VITA VMLC Primer

K251587 · Vita Zahnfabrik H. Rauter GmbH & Co. Kg. · KLE · Feb 23, 2026 · Dental

Device Facts

Record IDK251587
Device NameVITA VMLC Primer
ApplicantVita Zahnfabrik H. Rauter GmbH & Co. Kg.
Product CodeKLE · Dental
Decision DateFeb 23, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3200
Device ClassClass 2

Indications for Use

Adhesive bonding of metal substructure surfaces (non-precious metal alloys, PM alloys, titanium) with (meth-)acrylate-based resins/composites Adhesive bonding of ZrO2 substructure surfaces with (meth-)acrylate-based resins/composites Adhesive bonding of cross-linked composites/resins such as VITA CAD Temp or of high-performance polymers (PEEK, PEKK) with (meth-) acrylate-based resins/composites

Device Story

Two-component dental bonding system (Primer I and Primer II); used in dental laboratory settings by technicians. Primer I contains functional monomers in solvent for surface interaction; Primer II is light-curable methacrylate resin. Sequential application promotes adhesion between prosthetic substructures (metals, ZrO2, polymers) and light-curing composite veneering resins. Polymerization achieved via standard dental laboratory light-curing units. Enhances bond strength between substructure and veneer; improves prosthetic durability.

Clinical Evidence

No clinical data. Bench testing performed per ISO 10477 (including thermal cycling) demonstrated adequate bond strength across indicated substrates. Biocompatibility evaluated per ISO 10993-1 and ISO 7405.

Technological Characteristics

Two-component (Primer I/II) dental bonding system. Functional monomer-based chemical coupling. Light-curable methacrylate resin component. Polymerization via standard dental laboratory light-curing units. Biocompatibility per ISO 10993-1 and ISO 7405; bond strength testing per ISO 10477.

Indications for Use

Indicated for dental laboratory professionals requiring adhesive bonding of prosthetic substructures (metal alloys, titanium, ZrO2, cross-linked composites, PEEK, PEKK) to (meth-)acrylate-based veneering resins/composites.

Regulatory Classification

Identification

A resin tooth bonding agent is a device material, such as methylmethacrylate, intended to be painted on the interior of a prepared cavity of a tooth to improve retention of a restoration, such as a filling.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION February 23, 2026 Vita Zahnfabrik H. Rauter GmbH & Co. Kg. Lindsay Tilton Regulatory Affairs Consultant Spitalgasse 3 Bad Sackingen Baden-Wurttemberg, DE D-79713 GERMANY Re: K251587 Trade/Device Name: VITA VMLC Primer Regulation Number: 21 CFR 872.3200 Regulation Name: Resin Tooth Bonding Agent Regulatory Class: Class II Product Code: KLE Dated: January 20, 2026 Received: January 20, 2026 Dear Lindsay Tilton: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K251587 - Lindsay Tilton Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K251587 - Lindsay Tilton Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, MICHAEL E. ADJODHA -S Michael E. Adjodha, MChE, RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} | Indications for Use | | | | --- | --- | --- | | Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K251587 | ? | | Please provide the device trade name(s). | | ? | | VITA VMLC Primer | | | | Please provide your Indications for Use below. | | ? | | • Adhesive bonding of metal substructure surfaces (non-precious metal alloys, PM alloys, titanium) with (meth-)acrylate-based resins/composites • Adhesive bonding of ZrO2 substructure surfaces with (meth-)acrylate-based resins/composites • Adhesive bonding of cross-linked composites/resins such as VITA CAD Temp or of high-performance polymers (PEEK, PEKK) with (meth-) acrylate-based resins/composites | | | | Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? | {4} K251587 # 510(k) SUMMARY (21 CFR 807.92) **Submitter:** VITA ZAHNFABRIK H. RAUTER GMBH & CO. KG. **Address:** SPITALGASSE 3 BAD SACKINGEN Baden – Wurttemberg DE D – 79713 Germany **Contact Person:** Mrs. Lindsay Tilton **Telephone:** (925)699 – 9091 **Email:** ltilton@vitanorthamerica.com **Date Prepared:** February 19, 2026 ## Device Name **Trade Name:** VITA VM LC Primer **Common Name:** Resin tooth bonding agent **Classification Name:** Agent, Tooth Bonding, Resin **Regulation Number:** 21 CFR 872.3200 **Product Code:** KLE **Class:** II ## Predicate Devices **Primary predicate:** Kuraray Alloy Primer (K974089) **Secondary predicate:** Signum Universal Bond (K240660) ## Device Description VITA VM LC Primer is a two-component dental bonding system (Primer I and Primer II) used to promote adhesion between dental prosthetic substructures and light-curing (meth)acrylate-based composite veneering resins. Primer I contains functional monomers in a solvent carrier for interaction with substrate surfaces; Primer II is a light-curable methacrylate resin layer that polymerizes using standard dental laboratory light-curing units. The device is supplied as a two-bottle system for sequential application. ## Intended Use / Indications for Use - Adhesive bonding of metal substructure surfaces (non-precious metal alloys, PM alloys, titanium) with (meth-)acrylate-based resins/composites - Adhesive bonding of ZrO2 substructure surfaces with (meth-)acrylate-based resins/composites - Adhesive bonding of cross-linked composites/resins such as VITA CAD Temp or of high-performance polymers (PEEK, PEKK) with (meth-) acrylate-based resins/composites ## Technological Comparison The subject and predicate devices use functional monomer-based chemical coupling to promote bonding between prosthetic substrates and methacrylate resin materials. The subject device uses a two-component, sequential application approach including a light- {5} curable component; these differences do not alter the intended use and do not raise new or different questions of safety or effectiveness. ## Performance Data Non-clinical bench testing was performed in accordance with ISO 10477 (including thermal cycling) and demonstrated adequate bond strength across the indicated substrate materials. Biocompatibility was evaluated in accordance with FDA's Guidance "Use of International Standard ISO 10993-1, Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process" and ISO 7405, Dentistry — Evaluation of biocompatibility of medical devices used in dentistry. Clinical testing was not performed. ## Conclusion Based on the intended use, technological characteristics, and non-clinical performance and biocompatibility data, VITA VM LC Primer is substantially equivalent to the identified predicate devices.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...