K211854 · Vita Zahnfabrik GmbH H Rauter & CO · EBD · Nov 22, 2021 · Dental
Device Facts
Record ID
K211854
Device Name
VITA Akzent LC
Applicant
Vita Zahnfabrik GmbH H Rauter & CO
Product Code
EBD · Dental
Decision Date
Nov 22, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3310
Device Class
Class 2
Indications for Use
VITA AKZENT LC Indication range: - Restorations made of hybrid ceramic - · Restorations made of light-curing veneering material - · Restorations made of CAD/CAM composites - · Prefabricated Teeth - · Denture bases
Device Story
VITA Akzent LC is a light-curing, methacrylate-based stain and glaze system; used for extraoral surface characterization of dental restorations (hybrid ceramic, resin veneering, CAD/CAM composites, prefabricated teeth, denture bases) and internal characterization via layering technique. Operated by dental professionals in a clinical or laboratory setting. Device is applied to restoration surfaces; light-cured to achieve characterization, surface smoothness, and wear resistance. Output is a modified, characterized dental restoration surface. Benefits include improved aesthetic customization and surface properties of dental prosthetics.
Clinical Evidence
Bench testing only. No human clinical testing performed. Biocompatibility assessment conducted per ISO 10993-1:2009 and ISO 7405:2004. Adhesion performance testing compared to reference device demonstrated favorable efficacy.
Technological Characteristics
Light-curing methacrylate-based liquid stain/glaze system. Composition includes methacrylates, initiators, and pigments. Biocompatibility compliant with ISO 10993-1 and ISO 7405.
Indications for Use
Indicated for shade customization and surface characterization of dental restorations including hybrid ceramics, light-curing veneering materials, CAD/CAM composites, prefabricated teeth, and denture bases.
Regulatory Classification
Identification
A coating material for resin fillings is a device intended to be applied to the surface of a restorative resin dental filling to attain a smooth, glaze-like finish on the surface of the filling.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration. The Department of Health & Human Services logo is on the left, and the U.S. Food & Drug Administration logo is on the right. The FDA logo is in blue and white.
November 22, 2021
VITA Zahnfabrik GmbH H Rauter & Co % Lindsay Tilton Regulatory Affairs Consultant VITA North America 22705 Savi Ranch Parkway, Suite 100 Yorba Linda. California 92887
Re: K211854
Trade/Device Name: VITA Akzent LC Regulation Number: 21 CFR 872.3310 Regulation Name: Coating Material For Resin Fillings Regulatory Class: Class II Product Code: EBD Dated: August 25, 2021 Received: August 27, 2021
Dear Lindsay Tilton:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{2}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K211854
Device Name VITA Akzent LC
Indications for Use (Describe)
VITA AKZENT LC Indication range:
- Restorations made of hybrid ceramic
- · Restorations made of light-curing veneering material
- · Restorations made of CAD/CAM composites
- · Prefabricated Teeth
- · Denture bases
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
#### K211854
| Submitter | VITA Zahnfabrik H.Rauter GmbH Co.<br>Spitelgasse 3<br>Bad Sackingen, D-79713<br>Germany<br>Establishment Reg. No. 1000625496 |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact | Bernd Walker<br>Head of Regulatory Affairs and Quality System<br>Phone (+49) 7761 562-361<br>Fax (+49) 7761 562-384<br>B.Walker@vita-zahnfabrik.com |
| Official Correspondent | Lindsay Tilton<br>Regulatory Affairs Consultant<br>VITA North America, Inc.<br>22705 Savi Ranch Parkway, Suite 100<br>Yorba Linda, CA 92887<br>Establishment Reg. No. 2082832<br>Phone (800) 828-3839<br>Fax (714) 221-6759<br>E-mail: Itilton@vitanorthamerica.com |
Date Prepared
November 16,2021
- י Trade/Device Name Vita Akzent® LC
- י Classification Name Coating Material for Resin Filling
- Regulation Number 21 CFR 872.3310
- י Product Code EBD
# Predicate Devices
Vident's VITA Enamic Stains– K123761 – Primary Predicate Device
{5}------------------------------------------------
# GC America's Optiglaze - K133836 - Reference Device
#### Device Description
VITA Akzent® LC VITA AKZENT LC is a light-curing methacrylatebased stain/ glaze system for extraoral surface characterization of dental restorations made of hybrid ceramic, resin veneering materials, CAD/CAM composites, prefabricated teeth and denture base resins. It can also be used for internal characterization with the layering technique of veneering composites.
### Indications for Use
VITA AKZENT LC Indication range:
- Restorations made of hybrid ceramic
- Restorations made of light-curing veneering material
- Restorations made of CAD/CAM composites
- Prefabricated Teeth
- Denture bases
#### Technological Characteristics
VITA Akzent® LC provides a light cured glaze system similar to the predicate device formulas. The composition of the VITA Akzent®LC and the predicates are based on similar materials, therefore this difference does not impact the substantial equivalence.
#### Non-Clinical Performance Testing
VITA Akzent® LC meets the applicable requirements of the following FDA recognized standards:
- . ISO 10993-1:2009 – Biological evaluation of medical devices
- I ISO 7405:2004 – Dentistry – Evaluation of biocompatibility of medical devices
#### Biocompatibility
A biocompatibility assessment was performed on VITA Akzent® AC in accordance with ISO 10993-1:2009 – Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing within a Risk Management Process, and ISO 7405:2008 – Dentistry – Evaluation of Biocompatibility of Medical Devices Used in Dentistry. This assessment supports that VITA Akzent® LC is biocompatible and concludes that the device is substantially equivalent to the predicate device in terms of biocompatibility.
{6}------------------------------------------------
# Clinical Performance Data
Not applicable. No human clinical testing was performed to support the substantial equivalence of VITA Akzent® AC.
### Comparison to Predicate Devices
Below is a table showing the predicate device comparison.
| | Subject Device | Primary Predicate Device | Reference Device |
|----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | VITA Akzent® LC - K211854 | Vident's VITA Enamic<br>Stains- K123761 | GC America's Optiglaze<br>- K133836 |
| Device Classification | Coating, Filling Material,<br>Resin | Coating,Filling Material,<br>Resin | Coating,Filling<br>Material, Resin |
| Product Code | EBD | EBD | EBD |
| Indication for use | VITA AKZENT LC Indication<br>range:<br>• Restorations made of<br>hybrid ceramic<br>• Restorations<br>made of light curing<br>veneering material<br>• Restorations made of<br>CAD/CAM<br>Composites<br>• Prefabricated<br>Teeth<br>• Denture bases | VITA Enamic Stains are<br>indicated for shade<br>customization and<br>characterization of the<br>surface of dental<br>restorations made of<br>hybrid ceramic-resin and<br>resin materials. | For characterization<br>of direct and indirect<br>composite<br>restorations, acrylic<br>denture base and<br>artificial acrylic<br>teeth.<br>For obtaining surface<br>smoothness and<br>wear resistance of<br>restorations made of<br>composite<br>resin, acrylic denture<br>base and artificial<br>acrylic teeth. |
| Technological<br>Characteristics | Liquid | Powder and Liquid | Liquid |
| Principle of<br>operation | For characterization and<br>glazing restorations | For characterization and<br>glazing restorations | For characterization<br>and glazing<br>restorations |
| Material<br>composition | A mixture of | A mixture of | Multifuctional acrylate, |
{7}------------------------------------------------
| methacrylates, initiators and<br>other non-hazardous | methacrylates, initiators<br>and other non-hazardous | methyl methacrylate,<br>silica<br>filler, photo initiator,<br>pigment |
|------------------------------------------------------|------------------------------------------------------|-----------------------------------------------------------------------|
| additions | additions | |
Differences between Subject and Predicate Devices
The predicate device indication for use was submitted with a more generic indication. Below is how the predicate device is similar in indications for Akzent LC.
The predicate device was designed states that it's intended for customization and characterization of the surface of dental restorations made of hybrid ceramic-resin and resin materials. These terms are similar to the subject's device indication of Restorations made of hybrid ceramic, light-curing veneering material (resin materials), CAD/CAM composites (resin materials) and Denture bases (resin materials)
Both cover the same indications, and no additional indications are added to the subject device, therefore this does not affect the substantial equivalence or efficacy and safety.
The reference device was used since the subject device product is in liquid form as is the reference device. The primary predicate uses a liquid and a powder form. The same mechanism is used, and the subject device is considered a subset of the primary predicate device and equivalent to the reference device. Performance and biocompatibility testing prove that the subject device, in a liquid form, does not impact the safety and effectiveness of the product and shows. The adhesion performance testing conducted against the reference device, shows a more favorable efficacy during bench testing.
# Conclusion
Based on the above comparative tables, it is concluded that VITA Akzent®LC is equivalent to the predicate device. VITA Akzent® LC has similar indications for use and similar material composition as the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.