K152373 · Vita Zahnfabrik H.Rauter GmbH & Co. · EBF · Dec 30, 2015 · Dental
Device Facts
Record ID
K152373
Device Name
VITA VM LC Flow
Applicant
Vita Zahnfabrik H.Rauter GmbH & Co.
Product Code
EBF · Dental
Decision Date
Dec 30, 2015
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 872.3690
Device Class
Class 2
Indications for Use
Full and partial veneering of metal frameworks: crowns, bridges, telescopic crowns, implant superstructure Individualization of and layering on VITA ENAMIC® Layering-over long-term temporaries made from VITA CAD-Temp® Individualization of acrylic teeth Reproduction of gingival components Metal-free crowns and three-unit anterior bridges as long-term temporary restorations Veneering of removable and partially removable dentures (according to the manufacturer's information) Inlays Veneers
Device Story
VITA VM LC Flow is a light-cured microparticle indirect composite material; used for extraoral fabrication of fixed and removable dental restorations. Material consists of dimethacrylate resin mixture and filler particles. Fabricated by dental technicians in a laboratory setting; applied to frameworks or used for temporary restorations. Polymerization achieved via external light or UV radiation source. Output is a hardened, tooth-shaded dental restoration. Benefits include aesthetic restoration of dental structures and provision of long-term temporary solutions.
Clinical Evidence
Bench testing only. Compliance with ISO 10477:2004(E) demonstrated through physical property testing including flexural strength (135 MPa), E-modulus (6600-7500 MPa), water absorption (25.4 µg/mm³), and water solubility (0.1 µg/mm³). Biocompatibility assessment performed per ISO 10993-1:2009 and ISO 7405:2008; no clinical data provided.
Technological Characteristics
Type 2, Class 2 polymer-based crown and bridge material per ISO 10477:2004(E). Composition: dimethacrylate resins, inorganic fillers, pigments, and light/UV-sensitive initiators. Physical form: paste. Energy source: external light or UV radiation for polymerization. Biocompatibility assessed per ISO 10993-1 and ISO 7405.
Indications for Use
Indicated for dental patients requiring full/partial veneering of metal frameworks (crowns, bridges, telescopic crowns, implant superstructures), individualization/layering of VITA ENAMIC or acrylic teeth, layering over long-term temporaries (VITA CAD-Temp), reproduction of gingival components, metal-free crowns/three-unit anterior bridges as long-term temporaries, veneering of removable dentures, inlays, and veneers.
Regulatory Classification
Identification
Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol. The logo is simple and recognizable, and it is often used to represent the U.S. Department of Health & Human Services.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 30, 2015
Vita Zahnfabrik H. Rauter Gmbh & Co. c/o Ms. Nevine Erian Director, Regulatory Affairs & Compliance Vita North America, Inc. 22705 Savi Ranch Parkway, Suite 100 Yorba Linda. California 92887
Re: K152373
Trade/Device Name: VITA VM® LC Flow Regulation Number: 21 CFR 872.3690 Regulation Name: Tooth Shade Resin Material Regulatory Class: II Product Code: EBF Dated: December 1, 2015 Received: December 2, 2015
#### Dear Ms. Erian:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology. General Hospital. Respiratory. Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known): K152373
Device Name: VITA VM®LC Flow
#### Indications for Use:
- 그 Full and partial veneering of metal frameworks: crowns, bridges, telescopic crowns, implant superstructure
- 그 Individualization of and layering on VITA ENAMIC®
- Layering-over long-term temporaries made from VITA CAD-Temp®
- 트 Individualization of acrylic teeth
- Reproduction of gingival components
- 트 Metal-free crowns and three-unit anterior bridges as long-term temporary restorations
- Veneering of removable and partially removable dentures (according to the manufacturer's information)
- Inlays
- . Veneers
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (21 CFR Part 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR Part 807 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
# Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sian Off) Division of Dental, Infection Control and General Hospital Devices
510(k) Number
Prescription Use (Par. 21 CFR 801.109) OR
Over-The-Counter Use
{4}------------------------------------------------
# Section 5 – 510(k) Summary – K152373
| Submitter | Vita Zahnfabrik H.Rauter GmbH Co.<br>Spitelgasse 3<br>Bad Sackingen, D-79713<br>Germany<br>Establishment Reg. No. 1000625496 | | |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
| Contact | Bernd Walker<br>Head of Regulatory Affairs and Quality Systems<br>Phone (+49) 7761/562-361<br>Fax (+49) 7761/562-384 | | |
| Official Correspondent | Nevine Erian<br>Director, Regulatory Affairs & Compliance (Consultant)<br>VITA North America, Inc.<br>22705 Savi Ranch Parkway, Suite 100<br>Yorba Linda, CA 92887<br>Establishment Reg. No. 2082832<br>Phone (949) 370-7155<br>Fax (714) 221-6759<br>E-mail: nerian@vident.com | | |
| Date Prepared | December 29, 2015 | | |
| Trade/Device Name | VITA VM® LC Flow | | |
| Common Name | Dental Veneering Material | | |
| Classification Name | Tooth Shade Resin Material | | |
| Regulation Number | 21 CFR 872.3690 | | |
| Product Code | EBF | | |
# Predicate Devices
VITA Zeta® (VITA Zahnfabrik GmbH) – K961434 – Primary Predicate SR Nexco® (Ivoclar Vivadent, AG) – K123304 – Reference Predicate SINFONY® (ESPE Dental, AG) – K992645 – Reference Predicate
{5}------------------------------------------------
### Device Description
VITA VM LC Flow is a light-cured microparticle indirect composite material for fixed and removable restorations for extraoral fabrication.
#### Statement of Intended Use
- Full and partial veneering of metal frameworks: crowns, bridges, telescopic crowns, implant superstructure
- 트 Individualization of and layering on VITA ENAMIC®
- Layering-over long-term temporaries made from VITA CAD-Temp®
- 트 Individualization of acrylic teeth
- Reproduction of gingival components
- 트 Metal-free crowns and three-unit anterior bridges as long-term temporary restorations
- . Veneering of removable and partially removable dentures (according to the manufacturer's information)
- Inlays
- 트 Veneers
#### Material Composition
VITA VM LC Flow pastes consist of a mixture of dimethacrylates and filler particles.
#### Technological Characteristics
VITA VM LC Flow is classified as Type 2, Class 2, Polymer-based crown and bridge materials whose setting is effected by the application of energy from an external source, such as heat and/or light or UV radiation, Polymer-based crown and bridge materials that contain a light or UV-sensitive initiator , per ISO 10477:2004(E) - Dentistry -Polymer-based crown and bridge materials standard.
#### Biocompatibility
A biocompatibility assessment was performed on VITA VM LC Flow, in accordance with ISO 10933-1:2009 – Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing within a Risk Management Process, and ISO 7405:2008 – Dentistry – Evaluation of Biocompatibility of Medical Devices Used in Dentistry. Based on the assessment, biocompatibility testing was determined not necessary. This assessment supports that
{6}------------------------------------------------
VITA VM LC Flow is biocompatible.
#### Substantial Equivalence
VITA VM LC Flow shares the same intended use as the predicate devices. VITA VM LC Flow is comprised of same materials as the predicate devices. VITA VM LC Flow shares similar indications, chemical composition, technological characteristics and physical properties as the predicate devices. Any differences between VITA VM LC Flow and its predicates do not affect substantial equivalence as VITA VM LC Flow meets same biocompatibility requirements and ISO 10477 physical property requirements as the predicate devices.
Information provided in this application demonstrates that VITA VM LC Flow is substantially equivalent to the predicate devices.
| | VITA VM LC<br>Flow | VITA Zeta<br>(Primary<br>Predicate) | SR Nexco<br>(Reference<br>Predicate) | SINFONY<br>(Reference<br>Predicate) |
|--------------------------------------------------------------------------------------------------------------------|--------------------|-------------------------------------|--------------------------------------|-------------------------------------|
| Indications for Use | | | | |
| Full & Partial veneering of<br>metal frameworks: crowns,<br>bridges, telescopic crowns,<br>implant superstructures | Yes | Yes | Yes | Yes |
| Veneering of acrylic resin<br>frameworks | Yes | Yes | Yes | Yes |
| Layering-over long-term<br>temporaries | Yes | Yes | Yes | Yes |
| Individualization of acrylic<br>teeth | Yes | Yes | Yes | Yes |
| Metal-free crowns & three-<br>unit anterior bridges as<br>long-term temporary<br>restorations | Yes | Yes | Yes | Yes |
| Inlays | Yes | Yes | Yes | Yes |
| Onlays | No | Yes | Yes | Yes |
| Veneers | Yes | Yes | Yes | Yes |
| Technological Aspect | | | | |
| Light-Cured Material | Yes | Yes | Yes | Yes |
| Physical Configuration | Paste | Paste &<br>Powder | Paste | Paste |
#### Technical Comparison of VITA VM LC Flow to Predicate Devices
{7}------------------------------------------------
| Material Composition | All are made of Methacrylate resins, photo-initiators,<br>inorganic fillers and pigments | | | | |
|-------------------------------|------------------------------------------------------------------------------------------|-------------------|--------------------------|------------------------|--|
| ISO 10477 – Physical Property | All Pass | | | | |
| Flexural Strength | 135 MPa | 95 MPa | $90 \u00b1 10$ MPa | 105 MPa | |
| E-Modulus | 6600 - 7500<br>MPa | 3450 MPa | $6500 \u00b1 500$<br>MPa | 3100 MPa | |
| Water Absorption | 25.4 µg/mm³ | 28 - 30<br>µg/mm³ | $15 \u00b1 1$ µg/mm³ | $\approx 20$<br>µg/mm³ | |
| Water Solubility | 0.1 µg/mm³ | 0 µg/mm³ | $1 \u00b1 0.5$<br>µg/mm³ | 0.5 µg/mm³ | |
# Non-Clinical Performance Testing
Bench testing was performed in accordance with FDA recognized standard, ISO 10477:2004(E) – Dentistry – Polymer-based crown and bridge materials standard. VITA VM LC Flow meets applicable ISO 10477 criteria for the designated indications.
#### Clinical Performance Data
Not applicable. No human clinical testing was performed to support the substantial equivalence of VITA VM LC Flow.
#### Conclusion
Non-clinical performance testing of VITA VM LC Flow demonstrates that VITA VM LC Flow is as safe and effective for its intended use, and performs as well as the legally marketed predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.