Dynax® Dental Impression Material is intended to be placed on a preformed impression tray and used to reproduce the structure of a patient's teeth and gums.
Device Story
Dynax Dental Impression Material is a two-component vinylpolysiloxane silicone system; base and catalyst mixed in 1:1 ratio; placed in preformed impression tray; inserted into patient's mouth; conforms to dentition; vulcanizes to create accurate reproduction of teeth and occlusion. Used by dental professionals in clinical settings for restorative dentistry (crowns, bridges, implants). Output is a physical dental impression used by clinicians to guide restorative treatment planning and fabrication of dental prosthetics. Benefits include high dimensional accuracy and stability for precise dental restorations.
Clinical Evidence
Bench testing only. Performance verified through in-house testing against product specifications, including dimensional accuracy (99.2-99.9%), Shore A hardness (46-70), and setting times. Biocompatibility assessed per ISO 10993-1.
Technological Characteristics
Two-component room temperature vulcanizing (RTV) silicone; vinyl terminated polydimethylsiloxanes, filler materials, platinum catalyst, methylhydrosiloxane dimethylsiloxane copolymers. Form factors: putty (tubs) and light/heavy/mono/clear (cartridges). Standards: ISO 4823, ISO 10993-1, ISO 7405, ISO 14971, EN ISO 13485.
Indications for Use
Indicated for use by dental professionals to reproduce the structure of a patient's teeth and gums for the production of crowns, bridges, occlusal, and dental implant impressions.
Regulatory Classification
Identification
Impression material is a device composed of materials such as alginate or polysulfide intended to be placed on a preformed impression tray and used to reproduce the structure of a patient's teeth and gums. The device is intended to provide models for study and for production of restorative prosthetic devices, such as gold inlays and dentures.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Dreve Dentamid GmbH Rilchard Keen Consultant 1151 Hope Street Stamford, Connecticut 06907-1659 April 13, 2018
Re: K171562
Trade/Device Name: Dynax clear, Dynax putty, Dynax heavy body, Dynax light, Dynax mono Regulation Number: 21 CFR 872.3660 Regulation Name: Impression Material Regulatory Class: Class II Product Code: ELW Dated: February 22, 2018 Received: March 14, 2018
Dear Richard Keen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Mary S. Runner -S
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows a circular logo with the word "Dreve" in white letters. The word is centered within the circle. The background of the circle appears to be a dark color, possibly black or a dark shade of gray. The logo has a simple and clean design.
# Indications for Use
Dreve Dentamid
510(k) Number (if known): K171562
Device Name:
Dynax Dental Impression Material:
- . Dynax clear
- Dynax putty
- Dynax heavy body
- Dynax light
- Dynax mono
#### Indications for Use:
Dynax Dental Impression Material is intended to be placed on a preformed impression tray and used to reproduce the structure of a patient's teeth and gums.
| Prescription Use<br>(21 CFR Part 801 Subpart D) | <svg height="16" width="16"> <rect fill="white" height="16" stroke="black" stroke-width="1" width="16" x="0" y="0"></rect> <line stroke="black" stroke-width="2" x1="2" x2="14" y1="2" y2="14"></line> <line stroke="black" stroke-width="2" x1="14" x2="2" y1="2" y2="14"></line> </svg> |
|-------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
and/or
(Please do not write below this line - continue on another page if needed)
#### Concurrence of CDRH, Office of Device Evaluation (ODE)
Image /page/2/Picture/17 description: The image is a circular logo with the text "TUV NORD" at the top. Below that is the text "TUV Rheinland Group". The bottom of the logo contains the text "ISO 9001 < 119 15423". The logo is black and white.
(Division Sign-Off)
Image /page/2/Picture/19 description: The image shows a logo for VDDI Dental Solutions German Manufacture. The logo features the letters VDDI in bold, followed by the words "Dental Solutions" and "German Manufacture" in smaller fonts. To the left of the text is a graphic element that appears to be a stylized representation of a dental-related object. The overall design is simple and professional, conveying a sense of quality and reliability.
Dreve Dentamid GmbH Max-Planck-Straße 31 59423 Unna/Germany Tel .: +49 2303 8807 - 0 Fax: +49 2303 8807-55 E-Mail: dentamid@dreve.de
Internet: www.dreve.com/dentamid Lieferadresse: Einsteinstr. 36 Geschäftsführer: Dr. Volker Dreve, Dino Marchetti Sitz: 59423 Unna Reg .- Gericht: Hamm HRB 3712
Sparkasse Unna Kto .- Nr. 43 000 BLZ 443 500 60 BIC: WELADED1UNN IBAN: DE 26 44350060 0000043000
Commerzbank AG Dortmund Kto .- Nr. 03 222 354 00 BLZ. 440 800 50 BIC: DRESDEFF440 IBAN: DE 35 44080050 0322235400
Over-The-Counter Use
(21 CFR Part 801 Subpart C)
Page 1 of 1
Postbank Dortmund Kto .- Nr. 6 493 469 BLZ 440 100 46 BIC: PBNKDEFF IBAN: DE 42 44010046 0006493469
{3}------------------------------------------------
K171562
# 510(k) Summary
Dreve Dentamid
| Sponsor: | Dreve Dentamid GmbH<br>Dr. Reiner Altmann<br>(Head of Quality Management &<br>Regulatory Affairs)<br>Max-Planck-Str. 31<br>59423 Unna/Germany<br>Phone: +49 2303 8807 14<br>Email: reiner.altmann@dreve.de |
|--------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Consultant: | Mr. Richard Keen<br>Compliance Consultants<br>1151 Hope Street<br>Stamford, CT 06907-1659<br>Phone: 203 329 2700 F 203 329 2345<br>Email: rkeen@fda-complianceconsultants.com |
| Date: | April 12th, 2018 |
| 510(k) number: | K171562 |
| Proprietary Name: | Dynax® clear, models D6121 & D6122<br>Dynax® putty, model D6101<br>Dynax® heavy body, models D6112 & 6113<br>Dynax® light, models D6103 & 6104<br>Dynax® mono, models D6106 & 6107 |
| Common Name: | Dental Impression Material |
| Device Classification Name: | Impression material |
| Classification Number: | 21 CFR 872.3660 |
| Product Code: | ELW |
| Reviewing Group: | Dental |
| Device Classification: | Class 2 per 21 CFR 872.3660 |
| Establishment registration No. | 1000486347 |
| Predicate Device: | K053427 Fresh Bold, Impression material |
Trademark Notice: All Trademarks used other than those of Dreve DentamidTrademark Notice: All Trademarks used other than those of Dreve Dentamid GmbH are registered to their respective owners.
Image /page/3/Picture/5 description: The image is a circular logo with the text "TUV NORD" at the top. Below that is the text "TUV NORD CERT GmbH". The outer ring of the logo contains the text "ISO 9001 / ISO 13485".
Image /page/3/Picture/6 description: The image contains a logo with the text "VDDI" in bold letters. Below the text, there are smaller words that say "Dental Solutions German Manufactur". To the left of the text, there is a partial image of a circular design.
Image /page/3/Picture/7 description: The image shows the letters B, V, and D in a bold, sans-serif font. Below the letters, there is some smaller text that reads "Koop erative Mitglied im Bundesve stand Dentalhandel e.V." The text is slightly blurry, but the letters BVD are clear and prominent.
Dreve Dentamid GmbH Max-Planck-Straße 31 59423 Unna/Germany Tel.: +49 2303 8807-0 Fax: +49 2303 8807-55 E-Mail: dentamid@dreve.de Internet: www.dreve.com/dentamid Lieferadresse: Einsteinstr. 36 Geschäftsführer: Dr. Volker Dreve, Dino Marchetti Sitz: 59423 Unna Reg.-Gericht: Hamm HRB 3712
Sparkasse Unna Kto.-Nr. 43 000 BLZ 443 500 60 BIC: WELADED1UNN IBAN: DE 26 44350060 000043000
Commerzbank AG Dortmund Kto.-Nr. 03 222 354 00 BLZ 440 800 50 BIC: DRESDEFF440 IBAN: DE 35 44080050 0322235400
Page 1 of 5
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows a logo for Dreve Dentamid. The logo features a red circle with the word "Dreve" in white letters inside the circle. Below the circle, the words "Dreve Dentamid" are written in a larger, bold font.
# Device Description
Dynax® Dental Impression Materials are designed for dental applications to define and reproduce the structure of a patient's teeth and gums for producing crowns, bridges, occlusal and dental implant. Base and catalyst components are mixed in equal ratio 1:1, placed into an impression tray and inserted into the patient's mouth. The material will conform to the patient's dentition and when set will produce a reproduction of the patient's teeth and occlusion.
# Intended Use
Dental Impression Material on the basis of vinylpolysiloxane
## Indications for Use:
Dynax® Dental Impression Material is intended to be placed on a preformed impression tray and used to reproduce the structure of a patient's teeth and gums.
# Technological Characteristics
#### Design
The Dynax® Dental Impression Materials are similar in design to the predicate listed above. Same as the predicate the Dynax® Dental Impression Materials are intended to reproduce the structure of a patient's teeth and gums to make crowns, bridges, occlusal and dental implant impressions. They use similar technological characteristics and principles. Both materials are two component silicones being mixed in 1:1 ratio to start the vulcanizing process.
#### Material
As the predicate material the Dynax® Dental Impression Materials are based on mixtures of vinyl terminated polydimethylsiloxanes and filler materials with platinum catalyst and methylhydrosiloxane dimethylsiloxane copolymers. An assessment of the biocompatibility according to FDA Recognized Consensus Standard ISO 10993-1 is included in this application. As a result of this assessment/testing we conclude that the device is safe for its intended use and does not show any unacceptable risks for users, patients and third parties.
This is a prescription only material. The labeling and working instructions are designed for health care professionals.
#### Testing
The Dynax® Dental Impression Materials have been designed, developed, tested and produced according to ISO 13485; CAN/CSA ISO 13485 and European Medical Device Directive 93/42/EEC. The quality system is certified by a Notified Body.
Image /page/4/Picture/15 description: The image shows a logo with the text "VDDI Dental Solutions German Manufactu". The logo also includes a circular graphic element to the left of the text. The graphic element appears to be an abstract representation of dental equipment or a related symbol. The text is stacked vertically, with "VDDI" at the top in a larger font size, followed by "Dental Solutions" and "German Manufactu" in smaller fonts.
Image /page/4/Picture/16 description: The image shows the letters B, V, and D in bold font. The letters are vertically aligned and have a curved line above them. Below the letters is the text "Kooperatives Mitglied im Bundesverband des deutschen K".
Dreve Dentamid GmhH Max-Planck-Straße 31 59423 Unna/Germany Tel .: +49 2303 8807-0 Fax: +49 2303 8807-55 E-Mail: dentamid@dreve.de
Internet: www.dreve.com/dentamid Lieferadresse: Einsteinstr. 36 Geschäftsführer: Dr. Volker Dreve, Dino Marchetti Sitz: 59423 Unna Reg.-Gericht: Hamm HRB 3712
Sparkasse Unna Kto .- Nr. 43 000 BLZ 443 500 60 BIC: WELADED1UNN IBAN: DE 26 44350060 0000043000
Commerzbank AG Dortmund Kto.-Nr. 03 222 354 00 BLZ 440 800 50 BIC: DRESDEFF440 IBAN: DE 35 44080050 0322235400
Page 2 of 5
Postbank Dortmund Kto.-Nr. 6 493 469 BLZ 440 100 46 BIC: PBNKDEFF IBAN: DE 42 44010046 0006493469
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a red circle with the word "Dreve" written in white letters in the center. The circle has a thin white border around it. The background is white and slightly blurry.
Testing has confirmed that these devices meet their product specification. A series of in-house tests have been conducted to verify the intended signals are accurate and can maintain performance over its useful life.
# Substantial Equivalence
Information provided in this application shows that the product is substantially equivalent to the predicate device in intended use, performance, materials and application. The differences between these devices are incidental and not significant and do not raise new questions of safety or effectiveness.
# Safety and Effectiveness
There are no substantial differences between the Dynax® (clear, putty, heavy body, light, mono) and the predicate device: the K053427 Fresh Bold, Impression material. In fact, the Dynax® (clear, putty, heavy body, light, mono) and the predicate device: the K053427 Fresh Bold, Impression material are virtually identical with minor exceptions.
This device is safe and effective for the application for which it is intended and has been tested to confirm safety and efficacy. The Dynax® (clear, putty, heavy body, light, mono) has benefited from design, development, testing and production procedures that conform to our in-house quality systems.
# Harmonized Test
This device is the same (within the definition of design controls) as the device cleared in K053427 - with the addition of qualifying this device to the indicated harmonized standards. This device may be used interchangeably with the predicate.
| Comparable Parameter | Dynax® (clear, putty,<br>heavy body, light, mono) | K053427<br>Fresh<br>Bold.<br>Impression material |
|------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Dental Impression Material<br>on the basis of<br>vinylpolysiloxane | The intended use of this<br>material is for a trained dental<br>to reproduce the structure of<br>patient's teeth and gums to<br>make crowns, bridges,<br>occlusal and dental implant<br>impressions. |
| Indications for Use | Dynax® Dental Impression<br>Material is intended to be<br>placed on a preformed<br>impression tray and used to<br>reproduce the structure of a<br>patient's teeth and gums. | Fresh, Dental Impression<br>material is intended for use<br>with all crowns, bridges,<br>occlusal and dental implant<br>impression techniques to<br>reproduce the structure of a<br>patient's teeth and gums. |
| Working / Processing time | 90 sec | 90 - 120 sec |
| Setting time / Time in the<br>mouth | 90 sec | 120 - 140 sec |
| Hardness | 46 - 70 Shore A | 42 - 70 Shore A |
| Working humidity | 50 % | 50 % |
| Dimensional accuracy | 99.9 % - 99.2 % | 99.9 % - 99.2 % |
| Stability (linear<br>dimensional change) | < 0.2 % typical | < 0.1 % typical |
| Consistency | Type 0 - type 3<br>DIN EN ISO 4823 | Type 0 - type 3<br>DIN EN ISO 4823 |
| Design | Dynax® putty is sold in<br>tubs of 2 x 35 ml and<br>2 x 450 ml.<br>Dynax® light, mono,<br>heavy body and clear are<br>sold in 50 ml double<br>cartridges. | Fresh Bold liquid putty<br>impression system is sold<br>in tubs of 2 x 450 ml.<br>Fresh Bold light body,<br>monophase, heavy body<br>and clear bite are sold in<br>packs of 2 x 50 ml double<br>cartridges plus mixing<br>canulas |
| Material | Dynax® putty, light, mono<br>and heavy body:<br>Mixture of vinyl terminated<br>polydimethylsiloxanes and<br>filler materials with platinum<br>catalyst and<br>methylhydrosiloxane<br>dimethylsiloxane<br>copolymers.<br>Dynax® clear:<br>Mixture of vinyl terminated<br>polydi-methylsiloxanes and<br>silicic acid with platinum<br>catalyst and methyl-<br>hydrosiloxane<br>dimethylsiloxane<br>copolymers. | Fresh Bold liquid putty, light<br>body, monophase and heavy<br>body:<br>Mixture of vinyl terminated<br>polydimethylsiloxanes and<br>filler materials with platinum<br>catalyst and<br>methylhydrosiloxane<br>dimethylsiloxane copolymers,<br>Fresh Bold clear bite:<br>Mixture of vinyl terminated<br>polydi-methylsiloxanes and<br>silicic acid with platinum<br>catalyst and methyl-<br>hydrosiloxane<br>dimethylsiloxane copolymers. |
| Chemical Description | Room temperature<br>vulcanizing 2-components<br>silicone | Room temperature<br>vulcanizing 2-components<br>silicone |
# Summary of the Technological Characteristics of this Device compared to the Predicate:
Image /page/5/Picture/12 description: The image is a logo for TUV NORD CERT GmbH. The logo is circular with the text "TUV NORD" at the top, followed by "TUV NORD CERT GmbH" in smaller text. The outer ring of the circle contains the text "ISO 9001 / ISO 13485".
Image /page/5/Picture/13 description: The image shows a logo with the text "VDDI Dental Solutions German Manufacturing". The logo also includes a circular graphic element that is partially filled in with black, creating a crescent shape. The text is positioned to the right of the graphic, with "VDDI" on the top line and "Dental Solutions German Manufacturing" on the lines below.
Dreve Dentamid GmbH Max-Planck-Straße 31 59423 Unna/Germany Tel .: +49 2303 8807 - 0 Fax: +49 2303 8807-55
Internet: www.dreve.com/dentamid Lieferadresse: Einsteinstr. 36 Geschäftsführer: Dr. Volker Dreve, Dino Marchetti Sitz: 59423 Unna E-Mail: dentamid@dreve.de Reg.-Gericht: Hamm HRB 3712
Sparkasse Unna Kto .- Nr. 43 000 BLZ 443 500 60 BIC: WELADED1UNN IBAN: DE 26 44350060 0000043000
Commerzbank AG Dortmund Kto.-Nr. 03 222 354 00 BLZ 440 800 50 BIC: DRESDEFF440 IBAN: DE 35 44080050 0322235400
Page 3 of 5
Postbank Dortmund Kto .- Nr. 6 493 469 BLZ 440 100 46 BIC: PBNKDEFF IBAN: DE 42 44010046 0006493469
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the Dreve Dentamid logo. The logo features a red circle with the word "Dreve" written in white letters inside the circle. Below the circle, the words "Dreve Dentamid" are written in a larger, bold font.
Image /page/6/Picture/2 description: The image is a circular logo or certification mark. The text "TUV NORD" is prominently displayed at the top. Below that, the text "TUV NORD CERT GmbH" is present. The text "ISO 9001 / ISO 13485" is written around the bottom edge of the circle.
VDDI
Image /page/6/Picture/4 description: The image shows the letters B, V, and D in a row. Below the letters is the text "Kooperatives Mitglied Im Bundesverband Dentalhandel e.V." The letters are in a bold sans-serif font. There are vertical lines above each of the letters.
Dreve Dentamid GmbH Max-Planck-Straße 31 59423 Unna/Germany Tel.: +49 2303 8807-0 Fax: +49 2303 8807-55 E-Mail: dentamid@dreve.de
Internet: www.dreve.com/dentamid Lieferadresse: Einsteinstr. 36 Geschäftsführer: Dr. Volker Dreve, Dino Marchetti Sitz: 59423 Unna Reg.-Gericht: Hamm HRB 3712
Sparkasse Unna Kto.-Nr. 43 000 BLZ 443 500 60 BIC: WELADED1UNN IBAN: DE 26 44350060 0000043000
Commerzbank AG Dortmund Kto.-Nr. 03 222 354 00 BLZ 440 800 50 BIC: DRESDEFF440 IBAN: DE 35 44080050 032235400
Postbank Dortmund Kto.-Nr. 6 493 469 BLZ 440 100 46 BIC: PBNKDEFF
IBAN: DE 42 44010046 0006493469
Page 4 of 5
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the Dreve Dentamid logo. The logo features the word "Dreve" in white letters inside a red circle. Below the circle, the words "Dreve Dentamid" are written in a larger, bold font.
| Method of manipulation | Perforated impression trays | Perforated impression trays |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|
| Consensus standards | EN ISO 13485: 2012 +<br>AC:2012<br>ISO 7405: 2nd Edition<br>ISO 4823: 4th Edition<br>ISO 14971: 2nd Edition<br>ISO 10993-1: 2009 | EN ISO 13485: 2012 +<br>AC:2012<br>ISO 7405: 2nd Edition<br>ISO 4823: 4th Edition<br>ISO 14971: 2nd Edition<br>ISO 10993-1: 2009 |
## Conclusion
The technological characteristics of the subject device and the predicate are virtually identical and both devices have the same intended use. Any noted differences between. Dynax® Dental Impression Material and the predicate do not raise new questions of safety or effectiveness.
Dynax® Dental Impression Material is substantially equivalent to the predicate Fresh Bold, Impression material based on design, material, performance, chemical description and consensus standards.
Image /page/7/Picture/5 description: The image is a circular logo that is divided into two halves. The top left half is black and contains the text "TUV NORD" in white letters. Below that is the text "TUV NORD CERT" in smaller letters. The bottom half of the logo is white and contains the text "ISO 9001 / ISO 13485" in black letters. The text is arranged in a circular fashion around the edge of the logo.
Image /page/7/Picture/6 description: The image shows a logo with the text "VDDI Dental Solutions German Manufacturer". The logo features a stylized, abstract design resembling a partial circle or crescent shape on the left side. The text is arranged in a vertical stack, with "VDDI" in a larger, bolder font at the top, followed by "Dental Solutions" and "German Manufacturer" in smaller fonts below. The overall design is clean and professional, suggesting a company in the dental industry with a German manufacturing base.
Image /page/7/Picture/7 description: The image shows the letters BVD in a bold, sans-serif font. Below the letters, there is smaller text that reads "Kooperatives Mitglied Im Bundesverband Dentalhandel e V.". The letters are vertically aligned and appear to be part of a logo or heading.
Dreve Dentamid GmbH Max-Planck-Straße 31 59423 Unna/Germany Tel.: +49 2303 8807-0 Fax: +49 2303 8807-55 E-Mail: dentamid@dreve.de
Internet: www.dreve.com/dentamid Lieferadresse: Einsteinstr. 36 Geschäftsführer: Dr. Volker Dreve, Dino Marchetti Sitz: 59423 Unna Reg.-Gericht: Hamm HRB 3712
Sparkasse Unna Kto.-Nr. 43 000 BLZ 443 500 60 BIC: WELADED1UNN IBAN: DE 26 44350060 000043000
Commerzbank AG Dortmund Kto.-Nr. 03 222 354 00 BLZ 440 800 50 BIC: DRESDEFF440 IBAN: DE 35 44080050 0322235400
Page 5 of 5
Postbank Dortmund Kto.-Nr. 6 493 469 BLZ 440 100 46 BIC: PBNKDEFF IBAN: DE 42 44010046 0006493469
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.