TRIPLEPLAY-VT VASCULAR THERAPY SYSTEM

K103187 · Wildcat Medical, Inc. · JOW · Jan 26, 2011 · Cardiovascular

Device Facts

Record IDK103187
Device NameTRIPLEPLAY-VT VASCULAR THERAPY SYSTEM
ApplicantWildcat Medical, Inc.
Product CodeJOW · Cardiovascular
Decision DateJan 26, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.5800
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The TriplePlay-VT, model TPVT-01 (supplied with a pair of model TP-3333 "cuffs") is intended to be an easy to use portable system, prescribed by a physician, to help prevent the onset of DVT in patients by stimulating blood flow in the extremities (simulating muscle contractions). This device can be used to: - Aid in the prevention of DVT; - Enhance blood circulation; - Diminish post-operative pain and swelling; - Reduce wound healing time; - Aid in the treatment and healing of: stasis dermatitis, venous stasis ulcers, arterial and diabetic leg ulcers, chronic venous insufficiency and reduction of edema in the lower limbs. The unit can also be used as an aid in the prophylaxis for DVT by persons expecting to be stationary for long periods of time.

Device Story

Portable, battery-powered pneumatic compression system; uses electronically controlled air pump and solenoid valves to inflate single-chamber PVC cuffs. Input: user-selected output modes (leg 1, leg 2, auxiliary). Operation: pump fills cuffs to preset pressure (50 mmHg leg, 20-50 mmHg auxiliary); integrated pressure transducer monitors inflation; system software manages 60-second cycle (inflation followed by rest/deflation). Used in clinical or home settings under physician prescription. Output: intermittent pneumatic compression to stimulate blood flow. Healthcare providers use device to prevent DVT, reduce edema, and aid wound healing. Benefits: portable prophylaxis for stationary patients; aids post-operative recovery.

Clinical Evidence

No clinical testing performed on subject device. Substantial equivalence supported by literature review of predicate devices (Orthopedics 2001; Journal of Bone and Joint Surgery 1998). Studies confirm intermittent pneumatic compression efficacy in reducing postoperative DVT risk. Predicate (Venodyne) mean blood flow velocity (76.2 +/- 23.77) achieved at 43 mmHg, 11.6s fill time, and 60s cycle time, which are nearly identical to subject device default parameters.

Technological Characteristics

Microprocessor-controlled pneumatic pump; single-chamber PVC air bladders; non-woven medical fabric (sponge material) cuff covers. Powered by internal rechargeable battery or AC adapter. Connectivity: quick-disconnect air ports. Safety: integrated pressure transducer, high/low pressure alarms, low battery alarm, internal safety vent (100 mmHg). Dimensions/form factor: portable plastic case. Sterilization: supplied clean, non-sterile.

Indications for Use

Indicated for patients requiring DVT prophylaxis or treatment of venous/arterial/diabetic ulcers, stasis dermatitis, chronic venous insufficiency, and limb edema. Contraindicated in patients with suspected/active DVT, ischemic vascular disease, severe arteriosclerosis, pulmonary edema, congestive heart failure, thrombophlebitis, active infection, neuropathy, insensitive extremities, or where increased venous/lymphatic return is undesirable. Also contraindicated for use on legs with vein ligation, gangrene, dermatitis, open wounds, recent skin grafts, massive edema, or extreme deformity.

Regulatory Classification

Identification

A compressible limb sleeve is a device that is used to prevent pooling of blood in a limb by inflating periodically a sleeve around the limb.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # Wildcat Medical, Inc. 1638 South Main Street Tulsa, OK 74119 Telephone: 800-994-0464 JAN 2 6 2011 Section 5: 510(k) Summary Date Prepared: October 26, 2010 510(k) Number: K | 03 | 8 7 ## Sponsor Wildcat Medical, Inc. 1638 South Main Street Tulsa, OK 74119 Phone: 800-994-0464 Fax: 918-747-1078 ## Contact Person Joe Adkins, Senior Project Engineer Phone: 216-849-7847 Email: safeguardmd(@aol.com ## Device Name Trade Name: TriplePlay-VT Vascular Therapy System (including leg wrap cuffs) Model Number: TPVT-01 (with accessory cuffs model TP-3333) Common or Usual Name: Compressible Limb Sleeve Device Classification Name: Compressible Limb Sleeve Device Class II Product Code: JOW Regulation Number: 21 CFR 870.5800 > Model TPVT-01 Compressible Limb Sleeve Device 510(k) Submission for Wildcat Medical, Inc. October 26, 2010 {1}------------------------------------------------ ## Identification of Predicate Devices | ThermoTek, Inc., NanoTherm and VascuTherm | K061866 | |-------------------------------------------------------------|---------| | Doctor's Orders, Inc., DVTCare CA5 | K061125 | | Medical Compression Systems, LTD, WizAir DVT | K023573 | | Dynamic Air, Inc., TravelAir Portable Compression<br>System | K022340 | | Microtek Medical, Inc., Venodyne DVT Advantage<br>Plus | K011318 | ## Device Description The TriplePlay-VT Vascular Therapy System is a lightweight, portable, rechargeable battery powered prescriptive device that is intended to be used by or under the direction of a medical professional to help stimulate blood flow as an aid in the prevention of deep vein thrombosis (DVT). The system utilizes pneumatically controlled, single chamber cuffs actuated by an electronically controlled air pump unit and solenoid valves. All pump, battery and control components are protectively housed in a plastic case. A mylar control panel overlays 4 tactile touch control switches, an LED display for monitoring pressure, LED "selected function", "low battery" and "charging" indicators. There is also a port for connecting the battery charger/AC adapter plug. The leg wrap (cuff), model TP-3333, consists of a Polyvinyl Chloride (PVC) air bladder encased inside a soft, non-woven medical fabric made from common kitchen sponge material, which is adhered to the PVC air bladder. The cuffs are supplied clean, non-sterile, packaged in pairs. One set of cuffs are supplied with the TPVT-01 system, and are also available as accessory items. In operation, the user selects either 1 of 5 different output modes (leg 1, leg 2, auxiliary, or a combination of leg 1 + auxiliary or leg 1 + leg 2 + auxiliary). An LED corresponding to each output indicates the selected modes. Appropriate single user "cuffs" containing air bladders are connected to the unit via externally accessible plastic quick-disconnect air ports. The control unit then fills the cuffs to a pre-determined pressure (50 mmHg for the leg cuffs, with the auxiliary cuff setting being adjustable between 20 and 50 mmHg - default setting of 35 mmHg is preset for the auxiliary function). The LED that corresponds to the activated cuff flashes slowly while that cuff is being filled with air. Cuff pressure is monitored by an integrated pressure transducer and system software. Once the pressure reaches the proper level, the pump is turned OFF for a predetermined "rest" period, and the cuff deflates to ambient pressure through a valve inside the plastic case. After the "rest" period, the next cuff is sequenced, and so on. This cycle repeats until the unit is turned off. > Model TPVT-01 Compressible Limb Sleeve Device 510(k) Submission for Wildcat Medical, Inc. October 26, 2010 {2}------------------------------------------------ The "rest" period is internally preset to allow each cuff approximately 60 seconds between inflations in order to insure adequate relaxation time between compressions. When multiple cuff modes are selected, only one cuff is inflated at a time. ## Intended Use The TriplePlay-VT, model TPVT-01 is intended to be an easy to use portable system, prescribed by a physician, to help prevent the onset of DVT in patients by stimulating blood flow in the extremities (simulating muscle contractions). This device can be used to: - Aid in the prevention of DVT; . - . Enhance blood circulation; - . Diminish post-operative pain and swelling; - Reduce wound healing time; . - Aid in the treatment of: stasis dermatitis, venous stasis ulcers, arterial and diabetic leg . ulcers, chronic venous insufficiency and reduction of edema in the lower limbs; - As a prophylaxis for DVT by persons expecting to be stationary for long periods of time. . ## Contraindications The TriplePlay-VT, model TPVT-01 must not be used to treat the following conditions: - Persons with suspected, active or untreated: deep vein thrombosis, ischemic vascular . disease, severe arteriosclerosis, pulmonary edema, congestive heart failure, thrombophlebitis or an active infection; - . On a leg where cuffs would interfere with the following conditions: vein ligation, gangrene, dermatitis, open wounds, a recent skin graft, massive edema or extreme deformity of the leg; - . On patients with neuropathy: - On extremities that are insensitive to pain; . - Where increased venous or lymphatic return is undesirable; . #### Comparison of Indications The Indications for Use for the TriplePlay-VT Vascular Therapy system are the same as those for the predicate devices listed on the previous page. ## Substantial Equivalence of Technological Characteristics The TriplePlay-VT Vascular Therapy system is equivalent to the predicate devices listed in function and operating principals to achieve identical results. All systems (VascuTherm-K061866, DVTCare CA5-K061125, WizAir DVT-K023573, TravelAir-K022340 and Venodyne DVT Advantage-K011318) utilize microprocessor controlled pumps to deliver approximately 50 mmHg of pressurized air to bladders that are attached to the patient's lower limbs, using a cycle {3}------------------------------------------------ time of approximately 60 seconds / leg. Each cycle consists of inflation of a bladder, followed by a rest period during which the bladder deflates and the limb relaxes without any compression. If more than one wrap is being used, the additional bladders are cycled during the rest period of the first bladder, keeping a constant cycle time of approximately 60 seconds for each. Multiple audible and visual safety alarms are built into the system, similar to those built into the VascuTherm and DVTCare CA5, including; High and low pressure alarms, low battery alarm and system malfunction overpressure safety via an internal safety vent switch with a release pressure of 2 psi (approximately 100 mmHg). Cycle and maximum fill times are factory preset and cannot be changed. The default settings are similar to predicate devices in fill time, cycle time and pressure settings. Like the DVTCare CA5, the TriplePlay-VT includes provisions for an auxiliary third wrap to be used by physicians as may be needed in patient therapy. This third AUX outlet provides the same intermittent pneumatic compression as does the Leg 2 outlets. Both units allow the pressure setting for the auxiliary wrap to be adjusted from 20 mmHg to 50 mmHg (default setting 35 mmHg) for maximum patient comfort. The default pressure setting for the leg wraps is factory preset at approximately 50 mmHg for both units. This default setting cannot be adjusted on the TriplePlay-VT system. The TriplePlay-VT system uses similar means for pressure delivery to the cuffs as the predicate devices. Pressurized air is delivered by the pump to the cuffs via flexible, plastic air tubes connected to the plastic pump / control unit by means of locking, quick disconnect couplings. Like the DVTCare CA5 and VascuTherm, the TriplePlay-VT cuffs, model TP-3333, are comprised of single bladder PVC chambers encased in a covering of soft, non-latex, nonwoven medical fabric made from common kitchen sponge material for increased patient comfort and biocompatibility compliance. As with the WizAir DVT and TravelAir systems, the microprocessor and pump units are powered by internal rechargeable batteries, and can be connected to the main AC power line (through the battery charger / AC adaptor) while in use, allowing uninterrupted prolonged service. ## Non-Clinical Testing Non-clinical validation, including electrical safety, EMC, mechanical integrity, environmental and life cycle testing have shown that the TriplePlay-VT Vascular Therapy System has performance characteristics substantially equivalent to or surpassing those of the listed predicate devices. In-house bench testing has verified equivalent pressure delivery, cuff (bladder) fill time, cycle time and overall system operation as the predicate devices listed. {4}------------------------------------------------ ## Clinical Testing No clinical testing was performed on the TriplePlay-VT Vascular Therapy System, however test results of some predicate devices have been compared in the following published clinical studies: - Evaluation of Intermittant Pneumatic Compression Devices (Orthopedics 24(3):257-261, . 2001); - . Venous hemodynamics after total knee arthroplasty: Evaluation of active dorsal to planar flexion and several mechanical compression devices (Journal of Bone and Joint Surgery, November 1998) The summary conclusions of both studies state that the use of intermittent pneumatic compression devices is useful in decreasing the risk of postoperative DVT. The Venodyne model used in the studies produced a mean blood flow velocity of 76.2 +/- 23.77 using an average inflation pressure of 43 mmHg, a fill time of 11.6 seconds and a total cycle time of 60 seconds. In that these values are nearly identical to the corresponding default parameters used by the TriplePlay-VT Vascular Therapy System, and the methods of pressure delivery and operation are identical in both systems, substantial equivalence is concluded. ## Summary Conclusion Per the requirements of 21 CFR 807, surrogate clinical data, non-clinical validation testing and the information provided in the accompanying 510(k) submission, Wildcat Medical, Inc. concludes that the TriplePlay-VT model TPVT-01 (with accessory cuffs model TP-3333) is safe, effective and performs in a manner that is substantially equivalent to the predicate devices listed. {5}------------------------------------------------ Image /page/5/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" written around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Wildcat Medical, Inc. c/o Joe Adkins, Sr. Project Engineer Safeguard Manufacturing & Development Co. 284 Ridge Road Hinckley, Ohio 44233 JAN 2 6 201 Re: K103187 TriplePlay - VT (TPVT-01) and leg cuff (TP-3333) Regulation Number: 21 CFR 870.5800 Regulation Name: Compressible Limb Sleeve Regulatory Class: Class II Product Code: JOW Dated: October 26, 2010 Received: October 28, 2010 Dear Mr. Adkins: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. {6}------------------------------------------------ Page 2 – Mr. Joe Adkins If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that vour device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours Sincerely yours, Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {7}------------------------------------------------ #### Section 4: Indications for Use Statement ## 510(k) # K 103187 ## Indications for Use: The TriplePlay-VT, model TPVT-01 (supplied with a pair of model TP-3333 "cuffs") is intended to be an easy to use portable system, prescribed by a physician, to help prevent the onset of DVT in patients by stimulating blood flow in the extremities (simulating muscle contractions). This device can be used to: - Aid in the prevention of DVT; . - Enhance blood circulation; . - Diminish post-operative pain and swelling; . - . Reduce wound healing time; - Aid in the treatment and healing of: stasis dermatitis, venous stasis ulcers, arterial and . diabetic leg ulcers, chronic venous insufficiency and reduction of edema in the lower limbs. The unit can also be used as an aid in the prophylaxis for DVT by persons expecting to be stationary for long periods of time. Prescription Use: X (Part 21 CFR 801 Subpart D) AND/OR ## Over-The-Counter Use: (Part 21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) W. Wood iovascular Devices 510(k) Number K103687 Model TPVT-01 Compressible Limb Sleeve Device 510(k) Submission for Wildcat Medical, Inc. October 26, 2010
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...