COLLAGEN DENTAL WOUND DRESSINGS
K122115 · Collagen Matrix, Inc. · KGN · Nov 1, 2012 · SU
Device Facts
| Record ID | K122115 |
| Device Name | COLLAGEN DENTAL WOUND DRESSINGS |
| Applicant | Collagen Matrix, Inc. |
| Product Code | KGN · SU |
| Decision Date | Nov 1, 2012 |
| Decision | SESE |
| Submission Type | Traditional |
| Device Class | Class U |
| Attributes | Therapeutic |
Indications for Use
Collagen Dental Wound Dressings are intended for the management of oral wounds and sores, including: Denture sores . Oral ulcers (non-infected or viral) ● Periodontal surgical wounds ● Suture sites . - . Burns - Extraction sites - Surgical wounds - Traumatic wounds
Device Story
Collagen Dental Wound Dressings consist of absorbent, porous matrices derived from purified porcine Achilles tendon type I collagen. Applied directly to oral wounds, the dressing protects the wound site and supports delicate new tissue. Used in clinical settings by dental professionals for management of various oral lesions and surgical sites. Device is supplied sterile, non-pyrogenic, and for single use. Acts as a physical barrier to protect the wound environment.
Clinical Evidence
No clinical tests were performed on the device. Safety and effectiveness were established through in vitro characterization studies (material, biological, chemical, and physical properties) and in vivo biocompatibility testing. Clinical history of predicate devices was referenced.
Technological Characteristics
Absorbent, porous matrix composed of purified type I collagen from porcine Achilles tendon. Supplied sterile and non-pyrogenic. Single-use form factor. Biocompatibility verified per ISO 10993-1 and FDA Blue Book Memorandum G95-1.
Indications for Use
Indicated for management of oral wounds and sores including denture sores, non-infected or viral oral ulcers, periodontal surgical wounds, suture sites, burns, extraction sites, surgical wounds, and traumatic wounds in patients requiring oral wound care.
Predicate Devices
- Collacare Dental, Collagen Dental Matrix (K110388)
- Collagen Wound Dressing - Oral (K040403)
Submission Summary (Full Text)
{1}------------------------------------------------
# K122115
1 2012
NOV
## 510(k) SUMMARY
### 1. Applicant Information
Applicant Name: Address:
Telephone: Fax: Contact Person:
Date Prepared:
Collagen Matrix, Inc. 15 Thornton Road Oakland, New Jersey 07436 (201) 405-1477 (201) 405-1355 Peggy Hansen, RAC VP, Clinical, Regulatory, QA, and Marketing October 4, 2012
#### 2. Name of the Device
Device Common Name: Device Trade Name: Device Classification Name:
Collagen Wound Dressing Collagen Dental Wound Dressings Dressing, Wound, Collagen KGN Unclassified
#### Legally Marketed Devices to Which Substantial Equivalence is Claimed 3.
Predicate Device(s):
Collacare Dental, Collagen Dental Matrix K110388
Collagen Wound Dressing - Oral K040403
#### 4. Description of the Device
Collagen Dental Wound Dressings are absorbent, porous, collagen matrices engineered from highly purified type I collagen derived from porcine Achilles tendon. The Collagen Dental Wound Dressings are applied directly to the wound and protect the wound and delicate new tissue. Collagen Dental Wound Dressings are supplied sterile, non-pyrogenic, in various sizes, and for single use only.
#### Intended Use 5.
Collagen Dental Wound Dressings are intended for the management of oral wounds and sores, including:
- Denture sores .
- Oral ulcers (non-infected or viral) ●
- Periodontal surgical wounds ●
- Suture sites .
{2}------------------------------------------------
- . Burns
- Extraction sites
- Surgical wounds
- Traumatic wounds
### 6. Summary/Comparison of Technical Characteristics
Collagen Dental Wound Dressings and their predicates have similar technological characteristics. In particular, the Collagen Dental Wound Dressings and the predicates are similar with respect to intended use, purified starting material (type I collagen), form, sizes, physical integrity. The substantial equivalence of Collagen Dental Wound Dressings and their predicates was demonstrated based on in vitro characterization studies, biocompatibility studies, and clinical history of the predicate devices.
In vitro characterization studies included evaluation of material properties, biological properties, chemical and physical properties.
Collagen Dental Wound Dressing representative products have been evaluated in a number of in vitro and in vivo tests to assess its safety/biocompatibility. The products passed all applicable FDA Blue Book Memorandum G95-1and ISO 10993-1 testing for the biological evaluation of medical devices. No clinical tests were performed on the product, however clinical history of the predicate device was referenced in the submission.
Viral inactivation studies were performed to ensure the viral safety of the product.
#### Conclusion of Non-clinical Studies
The results of the in vitro product characterization studies as well as in vitro and in vivo biocompatibility studies show that Collagen Dental Wound Dressings are safe and substantially equivalent to the predicate devices.
{3}------------------------------------------------
Image /page/2/Picture/0 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird-like figure with three wing-like strokes above a wavy line, representing water or movement. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird-like figure.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
NOV
Collagen Matrix, Incorporated Ms. Peggy Hansen, RAC Vice President, Clinical, Regulatory, Quality Assurance, & Marketing 15 Thornton Road Oakland, New Jersey 07436
Re: K122115
Trade/Device Name: Collagen Dental Wound Dressings Regulation Number: Unclassified Regulation Name: Dressing, Wound, Collagen Regulatory Class: Unclassified Product Code: KGN Dated: October 4, 2012 Received: October 5, 2012
Dear Ms. Hansen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA) You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
1 2012
{4}------------------------------------------------
## Page 2- Ms. Hansen .
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known):
Collagen Dental Wound Dressings Device Name:
Indications for Use:
Collagen Dental Wound Dressings are intended for the management of oral wounds and sores, including:
- Denture sores �
- Oral ulcers (non-infected or viral) .
- Periodontal surgical wounds .
- Suture sites .
- Burns ●
- Extraction sites .
- Surgical wounds .
- Traumatic wounds .
Prescription Use _X (Part 21 CFR 801 Subpart D)
### AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Susan Runpe
(Division Sign-Off) (Division Sign-Off)
Division of Anesthesi-Jogy, General Hos;الان
Division of Anesthesi-Jogy, General Devices Division of Anesthesishogy - Devicess
Infection Control, Dental Devices
Infection Control, Dental Devices
Page 1 of _1
**510(k) Number:**