Visalys Restorative Primer

K191523 · Kettenbach GmbH & Co. KG · KLE · Oct 9, 2019 · Dental

Device Facts

Record IDK191523
Device NameVisalys Restorative Primer
ApplicantKettenbach GmbH & Co. KG
Product CodeKLE · Dental
Decision DateOct 9, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3200
Device ClassClass 2

Indications for Use

Restoration primer for treatment of: metal/silicate ceramics, oxide ceramic and composite surfaces - to cement prosthetic restorations - to repair fractured restorations with composite materials

Device Story

Visalys Restorative Primer is a single-component dental primer used to enhance adhesion between restorative materials (metal, composite, silicate/oxide ceramics) and luting materials. It contains adhesive monomers (10-MDP, silane methacrylate) in an ethanol base. The clinician applies the primer to the restoration surface; it does not require light-curing. The device is used in a dental clinic setting by a dentist or dental professional. By creating an adhesive surface, it enables the cementation of prosthetic restorations (crowns, bridges, inlays, onlays) and the repair of fractured restorations. The output is a chemically prepared surface that improves bond strength, ensuring the stability and longevity of the dental restoration.

Clinical Evidence

No clinical data provided. Substantial equivalence supported by bench testing, including shear bond strength testing on lithium disilicate, zirconium oxide, and non-precious metal, and biocompatibility testing per ISO 10993 and ISO 7405 (cytotoxicity, mutagenicity, sensitization, chemical analysis).

Technological Characteristics

Single-component liquid primer. Composition: ethanol, 10-MDP (acidic adhesive monomer), and silane methacrylate. No BPA or BPA precursors. Storage: 2–8 °C. Application: bottle system. Non-light-cured. Biocompatibility: ISO 10993, ISO 7405.

Indications for Use

Indicated for surface treatment of metal, silicate ceramic, oxide ceramic, and composite surfaces to facilitate cementation of prosthetic restorations or repair of fractured restorations.

Regulatory Classification

Identification

A resin tooth bonding agent is a device material, such as methylmethacrylate, intended to be painted on the interior of a prepared cavity of a tooth to improve retention of a restoration, such as a filling.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. October 9, 2019 Kettenbach GmbH & Co. KG Katja Simon Regulatory Affairs Manager Im Heerfeld 7 35713 Eschenburg, Germany Re: K191523 Trade/Device Name: Visalys Restorative Primer Regulation Number: 21 CFR 872.3200 Regulation Name: Resin Tooth Bonding Agent Regulatory Class: Class II Product Code: KLE Dated: September 4, 2019 Received: September 9, 2019 Dear Katja Simon: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {2}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, for Srinivas Nandkumar, Ph.D. Acting Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## Indications for Use #### 510(k) Number (if known) K191523 Device Name Visalys® Restorative Primer Indications for Use (Describe) Restoration primer for treatment of: metal/silicate ceramics, oxide ceramic and composite surfaces - · to cement prosthetic restorations - · to repair fractured restorations with composite materials | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for KETTENBACH. The logo consists of a blue square with six white circles arranged in a triangular pattern at the top. Below the circles, the word "KETTENBACH" is written in a bold, sans-serif font. Underneath the company name is the slogan "Simply intelligent" in a smaller, bold font. To the right of the blue square, there is a circled "R" symbol, indicating a registered trademark. # 510(k) Summary In accordance with the requirements of the Safe Medical Device Act, Kettenbach GmbH & Co KG herewith submits a Summary. ### A. Name and address of manufacturer: Kettenbach GmbH & Co KG Im Heerfeld 7 35713 Eschenburg Germany Establishment Registration No.: 9681356 Owner/Operator Number: 9022134 #### Name, title and phone number of official correspondent: Simon, Katja Regulatory Affairs Manager Im Heerfeld 7 35713 Eschenburg Germany Phone: + 49 277 4705 0 E-mail: katja.simon@kettenbach.com #### Name, title and phone number of U.S. Agent (Contact): Roggenbau, Wilfried InterGest North America LLC. 400 Oser Ave., Suite 1650 Hauppauge, NY 11788 Phone: 631 5010500 ext Fax: 631 5011060 Email: roggenbauw@intergestna.com Date of preparation: September 04th, 2019 {5}------------------------------------------------ Image /page/4/Picture/0 description: The image contains the logo for KETTENBACH. The logo features a blue square with six white circles arranged in a triangular pattern at the top. Below the circles, the word "KETTENBACH" is written in a bold, sans-serif font. Underneath "KETTENBACH", the phrase "Simply intelligent" is written in a smaller, sans-serif font. B. Device Identification: Device Trade Name: Common Name: # Visalys® Restorative Primer Visalys® Restorative Primer Universal adhesion primer for restoration materials #### Classification of the device: Device Classification Name: Product Code: Device Classification No.: Panel: Requlatory Status: Resin tooth bonding agent KLE Part 872.3200 Dental Class II #### C. Predicate device: Device Trade Name: Applicant: 510(k) No.: CLEARFIL™ CERAMIC PRIMER PLUS Kuraray Noritake Dental Inc. K150703 ### D. Device Description: Visalys® Restorative Primer with adhesive monomers (10-MDP, silane methacrylate) is for building up adhesive surfaces on restoration materials (metal, composite, as well as silicate and oxide ceramics). Visalys® Restorative Primer is a single component primer. Do not light cure Visalys® Restorative Primer. BPA or BPA precursors are not used in the manufacturing process of this device. ### E. Indications for use: Restoration primer for treatment of: metal/silicate ceramics, oxide ceramic and composite surfaces - · to cement prosthetic restorations - · to repair fractured restorations with composite materials {6}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for KETTENBACH. The logo consists of a blue square with six white circles arranged in a triangular shape. Below the square, the word "KETTENBACH" is written in white, followed by the tagline "Simply intelligent devices" in black. There is also a registered trademark symbol in the upper right corner of the blue square. # F. Comparison of technological characteristic with the predicate devices | | Predicate Devices | Substantial Equivalent Device | Conclusion | |------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------| | Product | CLEARFIL™ CERAMIC PRIMER PLUS | Visalys® Restorative Primer | | | Manufacturer | Kuraray Noritake Dental Inc. | Kettenbach GmbH & Co. KG | | | 510(k)<br>numbers | K150703 | K191523 | | | Product<br>Description | CLEARFIL™ CERAMIC PRIMER PLUS is a<br>dental universal prosthetic primer that<br>provides an enhanced adhesive surface to<br>ceramic (e.g. conventional porcelain, lithium<br>disilicate, zirconia), hybrid ceramics (e.g.<br>ESTENIA C&B), composite resin and metal. | Visalys® Restorative Primer with adhesive<br>monomers (10-MDP, silane methacrylate) is<br>for building up adhesive surfaces on<br>restoration materials (metal, composite, as well<br>as silicate and oxide ceramics). Visalys®<br>Restorative Primer is a single component<br>primer that does not have to be light-cured. | Similar | | Indication for<br>use | Indications:<br>CLEARFIL™ CERAMIC PRIMER PLUS is<br>indicated for the following uses:<br>[1] Surface treatment of prosthetic<br>restorations made of ceramic, hybrid<br>ceramics, composite resin or metal<br>[2] Intraoral repairs of fractured restorations<br>made of ceramics, hybrid ceramics,<br>composite resin or metal | Indications:<br>Restoration primer for treatment of<br>metal/silicate ceramics, oxide ceramic and<br>composite surfaces<br>• to cement prosthetic restorations<br>• to repair fractured restorations with composite<br>materials | Similar | | Storage<br>temperature | 2 – 8 °C/36 – 46 °F | 2 – 8 °C/35 – 46 °F | Similar | | Application<br>system | Bottle | Bottle | Similar | | Composition | • 3-Methacryloxypropyl trimethoxysilane<br>• 10-Methacryloyloxydecyl dihydrogen<br>phosphate (MDP)<br>• Ethanol | The main components of the Visalys®<br>Restorative Primer are<br>ethanol, acidic adhesive monomer (10-MDP)<br>and silane methacrylate. | Similar | | | Predicate Devices | Simply intelligent<br>Substantial Equivalent Device | Conclusion | | Product | CLEARFIL™ CERAMIC PRIMER PLUS | Visalys® Restorative Primer | | | Manufacturer | Kuraray Noritake Dental Inc. | Kettenbach GmbH & Co. KG | | | 510(k) numbers | K150703 | K191523 | | | Shear bond strengths on different restorative materials<br>(ISO 16506 / Chapter A 2.2.5): | | | | | Lithium<br>disilicate | (8.9 MPa) | (10.2 MPa) | Similar | | Zirconium<br>oxide | (8.4 MPa) | (7.5 MPa) | Similar | | Non-precious<br>metal | (8.3 MPa) | (12.5 MPa) | Similar | | Products used<br>in the testing | CLEARFIL™ Ceramic Primer Plus<br>(K150703) was tested with the resin<br>product Panavia V5 dental cement<br>(K150704), manufactured by Kuraray | Visalys® Restorative Primer (K191523)<br>was tested with the resin product Visalys<br>CemCore (K191527), manufactured by<br>Kettenbach GmbH & Co. KG. | | | Product description and composition for the predicate device has been extracted from the instruction for use. Values given in brackets<br>have been determined by Kettenbach | | | | {7}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the logo for KETTENBACH. The logo consists of a blue square with six white circles arranged in a triangular pattern at the top. Below the circles is the word "KETTENBACH" in white, and below that is the phrase "Simply intelligent" in black. There is also a registered trademark symbol in the upper right corner. Table 1: Visalys® Restorative Primer product description and characteristic to predicate device {8}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the logo for Kettenbach GmbH & Co. KG. The logo features a blue square with a white graphic of six circles arranged in a triangular shape at the top. Below the graphic, the word "KETTENBACH" is written in white, followed by the tagline "Simply intelligent" in black. ### G. Technological Characteristics Summary: Visalys® Restorative Primer is considered substantially equivalent to Clearfil Ceramic Primer Plus. There is no significant difference in intended use or technology. Therefore Visalys® Restorative Primer is substantially equivalent to the above listed predicate device. Differences in respect to chemical composition and resulting mechanical and physical properties have been evaluated. #### H. Summary of Non-Clinical Performance Testing: The following in vitro bench tests were performed on the Visalys® Restorative Primer to verify physical properties and performance in support of substantial equivalence: - . Shear Bond strength on restorative materials acc. to ISO 16506 "Dentistry -Polymer-based luting materials containing adhesive component shear bond strength test" (Chapter A 2.2.5). The performance of the Visalys® Restorative Primer satisfactorily met the requirements of the non-clinical bench testing conducted to support substantial equivalence. Biocompatibility tests have been performed to assure biological safety in accordance with the ISO 10993 family and ISO 7405. Tests in respect of cytotoxicity (ISO 10993-5), mutagenicity (OECD 487), sensitization (ISO 10993-10) and a chemical analysis (10993-18) showed, that Visalys® Restorative Primer, universal adhesion primer for restoration materials biocompatibility data is comparable or superior to the predicate device. Therefore no toxicological risks and resulting hazards for patients, users and third parties can be concluded. ### Rationale for using ISO 16506 (Chapter A 2.2.5) as test standard The Medical Device Product Classification for the stated product is KLE (Agent, Tooth Bonding, Resin). However, the stated product Visalys® Restorative Primer acts as a coupling agent between restoration material (e.g. ceramics, metals) and luting material (e.g. Visalys® CemCore) and not between luting material and tooth structure. For the later indication, Visalys® Tooth Primer can be used. The recognized test standards ISO 29022 and ISO/TS 11405 as well as ISO/TS 16506 describe different bonding strength measurements. Unfortunately, all three standards only show methods to determine bond strength to tooth structure. As no product-specific standard exists for the subject device, testing was conducted under ISO/TS 16506 due to the following reason: Shear bond strength is described in ISO 29022 and in ISO/TS 16506. The main difference is the loading of the test specimen after application of the luting material in ISO/TS 16506 (force of 150 N ± 2N, Ch 7.5.1.2), whereas no force is applied in the test method described in ISO 29022. Visalys® Restorative Primer is used on restorations such as crowns, bridges, inlays and onlays. After applying a luting material on the restoration, it is placed in the patient's mouth onto a tooth, {9}------------------------------------------------ Image /page/8/Picture/0 description: The image is a logo for KETTENBACH. The logo is set against a blue square background. The logo consists of seven white circles arranged in a triangular shape, with six circles forming the base and one circle at the apex. Below the circles, the word "KETTENBACH" is written in bold, white letters. Below the word, the phrase "Simply intelligent" is written in black letters. core or into a cavity by applying a force onto the restoration to maintain a small laver thickness of the luting material. Therefore, the shear bond test method described in ISO/TS 16506 is a more realistic measurement for this kind of product (Visalys® Restorative Primer) due to the loading of the test specimen, which simulates the above-mentioned clinical procedure. ISO/TS 11405 gives a general overview of techniques for examination of adhesion. Once aqain, only procedures for testing of the adhesion to tooth structure are included (title of the standard: "Testing of adhesion to tooth structure"). Annex A describes methods for the measurement of bond strength. As explained above, shear bond strength of Visalys® Restorative Primer was already tested according to ISO/TS 16506. An additional bond strength measurement is e.g. micro-tensile bond strength, as described in Annex A, A.2.4. Unfortunately, such measurement cannot be performed with restorative materials because there is no practicable way to cut restorative materials (e.g. metals like titanium) into the desired dimension of 1 mm x 1 mm. Additionally, no force can be applied on the luting material to simulate the clinical procedure (see above). For a detailed description of micro-tensile bond strength and the challenging test procedure to prepare test specimen, see S. Armstrong et al. (1) Armstrong, S. et al. Academy of Dental Materials quidance on in vitro 1. testing of dental composite bonding effectiveness to dentin / enamel using microtensile bond strength (µTBS) approach. Dent. Mater. 33, 133-143 (2016). #### -Clinical Performance Data No data from human clinical studies has been included to support the substantial equivalence of Visalys® Restorative Primer. ### J. Conclusion Regarding Substantial Equivalence: Kettenbach GmbH & Co KG believes that Visalys® Restorative Primer, universal adhesion primer for restoration materials is substantially equivalent to Clearfil Ceramic Primer Plus (K150703). It does not introduce new indications for use, has similar technological characteristics and does not introduce new potential hazards or safety risks.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...