K030532 · Cpca2000, Inc. · DRN · Mar 14, 2003 · Cardiovascular
Device Facts
Record ID
K030532
Device Name
CPCA2000 COUNTERPULSATION SYSTEM
Applicant
Cpca2000, Inc.
Product Code
DRN · Cardiovascular
Decision Date
Mar 14, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5225
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The CPCA2000™ Counterpulsation System is a non-invasive external counterpulsation device intended for use in the treatment of patients with stable angina pectoris, congestive heart failure, acute myocardial infarction and cardiogenic shock. Use of this device may reduce pain and impairment associated with angina pectoris, congestive heart failure or myocardial infarction and may enhance coronary function.
Device Story
Portable counterpulsation system; inputs include ECG and plethysmogram signals. Microprocessor-controlled pump inflates/deflates leg cuffs in synchronization with cardiac cycle. During diastole, sequential inflation (calves to buttocks) increases coronary perfusion pressure/venous return; rapid deflation before systole reduces afterload. Operated by clinician via touch screen; displays patient waveforms and treatment parameters. Data recorded on floppy disk; hard copies via printer. Used in clinical settings to reduce angina pain/impairment and enhance coronary function.
Clinical Evidence
Bench testing only. Evidence includes compliance with IEC60601-1-2 (EMC), UL 2601-1 (General Safety), and software verification/validation per FDA guidance. Device risk and hazard analysis performed per 21CFR820.30(g).
Technological Characteristics
Portable console with microprocessor, pump, reservoir, and strip chart recorder. External leg cuffs and hoses. Connectivity via floppy disk drive and printer. Software-controlled via touch screen interface. Complies with IEC60601-1-2 and UL 2601-1 standards.
Indications for Use
Indicated for patients with stable angina pectoris, congestive heart failure, acute myocardial infarction, and cardiogenic shock.
Regulatory Classification
Identification
An external counter-pulsating device is a noninvasive, prescription device used to assist the heart by applying positive or negative pressure to one or more of the body's limbs in synchrony with the heart cycle.
Special Controls
*Classification.* (1) Class II (special controls) when the device is intended for the treatment of chronic stable angina that is refractory to optimal anti-anginal medical therapy and without options for revascularization. The special controls for this device are:(i) Nonclinical performance evaluation of the device must demonstrate a reasonable assurance of safety and effectiveness for applied pressure, synchronization of therapy with the appropriate phase of the cardiac cycle, and functionality of alarms during a device malfunction or an abnormal patient condition;
(ii) Reliabilities of the mechanical and electrical systems must be established through bench testing under simulated use conditions and matched by appropriate maintenance schedules;
(iii) Software design and verification and validation must be appropriately documented;
(iv) The skin-contacting components of the device must be demonstrated to be biocompatible;
(v) Appropriate analysis and testing must be conducted to verify electrical safety and electromagnetic compatibility of the device; and
(vi) Labeling must include a detailed summary of the device-related and procedure-related complications pertinent to use of the device.
(2) Class III (premarket approval) for the following intended uses: Unstable angina pectoris; acute myocardial infarction; cardiogenic shock; congestive heart failure; postoperative treatment of patients who have undergone coronary artery bypass surgery; peripheral arterial disease associated with ischemic ulcers rest pain or claudication, threatened gangrene, insufficient blood supply at an amputation site, persisting ischemia after embolectomy or bypass surgery, and/or pre- and post-arterial reconstruction to improve runoff; diabetes complicated by peripheral arterial disease or other conditions possibly related to arterial insufficiency including nocturnal leg cramps and/or necrobiosis diabeticorum; venous diseases, including prophylaxis of deep vein thrombophlebitis, edema (e.g., chronic lymphedema) and/or induration (e.g., stasis dermatitis) associated with chronic venous stasis, venous stasis ulcers, and/or thrombophlebitis; athletic injuries, including Charley horses, pulled muscles and/or edematous muscles; necrotizing cellulitis.
(c)
*Date premarket approval application (PMA) or notice of completion of product development protocol (PDP) is required.* A PMA or notice of completion of a PDP is required to be filed with FDA on or before March 31, 2014, for any external counter-pulsating device, with an intended use described in paragraph (b)(2) of this section, that was in commercial distribution before May 28, 1976, or that has, on or before March 31, 2014, been found to be substantially equivalent to any external counter-pulsating device, with an intended use described in paragraph (b)(2) of this section, that was in commercial distribution before May 28, 1976. Any other external counter-pulsating device with an intended use described in paragraph (b)(2) of this section shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.
K020857 — EECP THERAPY SYSTEM MODEL TS3 WITH PULSE OXIMETRY AND ENHANCED EXTERNAL COUNTERPULSATION EECP MC-2 EXTERNAL COUNTER · Vasomedical, Inc. · Jun 14, 2002
K033657 — MOBILE EXTERNAL COUNTER PULSATION SYSTEM ANGIONEW-V · Living Data Technologies Corporation · Dec 19, 2003
K042413 — ACS MODEL NCP-2 · Applied Cardiac Systems · Dec 9, 2004
K033617 — EECP THERAPY SYSTEM MODEL TS4 WITH/WITHOUT PLUSE OXIMETRY · Vasomedical, Inc. · Mar 1, 2004
K050172 — CARDIASSIST COUNTER PULSATION SYSTEM SERIES 4000/5000 · Cardiomedics, Inc. · Mar 31, 2005
Submission Summary (Full Text)
{0}------------------------------------------------
1030532
# (510k) Summary
### CPCA2000, Inc. CPCA2000 Counterpulsation System
MAR 1 4 2003
- 1. Date Prepared:
| 2. Submitter's Name<br>and Address: | CPCA2000, Inc.<br>6336 17Th Street Circle East<br>Sarasota, FL 34243 |
|-------------------------------------|----------------------------------------------------------------------|
|-------------------------------------|----------------------------------------------------------------------|
- Rod Shipman, President 3. Contact Person: CPCA2000, Inc. Telephone (941) 727-4370 Facsimile: (941) 727-4371
- CPCA2000TM Counterpulsation System 4. Device Name:
CPCA2000TM Counterpulsation System Proprietary Name:
Classification Name: Device, Counter-pulsating, External
- The CPCA2000TM Counterpulsation System is substantially 5. Predicate Device: equivalent to the Enhanced External Counterpulsation Device -Model EECP-MC2. FDA granted 510(k) clearance for the predicate device on February 23, 1995 (K940264).
6. Device Description: The CPCA2000 Counter Pulsation System consists of a portable console containing computer, electrocardiogram. plethysmogram, pump and reservoir, integral strip chart recorder, and leg cuffs and hoses. The microprocessor is used to operate and monitor the system via proprietary software, with the operator using touch screen interface to control its operation. The screen displays information pertinent to operating the system, as well as treatment parameters and patient waveforms during use. An external floppy disk drive is used to record data onto transferable media, and a printer is used to produce a hard copy of site, patient and physiologic data. The system accommodates the use of any bed or treatment table.
> External pressure is applied via the patient cuff set to the lower extremities of the patient in synchronization with the heart, i.e., the cuffs compress vascular beds in the calves, lower thighs and upper though/buttocks on inflation. When the heart is in its relaxed state during the diastolic period, pressure is applied sequentially from the calves, to the lower thighs, to the upper thighs and buttocks, forcing blood back to the heart, increasing
{1}------------------------------------------------
coronary perfusion pressure and coronary flow (diastolic augmentation), as well as venous return. Immediately before the heart begins to eject blood during the next systolic phase, the cuffs are rapidly deflated and all externally applied pressure is eliminated. The vasculature in the lower extremities reconforms and is able to receive the output of the heart with lessened resistance, thereby reducing systolic pressure and the workload of the heart (decreased afterload). The CPCA2000™ Counterpulsation System is a non-invasive 7. Intended Use: external counterpulsation device intended for use in the treatment of patients with stable or unstable angina pectoris, congestive heart failure, acute myocardial infarction and cardiogenic shock. Use of this device may reduce pain and impairment associated with angina pectoris, congestive heart failure or myocardial infarction and may enhance coronary function. The principle technological and functional characteristics of the 8. Comparison of current device are same as the predicate devices. The treatment of Technological congestive heart failure patients with the CPCA2000 ECP system
#### 9. Non-Clinical Tests:
Characteristics:
not significantly change the safety and effectiveness of the device. Key non-clinical testing conducted on the CPCA2000™ Counterpulsation System includes the following:
treatment regimen. The treatment of this patient population does
does not require any changes in software, device design or
- IEC60601-1-2; 2001 Medical Electrical Equip., Part 1-. 2. General Requirements for Safety; Collateral Standard, Electromagnetic Compatibility
- . UL 2601-1 Medical Electrical Equipment, Part1: General Requirements for Safety
- Software Verification & Validation, per Guidance for . the Content of Premarket Submission for Medical Devices Containing Software, CDRH, ODE, FDA, May, 1998.
- Device Risk & Hazard Analysis per 21CFR820.30(g), . Design Validation
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles an eagle's head or a stylized representation of human profiles.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 1 4 2003
CPCA2000, Inc. c/o Mr. Michael P. Dayton BioMed Research, Inc. 4608 Rue Bordeaux Lutz. FL 33558
Re: K030532
Trade Name: CPCA2000™ Counterpulsation System Regulation Number: 21 CFR 870.5225 Regulation Name: External Counter-Pulsating Device Regulatory Class: Class III (three) Product Code: DRN Dated: February 14, 2003 Received: February 19, 2003
Dear Mr. Dayton:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 – Mr. Michael P. Dayton
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### PREMARKET NOTIFICATION
## INDICATIONS FOR USE STATEMENT
510k Number (if known):
Device Name: CPCA2000™ Counterpulsation System
Indications For Use:
The CPCA2000™ Counterpulsation System is a non-invasive external counterpulsation device intended for use in the treatment of patients with stable angina pectoris, congestive heart failure, acute myocardial infarction and cardiogenic shock. Use of this device may reduce pain and impairment associated with angina pectoris, congestive heart failure or myocardial infarction and may enhance coronary function.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
| | Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|------------------------------------------|--------------------------------------------------------|----------------------|
| | (Division Sign-Off | |
| | Division of Cardiovascular | |
| 510(k) Number | K030537 | |
| Prescription Use<br>(Per 21 CFR 801.109) | OR | Over-The-Counter Use |
| | (Optional Format 1-2- | |
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.