The NB 1 SA Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, screw retained, or overdenture restorations, and terminal or intermediate Abutment support for fixed bridgework. NB 1 SA Implant System is dedicated for two stage surgical procedures and for immediate loading when there is good primary stability and an appropriate occlusal load. Also, implants with diameters larger than 5mm are indicated for molar regions.
Device Story
NB 1 SA Implant System comprises endosseous dental implants (fixtures) and associated components (cover screws, healing abutments, cemented abutments, abutment screws). Fixtures are placed into the alveolar bone to replace missing teeth. Components are connected to fixtures to support dental prostheses. Used in dental clinics by trained clinicians. Fixtures feature SLA surface treatment. Components are either screw-retained or cemented. System facilitates gingival shaping and prosthetic restoration. Benefits include restoration of tooth function and aesthetics in edentulous sites.
Clinical Evidence
Bench testing only. Testing included biocompatibility (ISO 10993 series), gamma sterilization validation (ISO 11137-1), shelf-life testing (ASTM F1980), and LAL endotoxin testing. Surface characterization (roughness, composition, SEM) provided. Sterilization for non-sterile components validated per ISO 17665-1/2.
Technological Characteristics
Fixtures: Pure titanium (ASTM F67), SLA surface. Abutments/Screws: Ti-6Al-4V ELI (ASTM F136), some TiN coated. Connection: Internal hex. Sterilization: Gamma (fixtures/cover screws), non-sterile (other abutments). Dimensions: Fixture diameters 3.8-6.5mm, lengths 7.0-13.0mm.
Indications for Use
Indicated for partially or fully edentulous adult patients requiring single or multiple-unit dental restorations, including cemented, screw-retained, or overdenture restorations, and fixed bridgework support. Suitable for two-stage surgical procedures and immediate loading given sufficient primary stability and occlusal load. Implants >5mm diameter indicated for molar regions.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a seal with the words "HUMAN SERVICES-USA" around the edge. To the right of the seal, there is a blue square with the letters "FDA" in white. Next to the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.
Arumdentistry Co., Ltd. % April Lee Consultant Withus Group Inc 106 Superior Irvine, California 92620
Re: K213506
Trade/Device Name: NB 1 SA Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: March 4, 2022 Received: March 7, 2022
## Dear April Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K213506
Device Name NB 1 SA Implant System
#### Indications for Use (Describe)
The NB 1 SA Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, or overdenture restorations, and terminal or intermediate Abutment support for fixed bridgework. NB 1 SA Implant System is dedicated for two stage surgical procedures and for immediate loading when there is good primary stability and an appropriate occlusal load. Also, implants with diameters larger than 5mm are indicated for molar regions.
| Type of Use (Select one or both, as applicable) |
|--------------------------------------------------------------------------|
| <span></span> <span></span> Prescription Use (Part 21 CFR 801 Subpart D) |
| <span></span> <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(K) Summary
#### Submitter
Arumdentistry Co., Ltd. Hyang Mi Lee 1-dong, 44, Techno 8-ro, Yuseong-gu, Daejeon, Republic of Korea Email: arum ra@arumdentistry.com Tel. +82-42-935-3644 Fax. +82-42-935-3633
### Official Correspondent
Withus Group Inc. April Lee 106 Superior. Irvine, CA 92620 USA Email: withus6664@gmail.com Phone: 1-909-274-9971 Fax: 1-909-460-8122
### Device Information
- Trade Name: NB 1 SA Implant System ●
- Common Name: Endosseous Dental Implant
- Classification Name: Implant, Endosseous, Root-Form ●
- Primary Product Code: DZE
- Secondary Product Code: NHA
- Panel: Dental
- . Regulation Number: 21 CFR 872.3640
- Device Class: Class II
- . Date Prepared: 04/06/2022
## Predicate Devices:
The subject device is substantially equivalent to the following predicate devices:
## Primary Predicate
- . K181138, IS-III active System by Neobiotech Co., Ltd.
## Reference Device
- . K150060, J2A Dental Implant System manufactured by KJ Meditech Co., Ltd.
- K172100, URIS Implant System by Truabutment Inc. ●
- K193425, Pre-Milled Blank by Arumdentistry Co., Ltd.
#### Indication for Use:
The NB 1 SA Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, screw retained, or overdenture restorations, and terminal or intermediate Abutment support for fixed bridgework. NB 1 SA Implant System is dedicated for two stage surgical procedures and for immediate loading when there is good primary stability and an appropriate occlusal load. Also, implants with diameters larger than 5mm are indicated for molar regions.
{5}------------------------------------------------
### Device Description:
The NB 1 SA Implant System is composed of ARUM NB 1 SA Fixture, ARUM NB Cover Screw, ARUM NB Healing Abutment, ARUM Cemented Abutment (Hex, Non-Hex), and Abutment screw. ARUM NB 1 SA Fixture is a thread type implant body made of TI CP4 according to ASTM F67 which will be placed in the alveolar bone to replace the function of the missing tooth.
The surface of the fixture is treated with SLA (Sandblasted with Large grit and Acid-etching). The fixture is placed in the anterior or posterior site of maxillary or mandibular jawbone considering bone quality and bone quantity and it is connected with dental prostheses.
| No. | Device Name | Dimension Ranges |
|-----|--------------------------------|----------------------------------------------------------------------|
| 1 | ARUM NB 1 SA Fixture | Ø 3.8 (D) x 8.5, 10, 11.5, and 13 mm (L) |
| | | Ø 4.0, 4.5, 5.0 (D) x 7.0, 8.5, 10, 11.5, and 13 mm (L) |
| | | Ø 5.5, 6.0, 6.5 (D) x 7.0, 8.5, 10, 11.5, and 13 mm (L) |
| 2 | ARUM NB Cover Screw | Ø 3.6 (D) x 5.3, 6, 7, 8 mm (L) |
| 3 | ARUM NB Healing Abutment | Ø 4.2, 4.7, 5.7, 6.7, 7.7 (D) x 1.0, 2.0, 3.0, 4.0mm(Cuff<br>Height) |
| 4 | ARUM Cemented Abutment_Hex | Ø 4.5, 5.5, 6.5 (D) x 5.0, 5.5, 7.0 mm (Post Height) |
| | ARUM Cemented Abutment_Non Hex | Ø 4.5, 5.5, 6.5 (D) x 5.0, 5.5, 7.0 mm (Post Height) |
| 5 | Abutment Screw | Ø 2.35 (D) x 8.4 mm(L) |
The dimensions of the subject device are as following:
Tolerance of dimension shall be within ± 1% range.
#### The Abutments have below featured:
| Name | Uses | Surface | Connection |
|-----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|------------------------------|
| ARUM NB Cover<br>Screw | It is used for protecting inner hole and connecting<br>part with exposed upper part of structure during<br>the healing period after inserting dental implant<br>fixture | Non | Screw<br>Retained |
| ARUM NB Healing<br>Abutment | Used to formation appropriate gingival shape<br>during the soft tissue healing period combined<br>with implant | Non | Screw<br>Retained |
| ARUM Cemented<br>Abutment | The Abutment is connected with fixture and it<br>supports prosthesis which restores tooth function. | TiN Coating | Internal<br>Hex, Non-<br>hex |
| Abutment Screw | Abutment Screw is used to connect an abutment<br>to the fixture. | Non | Screw<br>Retained |
ARUM NB 1 SA Fixture and ARUM NB Cover Screw are provided sterilized. And the other Abutments are provided non-sterilized.
ARUM NB 1 SA Fixture is enclosed with ARUM NB Cover Screw in a packing. ARUM Cemented abutment is enclosed with Abutment Screw in a packing.
{6}------------------------------------------------
# Materials:
- The fixtures are fabricated from Pure titanium of ASTM F67 ●
- ARUM Cemented abutment, abutment screw, Healing Abutment, and cover screw are fabricated from . Ti-6Al-4V of ASTM F136
# Summaries of Technological Characteristics & Substantial Equivalence Discussion
| | Subject Device | Primary Predicate | Reference Device |
|----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company | ARUMDENTISTRY Co., Ltd. | Neobiotech Co., Ltd | KJ Meditech Co., Ltd. |
| Device Name | NB 1 SA Implant System | IS-III active System | J2A Dental Implant System |
| 510(k) Number | K213506 | K181138 | K150060 |
| Device<br>Classification | Implant, Endosseous,<br>Root-Form | Implant, Endosseous,<br>Root-Form | Implant, Endosseous,<br>Root-Form |
| Product Code | DZE | DZE | DZE |
| Regulation<br>Number | 872.3640 | 872.3640 | 872.3640 |
| Indications for<br>Use | The NB 1 SA Implant System<br>is indicated for use in partially<br>or fully edentulous mandibles<br>and maxillae, in support of<br>single or multiple-unit<br>restorations including;<br>cemented retained, screw<br>retained, or overdenture<br>restorations, and terminal or<br>intermediate Abutment<br>support for fixed bridgework.<br>NB 1 SA Implant System is<br>dedicated for two stage<br>surgical procedures and for<br>immediate loading when there<br>is good primary stability and<br>an appropriate occlusal load.<br>Also, implants with diameters<br>larger than 5mm are indicated<br>for molar regions. | The IS-III active System is<br>indicated for use in partially<br>or fully edentulous mandibles<br>and maxillae, in support of<br>single or multiple-unit<br>restorations including;<br>cemented retained, screw<br>retained, or overdenture<br>restorations, and terminal or<br>intermediate Abutment<br>support for fixed bridgework.<br>IS-III active System is<br>dedicated for two stage<br>surgical procedures and for<br>immediate loading when there<br>is good primary stability and<br>an appropriate occlusal load.<br>Also, implants with diameters<br>larger than 5mm are indicated<br>for molar regions. | The J2A Dental Implant<br>System is indicated for use in<br>partially or fully edentulous<br>mandibles and maxillae, in<br>support of single or multiple-<br>unit restorations including;<br>cemented retained, screw<br>retained, or overdenture<br>restorations, and terminal or<br>intermediate Abutment<br>support for fixed bridgework.<br>The J2A Dental Implant<br>System is for single and two<br>stage surgical procedures. The<br>system is intended for delayed<br>loading. |
| Material | TI CP4 of ASTM F67 | TI CP4 of ASTM F67 | Ti-6Al-4V ELI of ASTM<br>F136 |
| Design | Image: Design of NB 1 SA Implant System | Image: Design of IS-III active System | Image: Design of J2A Dental Implant System |
| Anti-Rotational<br>Feature | Internal Hex | Internal Hex | Internal Hex |
| Diameters( $\varnothing$ ) | 3.8/4.0/4.5/5.0/5.5/6.0/6.5 | 3.5/4.0/4.5/5.0/5.5/6.0/7.0 | 3.75/4.0/4.3/4.5/5.0/5.5/6.0 |
| Lengths(mm) | 7.0/ 8.5/ 10/ 11.5/ 13.0 | 7.3/8.5/10.0/11.5/13.0/15.0 | 7.0/8.5/10.0/11.5/13.0/15.0 |
| Surface<br>Treatment | SLA | SLA | RBM |
| Sterilization | Gamma Sterilization | Gamma Sterilization | Gamma Sterilization |
| Principle of<br>Operation | This product is a root-type<br>fixture which is inserted in the<br>alveolar bone. It replaces the<br>functions of the missing teeth<br>as a dental implant fixture. | This product is a root-type<br>fixture which is inserted in the<br>alveolar bone. It replaces the<br>functions of the missing teeth<br>as a dental implant fixture. | This product is a root-type<br>fixture which is inserted in the<br>alveolar bone. It replaces the<br>functions of the missing teeth<br>as a dental implant fixture. |
| Similarities | The NB 1 SA Implant System has same device characteristics with the Primary predicate device,<br>IS-III active System(K181138) such as intended use, material, functions, surface treatment,<br>general shape (Design), structure, anti-rotational feature, principle operation and applied<br>production method. | | |
| Differences | The differences between the subject device and the primary predicate device are the product<br>design and dimensional range. The design difference doesn't affect product's fundamental<br>function. To support the dimensional differences such as fixture with Ø4.0 X 7.0mm, we selected<br>K 150060 as the reference device, which covers the subject device's dimensional range.<br>Therefore, it is substantially equivalent. | | |
#### 1) ARUM NB 1 SA Fixture
{7}------------------------------------------------
# 2) ARUM NB Cover Screw
| | Subject Device | Primary Predicate |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company | ARUMDENTISTRY Co., Ltd. | Neobiotech Co., Ltd |
| Device Name | NB 1 SA Implant System | IS-III active System |
| 510(k) Number | K213506 | K181138 |
| Material | Ti-6Al-4V ELI of ASTM F136 | Ti-6Al-4V ELI of ASTM F136 |
| Design | Image: Implant design 1 | Image: Implant design 2 |
| Diameters(Ø) | 3.6 | 3.45/3.6 |
| Lengths(mm) | 5.3, 6, 7, 8 mm (L) | 5.85/6.85/7.45/<br>6.4/7.4/8.0/ |
| Surface Treatment | Non-Anodizing | Anodizing/ Non-Anodizing, |
| Sterilization | Gamma Sterilization | Gamma Sterilization |
| Principle of Operation | It is used for protecting inner hole and connecting part with exposed upper part of structure during the healing period after inserting dental implant fixture. When inserting the Abutment, Cover screw is removed. | It is used for protecting inner hole and connecting part with exposed upper part of structure during the healing period after inserting dental implant fixture. When inserting the Abutment, Cover screw is removed. |
| Similarities | The subject device has same intended use, material, functions, principle of operation, shelf life and similar design and dimensions. | |
| Differences | There are slightly different designs and dimension. These differences do not affect product's fundamental function; therefore, it is substantial equivalent. | |
{8}------------------------------------------------
| | Subject Device | Primary Predicate | Reference Device | |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Company | ARUMDENTISTRY Co.,<br>Ltd. | Neobiotech Co., Ltd | TruAbutment Korea Co.,<br>Ltd. | |
| Device Name | NB 1 SA Implant System | IS-III active System | URIS OMNI System | |
| 510(k) Number | K213506 | K181138 | K172100 | |
| Material | Ti-6Al-4V ELI of ASTM<br>F136 | Ti-6Al-4V ELI of ASTM<br>F136 | Ti-6Al-4V ELI (ASTM<br>F136) | |
| Design | Image: design | Image: design | Image: design | |
| Diameters (Ø) | 4.2/4.7/5.7/6.7/7.7 | 4.0/4.5/4.8/5.5/6.0/6.8/8.0/9.0 | 4.0/4.5/5.5/6.5/7.5mm | |
| Cuff height (mm) | 1.0,2.0, 3.0, 4.0mm | 2.3/2.8/3.3/3.8/4.3/4.8/5.3/5.8<br>/6.3/6.8/7.8/ | 1.0mm~5.0mm | |
| Surface<br>Treatment | N/A | N/A | Anodizing (Yellow, Green) | |
| Sterilization | Non-sterile | Gamma Sterilization | Non-sterile | |
| Principle of<br>Operation | This product is healing<br>Abutment to formation<br>appropriate gingival shape<br>during the soft tissue healing<br>period combined with<br>implant. This product should<br>be removed when the<br>superstructure is set up. | This product is healing<br>Abutment to formation<br>appropriate gingival shape<br>during the soft tissue healing<br>period combined with<br>implant. This product should<br>be removed when the<br>superstructure is set up. | This product is healing<br>Abutment to formation<br>appropriate gingival shape<br>during the soft tissue healing<br>period combined with<br>implant. This product should<br>be removed when the<br>superstructure is set up. | |
| Similarities | The subject device has same intended use, material, functions, principle of operation, and<br>similar design and dimensions. | | | |
| Differences | The differences between the subject device and the primary predicate device are the<br>sterilization, dimensions and design. To support the difference of the sterilization, K172100<br>was added and the differences of design and dimensions do not affect product's fundamental<br>function, therefore, it is substantially equivalent. | | | |
# 3) ARUM NB Healing Abutment
{9}------------------------------------------------
| | Subject Device | | Primary Predicate Device | | |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------|-------------------------------------------------------------------------------------------------------------|-----------------------|-----------------------|
| Company | ARUMDENTISTRY Co., Ltd. | | Neobiotech Co., Ltd | | |
| Device Name | NB 1 SA Implant System | | IS-III active System | | |
| 510(k) Number | K213506 | | K181138 | | |
| Material | Ti-6Al-4V ELI of ASTM F136 | | Ti-6Al-4V ELI of ASTM F136 | | |
| Design | Image: dental implant | Image: dental implant | Image: dental implant | Image: dental implant | Image: dental implant |
| | Hex | Non-Hex | Hex | Non-Hex | SCRP |
| Diameters (Ø) | 4.5 / 5.5 / 6.5 | | 4.5/5.2/5.7/6.5 | | |
| Post height<br>(mm) | 5.0/ 5.5 / 7.0 | | 4.0/4.5/5.5/7.0/8.0 | | |
| Surface<br>Treatment | TiN-Coating | | TiN-Coating | | |
| Principle of<br>Operation | It is indicated for screw-retained single<br>tooth or cement retained single tooth and<br>bridge restorations. | | It is indicated for screw-retained single tooth or cement<br>retained single tooth and bridge restorations. | | |
| Similarities | The subject and primary predicate have same indications for use, functions, materials, surface<br>treatment, and general shape (design) as the primary predicate. | | | | |
| Differences | The design of the devices is slightly different, but it doesn't affect device's fundamental functions;<br>therefore, it is substantial equivalent. | | | | |
# 4) ARUM Cemented Abutment
# 5) Abutment Screw
| | Subject Device | Primary Predicate | |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------|--|
| Company | ARUMDENTISTRY Co., Ltd. | Neobiotech Co., Ltd | |
| Device Name | NB 1 SA Implant System | IS-III active System | |
| 510(k) Number | K213506 | K181138 | |
| Material | Ti-6Al-4V ELI of ASTM F136 | Ti-6Al-4V ELI of ASTM F136 | |
| Design | Image: Screw | Image: Two screws | |
| Diameters (Ø) | 2.35 | 2.3 | |
| Length (mm) | 8.4 | 8.8/8.3 | |
| Surface Treatment | N/A | N/A | |
| Principle of Operation | This product is a screw for connected<br>with Abutment and fixture. | This product is a screw for connected with<br>Abutment and fixture. | |
| Similarities | The subject and primary predicate have same indications for use, functions, materials,<br>surface treatment, general shape (design) and diameters. | | |
| Differences | The design of the devices is slightly different, but it doesn't affect device's fundamental<br>functions; therefore, it is substantial equivalent. | | |
{10}------------------------------------------------
#### Non-Clinical Test Data
Below tests were performed on subject device:
- . Gamma Sterilization Validation Test on Fixtures according to ISO 11137-1.2.3
- . Shelf-Life Test on Fixtures according to ASTM F1980
- . Biocompatibility testing on fixtures according to ISO 10993-1:2009, ISO 10993-3:2014, ISO 10993-5:2009, ISO 10993-6:2007, ISO 10993-10:2010 and ISO 10993-11:2006
- . Biocompatibility testing on TiN Coating Abutments according to ISO 10993-1:2009, ISO 10993-5:2009 and ISO 10993-10:2010
- . LAL endotoxin testing according to AAMI / ANSI ST72:2011/(R)2016
Below tests were performed for predicate device, K193425 and leveraged for the subject device:
- End User Sterilization Validation Test Report on Abutments according to ANSI/AAMI ST79, ISO 17665-1,-2, ISO 11737-1,-2, and ISO 11138-1 referenced in K193425
- Biocompatibility testing on Abutments made with Ti-6Al-4V ELI according to ISO 10993-. 1:2009, ISO 10993-3:2014, ISO 10993-5:2009, ISO 10993-6:2007, ISO 10993-10:2010 and ISO 10993-11:2006 referenced in K193425
The results of the above tests have met the criteria of the standards and demonstrated the substantial equivalence with the predicate device.
The surface modification information such as surface roughness, surface composition analysis, and SEM imaging with SLA (Sandblasted with Large-grit and Acid-etching) for fixtures was provided.
For devices delivered sterile (ARUM NB 1 SA Fixture and Cover Screw) - a sterility assurance level (SAL) of 106 have been validated in accordance with ISO 11137-1:2006. Sterilization of health care products -Radiation - Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices.
Shelf Life Testing was performed in accordance with ASTM F1980, Standard Guide for Accelerated Aging of Sterile Medical Device Packages. The worst-case construct was tested, and results demonstrated equivalence to the predicate devices. The shelf life for devices provided sterile is 5 years. The devices will not be marketed as non-pyrogenic.
For all other subject devices delivered non-sterilized, the recommended sterilization has been validated according to ISO 17665-1 and ISO 17665-2 and to applicable recommendations in the FDA guidance document "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling, issued on March 17, 2015". The worst-case construct was tested, and results demonstrated equivalence to the predicate device.
Biocompatibility Testing was performed according to ISO 10993-1:2009. "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process," and to the FDA Guidance document, "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices- Part 1: Evaluation and testing within a risk management process', Guidance for Industry and Food and Drug Administration Staff", Document issued on: June 16, 2016", for each of the subject devices.
The non-clinical testing results demonstrate that the subject device is substantially equivalent to the predicate device.
{11}------------------------------------------------
# Conclusion
The documentation submitted in this premarket notification demonstrates the NB 1 SA Implant System is substantially equivalent to the primary predicate and reference devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.