WIZAIR COMPRESSION SYSTEM

K023573 · Medical Compression Systems (Dbn) , Ltd. · JOW · Nov 21, 2002 · Cardiovascular

Device Facts

Record IDK023573
Device NameWIZAIR COMPRESSION SYSTEM
ApplicantMedical Compression Systems (Dbn) , Ltd.
Product CodeJOW · Cardiovascular
Decision DateNov 21, 2002
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.5800
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The WizAir DVT™ is a prescriptive device that induces Continuous Enhanced Circulation Therapy of the lower limbs. The WizAir DVT™ is intended for use in: - Preventing Deep Vein Thrombosis (DVT). - Enhancing blood circulation. - Diminishing post-operative pain and swelling. - Reducing wound-healing time. - Treatment and assistance in healing: stasis dermatitis; venous stasis ulcers; arterial and diabetic leg ulcers. - Treatment of chronic venous insufficiency. - Reducing edema.

Device Story

WizAir DVT is a portable, prescriptive pneumatic compression system for lower limbs. Device consists of a control unit and inflatable cuffs (foot, calf, or thigh). Inputs: user-selected treatment program. Operation: control unit delivers sequential intermittent pressure pulses (foot) or sequential intermittent gradient pressure (calf/thigh) to cuffs. Output: mechanical compression of limbs to enhance circulation. Used in clinical or home settings; operated by patient or clinician. Benefits: DVT prevention, edema reduction, and wound healing. Healthcare providers use the device to manage venous insufficiency and post-operative recovery. System supports battery or mains power.

Clinical Evidence

Clinical validation performed on healthy volunteers to compare the Foot Booster cuff performance against the predicate device. Bench testing also conducted to verify specifications. No specific clinical trial data (e.g., patient outcomes, p-values) provided in the summary.

Technological Characteristics

Pneumatic compression system; portable control unit with battery/mains power. Cuffs: single-celled (foot) or three-celled (calf/thigh). Operation: sequential intermittent pressure pulses or gradient application. Software-controlled inflation cycles. No specific materials or ASTM standards listed.

Indications for Use

Indicated for patients requiring continuous enhanced circulation therapy of the lower limbs to prevent DVT, enhance circulation, reduce pain/swelling/edema, and assist in healing stasis dermatitis, venous stasis ulcers, and arterial/diabetic leg ulcers. Contraindicated in patients with fresh pre-existing DVT, pulmonary embolism, leg gangrene, recent skin graft, acute thrombophlebitis, or conditions where increased venous/lymphatic return is undesirable.

Regulatory Classification

Identification

A compressible limb sleeve is a device that is used to prevent pooling of blood in a limb by inflating periodically a sleeve around the limb.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 1023573 # NOV 2 1 2002 # 510(K) SUMMARY # WizAir DVT™M # 510(k) Number K_______________________________________________________________________________________________________________________________________________________________ ## Applicant's Name: Medical Compression Systems (DBN) Ltd. 2 Ha-Ilan Street, P.O. Box 75 Or-Akiva 30600, Israel Tel: +972 (4) 6266630 Fax: +972 (4) 6266640 mcs@mcsmed.com # Contact Person: Arava Hacohen Push-med Ltd. 117 Ahuzah St. Ra'anana 43373, Israel Tel: 972-9- 7718130 Fax: 972-9-7718131 #### Trade Name: # WizAir DVT™ #### Classification Name: Compressible Limb Sleeve #### Classification: Class II; Product Code JOW; Regulation No. 870.5800. {1}------------------------------------------------ # Device Description: The WizAir DVT™ is a prescriptive, pneumatic compression device designed to apply sequential compression to the lower limb. The control unit of the WizAir DVT™ is light and compact, thus making it a portable ambulant system. The WizAir DVT™ provides the user with an option of battery operation in addition to the operation from the mains option. The WizAir DVT™ is easy to use and provides the user with several treatment options: compression of the foot -- single or double (either regular foot or foot booster), compression of the calf - single or double, compression of the Thigh -- single or double, and combined compression of any combination of two cuffs. The foot compression program is sequential intermittent pressure pulse application to a single celled foot cuff. The calf and thigh compression program is a sequential intermittent gradient application of a pressure to a three-celled calf cuff. ## Indications: The WizAir DVT™ is a prescriptive device that induces Continuous Enhanced Circulation Therapy of the lower limbs. The WizAir DVT™ is intended for use in: - Preventing Deep Vein Thrombosis (DVT). - Enhancing blood circulation. - Diminishing post-operative pain and swelling. - Reducing wound-healing time. . - Treatment and assistance in healing: stasis dermatitis; ● venous stasis ulcers; arterial and diabetic leg ulcers. - Treatment of chronic venous insufficiency. - Reducing edema. # Contraindications: The WizAir™ system should not be used in the following cases: fresh pre-existing DVT, pulmonary embolism, leg gangrene, recent skin graft, acute thrombophlebitis and in medical situations where increased venous and lymphatic return is undesirable {2}------------------------------------------------ #### Substantial Equivalence: Statement The modified WizAir DVT™ System is substantially equivalent in all aspects. e.g., technological characteristics, modes of operation, performance characteristics, intended use, etc., to the commercially available WizAir DVT™. The modified WizAir DVT™ System includes an additional cuff -- the Foot Booster cuff for faster inflation of the cell. The pressure profile of the new cuff is similar to that of the A-V Impulse System. Moreover, the additional cuff and software modification were verified through bench testing and validated through clinical analysis that was performed on healthy volunteers. Testing results showed that the Foot Booster performs according to its specifications in safe and effective manner. #### Performance Data: A series of safety and performance testing including bench testing and clinical comparison between the Foot Booster cuff of the WizAir DVT™ and its predicate device demonstrated that the modified WizAir DVT™ System is substantially equivalent to its predicate devices without raising new safety and/or effectiveness issues {3}------------------------------------------------ Image /page/3/Picture/2 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized design of three human profiles facing to the right, stacked on top of each other. The words "HUMAN SERVICES - USA" are arranged in a circular fashion around the top of the design, and the word "DEPARTMENT" is written vertically along the left side of the design. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 NOV 2 1 2002 Medical Compression Systems (D.B.N.) Ltd. c/o Mr. Arava Hacohen Project Manager Push-med Ltd. 117 Ahuzah St. Ra'ananna 43373 ISRAEL Re: K023573 Trade Name: WizAir DVT™ Regulation Number: 21 CFR 870.5800 Regulation Name: Sleeve, Limb, Compressible Regulatory Class: Class II (two) Product Code: JOW Dated: October 21, 2002 Received: October 24, 2002 Dear Mr. Hacohen: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Page 2 – Mr. Arava Hacohen Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours. Dna Xlumberfu Bram D. Zuckerman. M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # INDICATIONS FOR USE 510(k) Number (if known): __ WizAir DVT™ Device Name: # Indications for Use: The WizAir DVT™ is a prescriptive device that induces Continuous Enhanced Circulation Therapy of the lower limbs. The WizAir DVT™ is intended for use in: - . Preventing Deep Vein Thrombosis (DVT). - Enhancing blood circulation. - Diminishing post-operative pain and swelling. - Reducing wound-healing time. - Treatment and assistance in healing: stasis dermatitis; venous stasis ulcers; arterial and diabetic leg ulcers. - Treatment of chronic venous insufficiency. - Reducing edema. ### (PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED) Division of C Respiratory Devices 510(k) Nu K023573 510(k) Number 11/20/02 Prescription Use (Per 21 CFR 801.109) Over the Counter U 8-7
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...