Device Story
PANAVIA V5 is an adhesive resin cement system used by dental professionals for permanent cementation of various dental restorations (crowns, bridges, inlays, onlays, veneers, posts, cores) and amalgam bonding. The system includes a dual-cure (light/self-cure) radiopaque resin cement paste, a Try-in Paste for shade matching, a self-etching Tooth Primer, and CLEARFIL CERAMIC PRIMER PLUS. The cement is delivered via an automix system. The Tooth Primer conditions tooth structures to accelerate polymerization. The clinician uses the Try-in Paste to verify shade/transparency before final cementation. The device provides secure bonding to tooth structure and various restorative materials (ceramics, hybrid ceramics, composite resins, metals). Benefits include reliable retention and esthetic restoration of dental structures.
Clinical Evidence
Bench testing only. No clinical data. Performance evaluated via ISO 4049:2009 (physical/mechanical properties) and ISO/TS 11405:2003 (shear bond strength to dentin, enamel, metal, composite, ceramic, and hybrid ceramic). Biocompatibility testing included cytotoxicity, sub-chronic systemic toxicity, oral mucosa irritation, skin sensitization, and genotoxicity (reverse mutation and chromosomal aberration tests).
Technological Characteristics
Adhesive resin cement system; dual-cure (light/self-cure) resin paste; automix delivery. Materials include resin cement, self-etching primer, and ceramic primer. Complies with ISO 4049:2009 (Class 1 or Class 3 depending on shade). Radiopaque. Biocompatible per ISO 10993/ISO 7405.
Indications for Use
Indicated for cementation of crowns, bridges, inlays, onlays, veneers, adhesion bridges, splints, prosthetic restorations on implant abutments/frames, posts, cores, and for amalgam bonding in dental patients.
Regulatory Classification
Identification
Zinc oxide-eugenol is a device composed of zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp. Dental cement other than zinc oxide-eugenol is a device composed of various materials other than zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized representation of three human profiles facing right, stacked on top of each other. The profiles are black and are set against a white background. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the left side of the seal.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 21, 2015
Kuraray Noritake Dental, Inc. Mr. Michio Takigawa Manager Ote Center Bldg. 7F, 1-1-3Otemachi Chiyoda-ku, Tokyo 100-0004 JAPAN
Re: K150704 Trade/Device Name: Panavia V5 Regulation Number: 21 CFR 872.3275 Regulation Name: Dental Cement Regulatory Class: II Product Code: EMA Dated: July 21, 2015 Received: July 23, 2015
Dear Mr. Takigawa:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina
Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Section 2: Indications for Use
# Indications for Use
510(k) Number (if known): K150704
Device Name: PANAVIA V5
Indications for Use:
[1] Cementation of crowns, bridges, inlays and onlays
- [2] Cementation of veneers
[3] Cementation of adhesion bridges and splints
[4] Cementation of prosthetic restorations on implant abutments and frames
[5] Cementation of posts and cores
[6] Amalgam bonding
| Prescription Use | ✓ |
|-----------------------------|---|
| (Part 21 CFR 801 Subpart D) | |
AND/OR
Over-The-Counter Use N/A (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{4}------------------------------------------------
[PANAVIA V5, Kuraray Noritake Dental Inc.] Section 3: Summary
> August 13, 2015 Date:
## 510(k) Summary
## 3-1. 510(k) owner (submitter)
| 1) | Name | Kuraray Noritake Dental Inc. | |
|----|-----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|
| 2) | Address | 1621 Sakazu, Kurashiki, Okayama 710-0801, Japan | |
| 3) | Contact person | Michio Takigawa<br>Quality Assurance Department | |
| 4) | Contact person in US | Shinichi Sato<br>KURARAY AMERICA, INC.<br>33 Maiden Lane, 6th Floor, New York, NY 10038<br>Tel: (212)-986-2230 (Ext. 115) or (800)-879-1676<br>Fax: (212)-867-3543 | |
| | 3-2. Name of Device | | |
| 1) | Trade / Proprietary name | PANAVIA V5 | |
| 2) | Classification name | Dental cement<br>(21 CFR section 872.3275. Product code: EMA) | |
| 3) | Common name | Adhesive resin cement system | |
| | 3-3. Predicate devices | | |
| 1) | PANAVIA F 2.0 (Primary predicate) | 510(k) Number:<br>Classification:<br>Product Code:<br>21 CFR Section:<br>Applicant: | K032455<br>Dental cement<br>EMA<br>872.3275<br>Kuraray Noritake Dental Inc. |
| 2) | PANAVIA SA Cement Plus Automix | 510(k) Number:<br>Classification:<br>Product Code:<br>21 CFR Section:<br>Applicant: | K142625<br>Dental cement<br>EMA<br>872.3275<br>Kuraray Noritake Dental Inc. |
| 3) | CLEARFIL AP-X | 510(k) Number:<br>Classification:<br>Product Code:<br>21 CFR Section:<br>Applicant: | K012740<br>Tooth shade resin material<br>EBF<br>872.3690<br>Kuraray Noritake Dental Inc. |
| 4) | CLEARFIL ESTHETIC CEMENT EX | 510(k) Number:<br>Classification:<br>Product Code:<br>21 CFR Section:<br>Applicant: | K062410<br>Dental cement<br>EMA<br>872.3275<br>Kuraray Noritake Dental Inc. |
| 5) | ESTENIA C&B | 510(k) Number:<br>Classification:<br>Product Code:<br>21 CFR Section:<br>Applicant: | K042929<br>Tooth shade resin material<br>EBF<br>872.3690<br>Kuraray Noritake Dental Inc. |
{5}------------------------------------------------
6) CLEARFIL SE Protect
510(k) Number: Classification: Product Code: 21 CFR Section: Applicant:
K033938 Resin Tooth Bonding Agent KLE 872.3200 Kuraray Noritake Dental Inc.
## 3-4. Device Description
The subject device is an adhesive resin cement system. It consists of the cement paste (Paste), Try-in Paste, Tooth Primer, CLEARFIL CERAMIC PRIMER PLUS and K-ETCHANT Syringe. The Paste is a dual-cure (light- and/or self-cure), fluoride-releasing, radiopaque resin cement for ceramics (lithium disilicate, zirconia, etc.), hybrid ceramics, composite resins, and metal restorations. It is supplied in an automix delivery system which mixes equal amounts of two components. It is available in the following 5 shades; Universal (A2), Clear, Brown (A4), White, and Opaque. The Opaque shade should be self-cured for final curing due to its strong opacity. The Try-in Paste is a shade matching material which has approximate color and transparency as the hardened mixture of Paste. The Tooth Primer is a self-etching primer to tooth structures that accelerates the polymerization of the Paste.
This is the new registration application for the subject device.
Concerning CLEARFIL CERAMIC PRIMER PLUS and K-ETCHANT Syringe, the specification of those 2 components were omitted for the reasons set forth below.
We already have had 510(k) clearance of K-ETCHANT Syringe (510(k) Number: 133078). And we submit the application for CLEARFIL CERAMIC PRIMER PLUS as separate application at the same time.
## 3-5. Statement of Intended Use
The subject device is indicated for the following uses:
- [1] Cementation of crowns, bridges, inlays and onlays
- [2] Cementation of veneers
- [3] Cementation of adhesion bridges and splints
- [4] Cementation of prosthetic restorations on implant abutments and frames
- [5] Cementation of posts and cores
- [6] Amalgam bonding
### 3-6. Substantial Equivalence Discussion
1) Intended uses
The INDICATIONS of the subject device and predicate devices, PANAVIA F2.0 which is dental resin cement and PANAVIA SA Cement Plus Automix which is self-adhesive resin cement, are as listed on the following table.
{6}------------------------------------------------
| | Trade name | Intended use |
|----------------------|-------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Subject<br>device | PANAVIA V5 | [1] Cementation of crowns, bridges, inlays and onlays<br>[2] Cementation of veneers<br>[3] Cementation of adhesion bridges and splints<br>[4] Cementation of prosthetic restorations on implant abutments and<br>frames<br>[5] Cementation of posts and cores<br>[6] Amalgam bonding |
| Predicate<br>devices | PANAVIA F2.0<br>(Primary predicate) | [1] Cementation of metal crowns and bridges, inlays and onlays<br>[2] Cementation of porcelain crowns, inlays, onlays and veneers<br>[3] Cementation of composite resin crowns, inlays, and onlays<br>[4] Cementation of adhesion bridges<br>[5] Cementation of endodontic cores and prefabricated posts<br>[6] Amalgam bonding |
| | PANAVIA SA<br>Cement Plus Automix | [1] Cementation of crowns, bridges, inlays and onlays<br>[2] Cementation of prosthetic restorations on implant abutments and<br>frames<br>[3] Cementation of adhesion bridges and splints<br>[4] Cementation of posts and cores<br>[5] Amalgam bonding |
The intended use of the subject device was written up based on those of the predicate devices. Therefore, the intended use of the subject device is substantially equivalent to those of the predicate devices
2) Chemical ingredients
Except for 5 chemical ingredients, all ingredients in the Paste and Tooth Primer are identical and have been used in the following predicate devices: K032455 Panivia 2.0, K142625 Panavia SACement, K012740 Clearfil AP-X, K062410 Clearfil Esthetic Cement and DC Bond, K033938 Clearfil Protect Bond and K042929 Estenia C&B.
Regarding these predicate devices, there have not been any reported problems or recalls according to the post market adverse event reporting requirements in the US.
5 chemical ingredients in the Paste and Tooth Primer are new ingredients.
Therefore, we evaluated the Paste and Tooth Primer referring to ISO 10993 series and ISO 7405. As a result of the tests, it was concluded that the Paste and Tooth Primer is substantially equivalent in biocompatibility to the predicate devices.
All the chemical ingredients of the Try-in Paste are identical to those of the predicate devices. Regarding the predicate devices, there have not been any reported problems or recalls according to the post-market adverse event reporting requirements in the US. Accordingly, it was considered that the Try-in Paste was substantially equivalent to the predicate devices.
- 3) Technological characteristics/ Effectiveness and Performance
Physical and mechanical properties of the subject device were evaluated according to ISO 4049: 2009 (Dentistry - Polymer-based restorative and materials). According to ISO 4049: 2009, the subject device is classified into the following:
Shade type: Universal (A2), Clear, Brown (A4), White
Class 3: materials that are cured by the application of external energy and also have a self-curing mechanism present
Shade type: Opaque
Class 1: materials whose setting is effected by mixing an initiator and activator.
{7}------------------------------------------------
| The results of comparative study performed according to ISO 4049: 2009 were indicated below. | | | | | | | |
|----------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------|------------------------------------------|----------------------------------|-----------------------------------|----------------------------------|----------------------------------|----------|
| Section | Requirement | PANAVIA V5 (Subject device) | | | PANAVIA F2.0 (Primary predicate) | | |
| | | Shade type:<br>Universal (A2)<br>Class 3 | Shade type :<br>White<br>Class 3 | Shade type :<br>Opaque<br>Class 1 | Shade type:<br>Brown<br>Class 3 | Shade type:<br>Opaque<br>Class 1 | |
| 5.2.2 Film<br>thickness, luting<br>materials | <50μm | COMPLIES | COMPLIES | COMPLIES | COMPLIES | COMPLIES | |
| 5.2.4 Working time,<br>Class 1 and Class<br>3 luting materials | >60 seconds | COMPLIES | COMPLIES | COMPLIES | COMPLIES | COMPLIES | |
| 5.2.5 Setting time,<br>Class 1 materials<br>or<br>5.2.6 Setting time,<br>Class 3 materials | <10 min | COMPLIES | COMPLIES | COMPLIES | COMPLIES | COMPLIES | |
| 5.2.9 Flexural<br>strength | ≥50 MPa | COMPLIES | COMPLIES | COMPLIES | COMPLIES | COMPLIES | |
| 5.2.10 | Water<br>sorption | ≤40<br>µg/mm³ | COMPLIES | COMPLIES | COMPLIES | COMPLIES | COMPLIES |
| | Solubility | ≤7.5<br>µg/mm³ | COMPLIES | COMPLIES | COMPLIES | COMPLIES | COMPLIES |
| 5.4 Color stability<br>after irradiation<br>and water sorption | No more<br>than a slight<br>change in<br>color shall be<br>observed | COMPLIES | COMPLIES | COMPLIES | COMPLIES | COMPLIES | |
| 5.5 Radio-opacity | Equal to or<br>greater than<br>that of the<br>same<br>thickness of<br>aluminum | COMPLIES | COMPLIES | COMPLIES | COMPLIES | COMPLIES | |
The results of comparative study nerformed according to ISO 4049; 2009 were indicated below
"COMPLIES" means that test values were in the acceptable range to pass the testing in compliance with the requirements of ISO 4049: 2009.
The results indicate that the subject device and the predicate device comply with the requirements of ISO 4049: 2009. From the above, it can be said that comparative study of the subject device is substantially equivalent to that of the predicate device.
| The shear bond strengths for the adherent surfaces which are matched to each indication in accordance | | |
|-------------------------------------------------------------------------------------------------------|--|--|
| with ISO/TS 11405: 2003 were indicated below. | | |
| Adherent surface<br>(Material composition) | PRIMER | Criteria | Subject device<br>(Universal (A2)) | Primary predicate<br>(PANAVIA F 2.0) |
|--------------------------------------------|-------------------------------|-------------------|------------------------------------|--------------------------------------|
| Dentin | Tooth Primer | In-house standard | COMPLIES | COMPLIES |
| Enamel | | | COMPLIES | COMPLIES |
| Metal | | | COMPLIES | COMPLIES |
| Composite resin | | | COMPLIES | COMPLIES |
| Ceramic | CLEARFIL<br>CERAMIC<br>PRIMER | | COMPLIES | COMPLIES |
| Hybrid ceramics | | | COMPLIES | COMPLIES |
| Metal | | | COMPLIES | COMPLIES |
| Composite resin | PLUS | | COMPLIES | COMPLIES |
| Dentin<br>Amalgam | Tooth Primer<br>N/A | | COMPLIES | COMPLIES |
"COMPLIES" means that test values were in the acceptable range to pass the testing in compliance with the requirements of in-house standard.
It was concluded that the bonding performance to all substrates of the subject device was equivalent to those of the predicate device.
{8}------------------------------------------------
Released fluorine ion test was performed to validate the substantial equivalence of the subject device with the predicate device in terms of performance for the intended uses.
It was confirmed that the released level from the cured one by the subject device was lower than that by PANAVIA F2.0.
### 3-7. Biocompatibility
Paste and Tooth Primer are categorized into the external communicating device (tissue/ bone/ dentin) and permanent contact device.
Except for 5 chemical ingredients, all ingredients in the Paste and Tooth Primer have been used in the predicate devices.
Regarding these predicate devices, there have not been any reported problems or recalls according to the nost market adverse event reporting requirements in the US.
5 chemical ingredients in the Paste and Tooth Primer are new ingredients.
Therefore, we evaluated the Paste and Tooth Primer referring to ISO 10993 series and ISO 7405. We decided to perform the following tests.
- Cytotoxicity test a)
- b) Sub-chronic systemic toxicity test
- Skin irritation and intra-cutaneous reactivity test (oral mucosa irritation test) c)
- d) Sensitization test (skin sensitization test)
- Genotoxicity test (reverse mutation test, chromosomal aberration test) e)
As a result of the tests, it was concluded that the Paste and Tooth Primer are substantially equivalent in biocompatibility to the predicate devices.
Try-in Paste is categorized into the external communicating device that may contact dentin and whose duration of contact is less than 24 hours.
All the chemical ingredients of the Try-in Paste are identical to those of the predicate devices.
Regarding the predicate devices, there have not been any reported problems or recalls according to the post-market adverse event reporting requirements in the US.
Accordingly, it was considered that the Try-in Paste was substantially equivalent to the predicate devices.
#### 3-8. Conclusion
The comparison for intended uses, chemical ingredients and performance data shows that the subject device is substantially equivalent to the predicate devices.
This submission information including the nonclinical testing provided supports that the subject device is as safe and as effective as the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.