The Straumann® CARES® Variobase™ Abutment is a two-piece dental abutment consisting of the Straumann® Variobase™ Abutment and the Straumann® CARES® Variobase™ Coping which is intended to be placed onto Straumann dental implants to provide support for prosthetic reconstruction such as crowns and bridges. Straumann® CARES® Variobase™ Abutments are indicated for screw-retained single tooth or cement-retained single tooth and bridge restorations. The Straumann® CARES® Variobase™ Coping polycon® ae in combination with the Straumann® Variobase™ Abutment is indicated for temporary (up to 180 days) dental restoration of a Straumann dental implant.
Device Story
System comprises premanufactured titanium-aluminum-niobium alloy bonding bases (Variobase™ Abutments) and patient-specific copings (ceramic or acrylic). Clinicians or technicians use Straumann® CARES® Scan CS2 3D scanner and Straumann® CARES® Visual CAD software to design copings via wax-up or digital 'CAD-up' procedures. Design data transmitted to Straumann milling center for CAM production. Copings cemented to bonding bases; assembly screw-fixed into dental implants. Used in dental clinics to provide support for prosthetic reconstructions (crowns/bridges). Benefits include patient-specific fit and restoration of dental function.
Clinical Evidence
Bench testing only. Dynamic fatigue testing performed in accordance with FDA guidance 'Root-form Endosseous Dental Implants and Endosseous Dental Abutments' to evaluate fatigue load limits. Specimens passed pre-defined acceptance criteria.
Technological Characteristics
Bonding bases: Titanium-aluminum-niobium alloy (ISO 5832-11). Copings: Ceramic (ZrO2) or Acrylic (PMMA). CAD/CAM manufacturing. Non-sterile; requires user sterilization via hot steam. Connectivity: Networked design data transfer to milling center.
Indications for Use
Indicated for patients requiring prosthetic rehabilitation of Straumann dental implants. Supports crowns and bridges via screw-retained or cement-retained single tooth and bridge restorations. Temporary use (up to 180 days) indicated for polycon® ae acrylic copings.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
Predicate Devices
ETKON ES1, MODEL 019.0001, ETKON VISUAL, STRAUMANN CADCAM ABUTMENT (K093113)
STRAUMANN CARES BRIDGE; STRAUMANN CARES DOLDER BAR (K101465)
WIELAND ZENO CAO TEMPORARY PMMA DISC, TOOTH-COLORED (K071548)
Submission Summary (Full Text)
{0}------------------------------------------------
K120822
Image /page/0/Picture/2 description: The image shows the Straumann logo. The logo consists of a stylized symbol on the left and the word "straumann" in bold, black letters on the right. The symbol appears to be two parallel lines that are cut by a third line.
5 Error! Reference source not found.
### 510(k) Summary of Safety and Effectiveness 5
# DEC 2 1 2012
#### Applicant's Name and Address 5.1
Institut Straumann AG Peter Merian-Weg 12 CH-4052 Basel Switzerland
Telephone Number: +41 61 965 1482 +41 61 965 1102 Fax Number: . Andreas Petermann, Director, Corporate Regulatory Affairs Contact Person: .
#### 5.2 Date of Submission
March 16, 2012
#### 5.3 Name of the Device
| Trade Name: | Straumann® CARES® Variobase™ Abutments<br>Straumann® CARES® Variobase™ Temporary abutments<br>Straumann® CARES® Variobase™ Copings<br>Straumann® Variobase™ Abutments |
|--------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Bonding Base<br>Patient-specific abutment coping |
| Classification Name:<br>Regulation Number: | Endosseous dental implant abutment<br>21CFR872.3630 |
#### 5.4 Predicate Devices
| ETKON ES1, MODEL 019.0001, ETKON VISUAL, STRAUMANN CADCAM ABUTMENT | K093113 |
|--------------------------------------------------------------------------------------------------------------------------|---------|
| STRAUMANN CARES BRIDGE; STRAUMANN CARES DOLDER BAR | K101465 |
| STRAUMANN DENTAL IMPLANT SYSTEM | K083550 |
| NNC CEMENTABLE ABUTMENT STRAIGHT, 15 DEGREE ANGLE, TYPE A, 15 DEGREE ANGLE, TYPE B, NNC BASAL SCREW CEMENTABLE ABUTMENTS | K113283 |
| STRAUMANN COMPUTER AIDED RESTORATION SERVICE | K052272 |
| STRAUMANN WN CARES TITANIUM ABUTMENT | K082764 |
| P.004 NC CARES TITANIUM AND CERAMIC ABUTMENTS | K081005 |
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the Straumann logo. The logo consists of a stylized "S" symbol on the left and the word "straumann" in bold, black letters on the right. The "S" symbol is made up of three parallel lines that are angled slightly upwards.
5 Error! Reference source not found.
| P.004 RC CARES TITANIUM AND CERAMIC ABUTMENTS | K072151 |
|-----------------------------------------------------|---------|
| LAVA FRAME, LAVA FRAME SHADE | K072055 |
| SIRONA CAD/CAM SYSTEM | K100152 |
| METOXIT CAM-BLANKS | K072569 |
| NACERA Z AND NACERA Z MEDIUM | K080195 |
| WIELAND ZENO CAO TEMPORARY PMMA DISC, TOOTH-COLORED | K071548 |
#### 5.5 Description of the Device
#### Straumann® CARES® Variobase™ portfolio 5.5.1
The Straumann® CARES® Variobase™ portfolio consists of different parts which are used to provide prosthetic rehabilitation of a dental implant.
Premanufactured Variobase™ Abutments are available for the different platforms of the Straumann® dental implant system. These serve as a bonding base to which a patientspecific coping can be cemented. The coping can be made from ceramics to result in a permanent restoration, or made from acrylics to result in a temporary restoration (up to 180 davs).
The coping can be fully anatomical, i.e. it is a replica of a tooth with incisal edge or occlusal surface. It may also be of a reduced tooth shape in which case a separate crown needs to be cemented onto the coping or direct veneering needs to be applied.
## 5.5.2 Straumann® Variobase™ Abutments
The premanufactured (stock) abutments of the Straumann® CARES® Variobase™ portfolio are sometimes also referred to as "bonding bases". Straumann® Variobase™ Abutments are available to fit to Straumann® dental implant platforms NNC (Narrow Neck CrossFit®). RN (Reqular Neck), WN (Wide Neck), NC (Narrow CrossFit®) and RC (Regular CrossFit®).
#### Straumann® CARES® Variobase™ Copings 5.5.3
Straumann® CARES® Variobase™ Copings for Straumann® Variobase™ Abutments are patient-specific. CADCAM manufactured copings to be cemented to the Straumann Variobase™ Abutments.
The copings are available in ceramic (zerion™) or in acrylic (polycon® ae) material. While the ceramic copings are intended to stay in the patient's mouth as a permanent restoration, the acrylic copings are intended for temporary restoration, i.e. use should be limited to 180 days.
Both materials, ceramic or acrylic, are available in different tooth colored shades. By CAD (Computer Aided Design) and CAM (Computer Aided Manufacturing), the copings are produced so that they exactly connect to the corresponding bases. The connection between the coping and the abutment is achieved by cementation. Straumann® CARES® Variobase™ Copings are CADCAM manufactured from 510(k) cleared milling blanks.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the Straumann logo. The logo consists of a stylized "S" symbol on the left and the word "straumann" in bold, sans-serif font on the right. The "S" symbol is made up of two parallel lines that curve and intersect.
Error! Reference source not found.
5
Ceramic (zerion™): K072569, K080195
Acrylic (polycon® ae): K071548
The Straumann® CARES® Variobase™ Copings for CAMLOG® and NobelReplace™ abutments are patient-specific and can be used to restore abutments manufactured by CAMLOG® and Nobel Biocare®. It is available in ceramic and acrylic material. A customer would need to carry out a 3D scan of a CAMLOG® or NobelReplace™ abutment. Based on the scan data the customer designs a coping that fits to the CAMLOG® or NobelReplace™ abutment. After CAM manufacturing the customer will receive the coping which he needs to cement to the abutment.
In such a case. Straumann only delivers the coping. The customer needs to get the abutment and screw from the original manufacturer (CAMLOG® or Nobel Biocare®).
510(k)s K072055 and K100152 describe similar design solutions, i.e. a patient-specific coping is designed to fit to a bonding base by a CAD system.
## 5.5.4 Straumann® CARES® Variobase™ Abutments
Straumann® CARES® Variobase™ Abutments zerion™ are a combination of the above mentioned components. Thus, a Straumann® CARES® Variobase™ Abutment consists of a Straumann® Variobase™ Abutment, the corresponding basal screw and a patientspecific Straumann® CARES® Variobase™ Coping zerion™.
Straumann® CARES® Variobase™ Temporary Abutments polycon® ae is comparable to the Straumann® CARES® Variobase™ Abutments zerion™, however, the coping in this case is made from acrylic.
The Straumann® CARES® Variobase™ Temporary Abutments polycon® ae is intended for temporary use (up to 180 days).
Straumann® CARES® Variobase™ Abutments are to be screw-fixed into a dental implant in the patient's mouth. The torque pressures of the screw are fully contained within the titanium interface.
#### 5.6 Intended Use of the Device
Straumann® Variobase™ Abutments and Straumann® CARES® Variobase™ Abutments are prosthetic components that are directly connected to the endosseous dental implant intended for use as an aid in prosthetic rehabilitations.
The Straumann® CARES® Variobase™ Coping is patient-specific and can be used to restore abutments by Straumann®, CAMLOG® and NobelReplace™ abutments by Nobel Biocare®. It is available in zerion™ and polycon® ae.
The coping can be fully anatomical, i.e. it is a replica of a tooth with incisal edge or occlusal surface. It may also be of a reduced tooth shape in which case a separate crown needs to be cemented onto the coping or direct veneering needs to be applied.
#### Technological Characteristics 5.7
Straumann® Variobase™ Abutments are bonding bases made from a titanium-aluminumniobium alloy. They are standard medical devices (stock produced).
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the Straumann logo. The logo consists of a stylized symbol on the left and the word "straumann" in bold, sans-serif font on the right. The symbol is made up of two parallel lines that are angled and slightly offset from each other.
5 Error! Reference source not found.
Straumann® CARES® Variobase™ Copings is patient-specific medical devices, i.e. they are designed for an individual patient.
Straumann® CARES® Variobase™ Copings are designed either by a wax-up or a "CADup" procedure. Either way is processed through Straumann's CAD system consisting of the table top 3D-scanner Straumann® CARES® Scan CS2 and the corresponding CAD software Straumann® CARES® Visual. (CAD: Computer Aided Design)
In a wax-up procedure a wax model of a coping is created and scanned to be able to CAM produce the coping from a selected material. (CAM: Computer Aided Manufacturing)
In a "CAD-up" procedure, scanned data is used as the source to digitally design a coping. The design data is then sent to a Straumann milling center. The Straumann® CARES® Variobase™ Copings are CAM produced at the Straumann milling center according to the design file received and from the selected material.
Validation of the Straumann® CARES® Visual CAD software provides evidence that design parameters for the Straumann® CARES® Variobase™ Copings have met their pre-determined acceptance criteria and that dental restorations meeting their design specifications can be manufactured by Straumann CAM milling devices.
The software Straumann® CARES® Visual has been subject of 510(k) review and clearance in 510(k)s K093113 and K101465.
#### 5.8 Performance Testing
The material used in the manufacture of Straumann® Variobase™ Abutments is a titanium-aluminum-niobium alloy which meets the requirements of ISO 5832-11.
Bench testing was performed to evaluate the fatigue load limits of the proposed Straumann® Variobase™ Abutments and related product portfolio.
Dynamic fatique tests were carried out in accordance to FDA's Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments.
Straumann® Variobase™ Abutments cemented to different Straumann® CARES® Variobase™ Copings passed the pre-defined acceptance criteria.
Al Response: K120822
Straumann Variobase Abutments July 6, 2012
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the Straumann logo. The logo consists of a stylized graphic to the left of the word "straumann" in a bold, sans-serif font. The graphic is composed of two parallel lines that are angled upwards and to the right, creating a sense of forward movement.
Error! Reference source not found.
5
#### Substantial Equivalence Comparison 5.9
The following table lists the proposed devices Straumann® CARES® Variobase™ Abutments and the respective predicate devices Straumann is claiming substantial equivalence to.
| Proposed Devices Straumann® CARES® Variobase™ Abutments | | | | |
|--------------------------------------------------------------------------------------------------------|--------------------------------------------------|-------------------------------------------------|--------------------------------------------------|--------------------------------------------------|
| NNC<br>Narrow Neck<br>CrossFit® | RN<br>Regular Neck | WN<br>Wide Neck | NC<br>Narrow CrossFit® | RC<br>Regular<br>CrossFit® |
| Patient specific coping Straumann® CARES® Variobase™ Coping ZrO₂ for permanent use | | | | |
| Image: Coping ZrO2 | | | | |
| Patient specific coping Straumann® CARES® Variobase™ Coping PMMA for temporary use<br>(up to 180 days) | | | | |
| Image: Coping PMMA | | | | |
| Image: NNC Abutment | Image: RN Abutment | Image: WN Abutment | Image: NC Abutment | Image: RC Abutment |
| Image: NNC Abutment with base | Image: RN Abutment with base | Image: WN Abutment with base | Image: NC Abutment with base | Image: RC Abutment with base |
| Predicate Devices | | | | |
| Straumann®<br>NNC<br>Cementable<br>Abutment<br>Straight | Straumann®<br>CADCAM<br>Abutment RN,<br>Titanium | Straumann® WN<br>CARES®<br>Titanium<br>Abutment | Straumann® NC<br>CARES®<br>Titanium<br>Abutments | Straumann® RC<br>CARES®<br>Titanium<br>Abutments |
| K113283 | K052272<br>K083550 | K082764 | K081005 | K072151 |
The proposed and the predicate devices share the following similarities.
| | Proposed device<br>Straumann® CARES® Variobase™ Abutments | Predicate device |
|------------------------------------|-------------------------------------------------------------------------|-------------------------------------------------------------------------|
| Anatomical<br>site | Oral cavity | Oral cavity |
| Implant-<br>abutment-<br>interface | Straumann Narrow Neck CrossFit®<br>(NNC)<br>Straumann Regular Neck (RN) | Straumann Narrow Neck CrossFit®<br>(NNC)<br>Straumann Regular Neck (RN) |
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the word "straumann" in bold, black letters. To the left of the word is a logo consisting of three parallel lines that are slanted and stacked on top of each other. The logo is also black, but the lines are filled with a gray pattern.
| 5 | Error! Reference source not found. |
|---|------------------------------------|
|---|------------------------------------|
| | Straumann Wide Neck (WN)<br>Straumann Narrow CrossFit® (NC)<br>Straumann Regular CrossFit® (RC) | Straumann Wide Neck (WN)<br>Straumann Narrow CrossFit® (NC)<br>Straumann Regular CrossFit® (RC) |
|------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Implant-<br>borne | Yes | Yes |
| Material of<br>abutment | Titanium-aluminum-niobium | Titanium |
| Material of<br>screws | Titanium-aluminum-niobium | Titanium-aluminum-niobium |
| Material of<br>permanent<br>coping | ZrO2 (YTZ-P) | ZrO2 (YTZ-P) K072569, K080195 |
| Material of<br>temporary<br>coping | PMMA (polymethyl methacrylate) | PMMA (polymethyl methacrylate)<br>K071548 |
| Patient-<br>specific<br>design | CAD design by software Straumann®<br>CARES® Visual | CAD design by software Straumann®<br>CARES® Visual K052272, K083550,<br>K082764, K081005, K072151 |
| Duration of<br>use | Permanent (coping made from ZrO2)<br>up to 180 days (coping made from<br>PMMA) | Permanent |
| Sterility | Non sterile, to be sterilized by user | Non sterile, to be sterilized by user |
| Sterilization | Hot steam | Hot steam |
| Perform-<br>ance testing | Dynamic fatigue tests of minimal body<br>designs (Connection testing)<br>The test specimens passed the pre-<br>defined acceptance criteria. | K113283: Dynamic fatigue test of<br>abutment (Connection testing)<br>K052272, K083550, K082764,<br>K081005, K072151: Dynamic fatigue<br>tests of minimal body designs<br>(Connection testing) |
| Indications<br>for Use | Abutments are used in connection with<br>the prosthetic restoration of<br>Straumann® dental implants.<br>Abutments are intended to be placed<br>into dental implants to provide support<br>for prosthetic reconstructions such as<br>crowns and bridges.<br>Straumann® Variobase™ Abutments<br>are indicated for screw-retained single<br>tooth or cement-retained single tooth<br>and bridge restorations. | K113283: Abutments are used in<br>connection with the prosthetic<br>restoration of Straumann dental<br>implants. Abutments are intended to be<br>placed into dental implants to provide<br>support for prosthetic reconstructions<br>such as crowns and bridges.<br>Narrow Neck CrossFit Cementable<br>Abutments are indicated for cement-<br>retained single tooth and bridge<br>restorations.<br>K052272, K083550: Abutments are<br>intended to be placed into dental<br>implants to provide support for<br>prosthetic reconstructions such as<br>crowns or bridges. The Straumann®<br>CARES® Titanium Abutment is<br>indicated for cemented restorations.<br>The abutment can be used in single<br>tooth replacements and multiple tooth<br>restorations.<br>K082764: Abutments are placed into<br>the dental implants to provide support<br>for prosthetic restoration such as<br>crowns, bridges or overdentures. |
| 5 | The Straumann® WN CARES®<br>Titanium Abutment is indicated for<br>single tooth replacements and multiple<br>tooth restorations. The prosthetic<br>restoration is cement-retained.<br>K081005: Abutments are placed into<br>the dental implants to provide support<br>for prosthetic restoration such as<br>crowns, bridges and overdentures.<br>The P.004 NC CARES® Titanium<br>Abutment is indicated for cemented<br>restoration. The abutment can be used<br>in single tooth replacements and<br>multiple tooth restorations.<br>K072151: Abutments are placed into<br>the dental implants to provide support<br>for prosthetic restoration such as<br>crowns or bridges. The P.004 RC<br>CARES® Titanium Abutment is<br>indicated for cemented restoration. The<br>abutment can be used in single tooth<br>replacements and multiple tooth<br>restorations. | |
·
.
Straumann Variobase Abutments July 6, 2012
•
.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the Straumann logo. The logo consists of two parallel lines that are cut by two other parallel lines, forming a parallelogram shape. To the right of the shape is the word "straumann" in a bold, sans-serif font.
5 Error! Reference source not found.
## 5.10 Conclusion
Non-clinical testing presented in this 510(k) premarket notification demonstrate that the components of the Straumann® CARES® Variobase™ portfolio met predefined
acceptance criteria and successfully passed testing. The information presented in this 510(k) demonstrated that the components of the Straumann® CARES® Variobase™ portfolio are substantially equivalent to the predicate devices.
Straumann believes that the components of the Straumann® CARES® Variobase™ portfolio are safe and effective for their intended use.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling a stylized caduceus or a representation of human figures, with three wavy lines forming a shape that suggests movement or flow.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
## December 21, 2012
Ms. Andreas Petermann Director, Corporate Regulatory Affairs Institut Straumann AG Peter Merian-Weg 12 Basel Switzerland CH-4052
Re: K120822
Trade/Device Name: Straumann® CARES® Variobase™ Abutments Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: December 14, 2012 Received: December 17, 2012
Dear Ms. Petermann:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{8}------------------------------------------------
## Page 2 - Ms. Petermann
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Kwame O. Ulmer
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices
Office of Device Evaluation
Center for Devices and
Radiological Health
Enclosure
{9}------------------------------------------------
## Indications for Use
# 510(k) Number (if known): K120822
Device Name: Straumann® CARES® Variobase™ Abutments
Indications for Use:
The Straumann® CARES® Variobase™ Abutment is a two-piece dental abutment consisting of the Straumann® Variobase™ Abutment and the Straumann® CARES® Variobase™ Coping which is intended to be placed onto Straumann dental implants to provide support for prosthetic reconstruction such as crowns dental limpiants to provide coppes® Variobase™ Abutments are indicated for and bridges. Othernally or cement-retained single tooth and bridge restorations.
The Straumann® CARES® Variobase™ Coping polycon® ae in combination with the Straumann® Variobase™ Abutment is indicated for temporary (up to 180 days) dental restoration of a Straumann dental implant.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-the-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
2012.12.19 Susan Runner DDS, MA 08:24:16 -05'00'
(Division Sign-Off) (Division Sign-On)
Division of Anestheslology, General Hospital Division Control, Dental Devices
K120822 510(k) Number;
> Response Letter: K120822 Straumann Variobase
Straumann USA, LLC
November 27, 2012 Page 19 of 129
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.