The Compression Care Flight is intended to be an easy to use, portable system that is prescribed by healthcare professionals, to help prevent the onset of DVT in patients, by stimulating blood flow in the legs (simulating muscle contractions). Furthermore, the unit can be used as an aid in the prophylaxis for DVT by persons traveling, or those expecting to be stationary for long periods of time (greater than 4 hours). This device can also be used to: aid in the prevention of DVT, enhance blood circulation, diminish post-operative pain and swelling, reduce wound healing time, and aid in the treatment and healing of: stasis dermatitis, venous stasis ulcers, arterial and diabetic leg ulcers, chronic venous insufficiency, and reduction of edema in the lower limbs.
Device Story
Portable pneumatic compression system; stimulates blood flow in legs to prevent DVT and treat venous/arterial conditions. System comprises microprocessor-controlled pump unit, solenoid valves, and single-chamber PVC/Nylon leg cuffs. Operates via pneumatic inflation/deflation cycles (approx. 60-75 seconds/leg; 10-12s inflation) to simulate muscle contractions. Powered by internal rechargeable battery or mains. Healthcare provider adjusts pressure (up to 65 mmHg) and cycle time. Used in clinical or home settings; patient-applied. Output is mechanical compression of limbs. Benefits include reduced DVT risk, pain/swelling reduction, and accelerated wound healing.
Clinical Evidence
Bench testing only. Validation included electrical safety (IEC 60601-1) and performance testing (pressure delivery, cuff fill time, cycle time) compared to predicate device.
Technological Characteristics
Microprocessor-controlled pneumatic pump; single-chamber PVC bladder cuffs with Nylon casing. Pressure up to 65 mmHg; cycle time 60-75 seconds. Powered by internal rechargeable battery or mains. Quick-disconnect air couplings. Electrical safety per IEC 60601-1.
Indications for Use
Indicated for patients requiring DVT prophylaxis, enhanced circulation, or treatment of venous/arterial/diabetic ulcers, stasis dermatitis, and lower limb edema. Suitable for travelers or stationary individuals (stationary >4 hours). Prescribed by healthcare professionals.
Regulatory Classification
Identification
A compressible limb sleeve is a device that is used to prevent pooling of blood in a limb by inflating periodically a sleeve around the limb.
K103187 — TRIPLEPLAY-VT VASCULAR THERAPY SYSTEM · Wildcat Medical, Inc. · Jan 26, 2011
K180389 — Venera 508 Deep Vein Thrombosis (DVT) Prevention System · Suzhou Minhua Medical Apparatus Supplies Co., Ltd. · Jun 29, 2018
K203310 — SyneCare 1000 Deep Vein Thrombosis Prevention Therapy System · Jiangsu Synecoun Medical Technology Co., Ltd. · Apr 2, 2021
K130174 — DVTCARE CA5 · Ossur Americas, Inc. · May 2, 2013
K151158 — ALP VasQcare Take Home System · Currie Medical Specialties, Inc. · Dec 10, 2015
Submission Summary (Full Text)
{0}------------------------------------------------
# Section 2 Summary
### JUL 2 4 2012
Date:
Submitter:
Contact Person:
510(k) Summary of Safety and Effectiveness
April 19, 2011 Medical Minds LLC
11491 78th. Ave Allendale, MI 49401
David Platt Vice President of Quality and Regulatory Medical Minds LLC 616-308-2859
Device: Trade Name: Common/Usual Name:
Compression Care Flight
Compression Limb Sleeve Device
Compression Limb Sleeve
Classification Names:
CFR Reference: 21CFR870.5800
Classification Name: Sieeve, Limb, Compressible
> Product Code: JOW
Predicate Device:
K061125 - DVTcare™ CA5
This predicate device submission was made by Doctor's Orders, Inc.
The Compression Care Flight system is a light weight, portable, prescriptive device that helps stimulate blood flow in the legs through the use of pneumatically controlled, single chamber leg cuff(s), actuated by an electronically controlled pump unit and solenoid valves. All pump control unit components are protectively housed in a plastic shell except the outer membrane switch (needed for user interface), 2 plastic quick disconnects for air tube connection, and an external power supply input jack. The option exists for the unit to be used in single leg or double leg modes.
The control unit is supplied with a non-serviceable, rechargeable battery, to allow user portability, and a power supply transformer for mains connection.
The Compression Care Flight is a pneumatic compression device intended to be an easy to use, portable system that is prescribed by healthcare professionals, to help prevent the onset of DVT in patients, by stimulating blood flow in the legs (simulating muscle contractions). Furthermore, the unit can be used as an aid in the prophylaxis for DVT by persons traveling, or those expecting to be stationary for long periods of time (greater than 4 hours). This device can also be used to: aid in the prevention of DVT, enhance blood circulation, diminish post-operative pain and swelling, reduce wound healing time, and aid in the treatment and healing of: stasis dermatitis, venous stasis ulcers, arterial and diabetic leg ulcers, chronic venous and arterial insufficiency, and reduction of edema in the lower limbs.
Device Description:
Intended Use:
5
{1}------------------------------------------------
## K121376 2092
#### Substantial Equivalence of Technological Characteristics:
The Compression Care Flight is similar to the predicate device listed above in the function and operating principles to achieve identical results. The Compression Care Flight offers greater treatment flexibility due to the portability, adjustable pressure, and cycle time capabilities. Our system uses a microprocessor controlled pump to deliver pressurized air (programmable up to 65 mmHg) to bladders that are attached to the patient's lower limbs, using a cycle of approximately 60 seconds / leg as prescribed by a healthcare provider. Each cycle consists of one inflation (fill time approximately 10-12 seconds) followed by deflation. The Compression Care Flight has the capability of allowing a healthcare professional the following feature adjustment options: pressure set point (programmable up to 65 mmHg), hold time (0 seconds, not adjustable), cycle time (60-75 seconds). The cycle time is the length of time for one complete cycle on one leg including fill time, exhaust, and idle time (relaxation).
The Compression Care Flight uses similar means for pressure delivery as the predicate device. Pressurized air is delivered by the pump to the leg cuffs via flexible, plastic air tubes connected to the plastic pump / control unit by locking, quick disconnect couplings. Like the DVT Care CA5, the Compression Care Flight leg cuffs are comprised of a single bladder PVC chambers encased in a soft Nylon material to increase patient comfort and compliance.
Like the DVT Care CAS, the microprocessor and pump units are powered by an internal battery supply and can be connected to the mains power supply for operation and recharging.
#### Non-Clinical Testing
Non-clinical validation including electrical safety and performance testing have shown that the Compression Care Flight has performance characteristics substantially equivalent to or superseding the listed predicate device. The Compression Care Flight has been validated by Medical Minds at a design validation level based on the requirements of IEC 60601-1. Additional bench testing has verified equivalent pressure delivery, cuff (garment) fill time, cycle time and system operation as the predicate device listed. Test reports can be found in Appendix 1.
#### Conclusion:
The Compression Care Flight is designed for the same intended use as the predicate device, the DVT Care CA5. This comparison of the specifications demonstrates the functional equivalence of the products. The differences discussed in this section do not raise new issues of safety and effectiveness. Verification and validation testing demonstrated that no adverse effects have been introduced by these differences.
Medical Minds LLC believes that the Compression Care Flight is as safe and effective and performs substantially equivalent to the predicate device.
6
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the bird symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
JUL 2 4 2012
Medical Minds LLC c/o Mr. David Platt Vice President of Quality & Regulation 11491 78th Avenue Allendale, MI 49401
Re: K121376
Trade/Device Name: Compression Care Flight Regulation Number: 21 CFR 870.5800 Regulation Name: Compressible Limb Sleeve Regulatory Class: Class II Product Code: JOW Dated: April 12, 2012 Received: May 8, 2012
#### Dear Mr. Platt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be
{3}------------------------------------------------
Page 2 - Mr. David Platt
found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may 1 publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
M.S. Hilleheme
? Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# K121376
## Section 11 Indications for Use
# Indications for Use
510(k) Number (if known): 《[2137L
Device Name: Compression Care Flight
Indications For Use:
The Compression Care Flight is intended to be an easy to use, portable system that is prescribed by healthcare professionals, to help prevent the onset of DVT in patients, by stimulating blood flow in the legs (simulating muscle contractions). Furthermore, the unit can be used as an aid in the prophylaxis for DVT by persons traveling, or those expecting to be stationary for long periods of time (greater than 4 hours). This device can also be used to: aid in the prevention of DVT, enhance blood circulation, diminish post-operative pain and swelling, reduce wound healing time, and aid in the treatment and healing of: stasis dermatitis, venous stasis ulcers, arterial and diabetic leg ulcers, chronic venous insufficiency, and reduction of edema in the lower limbs.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) | |
|------------------------------------|--|
| Division of Cardiovascular Devices | |
| 510(k) Number | K121376 |
|---------------|---------|
|---------------|---------|
31
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.