The NB Mini Implant System is indicated for use in surgical and restorative applications for placement in the mandibular central, lateral incisor and maxillary lateral incisor regions of partially edentulous jaws where the horizontal space is limited by the adjacent teeth and roots, to provide support for prosthetic devices, such as artificial teeth, in order to restore the patient's chewing function. The NB Mini Implant System is indicated also for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
Device Story
NB Mini Implant System comprises endosseous dental implants (fixtures) and associated abutments (cover screws, healing abutments, scan healing abutments, cemented/angled abutments, screws). Fixtures are placed into alveolar bone to replace missing teeth; abutments connect to fixtures to support prosthetic restorations. Used in dental clinics by trained clinicians. Fixtures feature SLA surface treatment; internal hex/morse taper connection. Scan healing abutments include machined markings for digital impression capture via intraoral scanners. System supports immediate loading. Benefits include restoration of chewing function in narrow spaces where standard implants may not fit.
Clinical Evidence
Bench testing only. Included fatigue testing (ISO 14801), Bone to Implant Contact (BIC) analysis, surface area analysis, and pullout force testing compared to predicate K190849. Biocompatibility demonstrated via reference to K213506. Sterilization validated per ISO 11137-1/2 and ANSI/AAMI ST79. MRI safety evaluated via scientific rationale and literature.
Indicated for partially edentulous patients requiring dental implants in mandibular central, lateral incisor, and maxillary lateral incisor regions where horizontal space is limited. Supports prosthetic devices to restore chewing function. Indicated for immediate loading if primary stability and occlusal conditions are met.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The text logo has the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Arum Dentistry Co., Ltd. Won-Yi Choi Official Correspondent 23, Gukjegwahak 11-ro, Yuseong-gu Daejeon, 34002 REPUBLIC OF KOREA
April 26, 2024
Re: K240091
Trade/Device Name: NB Mini Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: January 12, 2024 Received: April 23, 2024
Dear Won-Yi Choi:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products: and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{2}------------------------------------------------
3
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Andrew I. Steen -S
Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
Submission Number (if known)
#### K240091
Device Name
NB Mini Implant System
#### Indications for Use (Describe)
The NB Mini Implant System is indicated for use in surgical and restorative applications for placement in the mandibular central, lateral incisor and maxillary lateral incisor regions of partially edentulous jaws where the horizontal space is limited by the adjacent teeth and roots, to provide support for prosthetic devices, such as artificial teeth, in order to restore the patient's chewing function. The NB Mini Implant System is indicated also for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains a logo for "ARUM DENTISTRY". The logo consists of a geometric shape on the left and the text "ARUM DENTISTRY" on the right. The geometric shape is made up of two triangles, one red and one light orange. The text "ARUM" is in a larger, bold font, while the text "DENTISTRY" is in a smaller, bold font and is located below the word "ARUM".
## 7. 510(K) Summary
### Submitter
ARUM DENTISTRY Co., Ltd. Won-Yi Choi 23, Gukjegwahak 11-ro, Yuseong-gu Daejeon, 34002 Republic of Korea Email: arum ra@arumdentistry.com Tel. +82-42-935-3644 Fax. +82-42-935-3633
### Device Information
- Trade Name: NB Mini Implant System
- Common Name: Implant, Endosseous, Root-Form .
- . Classification Name: Endosseous Dental Implant
- . Primary Product Code: DZE
- . Secondary Product Code: NHA
- . Panel: Dental
- . Requlation Number: 21 CFR 872.3640
- Device Class: Class II
- Date Prepared: 04/26/2024 ●
### Predicate Devices
The subject device is substantially equivalent to the following predicate and reference devices:
#### Primary Predicate
- K190849, IS-III active System by Neobiotech Co., Ltd.
### Reference Device
- K220079, Magicore Narrow System by InnoBioSurg Co., Ltd.
- K161604, Osstem Implant System by Osstem Implant Co., Ltd. .
- K173374, TSV™ BellaTek® Encode® Healing Abutments by Biomet 3i .
- . K213506. NB 1 SA Implant System by ARUM DENTISTRY Co., Ltd.
- . K182091, Osstem Abutment System by Osstem Implant Co., Ltd.
- K222131, NB 1 SA Implant System by ARUM DENTISTRY Co., Ltd. .
- K230725, NB Implant System by ARUM DENTISTRY Co., Ltd. .
- K232560, Angled Abutment by ARUM DENTISTRY Co., Ltd.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Arum Dentistry. The logo consists of a red triangle and a yellow triangle on the left side, with the words "ARUM" and "DENTISTRY" in bold, dark brown letters on the right side. The word "ARUM" is above the word "DENTISTRY".
### General Description
The NB Mini Implant System consist of below:
### Fixture
- NB Fixture Mini
### Abutment
- Cover Screw
- . Healing Abutment
- Scan Healing Abutment
- Scan Healing Abutment Screw
- Cemented Abutment Mini
- . Angled Abutment Mini
- Abutment Screw
The NB Fixture Mini is made of Ti-6AI-4V Eli (Conforming to ASTM F136) which will be placed in the alveolar bone to replace the function of the missing tooth. The NB Mini Implant System consists of dental implants, abutments for use in one or two-stage dental implant placement and restorations.
The implant-abutment connection is tight and precise fitting with internal hex and morse taper bevel. The surface of the fixture is treated with SLA (Sandblasted with Large-grit and Acid-etching). It is only part to be implanted into bone, and to provide connection of prosthetic devices or other components of a dental implant set with human body (mandibular or maxillary bone).
The dimension ranges of the subject device are below:
| No. | Device Name | Dimension |
|-----|-----------------------|-----------------------------------------------------------------------|
| 1 | NB Fixture Mini | Ø 3.2, 3.5 (D) x 8.5, 10, 11.5, and 13 mm (L) |
| No. | Device Name | Dimension |
| 1 | Cover Screw | Ø2.8 (D) x 5.1, 5.8, 6.8, 7.8 mm (L) |
| 2 | Healing Abutment | Ø4.2, 4.7, 5.7, 6.7, 7.7 (D) x 1.0, 2.0, 3.0, 4.0 mm (Cuff<br>Height) |
| 3 | Scan Healing Abutment | Ø4.2, 4.7 (D) x 1.4, 3.4, 5.4, 7.4, 9.4 mm (Cuff Height) |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains a logo for "ARUM DENTISTRY". The logo consists of a geometric shape on the left, with a red triangle above a smaller, light orange triangle. To the right of the geometric shape is the text "ARUM" in a bold, dark brown font, with "DENTISTRY" in the same font directly below it. The overall design is clean and modern.
K240091
| | | Ø4.7, 5.7, 6.7, 7.7 (D) x 1.3, 3.3, 5.3, 7.3, 9.3 mm (Cuff Height) |
|---|-----------------------------|----------------------------------------------------------------------------------|
| 4 | Scan Healing Abutment Screw | Ø2.49 (D) x 9.7, 11.4, 11.7, 13.4, 13.7, 15.4, 15.7, 17.4, 17.7, 19.4 mm (L) |
| 5 | Cemented Abutment Mini | Ø4.0, 4.5 (D) x 5.0, 7.0 mm (Post Height) x 1.0, 2.0, 3.0, 4.0, 5.0(Cuff Height) |
| 6 | Angled Abutment Mini | Ø4.0, 4.5 (D) x 8 mm (Post Height) x 2, 4mm (Cuff Height) |
| 7 | Abutment Screw | Ø2.2 (D) x 10.8 mm (L) |
Below are the abutment's features:
| No. | Device Name | Uses | Surface treatment |
|-----|--------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|
| 1 | Cover Screw | The Cover Screw is Intended to be<br>temporarily connected to an endosseous<br>dental implant to protect the implant<br>connection interface during bone healing. | No surface<br>treatment |
| 2 | Healing Abutment | The Healing Abutment is designed to aid in<br>soft tissue contouring during the healing<br>period after implant placement, creating an<br>emergence profile for the final prosthesis. | No surface<br>treatment |
| 3 | Scan Healing<br>Abutment | The Scan Healing Abutment has the added<br>feature of machined markings for easy<br>identification when taking an abutment-level<br>impression or an intraoral scan/digital<br>impression from the healing abutment.<br>Identification and orientation information is<br>captured in the intraoral scan or model scan. | No surface<br>treatment |
| 4 | Scan Healing<br>Abutment Screw | The Scan Healing Abutment Screw is used<br>for connect fixture and Scan Healing<br>abutment. | No surface<br>treatment |
| 5 | Cemented<br>Abutment Mini | The Cemented Abutment Mini is used as a<br>support of prosthesis to restore the patient's<br>chewing function. | TiN Coating |
| 6 | Angled Abutment<br>Mini | The Angled Abutment Mini is used as a<br>support of prosthesis to restore the patient's<br>chewing function. | TiN Coating |
| 7 | Abutment Screw | The Abutment Screw is used for connect<br>fixture and abutment. | No surface<br>treatment |
The Cover Screw, Healing Abutment and Scan Healing Abutment is provided Sterile from manufacturer. Other abutments should be sterilized before use by end user sterilization. These devices are intended for single use only.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the logo for Arum Dentistry. The logo consists of a red triangle on top of a yellow triangle, with the text "ARUM" in bold, dark brown letters to the right of the triangles. Below "ARUM" is the word "DENTISTRY" in the same font and color, but slightly smaller.
#### K240091
### Indication for Use
The NB Mini Implant System is indicated for use in surgical and restorative applications for placement in the mandibular central, lateral incisor and maxillary lateral incisor regions of partially edentulous jaws where the horizontal space is limited by the adjacent teeth and roots, to provide support for prosthetic devices, such as artificial teeth, in order to restore the patient's chewing function. The NB Mini Implant System is indicated also for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
### Materials
The Fixtures are fabricated from Ti-6AI-4V Eli (Conforming to ASTM F136).
All Abutments and Abutment Screws are fabricated from Ti-6Al-4V Eli (Conforming to ASTM F136).
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image contains the logo for Arum Dentistry. The logo consists of a geometric shape on the left and the words "ARUM DENTISTRY" on the right. The geometric shape is made up of a red triangle on top and a smaller, light orange triangle on the bottom left. The words "ARUM DENTISTRY" are in a bold, dark brown font, with "ARUM" on top and "DENTISTRY" on the bottom.
# Summaries of Technology Characteristics
# 1) NB Fixture Mini
| | Subject Device | Primary Predicate | Reference Device |
|--------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | ARUM DENTISTRY Co., Ltd. | Neobiotech Co., Ltd. | InnoBioSurg Co., Ltd. |
| Device Name | NB Mini Implant System | IS-III active System | Magicore Narrow System |
| 510(k) Number | K240091 | K190849 | K220079 |
| Intended Use/<br>Indications for use | The NB Mini Implant System is<br>indicated for use in surgical and<br>restorative applications for<br>placement in the mandibular<br>central, lateral incisor and<br>maxillary lateral incisor regions<br>of partially edentulous jaws<br>where the horizontal space is<br>limited by the adjacent teeth<br>and roots, to provide support for<br>prosthetic devices, such as<br>artificial teeth, in order to<br>restore the patient's chewing<br>function. The NB Mini Implant<br>System is indicated also for<br>immediate loading when good<br>primary stability is achieved and<br>with appropriate occlusal<br>loading. | The IS-III active System_S-<br>narrow Type is indicated for use<br>in surgical and restorative<br>applications for placement in the<br>mandibular central, lateral incisor<br>and maxillary lateral incisor<br>regions of partially edentulous<br>jaws where the horizontal space<br>is limited by the adjacent teeth<br>and roots, to provide support for<br>prosthetic devices, such as<br>artificial teeth, in order to restore<br>the patient's chewing function.<br>The IS-III active System_S-<br>narrow Type is indicated also for<br>immediate loading when good<br>primary stability is achieved and<br>with appropriate occlusal<br>loading. | The The Magicore Narrow<br>System<br>(3.0, 3.5mm) may be used as an<br>artificial root structure for single<br>tooth replacement of mandibular<br>central and lateral incisors and<br>maxillary lateral incisors. The<br>implants may be restored<br>immediately 1) with a temporary<br>prosthesis that is not in<br>functional occlusion, 2) when<br>splinted together as an artificial<br>root structure for multiple tooth<br>replacement of mandibular<br>incisors, or 3) for denture<br>stabilization using multiple<br>implants in the anterior mandible<br>and maxilla. The implants may<br>be placed in immediate function<br>when good primary stability has<br>been achieved and with<br>appropriate occlusal loading. |
| Material | Ti-6Al-4V Eli (ASTM F136) | Ti-6Al-4V Eli (ASTM F136) | Ti-6Al-4V Eli (ASTM F136) |
| Anti-Rotational Feature | Internal Hex | Internal Hex | Internal Hex |
| Range of Diameters (Ø) | 3.2, 3.5 | 3.2 | 3.0, 3.5 |
| Range of Lengths (mm) | 8.5, 10, 11.5, 13.0, 15.0 | 8.5, 10, 11.5, 13.0, 15.0 | 11.0, 13.0, 15.0 |
| Surface treatment | SLA | SLA | RBM & SLA |
| Sterilization | Gamma Sterilization | Gamma Sterilization | Gamma Sterilization |
| Shelf life | 5 Years | 5 Years | 8 Years |
| Principle of Operation | This product is a root-type<br>fixture which is inserted in the<br>alveolar bone. It replaces the<br>functions of the missing teeth<br>as a dental implant fixture. | This product is a root-type fixture<br>which is inserted in the alveolar<br>bone. It replaces the functions of<br>the missing teeth as a dental<br>implant fixture. | This product is a root-type fixture<br>which is inserted in the alveolar<br>bone. It replaces the functions of<br>the missing teeth as a dental<br>implant fixture. |
| Substantial Equivalent<br>Discussion | 1. Similarities<br>The NB Fixture Mini has similar device characteristics with the Primary predicate such as Indications for<br>use, functions, material, surface treatment (SLA), fixture diameter, surgical technique, sterilization<br>method and shelf list, structure and applied production method.<br><br>2. Differences<br>The difference between the subject device and primary predicate are device design and the longest<br>implantable length of Ø3.5. To support the Ø3.5 X13mm long length, K220079 was added as reference<br>device. A surface area comparison and dynamic fatigue testing show that the subject device has | | |
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image contains the logo for Arum Dentistry. The logo consists of a geometric shape on the left and the words "ARUM DENTISTRY" on the right. The geometric shape is made up of two triangles, one red and one light orange. The words "ARUM DENTISTRY" are in a bold, sans-serif font.
substantially equivalent performance compared to the predicate and reference devices.
K240091
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image contains the logo for Arum Dentistry. The logo consists of a geometric shape on the left and the words "ARUM DENTISTRY" on the right. The geometric shape is made up of two triangles, one red and one light orange. The words "ARUM DENTISTRY" are in a bold, sans-serif font.
| | Subject Device | Reference Device | Reference Device |
|--------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------|------------------------|
| Manufacturer | ARUM DENTISTRY Co., Ltd. | ARUM DENTISTRY Co., Ltd. | InnoBioSurg Co., Ltd. |
| Device Name | Cover Screw | NB 1 SA Implant System | Magicore Narrow System |
| 510(k) No. | K240091 | K222131 | K220079 |
| Material | Ti-6Al-4V Eli | Ti-6Al-4V Eli | Ti-6Al-4V Eli |
| Range of Diameters (Ø) | 2.8 | 3.6 | 2.75 |
| Range of Length (mm) | 5.1, 5.8, 6.8, 7.8 | 5.3, 6.0, 7.0, 8.0 | 5.2 |
| Sterilization | Gamma Sterilization | Gamma Sterilization | Gamma Sterilization |
| Surface Treatment | Non-Anodizing | Non-Anodizing | Anodizing |
| Substantial Equivalent<br>Discussion | 1. Similarities<br>The subject device is similar in intended use, fundamental scientific technology, principle of operation,<br>design, technology, functions, and materials with the identified primary predicate.<br><br>2. Differences<br>The difference between the subject and primary predicate is the dimension. There are slightly different<br>dimensions. This dimensional difference doesn't affect device safety and effectiveness. | | |
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image contains the logo for Arum Dentistry. The logo consists of a geometric shape on the left and the words "ARUM DENTISTRY" on the right. The geometric shape is made up of two triangles, one red and one light orange. The words "ARUM DENTISTRY" are in a bold, sans-serif font.
## 3) Healing Abutment
| | Subject Device | Reference Device | Reference Device |
|--------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------|-----------------------------------|
| Manufacturer | ARUM DENTISTRY Co., Ltd. | ARUM DENTISTRY Co., Ltd. | Osstem Implant Co., Ltd |
| Device Name | Healing Abutment | NB 1 SA Implant System | Osstem Implant System |
| 510(k) No. | K240091 | K222131 | K161604 |
| Material | Ti-6Al-4V Eli | Ti-6Al-4V Eli | Ti-6Al-4V Eli |
| Range of Diameters (ø) | 4.2, 4.7, 5.7, 6.7, 7.7 | 4.2, 4.7, 5.7, 6.7, 7.7 | 3.8, 4.1, 5.0, 5.6, 6.0, 6.8, 7.5 |
| Range of Cuff (ø) | 1.0, 2.0, 3.0, 4.0 | 1.0, 2.0, 3.0, 4.0 | 3.0, 4.0, 6.0, 8.0 |
| Sterilization | Gamma Sterilization | Non-Sterilization | Gamma Sterilization |
| Shelf life | 5 Years | | 5 Years |
| Surface Treatment | Non-Anodizing | Non-Anodizing | Anodizing |
| Substantial Equivalent<br>Discussion | 1. Similarities<br>The Healing Abutment has same indication for use, principle of operation, functions, diameter and<br>material to the predicate K222131.<br>The intended use of the subject device as a healing abutment is designed to aid in soft tissue contouring<br>during the healing period after implant placement, creating emergence profile for the final prosthesis is<br>equivalent to the reference predicate K161604.<br>2. Differences<br>There are slightly different Sterilization, To support sterilization, K161604 were added. Therefore, this<br>sterilization method difference doesn't affect device safety and effectiveness. | | |
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image contains the logo for Arum Dentistry. The logo features a red triangle and a smaller tan triangle on the left side. To the right of the triangles, the word "ARUM" is written in a bold, dark brown font, with the word "DENTISTRY" written in the same font directly below it.
## 4) Scan Healing Abutment
| | Subject Device | Reference Device | Reference Device |
|--------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------|---------------------------------------------|
| Manufacturer | ARUM DENTISTRY Co., Ltd. | ARUM DENTISTRY Co., Ltd. | Biomet 3i |
| Device Name | Scan Healing Abutment | NB Implant System | TSV™ BellaTek® Encode®<br>Healing Abutments |
| 510(k) No. | K240091 | K230725 | K173374 |
| Material | Ti-6Al-4V Eli | Ti-6Al-4V Eli | Ti-6Al-4V Eli |
| Range of Diameters ( $∅$ ) | 4.2 ~ 7.7 | 4.2 ~ 7.7 | 3.8 ~ 6.8 |
| Range of Cuff ( $∅$ ) | 1.3 ~ 9.4 | 1.3 ~ 9.4 | 3.0, 5.0, 7.0 |
| Scanning feature | Machined marking | Machined marking | Machined marking |
| Surface Treatment | Non-Anodizing | Non-Anodizing | Anodizing |
| Sterilization | Gamma Sterilization | Non-Sterilization | Gamma Sterilization |
| Shelf life | 5 Years | | 5 Years |
| Substantial Equivalent<br>Discussion | 1. Similarities<br>The Scan Healing Abutment has same indication for use, principle of operation, functions, diameter and<br>material to the predicate K230725.<br>The intended use of the subject device as a scan healing abutment with scanning feature to transmit<br>position and angulation data of implant when taking a digital impression using an intra-oral scanner is<br>equivalent to the reference predicate K173374.<br>2. Differences<br>The difference between the subject and primary device is the sterilization method. To support<br>sterilization method, K173374 were added. Therefore, this sterilization method difference doesn't affect<br>device safety and effectiveness. | | |
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image contains a logo for Arum Dentistry. The logo consists of a geometric shape on the left and the text "ARUM DENTISTRY" on the right. The geometric shape is made up of two triangles, one red and one beige. The text "ARUM" is in a larger font than the text "DENTISTRY".
# 5) Cemented Abutment Mini
| | Subject Device | Reference Device |
|--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|
| Manufacturer | ARUM DENTISTRY Co., Ltd. | ARUM DENTISTRY Co., Ltd. |
| Device Name | Cemented Abutment Mini | NB 1 SA Implant System |
| 510(k) No. | K240091 | K213506 |
| Material | Ti-6Al-4V Eli | Ti-6Al-4V Eli |
| Diameters (Ø) | 4.0, 4.5 | 4.5, 5.5, 6.5 |
| Post Height (mm) | 5.0, 7.0 | 5.0, 5.5, 7.0 |
| Cuff Height (mm) | 1.0, 2.0, 3.0, 4.0, 5.0 | 1.0, 2.0, 3.0, 4.0, 5.0, 6.0 |
| Surface Treatment | TiN Coating…