K133836 · GC America, Inc. · EBD · Mar 6, 2014 · Dental
Device Facts
Record ID
K133836
Device Name
OPTIGLAZE COLOR
Applicant
GC America, Inc.
Product Code
EBD · Dental
Decision Date
Mar 6, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3310
Device Class
Class 2
Indications for Use
For characterization of direct and indirect composite restorations, acrylic denture base and artificial acrylic teeth. For obtaining surface smoothness and wear resistance of restorations made of composite resin, acrylic denture base and artificial acrylic teeth.
Device Story
Optiglaze Color is a light-cured resin material used by dental professionals to characterize and glaze dental restorations. It is applied to direct and indirect composite restorations, acrylic denture bases, and artificial acrylic teeth. The device is supplied in 17 shades and is applied using a brush to the surface of the restoration after contouring. Once applied, the material is polymerized using a visible light-curing device. The output is a hardened, characterized, and smooth surface layer that provides improved wear resistance and aesthetic customization. The device is intended for use in a clinical or laboratory setting by dental practitioners.
Clinical Evidence
Bench testing only. Performance testing confirmed compliance with specifications for appearance, depth of cure, color stability after irradiation, surface hardness, and application characteristics.
Technological Characteristics
Light-cured resin coating material. Composition: multifunctional acrylate, methyl methacrylate, amorphous fumed silica filler, photo-initiator, and pigments. Polymerization via photo-initiator system triggered by visible light. Form factor: liquid in bottles/kits with brushes. Biocompatibility: ISO 10993-1 compliant.
Indications for Use
Indicated for characterization and surface finishing of direct/indirect composite restorations, acrylic denture bases, and artificial acrylic teeth to improve smoothness and wear resistance.
Regulatory Classification
Identification
A coating material for resin fillings is a device intended to be applied to the surface of a restorative resin dental filling to attain a smooth, glaze-like finish on the surface of the filling.
{1}------------------------------------------------
KI33836
Section 5 - 510(k) Summary
Image /page/0/Picture/2 description: The image shows the text "MAR - 6 2014" in the upper portion of the image. Below the date is the text "CCI" in a bold, sans-serif font. The text is black and the background is white.
GC AMERICA INC..
3737 W 127th STREET
ALSIP, ILLINOIS 60803
TEL (708) 597-0900
FAX (708) 926-9100
www.gcamerica.com
- 1. Submitter Information:
GC AMERICA INC. 3737 W. 127th Street Alsip, IL 60803
| Contact Person: | Mark Heiss, D.D.S. |
|-----------------|--------------------|
| Phone: | (708) 926-3090 |
| Fax: | (708) 926-9100 |
Date Prepared: December 13, 2013
- 2. Device Name ..
| Proprietary Name: | Optiglaze Color |
|------------------------|----------------------------------|
| Classification Name: | Coating, Filling Material, Resin |
| Device Classification: | Class II, 872.3310 |
| Product Code: | EBD |
- Predicate Devices: 3.
| ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>A MIN OR LAND LE LE SELECT | ANNUAL COLLECT LE LAND<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>---------<br>.<br><br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | 20 - 1 - 1 - 1 - 1 - | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IN LAND SER E E E LAND SERVED LAUN<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>.<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>FINALERS CLASS CLASS COLLECT CAME CA<br>AND OF CHININGS CHARACT CHARACT COLLEGIAL COLLEGIAL COLLEGION COLLEGION CONTRACT COLLEGION CONTRACT COLLEGION CONTRACT COLLEGION CONTRACT COLLEGION CONTRACT CARD AND<br>. | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | C<br>---------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |
- Description of Device: 4.
- 5. Optiglaze Color is a light-cured glazing and for characterization of direct and indirect composite restorations, acrylic denture base and artificial acrylic teeth, and also for obtaining surface smoothness and wear resistance of restorations made of composite resin, acrylic denture base and artificial acrylic teeth. The material is available in 17 shades.
- 6. Indications for Use:
- For characterization of direct and indirect composite restorations, acrylic denture base and a. artificial acrylic teeth.
- b. For obtaining surface smoothness and wear resistance of restorations made of composite resin, acrylic denture base and artificial acrylic teeth.
## 7. Technological characteristics:
All the components of the applicant device, Optiglaze Color, have already been used in the predicate devices. The curing mechanism of the predicate devices is polymerization of uncured methacrylate ester monomers. This reaction is caused by photo initiator system.
- 8. Substantial equivalence:
The applicant device complies with all the requirements of ISO 10993-1 Biological evaluation of medical devices Part 1: Evaluation & testing.
The curing mechanism of the new and predicate devices is substantially equivalent in principle. Therefore, the new and predicate devices are the same in function, and similar in composition and intended use. This supports that the compatibility and safety of the applicant device are substantially equivalent to the predicate devices.
{2}------------------------------------------------
Page 2
- . 9. Differences
The following differences may be noted between the predicate device and Optiglaze Color: · Optiglaze Color has higher surface hardness and comes in multiple shades.
- 10. Performance Bench Tests
It is confirmed that the device conforms to the required specifications and is suitable for its intended use. Performance testing includes:
- Appearance .
- Depth of cure ●
- Color stability after irradiation �
- Surface hardness �
- Application characteristics .
- ਉਂ Packaging
Complete set: A-Plus 2.6ml, B-Plus 2.6ml, C-Plus 2.6ml, White 2.6ml, Ivory-White 2.6ml, Yellow 2.6ml, Orange 2.6ml, Pink-Orange 2.6ml, Pink 2.6ml, Red-Brown 2.6ml, Olive 2.6ml, Lavender 2.6ml, Grey 2.6ml, Blue 2.6ml, Red 2.6ml, Clear 5ml, Clear HV 5ml
Accessories Disposable Pallet Flat Brush Round Brush Brush Holder Box
- 10. Shelf Life Evaluation and Storage Conditions:
- Shelf Life 3 years -
- Store in a cool and dark place. 4-25°C (39.2 77.0°F)
| | Applicant device | Comparative device |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product<br>category | Coating, Filling Material, Resin, Class II | Coating, Filling Material, Resin, Class II |
| Trade name | OPTIGLAZE COLOR | G-Coat |
| Manufacturer | GC Corporation | GC Corporation |
| Intended use | For characterization of restorations<br>made of composite resins & acrylics.<br>For obtaining surface smoothness and<br>wear resistance of restorations made of<br>composite resins & acrylics. | Used to seal the surface of followings to protect<br>from moisture and dehydration, and to provide<br>an extra smooth surface gloss: direct composite<br>and glass-ionomer restorations, indirect<br>composite restorations, provisional restorations,<br>and their adhesive interface of margin to<br>enamel or dentin. |
| Product<br>description | OPTIGLAZE COLOR is a light-cured<br>glazing and characterizing material for<br>such restorations as direct and indirect<br>composite resins & acrylics. | G-COAT is a light-cured protective coating<br>agent to be used to seal the surface of such<br>restorations as direct or indirect composites,<br>glass-ionomer restorations, provisional and the<br>interface between those restorations and<br>enamel or dentine. The coated surface<br>presents high smoothness, and it may reduce<br>or even eliminate the need for manual<br>polishing. |
{3}------------------------------------------------
| | | polishing. |
|-------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Components | Multifunctional acrylate.<br>Methyl methacrylate.<br>Silica filler (amorphous, fumed).<br>Photo initiator.<br>Pigment. | Multifunctional methacrylate.<br>Methyl methacrylate.<br>Silica filler (amorphous, fumed).<br>Photo initiator. |
| Instructions for<br>use | 1) After contouring the resin surface,<br>finish with a carbide bur or coarse<br>silicone point.<br>2) Wash and dry the surface.<br>3) Apply a thin layer to the resin surface<br>using the brush provided.<br>4) Light cure with LaboLight LV-3 or<br>equivalent halogen light for 5 minutes. | a) Contour restoration according to the<br>manufacturer's instructions.<br>b) Clean the restoration and adjacent surfaces<br>to be coated with pumice and water.<br>c) Rinse and dry.<br>d) Using a cotton pellet or brush apply G-COA<br>to all exposed restoration and adjacent surface<br>e) Light cure for 20 seconds using suitable<br>visible light curing device.<br>f) Close bottle immediately after use. |
:
.
:
and the control of the county of the county of
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes representing the three levels of government: federal, state, and local. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 6, 2014
GC AMERICA INCORPORATED Dr. Mark Heiss, D.D.S. 3737 W. 127th Street Alsip, IL 60803
Re: K133836
Trade/Device Name: OPTIGLAZE COLOR Regulation Number: 21 CFR 872.3310 Regulation Name: Coating Material For Resin Fillings Regulatory Class: II Product Code: EBD Dated: December 13, 2013 Received: December 18, 2013
Dear Dr. Heiss:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 -- Dr. Heiss
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Section 4 - Indications for Use Statement
Indications for Use
135836 510(k) Number (if known):
Device Name: Optiglaze Color
Indications for Use:
- 1. For characterization of direct and indirect composite restorations, acrylic denture base and artificial acrylic teeth.
- 2. For obtaining surface smoothness and wear resistance of restorations made of composite. resin, acrylic denture base and artificial acrylic teeth.
Prescription Use _ ਮ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Runner -S -05,00,
Page 4.1 of 4.1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.