Porcine Mineral Collagen Composite Moldable is indicated for: - Augmentation or reconstructive treatment of alveolar ridge - Filling of infrabony periodontal defects - Filling of defects after root resection, apicoectomy, and cystectomy - Filling of extraction sockets to enhance preservation of the alveolar ridge - Elevation of maxillary sinus floor - Filling of periodontal defects in conjunction with products intended for Guided Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR) - Filling of peri-implant defects in conjunction with products intended for Guided Bone Regeneration.
Device Story
Porcine Mineral Collagen Composite Moldable is an osteoconductive bone graft material for periodontal, oral, and maxillofacial surgery. Composed of 90% porcine anorganic bone mineral granules and 10% porcine Achilles tendon collagen in a block form. Supplied sterile, non-pyrogenic, single-use. Used by clinicians in dental/oral surgery settings to fill bone defects or augment ridges. Acts as a scaffold for bone regeneration. Clinician places the moldable block into the surgical site; material supports bone growth. Benefits include enhanced ridge preservation and defect repair. Substantial equivalence established via bench testing and canine intrabony defect model comparison against predicate devices.
Clinical Evidence
No clinical studies performed. Substantial equivalence supported by bench testing and animal studies. Animal testing used a canine one-wall intrabony defect model comparing subject device to Bio-Oss Collagen reference device. Endpoints included radiographic, Micro CT, histology, and histomorphometry at 4, 8, and 13 weeks, demonstrating equivalent performance.
Technological Characteristics
Composite of 90% anorganic porcine bone mineral and 10% porcine Achilles tendon collagen. Block form. Biocompatible per ISO 10993. Sterilized via gamma irradiation (ISO 11137). Mineral component conforms to ASTM F1581. Non-pyrogenic. Single-use.
Indications for Use
Indicated for alveolar ridge augmentation/reconstruction, filling of infrabony periodontal defects, root resection/apicoectomy/cystectomy defects, extraction sockets, maxillary sinus floor elevation, and peri-implant defects in conjunction with GTR/GBR products. No specific age or gender restrictions stated.
Regulatory Classification
Identification
Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.
Special Controls
*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
September 11, 2020,
Collagen Matrix, Inc. Gloria Zuclich Director, Regulatory Affairs 15 Thornton Road Oakland, New Jersey 07436
Re: K201859
Trade/Device Name: Porcine Mineral Collagen Composite Moldable Regulation Number: 21 CFR 872.3930 Regulation Name: Bone Grafting Material Regulatory Class: Class II Product Code: NPM Dated: July 1, 2020 Received: July 6, 2020
Dear Gloria Zuclich:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Srinivas "Nandu" Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K201859
Device Name
Porcine Mineral Collagen Composite Moldable
Indications for Use (Describe)
Porcine Mineral Collagen Composite Moldable is indicated for:
- · Augmentation or reconstructive treatment of alveolar ridge
- · Filling of infrabony periodontal defects
- · Filling of defects after root resection, apicoectomy, and cystectomy
- · Filling of extraction sockets to enhance preservation of the alveolar ridge
- · Elevation of maxillary sinus floor
- · Filling of periodontal defects in conjuncts intended for Guided Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR)
- · Filling of peri-implant defects in conjunction with products intended for Guided Bone Regeneration.
| Type of Use (Select one or both, as applicable) |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <label><input checked="" type="checkbox"/> Reproduction Use (Part 31 CFR 301, Subpart C)</label> <label><input type="checkbox"/> One-Time Screening Use (31 CFR 301, Subpart C)</label> |
X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
### 510(k) SUMMARY
| 1. | Applicant Information | |
|----|----------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|
| | Applicant Name:<br>Address:<br>Telephone:<br>Fax: | Collagen Matrix, Inc.<br>15 Thornton Road<br>Oakland, New Jersey 07436<br>(201) 405-1477<br>(201) 405-1355 |
| | Contact Person: | Gloria Zuclich<br>Director of Regulatory Affairs<br>gzuclich@collagenmatrix.com |
| | Date Prepared: | September 08, 2020 |
| 2. | Name of the Device | |
| | Device Trade Name: | Porcine Mineral Collagen Composite Moldable |
| | Device Common Name(s): | Bone Grafting Material |
| | Device Classification Name: | Bone Grafting Material, Animal Source<br>872.3930<br>NPM<br>Class II |
| 3. | Legally Marketed Devices to Which Substantial Equivalence is Claimed | |
| | Primary Predicate Device: | Porcine Anorganic Bone Mineral<br>Collagen Matrix, Inc.<br>K140714 |
| | Reference Device(s): | Bio-Oss® Collagen<br>Geistlich-Pharma AG<br>K033815 |
| | | Collagen Dental Membrane - Porcine Type I<br>Collagen<br>Collagen Matrix, Inc.<br>K110600 |
| | | Collagen Dental Wound Dressings<br>Collagen Matrix, Inc.<br>K122115 |
# 4. Description of the Device
Porcine Mineral Collagen Composite Moldable is an osteoconductive bone mineral with collagen composite for bone grafting in periodontal, oral and maxillofacial surgery. The device is
{5}------------------------------------------------
composed of 90% anorqanic bone mineral granules derived from porcine cancellous bone and 10% collagen from porcine Achilles tendon in a composite matrix. The product is supplied sterile, non-pyrogenic and for single use only.
Porcine Mineral Collagen Composite Moldable is provided in a block form and is available in three sizes, 0.5cc, 1.0cc, and 2.0cc.
### 5. Intended Use
Porcine Mineral Collagen Composite Moldable is indicated for:
- . Augmentation or reconstructive treatment of alveolar ridge
- Filling of infrabony periodontal defects
- Filling of defects after root resection, apicoectomy, and cystectomy
- . Filling of extraction sockets to enhance preservation of the alveolar ridge
- . Elevation of maxillary sinus floor
- Filling of periodontal defects in conjunction with products intended for Guided Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR)
- . Filling of peri-implant defects in conjunction with products intended for Guided Bone Regeneration.
## 6. Summary/Comparison of Technical Characteristics
The subject device and the predicate device have the same indications for use.
The subject device has substantially equivalent technological characteristics as the cited legally marketed predicate device. Differences include the physical form, material composition and product range volumes. Differences in the physical form and material composition have been determined to be minor and are substantiated when compared to that of the cited reference devices. The difference in product range volume offered for the subject device falls within the range supplied for the primary predicate device.
| Feature | Porcine Mineral Collagen<br>Composite Moldable<br>(This Submission) | Porcine Anorganic Bone<br>Mineral<br>(K140714) | Bio-Oss® Collagen<br>(K033815) |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | Augmentation or<br>reconstructive treatment<br>of alveolar ridge Filling of infrabony<br>periodontal defects Filling of defects after<br>root resection,<br>apicoectomy, and<br>cystectomy Filling of extraction<br>sockets to enhance<br>preservation of the<br>alveolar ridge Elevation of maxillary<br>sinus floor Filling of periodontal | Augmentation or<br>reconstructive treatment<br>of alveolar ridge Filling of infrabony<br>periodontal defects Filling of defects after<br>root resection,<br>apicoectomy, and<br>cystectomy Filling of extraction<br>sockets to enhance<br>preservation of the<br>alveolar ridge Elevation of maxillary<br>sinus floor Filling of periodontal | Augmentation or<br>reconstructive treatment<br>of alveolar ridge Filling of periodontal<br>defects Filling of defects after<br>root resection,<br>apicoectomy, and<br>cystectomy Filling of extraction<br>sockets to enhance<br>preservation of the<br>alveolar ridge Elevation of maxillary<br>sinus floor Filling of periodontal |
{6}------------------------------------------------
| | defects in conjunction | defects in conjunction | defects in conjunction |
|---------------------------|-------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|
| | | | |
| | with products intended | with products intended | with products intended |
| | for Guided Tissue | for Guided Tissue | for Guided Tissue |
| | Regeneration (GTR) and | Regeneration (GTR) and | Regeneration (GTR) and |
| | Guided Bone | Guided Bone | Guided Bone |
| | Regeneration (GBR) | Regeneration (GBR) | Regeneration (GBR). |
| | Filling of peri-implant defects<br>in conjunction with products<br>intended for Guided Bone<br>Regeneration (GBR) | Filling of peri-implant defects<br>in conjunction with products<br>intended for Guided Bone<br>Regeneration (GBR) | Filing of peri-implant defects<br>in conjunction with products<br>intended for Guided Bone<br>Regeneration (GBR) |
| Physical Form | Block Shaped | Granules | Block Shaped |
| Color | White to off-white | White to off-white | White to off-white |
| Material<br>Composition | • Anorganic (porcine) bone<br>mineral | • Anorganic (porcine) bone<br>mineral | • Anorganic (bovine) bone<br>mineral |
| | • Purified (porcine) collagen | | • Purified (porcine) collagen |
| Product Range<br>(volume) | 0.5 cc to 2.0 cc | 0.5 cc to 4.0 cc | 0.2 cc to 1.2 cc |
| | Biocompatible | Biocompatible | Biocompatible |
| Biocompatibility | ISO 10993 | ISO 10993 | ISO 10993 |
| | Sterile, SAL 10-6 | Sterile, SAL 10-6 | Sterile, SAL 10-6 |
| Sterility | Gamma irradiation | Gamma irradiation | Gamma irradiation |
| | ISO 11137 | ISO 11137 | |
| Pyrogenicity | Non-pyrogenic | Non-pyrogenic | Non-pyrogenic |
| Single Use/<br>Reuse | Single use only | Single use only | Single use only |
# 7. Performance Data
In vivo and in vitro testing of the subject device was conducted to demonstrate substantial equivalence of the subject device to its predicate devices. The following performance data are provided in support of the substantial equivalence determination.
# Biocompatibility Testing
A series of in vitro and in vivo biocompatibility testing was performed to assess the safety of the subject device. Testing was determined in accordance with ISO 10993-1 and FDA Guidance on Use of International Standard ISO 10993-1 for the biological evaluation of medical devices within a risk management process. The biocompatibility testing performed is summarized in the table below.
| Test | Test Method | Results |
|--------------|---------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Cytotoxicity | L929 MEM Elution Test, ISO<br>10993-5 | Non-cytotoxic |
| Genotoxicity | Mouse Lymphoma Assay, ISO<br>10993-3 | No evidence of causing increase<br>in the mean mutant frequency of<br>the L5178Y/TK+/- cell line either in<br>the presence or absence of<br>metabolic inactivation. The test<br>article was not mutagenic |
| | Ames Assay | Non-mutagenic to Salmonella<br>typhimurium and to Escherichia |
{7}------------------------------------------------
| Test | Test Method | Results |
|---------------------------------------------|---------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
| | | <i>coli</i> strain WP2uvra |
| Sensitization | Guinea Pig Maximization,<br>ISO 10993-10 | No evidence of causing delayed<br>dermal contact sensitization in the<br>guinea pig |
| Irritation Intracutaneous<br>Reactivity | ISO Intracutaneous Reactivity<br>in Rabbits, ISO 10993-10 | No evidence of irritation or toxicity |
| Acute Systemic<br>Toxicity | Acute Systemic Toxicity in<br>Mice, ISO 10993-3 | No mortality or evidence of<br>systemic toxicity |
| Pyrogenicity | USP (151) Rabbit Pyrogen<br>Study<br>USP <85> Bacterial Endotoxin<br>Test | Non-pyrogenic |
| Implantation | Implantation in Canine<br>Intrabony Defect Model, ISO<br>10993-6 | Minimum tissue reaction up to 13<br>weeks of implantation and no<br>adverse tissue reaction to the host |
| Subacute / Subchronic /<br>Chronic Toxicity | Implantation in Canine<br>Intrabony Defect Model, ISO<br>10993-11 | Minimum tissue reaction up to 13<br>weeks of implantation and no<br>adverse tissue reaction to the host |
# Bench Testing
In vitro product characterization testing was performed to demonstrate substantial equivalence of the subject device to its predicate devices. A series of bench tests were conducted to evaluate material properties, biological properties, chemical and physical properties as indicated. Testing of the anorganic bone mineral component was conducted in accordance with ASTM F1581 Standard Specifications for Composition of Anorganic Bone for Surgical Implants.
| Test | Results |
|----------------------------------------------|--------------------------------------------------|
| Mineral Content | Mineral content similar to predicate device |
| Size | Volumes similar to predicate device |
| Calcium to Phosphate<br>Ratio (mineral only) | Ratio similar to predicate device |
| Scanning Electron<br>Micrograph (SEM) | Morphologies similar to reference device |
| X-Ray Diffraction | Similar diffraction patterns to reference device |
| IR Spectroscopy | Similar functional groups to reference device |
| Density | Appropriate density for sufficient porosity |
| pH | pH similar to predicate device |
| Absorbency | Absorbency similar to predicate device |
| Pyrogenicity | Non-pyrogenic |
{8}------------------------------------------------
# Animal Testing
The performance of the device in a canine one-wall intrabony defect model was compared to the performance of the reference device. Bio-Oss Collagen. Radiographic, Micro CT, Histology and Histomorphometry analyses were conducted following implantation at 4, 8, and 13 weeks for the subiect device, reference device and sham negative control. The results demonstrate performance substantially equivalent to the reference device Bio-Oss Collagen when used as intended.
### Animal Tissue Management
Animal tissues are managed in accordance with the following standards and quidance documents:
- ISO 22442-1 Animal Tissues and Their Derivatives Utilized in the Manufacture of Medical . Devices - Part 1: Analysis and Risk Management
- . ISO 22442-2 Animal Tissues and Their Derivatives Utilized in the Manufacture of Medical Devices - Part 2: Controls on Sourcing, Collection, and Handling
- ISO 22442-3 Animal Tissues and Their Derivatives Utilized in the Manufacture of Medical . Devices – Part 3: Validation of the Elimination and/or Inactivation of Viruses and Transmissible Agents
- Medical Devices Containing Materials Derived from Animal Sources (Except for In Vitro . Diagnostic Devices) Guidance for Industry and Food and Drug Administration Staff, CDRH, FDA, March 15, 2019
- FDA Guidance for Industry Q5A Viral Safety Evaluation of Biotechnology Products . Derived from Cell Lines of Human or Animal Origin, CDER, September 1998
### Sterilization
Sterilization validation was performed in accordance with ISO 11137-1 Sterilization of health care products - Radiation.
### Shelf Life and Stability
Product and packaging stability was determined using real-time aging data. Performance testing of packaging system was tested in accordance with ASTM D4169 Standard Practice for Performance Testing of Shipping Containers and Systems.
### Viral Inactivation
Viral inactivation studies were performed in accordance with ISO 22442-3 to ensure the viral safety of the product.
### Clinical Studies
Clinical performance data was not required to determine substantial equivalence.
#### 8. Conclusions Drawn from Non-clinical Studies
The conclusions drawn from the nonclinical tests demonstrate that the device is substantially equivalent to its predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.