Accusil Dental Impression Materials (Accusil Light Body Regular and Fast Set, Accusil Heavy Body Regular and Fast Set, Accusil Monophase Regular and Fast Set, Accusil Putty Regular and Fast Set, Accusil Bite Registration)
Accusil Dental Impression Materials (Accusil Light Body Regular and Fast Set, Accusil Heavy Body Regular and Fast Set, Accusil Monophase Regular and Fast Set, Accusil Putty Regular and Fast Set, Accusil Bite Registration)
Applicant
Prevest Denpro Limited
Product Code
ELW · Dental
Decision Date
Sep 9, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3660
Device Class
Class 2
Indications for Use
Accusil Dental Impression Material is intended to be placed on a preformed impression tray and used to reproduce the structure of a patient's teeth and gums.
Device Story
Accusil Dental Impression Materials are two-component (base and catalyst) silicone-based impression materials. Components are mixed in a 1:1 ratio, loaded into a preformed impression tray, and inserted into the patient's mouth by a dental professional. The material conforms to the patient's dentition and sets to create a physical reproduction of teeth and occlusion. Used in clinical dental settings to facilitate the fabrication of crowns, bridges, and restorative devices. The resulting impression provides the dentist with an accurate model of the patient's oral anatomy, aiding in the design and fit of dental prosthetics.
Clinical Evidence
Bench testing only. Performance evaluated against ISO 4823 and ISO 48-4 standards. Metrics included working time, detail reproduction, linear dimensional change, consistency, Shore hardness, compatibility with gypsum, elastic recovery, and strain in compression. All test results met established criteria.
Technological Characteristics
Room temperature vulcanizing (RTV) two-component silicone impression material. Composition: vinyl-terminated polydimethylsiloxanes, fillers, platinum catalyst, SIH capped polysiloxane. Form factors: 50ml double cartridges, 380ml 5:1 cartridges, or 150ml/300ml plastic jars. Non-sterile. Testing per ISO 4823 and ISO 48-4.
Indications for Use
Indicated for use in reproducing the structure of a patient's teeth and gums for dental applications including crowns, bridges, occlusal models, and dental implant restorative devices.
Regulatory Classification
Identification
Impression material is a device composed of materials such as alginate or polysulfide intended to be placed on a preformed impression tray and used to reproduce the structure of a patient's teeth and gums. The device is intended to provide models for study and for production of restorative prosthetic devices, such as gold inlays and dentures.
K143382 — DentiAnn Smart Sil · Seilglobal Co., Ltd. · Feb 23, 2015
K201483 — Impression Material · Zhengzhou Huaer Electro Optics Technology Co., Ltd. · Oct 8, 2020
K241924 — Elastic Impression Material · Beijing Okvd Biological Technology , Ltd. · Dec 27, 2024
K222741 — DENU Light Body(Regular, Fast), DENU Heavy Body(Regular, Fast), DENU Putty Set(Regular, Fast) · Hdi, Inc. · Nov 7, 2022
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of the square.
September 9, 2022
Prevest Denpro Limited % Angela Blackwell Senior Consultant Blackwell Device Consulting P.O. Box 718 Gresham, Oregon 97030-0172
Re: K213244
Trade/Device Name: Accusil Dental Impression Materials (Accusil Light Body Regular and Fast Set, Accusil Heavy Body Regular and Fast Set, Accusil Monophase Regular and Fast Set, Accusil Putty Regular and Fast Set, Accusil Bite Registration) Regulation Number: 21 CFR 872.3660 Regulation Name: Impression Material Regulatory Class: Class II Product Code: ELW Dated: September 29, 2021 Received: September 30, 2021
Dear Angela Blackwell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
## 510(k) Number (if known)
#### K213244
#### Device Name
Accusil Dental Impression Material: Accusil Light Body Regular and Fast Set, Accusil Heavy Body Regular and Fast Set, Accusil Monophase Regular and Fast Set, Accusil Putty Regular and Fast Set, Accusil Bite Registration
### Indications for Use (Describe)
Accusil Dental Impression Material is intended to be placed on a preformed impression tray and used to reproduce the structure of a patient's teeth and gums.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## K213244
# Accusil Dental Impression Materials (Accusil Light Body Regular and Fast Set, Accusil Heavy Body Regular and Fast Set, Accusil Monophase Regular and Fast Set, Accusil Putty Regular and Fast Set, Accusil Bite Registration) 510K Summary April 29, 2022
Name and Address:PrevestDenpro Limited Export Promotion Industrial Park Bari Brahmana, Jammu 181133 India Contact Person: Atul Modi Email:prevestindia@gmail.com Telephone: (941) 919 4280
Name of device: Accusil Dental Impression Materials (Accusil Light Body Regular and Fast Set, Accusil Heavy Body Regular and Fast Set, Accusil Monophase Regular and Fast Set, Accusil Putty Regular and Fast Set, Accusil Bite Registration)
Common name: impression material Classification Name: impression material CFR: 21 CFR 872.3660 Primary Product Code:ELW
# Submission Contact:
Angela Blackwell Blackwell Device Consulting P.O. Box 718 Gresham, OR 97030-0172 (704)450-9934 angela@blackwelldevice.com
# Device Description:
Accusil Dental Impression Materials are designed for dental applications to define and reproduce the structure of a patient's teeth and gums for producing crowns, bridges, occlusal models and dental implant restorative devices. Base and catalyst components are mixed in an equal ratio 1:1, placed into an impression tray and inserted into the patient's mouth. The material will conform to the patient's dentition and when set will produce a reproduction of the patient's teeth and occlusion.
{4}------------------------------------------------
# Indications for Use:
Accusil Dental Impression Material is intended to be placed on a preformed impression tray and used to reproduce the structure of a patient's teeth and gums.
# Testing Summary:
Accusil light, heavy, monophase and putty were tested with protocols meeting ISO 4823 and ISO 48-4 for working time, detail reproduction, linear dimensional change, consistency, Shore hardness, compatibility with gypsum, elastic recovery, and strain in compression. Setting time was tested using internal protocols (ISO 4823 does not have setting time). Accusil bite registration was tested with protocols meeting ISO 4823 and ISO 48-4 for working time, linear dimensional change, Shore hardness and compression and tested using an internal protocol for setting time.
Shelf life testing was completed on all devices using relevant protocols also used for the physical characteristics tested. Shelf life for impression materials is 3 years.
All tests results met the criteria in standards.
The impression materials contain ingredients found in the predicate and reference devices. They are used for the same type of contact, external communicating with contact to tissue/bone/dentin. The indications for use of the devices and the predicate devices are the risks incurred for use of the ingredients are the same. Since no new risks are introduced, no biocompatibility testing was done.
Predicate Device: Dynax Dental Impression Material K171562
Reference Devices: Fresh Impression Material K053427, President the Original, K220097, Aquasil Ultra + K152861
# Substantial Equivalence:
The impression materials have similar ingredients to the predicate and reference devices, the same indications for use, and similar physical parameter testing.
Accusil Dental Impression Materials
| Name | Accusil (light<br>body, heavy<br>body,<br>monophase,<br>putty and bite<br>registration) | Dynax (clear,<br>putty, heavy<br>body, light,<br>mono)<br>predicate<br>device | Fresh<br>Impression<br>Material<br>(reference<br>device for<br>ingredients) | President the<br>Original<br>(reference<br>device for<br>ingredients) | Aquasil Ultra +<br>(reference<br>device for<br>ingredients) |
|---------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510k Number | K213244 | K171562 | K053427 | K220097 | K152861 |
| Common Name | Impression<br>material | Impression<br>material | Impression<br>material | Impression<br>material | Impression<br>material |
| Classification<br>Name | Impression<br>material | Impression<br>material | Impression<br>material | Impression<br>material | Impression<br>material |
| Class | II | II | II | II | II |
| Product Code | ELW | ELW | ELW | ELW | ELW |
| CFR | 872.3660 | 872.3660 | 872.3660 | 872.3660 | 872.3660 |
| Indications for Use | Accusil dental<br>impression<br>material is<br>intended to be<br>placed on a<br>preformed<br>impression tray<br>and used to<br>reproduce the<br>structure of a<br>patient's teeth<br>and gums. | Dynax dental<br>impression<br>material is<br>intended to<br>be placed on a<br>preformed<br>impression<br>tray and used<br>to reproduce<br>the structure<br>of a patient's<br>teeth and<br>gums. | Fresh Dental<br>Impression<br>material is<br>intended for<br>use with all<br>crowns,<br>bridges,<br>occlusal and<br>dental implant<br>impression<br>techniques to<br>reproduce the<br>structure of a<br>patient's<br>teeth and<br>gums. | PRESIDENT<br>The Original<br>Xtra light body<br>/ light body /<br>regular body:<br>* Correction<br>material for<br>the corrective<br>impression<br>technique<br>* Injection<br>material for<br>the double<br>mix technique<br>* Injection<br>material for<br>the dual arch<br>technique))<br>* Lining<br>impression<br>material<br>PRESIDENT<br>The Original<br>heavy body:<br>* Impression<br>material for<br>pick-up<br>impression in<br>the double<br>mix technique<br>* Tray<br>material for<br>corrective<br>impression<br>technique<br>* Tray<br>material for<br>dual arch<br>technique<br>PRESIDENT<br>The Original<br>System 360:<br>* Impression<br>material for<br>pick-up | Aquasil® Ultra<br>+ Smart<br>Wetting®<br>Impression<br>Material is<br>indicated for<br>all dental<br>impression<br>techniques. |
| Material | | | | | |
| | | | | the double<br>mix technique | |
| | | | | * Tray<br>material for<br>corrective<br>impression<br>technique | |
| | | | | * Tray<br>material for<br>dual arch<br>technique | |
| | | | | PRESIDENT<br>The Original<br>putty / putty<br>soft / putty<br>super soft /<br>fast putty soft | |
| | | | | * Impression<br>material for<br>pick-up<br>impression in<br>the double<br>mix technique | |
| | | | | * Tray<br>material for<br>corrective<br>impression<br>technique | |
| Material | Accusil light,<br>heavy, mono,<br>and bite<br>registration:<br>Mixture of vinyl<br>terminated<br>polydimethylsilox<br>anes and filler<br>materials with<br>platinum catalyst<br>and SIH capped<br>polysiloxane<br><br>Accusil putty:<br>Mixture of vinyl<br>terminated<br>polydimethylsilox<br>anes and filler<br>materials with | Dynax putty,<br>light, mono,<br>and heavy<br>body: Mixture<br>of vinyl<br>terminated<br>polydimethylsi<br>loxanes and<br>filler materials<br>with platinum<br>catalyst and<br>methylhydrosi<br>loxane<br>dimethylsiloxa<br>ne copolymers<br><br>Dynax clear:<br>Mixture of<br>vinyl<br>terminated | Fresh Bold<br>liquid putty,<br>light body,<br>monophase<br>and heavy<br>body: Mixture<br>of vinyl<br>terminated<br>polydimethylsi<br>loxanes and<br>filler materials<br>with platinum<br>catalyst and<br>methylhydrosi<br>loxane<br>dimethylsiloxa<br>ne copolymers<br><br>Fresh Bold<br>clear bite: | Divinylpolydime<br>thylsiloxanes,<br>silicon dioxide,<br>zeolite,<br>Platinum(0)-<br>1,3-divinyl-<br>1,1,3,3-<br>tetramethyldisil<br>oxane complex<br>solution,<br>Titanium<br>dioxide, iron<br>oxides, organic<br>pigments, and<br>surfactants | Polydimethyl<br>siloxane polymer,<br>polymethylhydrog<br>en siloxane,<br>silicon dioxide,<br>sodium<br>aluminosiliicate<br>(zeolite), organic<br>platinum<br>complex, titanium<br>dioxide, iron<br>oxide pigments,<br>organic pigments,<br>other pigments.<br>Surfactants,<br>peppermint oil |
| | platinum catalyst<br>and SIH capped<br>polysiloxane plus<br>softener | polydi-<br>methylsiloxan<br>es and silicic<br>acid with<br>platinum<br>catalyst and<br>methylhydrosi<br>loxane<br>dimethylsiloxa<br>ne copolymers | Mixture of<br>vinyl<br>terminated<br>polydimethylsi<br>loxanes and<br>silic acid with<br>platinum<br>catalyst and<br>methylhydrosi<br>loxane<br>dimethylsiloxa<br>ne copolymers | | |
| Working/Processi<br>ng Time | 40-150 sec | 90 sec | 90-120 sec | unknown | 35-105 sec |
| Setting time/Time<br>in the mouth | 60-300 sec | 90 sec | 120-140 sec | unknown | 150-330 sec |
| Hardness | 63-70 Shore A | 46-70 Shore A | 42-70 Shore A | 50-75 Shore A | Unknown |
| Working humidity | 50% | 50% | 50% | unknown | Unknown |
| Dimensional<br>accuracy | 99.9%-99.2% | 99.9%-99.2% | 99.9% - 99.2% | 99.8%-99.2% | 98% |
| Stability (linear<br>dimensional<br>change) | <0.8% typical | <0.2% typical | <0.1% typical | unknown | <0.50% typical |
| Consistency | Type 0-type 3<br>ISO 4823 | Type 0-type 3<br>ISO 4823 | Type 0-type 3<br>ISO 4823 | Type 1-type 3<br>ISO 4823 | Type 1-type 3<br>ISO 4823 |
| Chemical<br>Description | Room<br>temperature<br>vulcanizing 2-<br>components<br>silicone | Room<br>temperature<br>vulcanizing 2-<br>components<br>silicone | Room<br>temperature<br>vulcanizing 2-<br>components<br>silicone | Room<br>temperature<br>vulcanizing 2-<br>components<br>silicone | Room<br>temperature<br>vulcanizing 2-<br>components<br>silicone |
| Package | Accusil bite<br>registration,<br>heavy, light, and<br>monophase<br>come in 2 x 50 ml<br>double cartridges<br>Monophase<br>regular also<br>comes in a 5:1<br>380ml double<br>cartridges.<br>Accusil putty<br>comes in plastic<br>jars of 150ml and<br>300ml for both<br>base and<br>catalyst. | Dynax putty in<br>tubs of 2x35<br>ml and 2 x 450<br>ml<br>Dynax light,<br>mono, heavy<br>body, and<br>clear are sold<br>in 50 ml<br>double<br>cartridges | Fresh Bold<br>liquid putty in<br>tubs of 2 x<br>450 ml<br>Fresh Bold<br>light body,<br>monophase,<br>heavy body<br>and clear bite<br>in 2x 50 ml<br>double<br>cartridges | Primary<br>packaging<br>Tubes, Pots or<br>cartridges<br>Secondary<br>packaging:<br>Folding carton | Primary packaging<br>Tubes, Pots or<br>cartridges<br>Secondary<br>packaging:<br>Folding carton |
{5}------------------------------------------------
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
| Method of<br>manipulation | Preformed<br>impression tray | Preformed<br>impression<br>tray | Preformed<br>impression<br>tray | Preformed<br>impression<br>tray | Preformed<br>impression tray |
|---------------------------|------------------------------|---------------------------------|---------------------------------|---------------------------------|------------------------------|
| Sterility | Non-sterile | Non-sterile | Non-sterile | Non-sterile | Non-sterile |
Conclusion: Accusil dental impression materials are substantially equivalent to the predicate device, Dynax dental impression materials. They have the same indications, similar testing, and very similar ingredients. References devices contain ingredients not found in the predicate device but found in Accusil. Setting time, working time, hardness, working humidity, are similar to the predicate device and/or reference devices. The testing for these parameters for the subject device, predicate device, and reference devices are according to ISO 4823 so there are no differences in test methods. Since the test results are in the same ranges and the test methods are according to the same standard these results support the substantial equivalence of Accusil to Dynax. The subject devices, predicate devices, and reference devices all have the same method of manipulation, same working humidity, and same ISO 4823 types. Both the subject devices and the predicate device have physical parameters which meet requirements of the relevant ISO standards. Any differences in ingredients are minor and do not change the substantial equivalence.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.