Signum composite flow; Signum universal bond; Signum metal bond ; Signum composite; Signum matrix; Signum liquid; Signum cre-active
Applicant
Kulzer, LLC
Product Code
EBF · Dental
Decision Date
Sep 17, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3690
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K240660 · Sep 17, 2024
Signum composite flow; Signum universal bond; Signum metal bond ; Signum composite; Signum matrix; Signum liquid; Signum cre-active
Kulzer, LLC
Post-market surveillance (PMS) data; Scientific literature regarding comparable medical devices
The sponsor utilized post-market surveillance data and existing scientific literature to support the clinical evaluation and benefit-risk profile of the devices, citing their long-term market history as evidence of safety and performance.
Post-market surveillance; Clinical evaluation; Literature review; Long-term market history
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Clinical evaluation of Signum veneering composites; Literature-based clinical evaluation; Follow-up/Duration: Marketed since 1999/2011; Study Period: Historical market data
Patients receiving dental restorations using veneering composites; Number of Sites: Not applicable
Not applicable for this study
Safety and performance
Clinical evaluation of Signum bondings and cre-active; Literature-based clinical evaluation and PMS review; Follow-up/Duration: Marketed since 1983/1991/2007; Study Period: Historical market data
Patients receiving dental restorations; Number of Sites: Not applicable
Not applicable for this study
Safety and performance
Indications for Use
Signum metal bond: - Bonding agent between framework surfaces made of metal alloys (precious metal, non-precious metal or titanium) and (meth-) acrylate-based veneering materials in conjunction with micromechanical retention. - Bonding agent between partial framework surfaces made of metal alloys (precious metal, nonprecious metal or titanium) and (meth-) acrylate-based veneering materials or denture base materials (hot or auto-polymerizates). - Bonding agent for extraoral repair of metal supported composite veneers. Signum universal bond: Metal - Bonding agent between framework surfaces made of metal alloys (precious metal, non-precious metal or titanium) and (meth-) acrylate-based veneering materials in conjunction with macromechanical retentions. - Bonding agent between partial framework surfaces made of metal alloys (precious metal, nonprecious metal or titanium) and (meth-) acrylate-based veneering materials or denture base materials (hot or auto-polymerizates). - Bonding agent for the extraoral repair of metal-supported composite veneers. Zirconium dioxide - Bonding agent between frameworks, implant abutments and single crowns made of ZrO3 and lightcuring veneering materials. - Bonding agent for the extraoral repair of composite veneers on ZrO2 frameworks. High performance polymers (PEEK) - Bonding agent between frameworks, implant abutments and individual crowns made of high performance polymers and light-curing veneering materials. - Bonding agent between partial framework surfaces made of high-performance polymers and lightcuring veneering materials or denture base materials (hot or auto-polymerizate). - Bonding agent for the extraoral repair of composite veneers on PEEK frameworks. Signum liquid: For light-curing veneering materials from Kulzer: - Restoring the dispersion layer of light curing veneering materials. Signum composite / Signum composite flow / Signum matrix: Suitable for veneering the following dental framework materials: Metal alloys (precious metal, non-precious or titanium), zirconia, high-performance polymers (PEEK) Suitable for partial alteration of the colour and shape of the following dental materials: • PMMA • Photopolymers Suitable for the following prosthetic restorations: • full and partial veneers for permanent framework-supported crowns and bridges • veneering removable combination prosthetics (telescopic and tapered crowns and at-tachment prosthetics) • veneering of superstructures/tertiary structures on implants • metal-free front tooth and posterior tooth temporary restorations • alteration of the colour and shape of Kulzer acrylic denture teeth • alteration of the colour and shape of acrylic-based frameworks (PMMA). • alteration of the colour and shape of Kulzer photopolymers that are approved for char-acterisation with Signum components (according to the photopolymer instructions for use) Signum cre-active: Individual characterization (e.g., of enamel cracks, white spots, imitation fillings, enamel wedges, intermediate colour layers, abrasion surfaces, fissures) of • light-curing Kulzer veneering materials, • Kulzer denture acrylic teeth and PMMA, • (meth-)acrylate-based 3D printing photopolymers in accordance with the photopolymer instructions for use
Device Story
System of dental laboratory materials including bonding agents, veneering composites, modeling liquids, and characterization fluids. Used by dental technicians to fabricate or repair custom dental prosthetics (crowns, bridges, implant superstructures). Bonding agents (Signum metal bond, universal bond) applied to framework surfaces (metal, zirconia, PEEK) to promote adhesion of (meth-)acrylate-based veneering materials. Signum liquid used to restore dispersion layers during modeling. Signum composite/matrix used for veneering; Signum cre-active used for aesthetic characterization (enamel cracks, white spots). All components applied manually in laboratory settings; polymerized via light-activation. Final custom devices inserted into patient oral cavity by clinicians. Benefits include restoration of tooth function, aesthetics, and structural integrity of dental frameworks.
Clinical Evidence
No clinical trials performed. Evidence based on literature review, historical market data (some components marketed since 1983/1991/1999), and post-market surveillance. Bench testing included depth of cure, flexural strength, shear bond strength (to metal, zirconia, PEEK), and viscosity. Benefit-risk ratio determined to be positive based on established state-of-the-art dental laboratory practices.
Indicated for dental laboratory technicians to perform veneering of framework-supported dental restorations (crowns, bridges, implant superstructures), repair of composite veneers, and characterization of dental prosthetics (PMMA, photopolymers, acrylic teeth) using light-curing resin materials.
Regulatory Classification
Identification
Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.
Predicate Devices
VITA ZETA CROWN AND BRIDGE ACRYLIC VENEERING SYS (K961434)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 17, 2024
Kulzer, LLC Lucas Harmon Regulatory & Quality Engineer 4315 S Lafayette Blvd South Bend, Indiana 46614
Re: K240660
Trade/Device Name: Signum composite flow: Signum universal bond: Signum metal bond : Signum composite; Signum matrix; Signum liquid; Signum cre-active Regulation Number: 21 CFR 872.3690 Regulation Name: Tooth Shade Resin Material Regulatory Class: Class II Product Code: EBF Dated: February 9, 2024 Received: March 8, 2024
Dear Lucas Harmon:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{2}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{3}------------------------------------------------
Sincerely,
Bobak Shirmohammadi -S
For Michael E. Adjodha, M.ChE., RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
### Indications for Use
Submission Number (if known)
| K240660 |
|------------------------|
| Device Name |
| Signum composite flow; |
| Signum universal bond; |
| Signum metal bond ; |
| Signum composite; |
| Signum matrix; |
| Signum liquid; |
| Signum cre-active |
Indications for Use (Describe)
Signum metal bond:
- Bonding agent between framework surfaces made of metal alloys (precious metal, non-precious metal or titanium) and (meth-) acrylate-based veneering materials in conjunction with micromechanical retention.
- Bonding agent between partial framework surfaces made of metal alloys (precious metal, nonprecious metal or titanium) and (meth-) acrylate-based veneering materials or denture base materials (hot or auto-polymerizates).
- Bonding agent for extraoral repair of metal supported composite veneers.
Signum universal bond:
Metal
- Bonding agent between framework surfaces made of metal alloys (precious metal, non-precious metal or titanium) and (meth-) acrylate-based veneering materials in conjunction with macromechanical retentions.
- Bonding agent between partial framework surfaces made of metal alloys (precious metal, nonprecious metal or titanium) and (meth-) acrylate-based veneering materials or denture base materials (hot or auto-polymerizates).
- Bonding agent for the extraoral repair of metal-supported composite veneers.
Zirconium dioxide
- Bonding agent between frameworks, implant abutments and single crowns made of ZrO3 and lightcuring veneering materials.
- Bonding agent for the extraoral repair of composite veneers on ZrO2 frameworks.
High performance polymers (PEEK)
- Bonding agent between frameworks, implant abutments and individual crowns made of high performance polymers and light-curing veneering materials.
- Bonding agent between partial framework surfaces made of high-performance polymers and lightcuring veneering materials or denture base materials (hot or auto-polymerizate).
- Bonding agent for the extraoral repair of composite veneers on PEEK frameworks.
Signum liquid:
For light-curing veneering materials from Kulzer:
- Restoring the dispersion layer of light curing veneering materials.
{5}------------------------------------------------
| Signum composite / Signum composite flow / Signum matrix: | Suitable for veneering the following dental framework materials: Metal alloys (precious metal, non-precious or titanium), zirconia, high-performance polymers (PEEK) |
|--------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Suitable for partial alteration of the colour and shape of the following dental materials: | • PMMA<br>• Photopolymers |
| Suitable for the following prosthetic restorations: | • full and partial veneers for permanent framework-supported crowns and bridges<br>• veneering removable combination prosthetics (telescopic and tapered crowns and at-tachment prosthetics)<br>• veneering of superstructures/tertiary structures on implants<br>• metal-free front tooth and posterior tooth temporary restorations<br>• alteration of the colour and shape of Kulzer acrylic denture teeth<br>• alteration of the colour and shape of acrylic-based frameworks (PMMA).<br>• alteration of the colour and shape of Kulzer photopolymers that are approved for char-acterisation with Signum components (according to the photopolymer instructions for use) |
| Signum cre-active: | Individual characterization (e.g., of enamel cracks, white spots, imitation fillings, enamel wedges, intermediate colour layers, abrasion surfaces, fissures) of<br>• light-curing Kulzer veneering materials,<br>• Kulzer denture acrylic teeth and PMMA,<br>• (meth-)acrylate-based 3D printing photopolymers in accordance with the photopolymer instructions for use |
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{6}------------------------------------------------
| 510(k) #: K240660 | 510(k) Summary | | Prepared on: 2024-09-16 |
|-------------------|----------------|--|-------------------------|
|-------------------|----------------|--|-------------------------|
| Contact Details | 21 CFR 807.92(a)(1) |
|-----------------------------|---------------------------------------------------------|
| Applicant Name | Kulzer, LLC |
| Applicant Address | 4315 S Lafayette Blvd South Bend IN 46614 United States |
| Applicant Contact Telephone | (574) 299-5444 |
| Applicant Contact | Mr. Lucas Harmon |
| Applicant Contact Email | Lucas.Harmon@kulzer-dental.com |
| Device Name | 21 CFR 807.92(a)(2) |
|-------------|---------------------|
|-------------|---------------------|
| Device Trade Name | Signum composite flow;<br>Signum universal bond;<br>Signum metal bond ;<br>Signum composite;<br>Signum matrix;<br>Signum liquid;<br>Signum cre-active |
|---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name | Tooth shade resin material |
| Classification Name | Material, Tooth Shade, Resin |
| Regulation Number | 872.3690 |
| Product Code(s) | EBF |
| Legally Marketed Predicate Devices | 21 CFR 807.92(a)(3) |
|------------------------------------|---------------------|
|------------------------------------|---------------------|
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
|-------------|----------------------------------------------------------|--------------|
| K961434 | VITA ZETA CROWN AND BRIDGE ACRYLIC VENEERING SYS | EBF |
| K152373 | VITA VM LC Flow | EBF |
| K211854 | VITA Akzent LC | EBD |
| Device Description Summary | 21 CFR 807.92(a)(4) |
|----------------------------|---------------------|
|----------------------------|---------------------|
#### Signum Bondings (Signum metal bond) are intended for dental works in the laboratory as an adhesion promoter between a given framework made of zirconium dioxide, dental alloys, high performance composites, PMMA, and a second material (e.g. composites or acrylics). These primers are only applied in the laboratory. After restorative work, Signum bondings are encapsulated between framework and the restoration/veneering material and a direct contact to the patient is very unlikely. Signum metal bond is a liquid stored in a 5 mL bottle with dropper and red cap.
Signum universal bond is a liquid stored in a 5 mL bottle with dropper and blue cap.
Signum liquid is indicated for light-curing materials from Kulzer to simplify modeling procedures and restore the dispersion layer of light-curing veneering materials.
Signum liquid is foreseen for dental works in the laboratory, where an adhesion promoter between a given composite layer and a second acrylate based material, including procedures is necessary. After completion of the restorative work, Signum
{7}------------------------------------------------
liquid is encapsulated between both layers and a direct contact to the patient is very unlikely. Signum liquid is a liquid stored a 5 mL bottle with a spout and red cap.
Signum composite, Signum composite flow, and Signum matrix are veneering composites for crowns and bridges. The veneering composites are processed by dental technicians for dental restoration by veneering framework supported restorations (for example metal or zirconia frameworks).
After light activated polymerization is inserted into the oral cavity of a specific patient as a custom made device, adapted and connected to the remaining residual teeth or functionally adapted.
The veneering composites are pastes stored in a syringe featuring a plunger and a closing cap.
Signum cre-active are a light-curing color fluids for crown and bridge techniques and prosthetics.
Signum cre-active is used in the dental laboratory to create customized characterizations of dental prosthetic works, that are then inserted into the oral cavity of a specific patient as individualized medical devices, adapted, and connected to the remaining residual teeth or functionally adapted to the edentulous alved Signum cre-activelayers must be covered with veneering material, there is no direct patient contact.
Signum cre-active are liquids stored in a syringe featuring a plunger and a closing cap.
## Intended Use/Indications for Use
### 21 CFR 807.92(a)(5)
Signum metal bond:
- Bonding agent between framework surfaces made of metal, non-precious metal or titanium) and (meth-) acrylate-based veneering materials in conjunction with micromechanical retention.
- Bonding agent between partial framework surfaces made of metal, non-precious metal or titanium) and (meth-) acrylate-based veneering materials or denture base materials (hot or auto-polymerizates).
- Bonding agent for extraoral repair of metal supported composite veneers.
Signum universal bond:
Metal
- Bonding agent between framework surfaces made of metal alloys (precious metal, non-precious metal or titanium) and (meth-) acrylate-based veneering materials in conjunction with macro-mechanical retentions.
- Bonding agent between partial framework surfaces made of metal, non-precious metal or titanium) and (meth-) acrylate-based veneering materials or denture base materials (hot or auto-polymerizates).
- Bonding agent for the extraoral repair of metal-supported composite veneers.
Zirconium dioxide
- Bonding agent between frameworks, implant abutments and single crowns made of ZrC, and light-curing veneering materials.
- Bonding agent for the extraoral repair of composite veneers on ZrQz frameworks.
High performance polymers (PEEK)
- Bonding agent between frameworks, implant abutments and individual crowns made of high performance polymers and light-curing veneering materials.
- Bonding agent between partial framework surfaces made of high-performance polymers and light-curing materials or denture base materials (hot or auto-polymerizate).
- Bonding agent for the extraoral repair of composite veneers on PEEK frameworks.
#### Signum liquid:
For light-curing veneering materials from Kulzer:
- Restoring the dispersion layer of light curing veneering materials.
Signum composite / Signum composite flow / Signum matrix: Suitable for veneering the following dental framework materials: Metal alloys (precious metal, non-precious or titanium), zirconia, high-performance polymers (PEEK)
Suitable for partial alteration of the colour and shape of the following dental materials:
• PMMA
• Photopolymers
Suitable for the following prosthetic restorations:
• full and partial veneers for permanent framework-supported crowns and bridges
· veneering removable combination prosthetics (telescopic and tapered crowns and at-tachment prosthetics)
{8}------------------------------------------------
• veneering of superstructures/tertiary structures on implants
• metal-free front tooth and posterior tooth temporary restorations
• alteration of the colour and shape of Kulzer acrylic denture teeth
· alteration of the colour and shape of acrylic-based frameworks (PMMA).
• alteration of the colour and shape of Kulzer photopolymers that are approved for char-acterisation with Signum components (according to the photopolymer instructions for use)
Signum cre-active:
Individual characterization (e.g., of enamel cracks, white spots, imitation fillings, enamel wedges, intermediate colour layers, abrasion surfaces, fissures) of
· light-curing Kulzer veneering materials,
• Kulzer denture acrylic teeth and PMMA,
· (meth-)acrylate-based 3D printing photopolymers in accordance with the photopolymer instructions for use
## Indications for Use Comparison
### 21 CFR 807.92(a)(5)
#### Signum Metal Bond/Signum Universal Bond:
In comparison to VITA VM LC Primer (K961434), Signum Metal Bond have the same indications for use in which the device may act as a bonding agent between metal, titanium), (meth-)acrylate-based materials, ZrO2 surfaces, and high-performance polymers (PEEK). While the verbiage and the scope of indications for use slightly differs between the products, they are equally indicated to act as an adhesive bonding agent for various framework surfaces.
#### Signum Liquid:
In comparison to VITA VM LC Modelling Liquid (K961434), Signum Liquid has the same indications for use in which the device may act in the layering process of dental materials. While the verbiage and the scope of indications for use slightly differs between the products, they are equally indicated to act in the layering process.
#### Signum Composite:
In comparison to VITA VM LC (K961434), Signum Composite has the same indications for use in which the device may be used for full and partial frameworks, including crowns, bridges, telescopic crowns, and implant suprastructures. While the verbiage and the scope of indications for use slightly differs between the products, they are equally indicated for the veneering of frameworks.
#### Signum Composite Flow/Signum Matrix:
In comparison to VITA VM LC Flow (K961434 & K152373), Signum Composite Flow/Signum Matrix has the same indications for use in which the device may be used frameworks, including crowns, bridges, telescopic crowns, and implant suprastructures. While the verbiage and the scope of indications for use slightly differs between the products, they are equally indicated for the veneering of frameworks.
#### Signum Cre-active:
In comparison to VITA Akzent LC (K961434 & K211854), Signum Creactive has the same indications for use in which the device may be used for the characterization of veneering materials, and acrylic teeth, 3D-print polymers. While the verbiage and scope of indications for use slightly differs between the products, they are equally indicated for the listed materials.
# Technological Comparison
## 21 CFR 807.92(a)(6)
#### Signum Metal Bond/Signum Universal Bond:
Signum Metal Bond/Signum Universal Bond and the predicate device, VITA VM LC Primer (K961434), are composed of two parts (I and II). Signum Metal Bond I/Signum Universal Bond I and VITA VM LC PRIMER I are added to a mixing dish applied with a disposable brush to serve as a primer. Signum Metal Bond II/Signum Universal Bond II and VITA VM LC PRIMER II are then applied with a disposable brush and followed with an opaque material for coverage. The design, principle of operation, etc. are very similar between Signum Metal Bond/Signum Universal Bond and the predicate device VITA VM LC Primer with minor difference in technological characteristics.
#### Signum Liquid:
Signum Liquid and the predicate device, VITA VM LC Modeling Liquid (K961434), are both utilized in the layering process and are applied to the veneering material surface sparingly. The design, principle of operation, etc. are very similar between Signum Liquid and the predicate device VITA VM LC Modelling Liquid with minor difference in technological characteristics.
#### Signum Composite:
Signum Composite and the predicate device, VITA VM LC (K961434), are both utilized after the preparation of the framework. Both products are also utilized with the necessary other system products during the veneering of frameworks. The design, principle of
{9}------------------------------------------------
operation, etc. are very similar between Signum Composite device VITA VM LC with minor difference in technological characteristics.
Signum Composite Flow/Signum Matrix:
Signum Composite Flow/Signum Matrix and the predicate device, VITA VM LC Flow (K961434 & K152373), are packaged in a syringe with a plunger to extrude product. They are all applied in layers and polymerized.
The design, principle of operation, etc. are very similar between Signum Matrix and the predicate device VITA VM LC Flow with minor difference in technological characteristics.
Signum Cre-active:
Signum Cre-active and the predicate device, VITA Akzent LC (K961434 & K211854), are both fluids utilized after the proper preparation of the surface. Both products are applied in layers and then polymerized. The design, principle of operation, etc. are very similar between Signum Cre-active and the predicate device VITA Akzent LC with minor difference in technological characteristics.
#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
For Signum composite, Signum composite flow and Signum matrix mechanical testing was performed in regard to depth of cure, flexural strength and bond strength.
For Singum liquid, Signum metal bond and Signum universal bond mechanical testing was performed in regard to shear bond strength with precious metal non-precious metal, zirconium and PEEK (high performance polymers).
For Signum cre-active mechanical testing was performed in regard to viscosity and curing depth.
Thermal testing was preformed in the shelf-life report, utilizing testing methods at differing temperatures.
The clinical evaluation for Signum composite flow and Signum matrix is based on internal data and scientific literature currently available about venering comparable medical devices. Signum composite is marketed since 1999, Signum composite flow since 2011 and Signum matrix since 1999. The benefit-risk-ratio for the use of Signum veneering composites c&b in the intended purpose is positive.
The application of veneering composites c&b is state-of-the-art both technologically and concerning the medical application. Specific questions of this clinical evaluation were answered, therefor a clinical investigation with veneering composites c&b is not necessary and the route for this clinical evaluation is literature based.
The clinical evaluation for Signum metal bond and Signum liquid (summed up by the term "Signum bondings") is based on currently available scientific literature. Clinical studies with similar medical devices were identified and PMS data were considered as well. Signum metal bond is marketed since 2007, Signum universal bond since 2022 and Signum liquid since 1983 (for Dentacolor / Artglass liquid).
Specific questions of this clinical evaluation were and suitable data from the literature. Therefor a clinical investigation with Signum bondings is not necessary and the route for this clinical evaluation is literature based. The benefit-risk-ratio of Signum bondings is positive.
The clinical evaluation for Signum cre-active is based on currently available scientific literature.
Signum cre-active is marketed since 1991 (formerly Dentacolour cre-active).
Specific questions of this clinical evaluation were answered with adequate and suitable data from liternal documentation. Color-Fluids can be considered as well established technology. The adequate safety and performance of the evaluated devices was proven within this clinical evaluation with data from product surveillance and from scientific literature. The benefit risk ratio for Signum cre-active is positive.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.