-sterile, intended for clinician sterilization. Compatible with eleven specific dental implant systems. Design involves precision implant/abutment interfaces matching predicate systems. Clinicians use CAD/CAM software to design the final abutment, which is then manufactured at a Straumann CARES validated milling center. Output is a custom abutment providing structural support for dental prostheses, aiding in patient oral rehabilitation.

Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing: engineering/dimensional analysis for implant compatibility, static and dynamic compression-bending fatigue testing per ISO 14801, sterilization validation per ISO 17665-1/2, and in vitro cytotoxicity testing per ISO 10993-5.

Technological Characteristics

Materials: Titanium alloy (abutment and screw). Design: CAD/CAM abutment blanks and TiBase two-piece systems. Dimensions: Diameters 3.0-7.0 mm, max angle 30°, max gingival height 6 mm. Connectivity: Compatible with specific dental implant systems via precision interfaces. Sterilization: Non-sterile, moist heat sterilization by clinician. Software: CAD/CAM design workflow.

Indications for Use

Indicated for partially or fully edentulous patients requiring single or multiple tooth prostheses supported by dental implants in the maxilla or mandible.

Regulatory Classification

Identification

An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.

Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 October 23, 2015 Medentika GmbH c/o Ms. Linda Schulz Regulatory Affairs PaxMed International, LLC 12264 El Camino Real, Suite 400 San Diego, California 92130 Re: K150203 Trade/Device Name: Medentika CAD/CAM Abutments Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: September 23, 2015 Received: September 25, 2015 Dear Ms. Schulz: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {2}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, >Tina Kiang -S for Erin I. Keith, M.S. Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K150203 Device Name Medentika CAD/CAM Abutments Indications for Use (Describe) Medentika TiBase CAD/CAM Abutments are intended for use with dental implants as a support for single or multiple tooth prostheses in the maxilla or mandible of a partially or fully edentulous patient. | Implant System Compatibility | Series | Implant Diameter (mm) | Platform Diameter (mm) | |----------------------------------|--------|-------------------------|---------------------------| | Nobel Biocare Replace™ Select | E | 3.5, 4.3, 5.0, 6.0 | 3.5, 4.3, 5.0, 6.0 | | Nobel Biocare NobelActive™ | F | 3.5, 4.3, 5.0 | 3.5, 3.9 (4.3), 3.9 (5.0) | | Biomet 3i Osseotite® Certain® | H | 3.25, 4.0, 5.0 | 3.4, 4.1, 5.0 | | Biomet 3i Osseotite® | I | 3.25, 3.75, 4.0, 5.0 | 3.4, 4.1, 5.0 | | Nobel Biocare Brånemark | K | 3.3, 3.75, 4.0, 5.0 | 3.5, 4.1, 4.1, 5.1 | | Straumann Bone Level | L | 3.3, 4.1, 4.8 | 3.3, 4.1, 4.8 | | Straumann Standard | N | 3.3, 4.1, 4.8 | 3.5(NNC), 4.8, 6.5 | | Zimmer Tapered Screw-vent® | R | 3.3, 3.7, 4.1, 4.7, 6.0 | 3.5, 4.5, 5.7 | | Astra Tech OsseoSpeed™ | S | 3.5, 4.0, 4.5, 5.0 | 3.5, 4.0, 4.5, 5.0 | | Dentsply Friadent® Frialit/XiVE® | T | 3.4, 3.8, 4.5, 5.5 | 3.4, 3.8, 4.5, 5.5 | | Dentsply Friadent® Ankylos® | Y | 3.5, 4.5, 5.5, 7.0 | 3.5, 4.5, 5.5, 7.0 | Medentika TiBase is intended for use with the Straumann® CARES® System. All digitally designed abutments for use with Medentika CAD/CAM Abutments are intended to be manufactured at a Straumann® CARES® validated milling center. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ # Indications for Use 510(k) Number (if known) K150203 Device Name Medentika CAD/CAM Abutments Indications for Use (Describe) Medentika Preface CAD/CAM Abutments are intended for use with dental implants as a support for single or multiple tooth prostheses in the maxilla or mandible of a partially or fully edentulous patient. | Implant System Compatibility | Series | Implant Diameter (mm) | Platform Diameter (mm) | |----------------------------------|--------|-------------------------|--------------------------------| | Nobel Biocare Replace™ Select | E | 3.5, 4.3, 5.0, 6.0 | 3.5, 4.3, 5.0, 6.0 | | Nobel Biocare NobelActive™ | F | 3.0, 3.5, 4.3, 5.0 | 3.0, 3.5, 3.9 (4.3), 3.9 (5.0) | | Biomet 3i Osseotite® Certain® | H | 3.25, 4.0, 5.0 | 3.4, 4.1, 5.0 | | Biomet 3i Osseotite® | I | 3.25, 3.75, 4.0, 5.0 | 3.4, 4.1, 5.0 | | Nobel Biocare Brånemark | K | 3.3, 3.75, 4.0, 5.0 | 3.5, 4.1, 4.1, 5.1 | | Straumann Bone Level | L | 3.3, 4.1, 4.8 | 3.3, 4.1, 4.8 | | Straumann Standard | N | 3.3, 4.1, 4.8 | 3.5(NNC), 4.8, 6.5 | | Zimmer Tapered Screw-vent® | R | 3.3, 3.7, 4.1, 4.7, 6.0 | 3.5, 4.5, 5.7 | | Astra Tech OsseoSpeed™ | S | 3.0, 3.5, 4.0, 4.5, 5.0 | 3.0, 3.5, 4.0, 4.5, 5.0 | | Dentsply Friadent® Frialit/XiVE® | T | 3.4, 3.8, 4.5, 5.5 | 3.4, 3.8, 4.5, 5.5 | Medentika PreFace is intended for use with the Straumann® CARES® System. All digitally designed abutments for use with Medentika CAD/CAM Abutments are intended to be manufactured at a Straumann® CARES® validated milling center. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW! * The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An aqency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {5}------------------------------------------------ # 510(k) Summary Medentika GmbH Medentika CAD/CAM Abutments # K150203 October 22, 2015 # ADMINISTRATIVE INFORMATION | Manufacturer Name | Medentika GmbH | |---------------------------|-------------------------------------------------| | | Hammweg 8-10 | | | 76549 Hügelsheim, Germany | | Telephone | +49 (0)7229-69912-0 | | Fax | +49 (0)7229-69912-20 | | Official Contact | Gerhard Polzer | | | Head of Quality Management / Regulatory Affairs | | Representative/Consultant | Linda K. Schulz | | | Floyd G. Larson | | | PaxMed International, LLC | | | 12264 El Camino Real, Suite 400 | | | San Diego, CA 92130 | | Telephone: | +1 (858) 792-1235 | | Fax: | +1 (858) 792-1236 | | Email: | LSchulz@paxmed.com | | | FLarson@paxmed.com | # DEVICE NAME AND CLASSIFICATION | Trade/Proprietary Name | Medentika CAD/CAM Abutments | |----------------------------|------------------------------------| | Common Name | Endosseous dental implant abutment | | Classification Name | Endosseous dental implant abutment | | Classification Regulations | 21 CFR 872.3630, Class II | | Product Code | NHA | | Classification Panel | Dental Products Panel | | Reviewing Branch | Dental Devices Branch | {6}------------------------------------------------ #### PREDICATES The primary predicate for the Medentika TiBase is K120822, Straumann® CARES® Variobase™ Abutment. The primary predicate for the Medentika PreFace is K052272, Straumann C.A.R.E.S. Titanium Abutment. Reference Predicates: | K111935 | Ti-Base for individual milled Zirconium Abutment, 2-CONnect Abutment for Bridges | |---------|----------------------------------------------------------------------------------| | K082545 | Straumann® NN CARES® Titanium Abutment and NN CARES® Ceramic Abutment | | K083192 | Inclusive® Titanium Abutment Blanks | | K073713 | Blue Sky Bio Dental Implant System | | K120822 | Straumann® CARES® Variobase™ Abutments | | K111935 | Ti-Base for individual milled Zirconium Abutment, 2-CONnect Abutment for Bridges | | K052272 | Straumann C.A.R.E.S. Titanium Abutment | | K082545 | Straumann® NN CARES® Titanium Abutment and NN CARES Ceramic Abutment | | K083192 | Inclusive® Titanium Abutment Blanks | | K073713 | Blue Sky Bio Dental Implant System | | K020646 | Replace™ HA Coated Implant | | K071370 | NobelActive™ Internal Connection Implant | | K102436 | NobelActive™ 3.0 | | K063341 | 3i OSSEOTITE® Certain® Dental Implants | | K063286 | OSSEOTITE® Dental Implants | | K022562 | Various Brånemark System Implants – Immediate Function Indication | | K062129 | P.004 Implants | | K130222 | Straumann® Dental Implant System SLActive and Roxoid Product Families | | K061410 | Zimmer Dental Implant System | | K101732 | Astra Tech Implant System | | K073075 | FRIADENT® Implant Systems | | K041509 | ANKYLOS® Dental Implant System | #### DEVICE DESCRIPTION The subject device includes two CAD/CAM abutment designs, the Medentika TiBase and the Medentika PreFace. The TiBase is a two-piece abutment used as a base when fabricating a zirconia superstructure and the PreFace is an abutment used in fabricating a full patient-specific abutment in titanium alloy. Both abutment designs are provided non-sterile and are intended to be sterilized by the clinician. Medentika Preface Abutment is available in diameters 3.0 mm to 7.0 mm. Medentika TiBase Abutment is available in diameters 3.25 mm to 7.0 mm. The specific diameters for each Series coordinate with the compatible implant systems and sizes listed below. TiBase is available in two post designs. TiBase Generation 1 has a conically shaped post that is 4.0 mm high and TiBase Generation 2 has a parallel walled post shape that is 5.5 mm high. PreFace is available in one cylinder height of 20 mm. The maximum angle for abutments fabricated using TiBase or PreFace is 30°, the maximum gingival height is 6 mm and the minimum post height is 4 mm. Medentika CAD/CAM Abutments are compatible with eleven dental implant systems. Each Medentika abutment series has a precision implant/abutment interface corresponding to the implant system predicate for that series. Compatible sizes are as follows: {7}------------------------------------------------ | Nobel Biocare Replace™ Select | E-Series | 3.5, 4.3, 5.0, 6.0 | |----------------------------------|----------|-------------------------------------------| | Nobel Biocare NobelActive™ | F-Series | 3.0, 3.5, 4.3, 5.0 (3.0 PreFace Only) | | Biomet 3i Osseotite® Certain® | H-Series | 3.25, 4.0, 5.0 | | Biomet 3i Osseotite® | I-Series | 3.25, 3.75, 4.0, 5.0 | | Nobel Biocare Brånemark | K-Series | 3.3, 3.75, 4.0, 5.0 | | Straumann Bone Level | L-Series | 3.3, 4.1, 4.8 | | Straumann Standard | N-Series | 3.3, 4.1, 4.8 | | Zimmer Tapered Screw-Vent® | R-Series | 3.3, 3.7, 4.1, 4.7, 6.0 | | Astra Tech OsseoSpeed™ | S-Series | 3.0, 3.5/4.0, 4.5/5.0 (3.0 PreFace Only) | | Dentsply Friadent® Frialit/XiVE® | T-Series | 3.4, 3.8, 4.5, 5.5 | | Dentsply Friadent® Ankylos® | Y-Series | 3.5, 4.5, 5.5, 7.0 (Y-Series TiBase Only) | #### PERFORMANCE DATA Non-clinical testing data submitted, referenced, or relied upon to demonstrate substantial equivalence included engineering analysis and dimensional analysis for determination of compatibility using critical dimension measurements of the original manufacturers' components, static and dynamic compression-bending festing for the Medentika TiBase and Medentika PreFace according to ISO 14801 Dentistry – Implants – Dynamic fatigue test for endosseous dental implants, sterilization according to ISO 17665-1 Sterilization of health care products – Moist heat – Part 1: Requirements for the development, validation, and routine control of a sterilization process for medical devices and ISO 17665-2 Sterilization of health care products - Moist heat -Part 2: Guidance on the application of ISO 17665-1 to demonstrate an SAL of 10*, and biocompatibility testing for cytotoxicty according to ISO 10993-5 Biological evaluation of medical devices -- Part 5: Tests for in vitro c vtotoxicity. All materials conform to FDA recognized standards and no further biocompatibility testing was performed. The performance data included in this submission demonstrate substantial equivalence to the predicate devices listed above. {8}------------------------------------------------ ### INTENDED USE TiBase Medentika TiBase CAD/CAM Abutments are intended for use with dental implants as a support for single or multiple tooth prostheses in the maxilla or mandible of a partially or fully edentulous patient. | Implant System Compatibility | Series | Implant Diameter (mm) | Platform Diameter (mm) | |----------------------------------|--------|-------------------------|---------------------------| | Nobel Biocare Replace™ Select | E | 3.5, 4.3, 5.0, 6.0 | 3.5, 4.3, 5.0, 6.0 | | Nobel Biocare NobelActive™ | F | 3.5, 4.3, 5.0 | 3.5, 3.9 (4.3), 3.9 (5.0) | | Biomet 3i Osseotite® Certain® | H | 3.25, 4.0, 5.0 | 3.4, 4.1, 5.0 | | Biomet 3i Osseotite® | I | 3.25, 3.75, 4.0, 5.0 | 3.4, 4.1, 5.0 | | Nobel Biocare Brånemark | K | 3.3, 3.75, 4.0, 5.0 | 3.5, 4.1, 4.1, 5.1 | | Straumann Bone Level | L | 3.3, 4.1, 4.8 | 3.3, 4.1, 4.8 | | Straumann Standard | N | 3.3, 4.1, 4.8 | 3.5(NNC), 4.8, 6.5 | | Zimmer Tapered Screw-vent® | R | 3.3, 3.7, 4.1, 4.7, 6.0 | 3.5, 4.5, 5.7 | | Astra Tech OsseoSpeed™ | S | 3.5, 4.0, 4.5, 5.0 | 3.5, 4.0, 4.5, 5.0 | | Dentsply Friadent® Frialit/XiVE® | T | 3.4, 3.8, 4.5, 5.5 | 3.4, 3.8, 4.5, 5.5 | | Dentsply Friadent® Ankylos® | Y | 3.5, 4.5, 5.5, 7.0 | 3.5, 4.5, 5.5, 7.0 | Medentika TiBase is intended for use with the Straumann® CARES® System. All digitally designed abutments for use with Medentika CAD/CAM Abutments are intended to be manufactured at a Straumann® CARES® validated milling center. PreFace Medentika Preface CAD/CAM Abutments are intended for use with dental implants as a support for single or multiple tooth prostheses in the maxilla or mandible of a partially or fully edentulous patient. | Implant System Compatibility | Series | Implant Diameter (mm) | Platform Diameter (mm) | |----------------------------------|--------|-------------------------|--------------------------------| | Nobel Biocare Replace™ Select | E | 3.5, 4.3, 5.0, 6.0 | 3.5, 4.3, 5.0, 6.0 | | Nobel Biocare NobelActive™ | F | 3.0, 3.5, 4.3, 5.0 | 3.0, 3.5, 3.9 (4.3), 3.9 (5.0) | | Biomet 3i Osseotite® Certain® | H | 3.25, 4.0, 5.0 | 3.4, 4.1, 5.0 | | Biomet 3i Osseotite® | I | 3.25, 3.75, 4.0, 5.0 | 3.4, 4.1, 5.0 | | Nobel Biocare Brånemark | K | 3.3, 3.75, 4.0, 5.0 | 3.5, 4.1, 4.1, 5.1 | | Straumann Bone Level | L | 3.3, 4.1, 4.8 | 3.3, 4.1, 4.8 | | Straumann Standard | N | 3.3, 4.1, 4.8 | 3.5(NNC), 4.8, 6.5 | | Zimmer Tapered Screw-vent® | R | 3.3, 3.7, 4.1, 4.7, 6.0 | 3.5, 4.5, 5.7 | | Astra Tech OsseoSpeed™ | S | 3.0, 3.5, 4.0, 4.5, 5.0 | 3.0, 3.5, 4.0, 4.5, 5.0 | | Dentsply Friadent® Frialit/XiVE® | T | 3.4, 3.8, 4.5, 5.5 | 3.4, 3.8, 4.5, 5.5 | Medentika PreFace is intended for use with the Straumann® CARES® System. All digitally designed abutments for use with Medentika CAD/CAM Abutments are intended to be manufactured at a Straumann® CARES® validated milling center. {9}------------------------------------------------ # EQUIVALENCE TO MARKETED DEVICE Medentika CAD/CAM Abutments are compatible with eleven dental implant systems. Each Medentika abutment series has an implant/abutment interface corresponding to the implant system predicate for that series. | Indications for Use | | | | | |--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|-------------------------|--------------------------------| | Subject Device | Medentika TiBase CAD/CAM Abutments are intended for use with dental implants as a support for single or multiple tooth prostheses in the maxilla or mandible of a partially or fully edentulous patient. | | | | | Medentika GmbH<br>Medentika CAD/CAM Abutments | Implant System Compatibility | Series | Implant Diameter (mm) | Platform Diameter (mm) | | | Nobel Biocare Replace™ Select | E | 3.5, 4.3, 5.0, 6.0 | 3.5, 4.3, 5.0, 6.0 | | | Nobel Biocare NobelActive™ | F | 3.5, 4.3, 5.0 | 3.5, 3.9 (4.3), 3.9 (5.0) | | | Biomet 3i Osseotite® Certain® | H | 3.25, 4.0, 5.0 | 3.4, 4.1, 5.0 | | | Biomet 3i Osseotite® | I | 3.25, 3.75, 4.0, 5.0 | 3.4, 4.1, 5.0 | | | Nobel Biocare Brånemark | K | 3.3, 3.75, 4.0, 5.0 | 3.5, 4.1, 4.1, 5.1 | | | Straumann Bone Level | L | 3.3, 4.1, 4.8 | 3.3, 4.1, 4.8 | | | Straumann Standard | N | 3.3, 4.1, 4.8 | 3.5(NNC), 4.8, 6.5 | | | Zimmer Tapered Screw-vent® | R | 3.3, 3.7, 4.1, 4.7, 6.0 | 3.5, 4.5, 5.7 | | | Astra Tech OsseoSpeed™ | S | 3.5, 4.0, 4.5, 5.0 | 3.5, 4.0, 4.5, 5.0 | | | Dentsply Friadent® Frialit/XiVE® | T | 3.4, 3.8, 4.5, 5.5 | 3.4, 3.8, 4.5, 5.5 | | | Dentsply Friadent® Ankylos® | Y | 3.5, 4.5, 5.5, 7.0 | 3.5, 4.5, 5.5, 7.0 | | Medentika TiBase is intended for use with the Straumann® CARES® System. All digitally designed abutments for use with Medentika CAD/CAM Abutments are intended to be manufactured at a Straumann® CARES® valid milling center. | | | | | | Subject Device | Medentika Preface CAD/CAM Abutments are intended for use with dental implants as a support for single or multiple tooth prostheses in the maxilla or mandible of a partially or fully edentulous patient. | | | | | Medentika GmbH<br>Medentika CAD/CAM Abutments | Implant System Compatibility | Series | Implant Diameter (mm) | Platform Diameter (mm) | | | Nobel Biocare Replace™ Select | E | 3.5, 4.3, 5.0, 6.0 | 3.5, 4.3, 5.0, 6.0 | | | Nobel Biocare NobelActive™ | F | 3.0, 3.5, 4.3, 5.0 | 3.0, 3.5, 3.9 (4.3), 3.9 (5.0) | | | Biomet 3i Osseotite® Certain® | H | 3.25, 4.0, 5.0 | 3.4, 4.1, 5.0 | | | Biomet 3i Osseotite® | I | 3.25, 3.75, 4.0, 5.0 | 3.4, 4.1, 5.0 | | | Nobel Biocare Brånemark | K | 3.3, 3.75, 4.0, 5.0 | 3.5, 4.1, 4.1, 5.1 | | | Straumann Bone Level | L | 3.3, 4.1, 4.8 | 3.3, 4.1, 4.8 | | | Straumann Standard | N | 3.3, 4.1, 4.8 | 3.5(NNC), 4.8, 6.5 | | | Zimmer Tapered Screw-vent® | R | 3.3, 3.7, 4.1, 4.7, 6.0 | 3.5, 4.5, 5.7 | | | Astra Tech OsseoSpeed™ | S | 3.0, 3.5, 4.0, 4.5, 5.0 | 3.0, 3.5, 4.0, 4.5, 5.0 | {10}------------------------------------------------ | Predicate Devices | | |----------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Institut Straumann<br>AG<br>Straumann®<br>CARES®<br>Variobase ™<br>Abutment<br>K120822 | The Straumann® CARES® Variobase™ Abutment is a two-piece dental abutment consisting of the Straumann®<br>Variobase ™ Abutment and the Straumann® CARES® Variobase™ Coping which is intended to be placed onto<br>Straumann dental implants to provide support for prosthetic reconstruction such as crowns and bridges.<br>Straumann® CARES® Variobase ™ Abutments are indicated for screw-retained single tooth or cement-retained<br>single tooth and bridge restorations.<br>The Straumann® CARES® Variobase ™ Coping polycon® ae in combination with the Straumann® Variobase ™<br>Abutment is indicated for temporary (up to 180 days) dental restoration of a Straumann dental implant. | | Straumann USA<br>Straumann<br>C.A.R.E.S.<br>Titanium<br>Abutment<br>K052272 | Abutments are intended to be placed into dental implants to provide support for prosthetic reconstructions such as<br>crowns or bridges. The Straumann C.A.R.E.S. Titanium Abutment is indicated for cemented restorations. The<br>abutment can be used in single tooth replacements and multiple tooth restorations. | The specific language of the Indications for Use Statement is not identical. However, the subject device and corresponding primary predicates are all abutments intended for use with endosseous dental implants for support of single-tooth or multiple-tooth restorations in the upper or lower arch, and therefore, have the same intended use. | | Subject Device | Primary Predicate Devices | | |-----------------------------------|-----------------------------------|---------------------------------------------|-------------------------------------------| | | Medentika GmbH | Institut Straumann AG | Straumann USA | | | Medentika<br>CAD/CAM<br>Abutments | Straumann® CARES®<br>Variobase™<br>Abutment | Straumann C.A.R.E.S.<br>Titanium Abutment | | | | K120822 | K052272 | | Design | | | | | Abutment<br>Design | Titanium Base,<br>CAD/CAM Blank | Titanium Base | CAD/CAM Blank | | Prosthesis<br>Attachment | Cement-retained | Cement-retained | Cement-retained | | Restoration | Single-unit<br>Multi-unit | Single-unit<br>Multi-unit | Single-unit<br>Multi-unit | | Abutment<br>Diameter<br>(mm) | 3.0 - 7.0 | 3.3 - 6.5 | 3.3 - 6.5 | | Abutment<br>Angle | Up to 30° | Up to 30° | Up to 30° | | Abutment/<br>Implant<br>Interface | Indexed,<br>non-indexed | Indexed,<br>non-indexed | Indexed,<br>non-indexed | Summary: Table of Substantial Equivalence – Abutment Design {11}------------------------------------------------ | | Subject Device | Primary Predicate Devices | | |-----------|-----------------------------------|---------------------------------------------------------|------------------------------------------------------| | | Medentika GmbH | Institut Straumann AG | Straumann USA | | | Medentika<br>CAD/CAM<br>Abutments | Straumann® CARES®<br>VariobaseTM<br>Abutment<br>K120822 | Straumann C.A.R.E.S.<br>Titanium Abutment<br>K052272 | | Material | | | | | Abutment | Titanium Alloy | Titanium Alloy | Titanium Alloy | | Screw | Titanium Alloy | Titanium Alloy | Titanium Alloy | | Sterility | | | | | Abutment | Provided Non-Sterile | Provided Non-Sterile | Provided Non-Sterile | | Screw | Provided Non-Sterile | Provided Non-Sterile | Provided Non-Sterile | ## CONCLUSION The data included in this submission demonstrate substantial equivalence to the predicate devices listed above. Overall, the subject device has the following similarities to the predicate devices: - . has the same intended use, - uses the same operating principle, - incorporates the same basic design, - incorporates the same or very similar materials, and - . has similar packaging and is to be sterilized using the same processes.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...